The Advantest R3132/3162 is a series of swept spectrum analyzers built for versatile measurement across a broad frequency range. The R3132 covers 9 kHz to 3 GHz and the R3162 extends coverage to 8 GHz, and the series can be tailored to a given task through a selection of available options.
The series is positioned for applications such as Cable TV, EMC measurement, and digital mobile communications. For EMI precompliance work, 6-dB bandwidth filters for 9 kHz, 120 kHz, and 1 MHz are equipped as standard along with a QP detector, and an optional 200 Hz narrow resolution bandwidth filter can be added. ACP and OBW functions support evaluation of the transmission characteristics of radio systems, and the instrument can also measure the attenuation characteristic of filters or the frequency characteristic of cables when the optional tracking generator is fitted.
ValueTronics stocks and holds its own inventory in a 20,000 square-foot secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. Every used unit is inspected and functionally verified in-house before it ships, while new units ship factory-sealed exactly as received, so buyers know precisely what they are getting either way.
Advantest Corporation traces its roots to Takeda Riken, the Japanese instrument maker that adopted the Advantest name in 1985.
The R3132/3162 series comprises swept spectrum analyzers that share a common platform while differing chiefly in frequency coverage. The R3132 spans 9 kHz to 3 GHz, while the R3162 extends the upper limit to 8 GHz across multiple frequency bands. A 75Ω input variant, the R3132N, is offered for applications calling for 75Ω matching.
All models can be customized through the same set of options and share the series' core measurement architecture, including internal automatic calibration, the 6.5-inch TFT color LCD, a built-in frequency counter, and standard GPIB and RS-232 interfaces.
Each pre-owned model listed below is its own product page with condition-matched pricing. Select the model whose frequency coverage and input impedance match your application, then choose the pre-owned listing that fits your requirements.
The primary difference across the series is frequency range: the R3132 reaches 3 GHz, while the R3162 reaches 8 GHz using banded coverage (Band 0 from 9 kHz to 3.4 GHz, Band 1- from 3.2 GHz to 7.5 GHz, and Band 1+ from 6.5 GHz to 8 GHz). The input attenuator range also differs, spanning 0 to 50 dB on the R3132 and 0 to 75 dB on the R3162.
Input impedance is the other distinguishing factor: the R3132 and R3162 present a nominal 50Ω input, while the R3132N provides a nominal 75Ω input. The comparison table below details the per-model frequency, attenuation, and impedance figures.
| Model | Frequency Range | Input Impedance | Input Attenuation Range |
|---|---|---|---|
| R3132N | 9 kHz to 3 GHz | 75 Ω | 0 to 50 dB |
| R3132 | 9 kHz to 3 GHz | 50 Ω | 0 to 50 dB |
| R3162 | 9 kHz to 8 GHz | 50 Ω | 0 to 75 dB |
Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.
| Specification | Value |
|---|---|
| Frequency | |
| Frequency range | 9 kHz to 3 GHz |
| Frequency reading accuracy (start, stop, center, marker frequency) | ± (Reading of frequency × Frequency reference accuracy + Span × 1% + Resolution bandwidth × 15% + 60 Hz) |
| Counter | |
| Resolution | 1 Hz to 1 kHz |
| Accuracy | ± (Marker frequency × Frequency reference accuracy + 1 LSD) |
| Frequency reference accuracy | |
| Stability | ±2 × 10⁻⁶/year, ±1 × 10⁻⁷/year (Option) |
| Temperature stability | ±1 × 10⁻⁵ (0 °C to 50 °C) |
| Frequency span | |
| Range | 1 kHz to 3 GHz, 0 Hz |
| Accuracy | ≤ ±1% |
| Residual FM | ≤ 60 Hzp-p × 0.1 s |
| Signal purity | ≤ -105 dBc/Hz (20 kHz offset), ≤ -100 dBc/Hz (10 kHz offset) |
| Resolution bandwidth (3 dB) | |
| Range | 1 kHz to 3 MHz, 1-3-10 sequence, 30 Hz to 300 Hz (Option) |
| Accuracy | < ±20%, 1 kHz to 1 MHz; < ±25%, 3 MHz |
| 6 dB bandwidth | 1 MHz, 120 kHz, 9 kHz, 200 Hz (Option) |
| Video bandwidth | 10 Hz to 3 MHz, 1-3-10 sequence |
| Amplitude range | |
| Measuring range | +30 dBm to the average of displayed noise level |
| Maximum input level (Input ATT ≥ 5 dB), Preamplifier OFF | +30 dBm, ±50 VDC max. |
| Maximum input level (Input ATT ≥ 5 dB), Preamplifier ON | +13 dBm, ±50 VDC max. |
| Indication range | 10 × 10 div; Log: 10, 5, 2, 1 dB/div; Linear: 10% of the reference level/div. |
| Reference level range, Preamplifier OFF (Input ATT: 0 to 50 dB) | Log: -64 dBm to +40 dBm (0.1 dB step); Linear: 141.1 µV to 22.36 V |
| Reference level range, Preamplifier ON (Input ATT: 0 to 10 dB) | Log: -84.4 dBm to -20.4 dBm (0.1 dB step); Linear: 13.47 µV to +22.35 mV |
| Input ATT range | 0 to 50 dB (5 dB step) |
| Dynamic range | |
| Average noise level (RBW 1 kHz, VBW 10 Hz, input ATT 0 dB, for 1 MHz or more), Preamplifier OFF | -118 dBm + 2f (GHz) dB |
| Average noise level, Preamplifier ON | -132 dBm + 3f (GHz) dB |
| 1 dB gain compression (at 10 MHz or more), Preamplifier OFF | > 0 dBm (mixer input level) |
| 1 dB gain compression, Preamplifier ON | > -30 dBm (RF input level) |
| 2nd-order harmonic distortion (Preamplifier OFF) | ≤ -80 dBc (Mixer input -30 dBm, f ≥ 100 MHz) |
| 2 signal, 3rd-order intermodulation distortion (Preamplifier OFF) | ≤ -80 dBc (Mixer input -30 dBm, f ≥ 100 MHz, Offset > 10 kHz) |
| Residual response (input ATT 0 dB, 50 Ω terminated), Preamplifier OFF | ≤ -100 dBm |
| Residual response, Preamplifier ON | ≤ -105 dBm |
| Amplitude accuracy | |
| Frequency response after auto calibration, Preamplifier OFF | ≤ ±0.5 dB (100 kHz to 3 GHz, ATT=0 dB); ≤ ±1 dB (100 kHz to 2.7 GHz); ≤ ±2 dB (9 kHz to 3 GHz) |
| Frequency response, Preamplifier ON | ≤ ±1 dB (100 kHz to 2.7 GHz); ≤ ±2 dB (9 kHz to 3 GHz) |
| Calibration signal level accuracy | -20 dBm ±0.3 dB |
| IF gain error | ≤ ±0.5 dB after auto calibration |
| Scale indication accuracy (Log) | ≤ ±0.5 dB (0 to -20 dB); ≤ ±1.5 dB/90 dB; ≤ ±1 dB/10 dB; ≤ ±0.2 dB/1 dB |
| Scale indication accuracy (Linear) | ±5% of reference level |
| Input ATT switching error | ≤ ±0.3 dB (for 0 to 50 dB, with respect to 30 MHz/10 dB) |
| Resolution bandwidth switching level error | ≤ ±0.5 dB after auto calibration |
| Total level accuracy | ±1.5 dB (REF=-50 to 0 dBm, ATT=10 dB, 2 dB/div, RBW=300 kHz, f > 100 kHz, after auto calibration) |
| Sweep | |
| Sweep time | 20 ms to 1000 s, 50 µs to 1 s (Option) |
| Accuracy | < ±1% |
| Trigger mode | FREE RUN, LINE, VIDEO, EXT, TV |
| Sweep mode | REPEAT, SINGLE |
| I/O | |
| RF input connector | N type female |
| Impedance | 75 Ω (nominal) |
| VSWR, Preamplifier OFF (Input ATT 5 to 50 dB) | < 1.5 : 1 (100 kHz to 2 GHz); < 2 : 1 (9 kHz to 3 GHz) |
| VSWR, Preamplifier ON | < 2.5 : 1 (9 kHz to 3 GHz) |
| Probe power | ±12 V, 4 pin connector |
| Calibration output signal | BNC female, 30 MHz, -20 dBm |
| 10 MHz reference input | BNC female, 50 Ω, 0 dBm to +16 dBm |
| External trigger input | BNC female |
| Sound output (demodulated audio) | Small monophonic jack |
| GPIB interface | IEEE-488 specification BUS connector |
| Serial interface | D-sub 9 pin |
| Printer interface | D-sub 25 pin, ESC/P, ESC/P-R, PCL |
| Video out | VGA (15 pin, female) |
| Floppy disc | 3.5 inch, MS-DOS format |
| General | |
| Operating temperature | 0 to +50 °C |
| Humidity | RH 85% or less |
| Storage temperature | -20 to +60 °C, RH 85% or less |
| Power supply | 100/200 VAC switchable; 100 VAC: 100 to 120 VAC, 50 to 60 Hz; 200 VAC: 200 to 240 VAC, 50 to 60 Hz |
| Dimensions | 424 (W) × 177 (H) × 300 (D) mm (without feet and connectors) |
Important: Maximum input level: +30 dBm (input ATT ≥ 5 dB).
Important: For EMI precompliance measurement, 6-dB bandwidth filters for 9 kHz, 120 kHz and 1 MHz as well as a QP detector are equipped as standard; the 200 Hz narrow RBW filter is available as an option.
Important: A tracking generator (100 kHz to 3 GHz) is available only as an option for measuring filter attenuation or cable/amplifier frequency response; it is not included in the standard configuration.
Important: High-speed time domain measurement (50 µs sweep) for mobile communications is an optional function and is not part of the standard sweep range (standard sweep time 20 ms to 1000 s).
This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.
ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.
Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.
Model No
Advantest
Condition
Pre-Owned
Manufacturer
Advantest
Spec_ana_freq
9 kHz to 3 GHz
Request A Quote