Key Features At A Glance

  • Three-in-one handheld: dual-channel oscilloscope, 6000-count true RMS DMM, and real-time data logger in one unit
  • Bandwidth from 20 MHz to 40 MHz across the two-model series
  • 200 MSa/s real-time sampling (interleaved), 100 MSa/s per channel, with 125 kB/channel memory depth
  • 4.5-inch 320 x 240 color LCD for clear two-channel waveform distinction
  • Advanced triggering: edge, pulse-width, pattern, and video standard across the series
  • USB 2.0 full speed client standard; USB host port optional (#001) for flash-drive storage
  • 11 DMM functions and 22 automatic measurements; Dual Waveform Math standard, FFT on the 40 MHz model
  • PC Link software bundled for remote control and waveform data transfer over USB
View full details

The Agilent U1600A Series is a family of handheld digital oscilloscopes built around a three-in-one design: a dual-channel oscilloscope, a true RMS digital multimeter, and a real-time data logger combined in a single portable instrument. The series consists of two models — the U1602A with 20 MHz bandwidth and the U1604A with 40 MHz bandwidth — each offering a real-time sampling rate up to 200 MSa/s and a 4.5-inch color LCD that clearly distinguishes waveforms between two channels.

Positioned as a high-performance troubleshooting and quality-assurance tool for technical professionals, the U1600A Series targets the installation, maintenance, service, and automotive industries. Its handheld form factor and combined scope-plus-meter capability suit work in demanding industrial environments where capturing signal deviations, glitches, and dropouts quickly matters. The Dual Waveform Math function supports signal addition and subtraction across channels for quick analysis in the time domain, with Fast Fourier Transform analysis available on the 40 MHz model for frequency-domain work.

ValueTronics stocks and holds its own inventory in a 20,000-square-foot secure warehouse at 1675 Cambridge Drive in Elgin, Illinois, giving buyers a single accountable source for both new and pre-owned test instruments. Every used unit is inspected and functionally verified in-house before it ships, while new units leave factory-sealed exactly as received from the manufacturer. Holding stock on our own shelves means faster fulfillment and direct visibility into the equipment we sell.

Brand Heritage

This instrument carries Agilent Technologies branding. Agilent Technologies was spun off from Hewlett-Packard in 1999, and Agilent's electronic measurement business was later established as Keysight Technologies in 2014. Instruments from this period are commonly encountered under any of these names depending on their production date.

Compare Other Models in This Series

The U1600A Series comprises two handheld digital oscilloscope models that share a common three-in-one architecture — oscilloscope, true RMS DMM, and real-time data logger — and differ primarily in bandwidth. The U1602A is the 20 MHz model and the U1604A is the 40 MHz model, with each delivering real-time sampling up to 200 MSa/s and 125 kilobytes of memory per channel.

Both models share the same 4.5-inch color display, 6000-count DMM with 11 measurement functions, advanced triggering set, and USB 2.0 connectivity. The choice between them centers on bandwidth and the additional frequency-domain analysis the higher-bandwidth model provides.

Each model below is its own dedicated product page with condition-matched pricing. Compare the bandwidth tiers, then select the pre-owned unit that fits your application and budget.

The U1604A is the 40 MHz model in the series and is the only model equipped with the Fast Fourier Transform function, which displays waveforms in the frequency domain using four windowing techniques (Rectangular, Hanning, Hamming, and Blackman-Harris). It also provides an equivalent-time sampling rate of 2.5 GSa/s and a finer horizontal resolution of 400 ps.

Across the series, the defining difference is bandwidth: the U1602A is specified at DC to 20 MHz, while the U1604A reaches DC to 40 MHz. This is reflected in rise time — under 17.5 ns on the U1602A versus under 8.8 ns on the U1604A — and in single-shot bandwidth of 20 MHz and 40 MHz respectively.

The U1604A additionally offers the FFT frequency-domain analysis function, a 2.5 GSa/s equivalent sampling rate, and a finer timebase resolution (400 ps versus 2 ns) with a horizontal range starting at 10 ns/div rather than 50 ns/div. Both models otherwise share the same 200 MSa/s real-time sampling, 125 kB/channel memory, DMM, and data-logging capabilities. See the comparison table for the full model-by-model breakdown.

Model Bandwidth Rise Time Timebase Resolution
U1604A 40 MHz < 8.8 ns 400 ps
U1602A 20 MHz < 17.5 ns 2 ns

Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.

Accessories Supplied

  • U1560A Scope probe (1:1) CAT III 300 V
  • U1561A Scope probe (10:1) CAT III 600 V
  • U1571A Ni-MH battery pack 7.2 V, 4500 mA
  • U1580A DMM test lead
  • Ground alligator clip
  • Medium jaw alligator clip
  • Hook Clip
  • USB cable
  • Power cord and AC adapter
  • Quick Start Guide
  • Product Reference CD-ROM containing User's and Service Guide, Quick Start Guide and PC Link Application Software
  • Certificate of Calibration (CoC)
  • Test Report

Product Core & Specifications

Specification Value
Vertical System: Scope Channels
Analog channels Channel 1 and Channel 2 simultaneous acquisition
Bandwidth (–3 dB) DC to 40 MHz
AC coupled < 10 Hz without probe; < 1 Hz with 10 MΩ 10:1 probe
Rise time < 8.8 ns
Single shot bandwidth 40 MHz
Vertical sensitivity 5 mV/div to 100 V/div (1:1 scope probe); 50 mV/div to 1 kV/div (10:1 scope probe); 500 mV/div to 10 kV/div (100:1 scope probe)
Maximum input CAT III 300 Vrms (up to 400 Hz) from terminal to ground
Offset/Dynamic range ± 5 div
Input impedance 1 MΩ || < 20 pF
Coupling AC, DC, GND
DC vertical gain accuracy 5 mV/div to 20 mV/div: ± 5% full scale; 50 mV/div to 100 V/div: ± 3% full scale
Noise peak-to-peak 3% of full scale or 5 mV, whichever is greater
DC vertical offset accuracy ±0.1 div ±2 mV ±0.5% offset value
Single cursor accuracy 4% full scale
Dual cursor accuracy 4% full scale
Probes
Probes U1560-60001: 1:1 passive probe; U1561-60001: 10:1 passive probe; U1562-60001: 100:1 passive probe
Probe attenuation factors 1x, 10x, 100x
Coupling 3 Vp-p, ~ 1 kHz
Maximum probe input 1x CAT III 300 VAC; 10x, 100x CAT III 600 VAC
Acquisition: Scope Channels
Maximum sample rate 200 MSa/s interleaved, 100 MSa/s each channel (50 s/div to 250 ns/div)
Equivalent sample rate 2.5 GSa/s (125 ns/div to 10 ns/div)
Vertical resolution 8 bits
Maximum memory depth 125 kilobytes/channel
Peak detection 5 ns
Average Selectable in average number of 2, 4, 8, 16, 32, 64, 128, 256
Horizontal System
Range 10 ns to 50 s/div
Resolution 400 ps
Reference position Left, center, right
Delay range (pre-trigger) 15 divisions
Delay range (post-trigger) 1000 divisions
Analog Δt accuracy ± 3% reading ± 0.4% screen
Modes Main, XY, Roll
RMS Jitter 5% of horizontal scale or 5 ns whichever is higher
Trigger System
Source Channel 1 and Channel 2
Modes Auto, normal, single
Selections Edge, pulse width, pattern, video
Edge Trigger on a rising or falling edge of any source.
Pattern Trigger at the beginning of a pattern of high, low levels and rising or falling edge established conditions of AND, OR, NOR and NAND between the channels.
Pulse width 200 ns to 10 s. Trigger when a positive or negative pulse width of any source larger than, less than, equal to or not equal to duration.
Video Video trigger sensitivity: 0.7 division trigger level. Available to both Channel 1 and Channel 2. Analog progressive and interlaced video standards including NTSC, PAL and SECAM. Positive or negative sync pulse polarity. Modes — all fields, even fields, odd fields or line 5 – 263 within a field.
Range ± 4 divisions from center screen
Level accuracy ± 0.5 divisions
Trigger sensitivity DC to 5 MHz: 0.8 divisions; 5 MHz to 40 MHz: 1 division
Coupling DC, AC (< 1 Hz), HF reject (> 50 kHz), LF reject (< 30 kHz), Noise reject
Measurement System
Autoscale Finds and displays all active scope channels, sets edge trigger mode on highest numbered channel, sets vertical sensitivity on scope channel. Requires voltage > 20 mVp-p, 0.5% duty cycle and frequency > 100 Hz.
Automatic measurement Measurements continuously updated.
Voltage Peak-to-peak, maximum, minimum, amplitude, top, base, +overshoot, –overshoot, preshoot, RMS, mean and one cycle mean.
Time Frequency, period, +width, –width and +duty cycle and –duty cycle on any channel. Rise time, fall time, delay and phase shift.
Cursors Manually place readout of horizontal (X, DX) and vertical (Y, DY).
Waveform math CH1 + CH2, CH1 – CH2, CH2 – CH1
FFT (U1604A only)
Window Rectangular, Hamming, Hanning, Blackman-Harris
Amplitude Display Selectable in amplitude display of 1 dB, 2 dB, 5 dB, 10 dB
Digital Multimeter Specifications ± (% of reading + % of range)
DC Voltage 600.0 mV 0.3 + 0.08
DC Voltage 6.000 V 0.3 + 0.08
DC Voltage 60.00 V 0.3 + 0.08
DC Voltage 600.0 V 0.3 + 0.08
AC Voltage 600.0 mV – 600.0 V (50 Hz – 1 kHz) 1.0 + 0.2
AC Voltage 600.0 mV – 600.0 V (1 kHz – 30 kHz) 3.0 + 0.2
AC + DC Voltage 6.0000 V – 600.0 V (50 Hz – 1 kHz) 1.0 + 0.2
AC + DC Voltage 6.0000 V – 600.0 V (1 kHz – 30 kHz) 3.0 + 0.2
Resistance 600.0 Ω 0.5 + 0.2
Resistance 6.000 kΩ 0.5 + 0.2
Resistance 60.00 kΩ 0.5 + 0.2
Important: Maximum scope input: CAT III 300 Vrms (up to 400 Hz) from terminal to ground. DMM maximum input: 600 Vrms CAT II, 300 Vrms CAT III.
Important: This instrument is rated for indoor use only (Pollution degree 2).
Important: The optional USB flash drive can only be connected via the USB host port, which is option #001 (the standard configuration provides only the USB 2.0 full speed client port).
Important: FFT (Fast Fourier Transform) function is only available on the U1604A model; the U1602A provides Dual Waveform Math only.
Recommended pairing: add the Agilent U1562A 100:1 scope probe (CAT III 600 V) for higher-voltage measurements and the U1590A soft casing for field protection, both listed as optional accessories.

About this Pre-Owned unit

This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.

  • Warranty included
  • Functional verification included
  • Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
  • Note that unnecessary accessories may not be included (contact Test Architect to confirm).

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.

Model No

Agilent

Condition

Pre-Owned

Manufacturer

Agilent Technologies

Osc_ch

2

Osc_freq

40 MHz

1 mω

Osc_sampl

200 MSa/s