Skip to content

The Tektronix TCP0020, TCP2020, and TCP202A are a family of high-performance, easy-to-use AC/DC current probes designed for use with a variety of oscilloscopes. Each clamps around a conductor to measure both AC and DC current, presenting low insertion impedance to the circuit under test. The three models differ primarily in their oscilloscope interface and maximum current range.
A current probe lets an engineer measure current without breaking into the circuit, displaying the result as a waveform on a connected oscilloscope. Tektronix lists the intended applications for this family as power supplies, semiconductor devices, power inverters and converters, electronic ballasts, industrial and consumer electronics, mobile communications, motor drives, and transportation systems.
ValueTronics stocks this equipment in its own 20,000 square-foot secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. Every pre-owned probe is inspected and functionally verified in-house before it ships, while new units ship factory-sealed exactly as received from the manufacturer.
The TCP0020, TCP2020, and TCP202A make up a single Tektronix family of AC/DC current probes built for use with a variety of oscilloscopes. They share a 50 MHz-class measurement capability and a common AC/DC current-measurement approach, while differing in how they connect to the host instrument and in their maximum current range.
Tektronix designs the TCP0020 for direct connection to oscilloscopes with the TekVPI interface and the TCP202A for direct connection to oscilloscopes with the TekProbe interface, while the TCP2020 is built for any instrument with BNC inputs and greater-than-100 kΩ termination. This lets a buyer match the probe to the oscilloscope or measurement equipment already on the bench.
Each pre-owned model below is its own dedicated product page with condition-matched pricing. Select the model that fits your oscilloscope interface and current range, then choose the pre-owned listing that matches your budget and availability needs.
The TCP0020 connects directly to oscilloscopes with the TekVPI probe interface and is powered through that interface, with no separate adapter required. It provides automatic units scaling and readout on the oscilloscope display, one-button degauss and AutoZero control, an on-screen probe menu for setup and operating-status information, and remote-control capability through the TekVPI oscilloscope. It offers 20 ARMS maximum current capability and 100 A peak pulse current capability.
The clearest differences across the family are interface and current handling. The TCP0020 (TekVPI) and TCP2020 (BNC output, AC-adapter powered) both provide 20 ARMS maximum current and 100 A peak pulse current, while the TCP202A (TekProbe interface) provides 15 ADC + Peak AC maximum current and 50 A peak pulse current. Maximum amp-second product is 1000 A·µs for the TCP0020 and TCP2020 and 500 A·µs for the TCP202A.
Power and control also vary: the TCP0020 draws power from the TekVPI interface, the TCP202A draws power from the TekProbe interface (or through TekVPI with the TPA-BNC adapter), and the TCP2020 is powered by an AC adapter whose plug option is specified at order time. The TCP2020 additionally offers 10 mA/mV sensitivity. The comparison table below summarizes these per-model differences.
| Model | Max Current | Max Peak Pulse Current | Probe Interface |
|---|---|---|---|
| TCP2020 | 20 ARMS | 100 A | BNC Output |
| TCP0020 | 20 ARMS | 100 A | TekVPI |
| TCP202A | 15 ADC + Peak AC | 50 A | TekProbe |
Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.
| Specification | Value |
|---|---|
| General | |
| Bandwidth | DC to ≥50 MHz |
| Rise Time | ≤7 ns |
| Max Current | 20 ARMS (TCP0020 and TCP2020); 15 ADC + Peak AC (TCP202A) |
| Max Peak Pulse Current | 100 A (TCP0020 and TCP2020); 50 A (TCP202A) |
| Maximum Sensitivity | 10 mA (on oscilloscopes that support 1 mV/div setting) |
| Coupling | DC coupling only |
| Max Bare-wire Voltage | 150 V CAT II (300 V CAT II Insulated Wire) |
| Typical Characteristics | |
| DC Accuracy | ±1% typical, ±3% warranted |
| Max Amp-second Product | 1000 A·µs (TCP0020 and TCP2020); 500 A·µs (TCP202A) |
| Insertion Impedance | 0.036 Ω at 1 MHz; 0.117 Ω at 10 MHz; 0.54 Ω at 50 MHz |
| Signal Delay | 17 ns |
| Environmental | |
| Temperature, Operating | 0 °C to +50 °C |
| Temperature, Nonoperating | –40 °C to +75 °C |
| Humidity, Operating | 5% to 95% RH, tested up to +30 °C (+86 °F); 5% to 85% RH, tested at +30 °C to +50 °C (+86 °F to +122 °F) |
| Humidity, Nonoperating | 5% to 95% RH, tested up to +30 °C (+86 °F); 5% to 85% RH, tested at +30 °C to +75 °C (+86 °F to +167 °F) |
| Regulatory | |
| Safety | CSA1010.1:1997, CSA1010.2.032-96, IEC61010-1:2001, IEC61010-2-032 |
| Physical Characteristics | |
| Probe Head Height | 30.5 mm (1.2 in.) |
| Probe Head Width | 15.2 mm (0.6 in.) |
| Probe Head Length | 148 mm (5.8 in.) |
| Cable Length | 2 m (79 in.) |
| Weight, Probe Only | 0.227 kg (0.5 lb.) |
| Weight, Shipping | 1.22 kg (2 lb. 11 oz.) |
| Power Requirements | |
| TCP0020 | Powered directly by oscilloscopes with the TekVPI probe interface |
| TCP2020 | Powered by AC adapter (specify power plug option when ordering) |
| TCP202A | Powered directly by oscilloscopes with the TekProbe probe interface, or through the TekVPI probe interface when used with the TPA-BNC adapter |
| Warranty | |
| Standard Warranty | One-year parts and labor |
Important: The TCP2020's BNC output requires greater-than-100 kΩ input termination.
Important: The TCP2020 is powered by an AC adapter; specify the power plug option when ordering.
Important: The TCP202A is powered directly by oscilloscopes with the TekProbe interface, or through the TekVPI interface when used with the TPA-BNC adapter.
Important: The maximum bare-wire voltage is 150 V CAT II (300 V CAT II with insulated wire).
Recommended pairing: the 067-1686-xx Deskew/Calibration Fixture for probe deskew, and the 067-2396-xx Current Loop (1 turn, 50 Ω, with BNC connector) for performance verification.
This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.
ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.
Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.
Model No
Tektronix
Condition
Pre-Owned
Manufacturer
Tektronix
Curr_prob_freq
50 MHz
Tektronix TCP404XL 500 A, 2 MHz AC/DC Current Probe (Requires TCPA400 Amplifier) (Pre-Owned)
TCP404XL
TCP0030A Tektronix Current Probe Used
TCP0030A
TDP0500 Tektronix Differential Probe Used
TDP0500
THDP0100 Tektronix Differential Probe Used
THDP0100
MSO2024B Tektronix Mixed Signal Oscilloscope Used
MSO2024B
MSO2014B Tektronix Mixed Signal Oscilloscope Used
MSO2014B
TAP1500 Tektronix Active Probe Used
TAP1500
MSO2004B Tektronix Mixed Signal Oscilloscope Used
MSO2004B
DPO2024B Tektronix Digital Oscilloscope Used
DPO2024B
ADA400A Tektronix Probe Amplifier Used
ADA400A
TMDP0200 Tektronix Differential Probe Used
TMDP0200
DPO2014B Tektronix Digital Oscilloscope Used
DPO2014B
DPO2022B Tektronix Digital Oscilloscope Used
DPO2022B
TCP305A Tektronix Current Probe Used
TCP305A
THDP0200 Tektronix Differential Probe Used
THDP0200
DPO2004B Tektronix Digital Oscilloscope Used
DPO2004B
MSO2002B Tektronix Mixed Signal Oscilloscope Used
MSO2002B
P5205A Tektronix Differential Probe Used
P5205A
DPO2012B Tektronix Digital Oscilloscope Used
DPO2012B
TBS2104B Tektronix Digital Oscilloscope Used
TBS2104
Tektronix P5100A 100X High-Voltage Passive Probe, 500 MHz (Pre-Owned)
P5100A
DPO2002B Tektronix Digital Oscilloscope Used
DPO2002B
TBS2102 Tektronix Digital Oscilloscope Used
TBS2102
Request A Quote