Skip to product information
1 of 1
Used
lab tested

PFV-33100 Preen AC Source USED

PFV-33100 Preen AC Source USED

Calibrations

🎓 Request a Quote To Get An EDU Discount

Key Features At A Glance

The Preen PFV-33100 is a PFV Series three-phase high power programmable AC power source with regenerative function, rated 100kVA, with output voltage ranges of 0V-150.0V(L-N) and 0V-300.0V(L-N) at 45-65Hz.

  • Maximum output current (RMS), PFV Series low range (0V-150.0V): 277.8A
  • Maximum output current (RMS), PFV Series high range (0V-300.0V): 138.9A
  • Output phase: 3Ø / 4 Wire + G
  • Output voltage range, PFV Series low range: 0V-150.0V(L-N)
  • Output voltage range, PFV Series high range: 0V-300.0V(L-N)
  • Output frequency range, standard: 45-65Hz
  • Total harmonic distortion (THD): ≤ 2%
  • Crest factor: ≥ 3
  • Regenerative function: YES
  • Efficiency: ≥ 92%
  • Remote interfaces, standard: RS-485, RS-232
View full details

What the Preen PFV-33100 is and what it is used for

The Preen PFV-33100 is a programmable AC power supply, with four quadrant energy feedback function. The Preen PFV-33100 not only provides power to the EUT, but also sinks the power back to the grid system, which the data sheet calls very useful for grid-tied devices testing applications. Efficiency is ≥ 92% at maximum power.

Programmable functions on the Preen PFV-33100 simulate power line disturbances such as voltage or frequency ramp up or ramp down, transient and step changes. Three-phase independent adjustment and phase angle setting are listed for the Preen PFV-33100; voltage ride through (VRT) is listed as NO for the PFV Series, so LVRT and HVRT testing belongs to the PAS-F Series.

Two output ranges let the Preen PFV-33100 trade voltage for current. The low range, 0V-150.0V(L-N), delivers a maximum of 277.8A (RMS), and the high range, 0V-300.0V(L-N), delivers a maximum of 138.9A (RMS); the rating is 100kVA.

Preen PFV-33100 key specifications

Output ratings

Preen PFV-33100: Output ratings (technical document PAS/PFV catalog section (catalog p.52-57) p.5)
Parameter Value Conditions
Output power (VA) 100kVA —
Maximum output current (RMS), PFV Series low range (0V-150.0V) 277.8A —
Maximum output current (RMS), PFV Series high range (0V-300.0V) 138.9A —
Output phase 3Ø / 4 Wire + G —
Output voltage range, PFV Series low range 0V-150.0V(L-N) —
Output voltage range, PFV Series high range 0V-300.0V(L-N) —
Output voltage resolution 0.1V —
Output voltage accuracy 0.5% F.S.+4 counts —
Output frequency range, standard 45-65Hz —
Output frequency range, option 40-70Hz listed as an option in the specification table; the PFV ordering list names no option code for it
Output frequency resolution 0.1Hz —
Output frequency accuracy ±0.02% F.S —
Line regulation ≤ 1% —
Load regulation ≤ 1% Resistive Load
Response time ≤ 2ms the data sheet does not state the conditions of this figure
Total harmonic distortion (THD) ≤ 2% Resistive Load
Crest factor ≥ 3 —

Measurement

Preen PFV-33100: Measurement (technical document PAS/PFV catalog section (catalog p.52-57) p.5)
Parameter Value
Voltage measurement range 0V-300.0V
Voltage measurement resolution 0.1V
Voltage measurement accuracy 0.5%F.S.+4counts
Frequency measurement range Standard : 45-65Hz Option : 40-70Hz
Frequency measurement resolution 0.01Hz
Frequency measurement accuracy ±0.02% F.S
Current measurement range (RMS) 0-9999A
Current measurement resolution (RMS) 0.1A
Current measurement accuracy (RMS) 0.5% F.S.+4counts
Power measurement range 0-1000kW
Power measurement resolution 0.1kW
Power measurement accuracy 1% F.S.+6 counts

General and interfaces

Preen PFV-33100: General and interfaces (technical document PAS/PFV catalog section (catalog p.52-57) p.5)
Parameter Value Conditions
Regenerative function YES —
Voltage ride through (VRT) NO PFV Series
Three-phase independent adjustment YES —
Phase angle setting YES —
Efficiency ≥ 92% at Max. Power
Operator interface (HMI) Touch Screen —
Input protection Input N.F.B, Over Voltage, Under Voltage —
Output protection Over Voltage, Over Current, Reverse Current, Over Temperature —
Remote interfaces, standard RS-485, RS-232 —
Remote interfaces, optional GPIB, Ethernet, USB —

Compare the Preen PFV-33100 with other models

Preen PFV-33100: compare PFV Series Three-Phase Output (30kVA-400kVA) (technical document PAS/PFV catalog section (catalog p.52-57) p.5)
Model Output power (VA) Maximum output current (RMS), PFV Series high range (0V-300.0V) Maximum input current Weight
PFV-33030 30kVA 41.6A 58.7A 942kg
PFV-33045 45kVA 62.5A 88.1A 1050kg
PFV-33060 60kVA 83.3A 117.4A 1185kg
PFV-33075 75kVA 104.1A 146.8A 1485kg
PFV-33100 (this page) 100kVA 138.9A 195.7A 1919kg
PFV-33120 120kVA 166.6A 234.9A 2300kg
PFV-33150 150kVA 208.3A 293.6A 2700kg
PFV-33200 200kVA 277.8A 391.4A 3400kg
PFV-33300 300kVA 416.7A 587.1A 4500kg
PFV-33400 400kVA 555.6A 782.8A 5600kg

What decides whether the Preen PFV-33100 fits

Power and current: does the Preen PFV-33100 cover your EUT?

The Preen PFV-33100 is rated 100kVA, with a maximum output current of 277.8A (RMS) on the low range and 138.9A (RMS) on the high range, and a crest factor of ≥ 3. Pick the range that holds your EUT voltage, then check its RMS and peak current against that range.

Voltage and frequency: does the Preen PFV-33100 reach your test points?

The Preen PFV-33100 output is 3Ø / 4 Wire + G, with voltage ranges of 0V-150.0V(L-N) and 0V-300.0V(L-N) and a standard frequency range of 45-65Hz (40-70Hz is listed as an option in the specification table). Output voltage accuracy is 0.5% F.S.+4 counts, total harmonic distortion ≤ 2%, and line and load regulation ≤ 1% and ≤ 1% with a resistive load.

Site supply: what does the Preen PFV-33100 need?

The Preen PFV-33100 input is 3Ø / 3 Wire + G, 380V±15% at 47-63Hz, with a maximum input current of 195.7A (based on a rated input voltage of 380V minus 15%) and a power factor of ≥ 0.99 at maximum power. The catalog refers other input voltages to the manufacturer.

Installation: is there room for the Preen PFV-33100?

The Preen PFV-33100 measures 2200 x 1600 x 800 mm / 86.61 x 62.99 x 31.49 inch (H x W x D) and weighs 1919kg; the catalog notes that the weight of certain models is estimated. Operating temperature is 0℃~45℃, humidity 0-90% non condensing, at an altitude below 1,500m.

Control and protection: how is the Preen PFV-33100 operated and guarded?

The Preen PFV-33100 has a Touch Screen operator interface. RS-485 and RS-232 remote interfaces are standard, and GPIB, Ethernet and USB are options. Input protection covers Input N.F.B, Over Voltage, Under Voltage; output protection covers Over Voltage, Over Current, Reverse Current, Over Temperature.

Preen PFV-33100 options and what they change

The soft start, interface and other ordering options for the Preen PFV-33100 are listed on the Options tab.

What you need with the Preen PFV-33100

Preen PFV-33100: AC input (technical document PAS/PFV catalog section (catalog p.52-57) p.5)
Parameter Value Conditions
Maximum input current 195.7A based on a rated input voltage of 380V minus 15% (footnote *2)
Input phase 3Ø / 3 Wire + G —
Input voltage 380V±15% the data sheet refers other input voltage specifications to the manufacturer (footnote *1)
Input frequency 47-63Hz —
Input power factor ≥ 0.99 Max. Power

Other Preen PFV-33100 specifications

Environment and physical

Preen PFV-33100: Environment and physical (technical document PAS/PFV catalog section (catalog p.52-57) p.5)
Parameter Value Conditions
Weight 1919kg The weight of certain models is estimated (footnote *4)
Weight (lbs) 4230.7 lbs The weight of certain models is estimated (footnote *4)
Operating temperature 0℃~45℃ —
Humidity 0-90% Non condensing
Altitude < 1,500m —
Dimensions (H x W x D) 2200 x 1600 x 800 mm / 86.61 x 62.99 x 31.49 inch H x W x D; the data sheet refers detailed dimensions to the manufacturer (footnote *3)

Overview

Preen PFV-33100: Overview (technical document PAS/PFV catalog section (catalog p.52-57) p.2, 3)
Parameter Value
Gradual and step functions voltage or frequency ramp up or ramp down, transient and step changes
Product type programmable AC power supply, with four quadrant energy feedback function
Energy feedback not only provides power to the EUT, but also sinks the power back to the grid system
Applications grid-tied devices testing applications

Ordering information

Preen PFV-33100: Ordering information (technical document PAS/PFV catalog section (catalog p.52-57) p.6)
Parameter Value
Ordering description 100kVA/300V/45-65Hz, Including Regenerative Function

Preen PFV-33100 questions and answers

What is the Preen PFV-33100?

The Preen PFV-33100 is a PFV Series high power programmable AC power source rated 100kVA with three-phase output and regenerative function, a programmable AC power supply, with four quadrant energy feedback function.

What is the Preen PFV-33100 used for?

The Preen PFV-33100 supplies three-phase AC to an EUT and sinks the power back to the grid, which the data sheet calls very useful for grid-tied devices testing applications.

What is the AC output voltage range of the Preen PFV-33100?

The Preen PFV-33100 has two output voltage ranges, 0V-150.0V(L-N) and 0V-300.0V(L-N), set with a resolution of 0.1V and an accuracy of 0.5% F.S.+4 counts.

What is the AC output frequency range of the Preen PFV-33100?

The Preen PFV-33100 output frequency range is 45-65Hz as standard, with a resolution of 0.1Hz and an accuracy of ±0.02% F.S. The specification table lists 40-70Hz as an option; the PFV ordering list does not name a frequency option code.

What is the output apparent power of the Preen PFV-33100?

The Preen PFV-33100 output power is 100kVA, with a maximum output current of 277.8A (RMS) on the low range and 138.9A (RMS) on the high range.

Which phase configurations does the Preen PFV-33100 support?

The Preen PFV-33100 has a three-phase output, 3Ø / 4 Wire + G, with three-phase independent adjustment and phase angle setting.

Is the Preen PFV-33100 regenerative?

The Preen PFV-33100 is regenerative (the data sheet lists the regenerative function as YES), and its efficiency is ≥ 92% at maximum power.

Which options change the waveform and disturbance synthesis of the Preen PFV-33100?

The PFV Series ordering list for the Preen PFV-33100 has soft start and interface options only (PFV-001 to PFV-004); the harmonics editor is listed as a PAS series option, and voltage ride through is not listed for the PFV Series. Gradual and step functions program voltage or frequency ramp up or ramp down, transient and step changes.

What does the Preen PFV-33100 measure?

The Preen PFV-33100 measures voltage over 0V-300.0V with an accuracy of 0.5%F.S.+4counts, current over 0-9999A (RMS) with an accuracy of 0.5% F.S.+4counts, and power over 0-1000kW with an accuracy of 1% F.S.+6 counts.

How is an AC power source different from a DC power supply?

The Preen PFV-33100 produces three-phase AC at a set voltage and frequency and programs line disturbances such as ramps and steps; a DC power supply produces a DC output. The Preen PFV-33100 also sinks power from the EUT back to the grid.

How is the Preen PFV-33100 different from the PAS-F Series?

The Preen PFV-33100 adds a low output range (0V-150.0V(L-N)) to the 0V-300.0V(L-N) range, while voltage ride through is listed as NO for the PFV Series and the harmonics editor as a PAS series option. The compare table on this page lists every PFV model.

Specifications are from the manufacturer's technical document (Preen PAS/PFV Series Regenerative Grid Simulator (catalog pages 52-57), PAS/PFV catalog section (catalog p.52-57), no date printed). Questions? Contact your Test Architect.

Updated 2026-11-26

Every used unit is fully tested in our lab, and its performance is guaranteed to pass an optional calibration.

Model No

PFV-33100

Condition

Used

Manufacturer

Preen

Frequency

65 Hz

Phase

3

Power (va)

100 kVA

Volts

300 V

PFV-001

Soft Start Function.

PFV-002

GPIB Interface.

PFV-003

Ethernet Interface.

PFV-004

USB Interface.