Key Features At A Glance

  • Frequency range from 9 kHz to 1.1 GHz or 3.2 GHz, set by the installed RF path option (SMC-B101 or SMC-B103)
  • Typical maximum output level greater than +17 dBm
  • Low SSB phase noise of typ. –111 dBc (f = 1 GHz, 20 kHz offset, 1 Hz measurement bandwidth)
  • Wideband noise of typ. < –146 dBc (f > 1 MHz, carrier offset > 10 MHz, 1 Hz bandwidth)
  • Level error < 0.9 dB (ALC ON, AUTO mode, +18 °C to +33 °C, specified level range)
  • AM, FM, φM and pulse modulation integrated as standard; frequency and level setting times < 5 ms
  • LAN (10/100BaseT) and USB 2.0 interfaces standard; GPIB (IEEE-488) via optional SMC-K4 for ATE integration
  • Wear-free electronic attenuator with integrated overvoltage protection in a ½ × 19", 2-height-unit chassis
View full details

The Rohde & Schwarz R&S®SMC100A is an analog signal generator built around the design goal of high performance at an attractive price. It is offered with a frequency range of 9 kHz to 1.1 GHz or 9 kHz to 3.2 GHz depending on the installed RF path option, and is positioned as a flexible, universal all-purpose signal generator. It is described as the smallest signal generator in its economy class, occupying a ½ × 19", 2-height-unit form factor, and is lightweight for both bench and portable work.

Rohde & Schwarz identifies the SMC100A as an ideal tool across several application fields: service and maintenance, research and education, field use, secure areas, and simple production applications. Its compact dimensions, integrated analog modulation, and straightforward operation suit these general-purpose stimulus tasks.

ValueTronics holds and stocks its own test-and-measurement inventory in a 20,000 sq ft secure warehouse at 1675 Cambridge Drive, Elgin, Illinois. Used units are inspected and functionally verified in-house against the original manufacturer's specifications before they ship, while new units leave our facility factory-sealed exactly as received from the manufacturer. Handling inventory directly in our own warehouse lets our team stand behind every instrument we sell.

Brand Heritage

Rohde & Schwarz is an independent group of companies specializing in electronics, established more than 75 years ago with company headquarters in Munich, Germany. It has operated under the Rohde & Schwarz name throughout its history, with no major corporate name change or acquisition affecting the brand.

Accessories Supplied

  • Power cable
  • Quick Start Guide
  • CD-ROM (with operating and service manual)

Product Core & Specifications

Specification Value
Frequency
Range (SMC-B101 option) 9 kHz to 1.1 GHz
Range (SMC-B103 option) 9 kHz to 3.2 GHz
Resolution of setting 0.001 Hz
Resolution of synthesis (f = 1 GHz) 4.71 µHz (nom.)
Setting time (with SMC-K4, after IEC/IEEE bus delimiter) < 5 ms, < 2 ms (typ.)
Setting time (with SMC-K4, in ALC OFF mode S&H) < 7 ms, < 3 ms (typ.)
Resolution of phase offset setting 0.1°
Frequency Sweep
Reference Frequency
Frequency error at time of calibration in production < 1 × 10⁻⁷
Frequency error (with SMC-B1 option) < 1 × 10⁻⁸
Aging (after 30 days of uninterrupted operation) < 1 × 10⁻⁶/year
Aging (with SMC-B1 option) < 1 × 10⁻⁹/day, < 1 × 10⁻⁷/year
Temperature effect (0 °C to +55 °C) < 2 × 10⁻⁶
Temperature effect (with SMC-B1 option) < 1 × 10⁻⁷
Warm-up time to nominal thermostat temperature (with SMC-B1 option) ≤ 10 min
Reference output (REF OUT, BNC female) 10 MHz sine wave
Reference output level +6 dBm to +12 dBm, 9 dBm (typ.)
Reference output source impedance 50 Ω (nom.)
Reference input (REF IN, BNC female) 10 MHz
Frequency locking range ±3 × 10⁻⁶
Input level range 0 dBm to +16 dBm
Input impedance 50 Ω (nom.)
Level
Setting range –120 dBm to +19 dBm
Specified level range (200 kHz ≤ f ≤ 3.2 GHz) –120 dBm to +13 dBm (PEP)
Resolution 0.01 dB
Level error (ALC ON, AUTO mode, +18 °C to +33 °C, specified level range) < 0.9 dB
Additional level error with ALC OFF S&H < 0.5 dB
Output impedance VSWR in 50 Ω system (specified level range) < 1.8
Setting time to < 0.1 dB deviation from final value (with SMC-K4, after IEC/IEEE bus delimiter) < 5 ms, < 2 ms (typ.)
Setting time (with SMC-K4, in ALC OFF mode S&H) < 7 ms, < 3 ms (typ.)
Interruption-free level setting range (FIXED MODE, ALC ON) 0 dB to 20 dB
Maximum permissible RF power (1 MHz ≤ f ≤ 1 GHz) 50 W
Maximum permissible RF power (1 GHz < f ≤ 2 GHz) 25 W
Maximum permissible RF power (2 GHz < f ≤ 3.2 GHz) 10 W
Maximum permissible DC voltage 50 V
Level Sweep
Operating mode digital sweep in discrete steps
Sweep range full specified level range
Interruption-free level sweep (attenuator mode FIXED) +0.01 dB to +20 dB
Sweep shape sawtooth, triangle
Step size setting resolution 0.01 dB
Dwell time setting range 10 ms to 10 s
Dwell time setting resolution 0.1 ms
Spectral Purity
Harmonics (f > 1 MHz, level ≤ 8 dBm) < –30 dBc
Nonharmonics (CW, level > –10 dBm, offset > 10 kHz from carrier, f ≤ 1600 MHz) < –60 dBc, < –72 dBc (typ.)
Nonharmonics (1600 MHz < f ≤ 3200 MHz) < –54 dBc
Wideband noise (f > 1 MHz, carrier offset > 10 MHz, level > 5 dBm, 1 Hz BW, CW) < –138 dBc, < –146 dBc (typ.)
SSB phase noise (carrier offset 20 kHz, 1 Hz BW, f = 1 GHz) < –105 dBc, –111 dBc (typ.)
SSB phase noise (f = 2 GHz) < –99 dBc, –105 dBc (typ.)
SSB phase noise (f = 3.2 GHz) < –95 dBc, –101 dBc (typ.)
Residual FM (RMS at f = 1 GHz, CW, 0.3 kHz to 3 kHz) < 5 Hz, 1 Hz (typ.)
Residual FM (0.03 kHz to 23 kHz) < 30 Hz, 8 Hz (typ.)
Residual AM (RMS value 0.03 kHz to 20 kHz, level = 8 dBm) < 0.02 %, < 0.005 % (typ.)
Amplitude Modulation
AM depth setting range 0 % to 100 %
Resolution of setting 1 %
AM depth error (fmod = 1 kHz, m < 80 %, f ≤ 23.4375 MHz) < (1 % of setting + 1 %)
AM depth error (f > 23.4375 MHz) < (4 % of setting + 1 %)
AM distortion (fmod = 1 kHz, m = 30 %, f ≤ 23.4375 MHz) < 1 %
AM distortion (f > 23.4375 MHz) < 3 %
Modulation frequency response (m = 60 %, DC/10 Hz to 50 kHz) < 3 dB
Synchronous φM at AM (m = 30 %, fmod = 1 kHz, peak) < 0.3 rad
Frequency Modulation
External coupling AC, DC
Maximum deviation (f ≤ 23.4375 MHz) 500 kHz
Maximum deviation (23.4375 MHz < f ≤ 25 MHz) 31.25 kHz
Maximum deviation (25 MHz < f ≤ 50 MHz) 62.5 kHz
Maximum deviation (50 MHz < f ≤ 100 MHz) 125 kHz
Maximum deviation (100 MHz < f ≤ 200 MHz) 250 kHz
Maximum deviation (200 MHz < f ≤ 400 MHz) 500 kHz
Maximum deviation (400 MHz < f ≤ 800 MHz) 1 MHz
Maximum deviation (800 MHz < f ≤ 1.6 GHz) 2 MHz
Maximum deviation (1.6 GHz < f ≤ 3.2 GHz) 4 MHz
Resolution < 1 % of maximum deviation, minimum 10 Hz
FM deviation error (fmod = 1 kHz) < (4 % of setting + 20 Hz)
FM distortion (fmod = 2 kHz, half the max. deviation) < 0.2 %
Synchronous AM at FM (40 kHz deviation, fmod = 1 kHz, f > 10 MHz) < 0.2 %
Carrier frequency offset with FM DC (after FM offset adjustment) < 0.4 % of deviation setting
Phase Modulation
Modulation source internal, external
External coupling AC, DC
Modulation frequency response (DC/10 Hz to 100 kHz) < 3 dB
Modulation source internal, external
Operating mode digital sweep in discrete steps
Sweep range full frequency range
Sweep shape sawtooth, triangle
Step size (linear) full frequency range
Step size (logarithmic) 0.01 % to 100 % per step
Dwell time setting range 10 ms to 10 s
Dwell time setting resolution 0.1 ms
Important: The R&S®SMC100A base unit must be ordered together with a frequency path option — either the R&S®SMC-B101 (9 kHz to 1.1 GHz) or the R&S®SMC-B103 (9 kHz to 3.2 GHz).
Important: The R&S SMC-DKD DKD (ISO 17025) calibration including ISO 9000 calibration can only be ordered together with the device.
Important: When pulse modulation is activated, the R&S SMC100A automatically switches to the ALC mode S&H; the ALC loop is opened and the output level is set directly, with a Sample & Hold measurement performed prior to each frequency and level setting.
Recommended pairing: the R&S®NRP-Z91 or R&S®NRP-Z92 power sensors, listed as recommended extras, support level correction and external power measurement with the SMC100A.

About this Pre-Owned unit

This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.

  • Warranty included
  • Functional verification included
  • Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
  • Note that unnecessary accessories may not be included (contact Test Architect to confirm).

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.

Model No

Rohde

Condition

Pre-Owned

Manufacturer

Rohde & Schwarz

Rf_gen_freq

9 kHz to 3.2 GHz

B103

9 kHz to 3.2 GHz