Skip to product information
1 of 1

Agilent N5264A PNA-X Measurement Receiver (Pre-Owned)

Agilent N5264A PNA-X Measurement Receiver (Pre-Owned)

🎓 Request a Quote To Get An EDU Discount

Login & Get 2% Back on Purchases
Login or Sign Up for Loyalty

Agilent N5264A PNA-X Measurement Receiver (Pre-Owned)

Key Features At A Glance

  • Five simultaneous receiver inputs and five measurement receivers
  • Measurement speed up to 400,000 points/sec at 600 kHz IFBW, CW (Fast CW mode, no point triggering)
  • Selectable IF bandwidth from 1 Hz to 5 MHz
  • 4-billion-byte data buffer; up to 500 million points per single cut (single parameter)
  • Ethernet (LAN), USB, GPIB and RS-232 host interfaces for automated test systems
  • Removable hard drive for secure environments
  • Optional internal LO output (Option 108) covering 10 MHz to 26.5 GHz
  • Five external IF inputs (A, B, C, D, R) on 50 Ohm SMA (female) connectors
View full details

The Agilent N5264A is a PNA-X measurement receiver that provides five simultaneous receiver inputs and five measurement receivers in a single instrument. It is built around external IF inputs that accept signals from remote mixers or frequency converters, with a selectable IF bandwidth spanning 1 Hz to 5 MHz.

The external IF inputs allow the use of external IF signals from remote mixers or frequency converters, presented as five inputs labeled A, B, C, D, and R on 50 Ohm SMA (female) connectors. The instrument exposes Ethernet, USB, and GPIB host computer interfaces, and its frequency source control interface uses a TTL handshake for coordinating an external source.

ValueTronics International stocks and holds its own inventory in a 20,000 sq ft secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. Every used unit is inspected and functionally verified in-house before it ships, while new units ship factory-sealed exactly as received from the manufacturer.

Brand Heritage

Hewlett-Packard's test and measurement business was spun out as Agilent Technologies in 1999. Agilent's electronic measurement business was subsequently established as Keysight Technologies in 2014.

Product Core & Specifications

Specification Value
Key Specifications
Measurement Speed (max) points/sec @ 600 KHz IFBW, CW frequency 400,000 points/sec (Fast CW mode - no point triggering)
Receiver Inputs 5 (simultaneously)
Measurement Receivers 5 (simultaneously)
Data Buffer Size 4 billion bytes
Data Buffer size (max. points for single cut) 500 million points (for single parameter; two parameters are 250 million points each)
IF Bandwidth 1 Hz to 5 MHz
Frequency Source Control Interface TLL hand shake
Trigger In / Out Three pairs
Host Computer Interface Ethernet, USB and GPIB
Security Hard drive removable
External IF Inputs
Function Allows use of external IF signals from remote mixers or frequency converters
Connectors SMA (female); A, B, C, D, R
Input Frequency 7.605634 MHz
Input Impedance 50 Ω
RF Damage Level +23 dBm
DC Damage Level 1 VDC
0.1 dB Compression Point -9.0 dBm
Compression @ -10 dBm, Magnitude 0.03 dB
Compression @ -10 dBm, Phase 0.23°
Noise Floor, 10 Hz IF BW -143 dBm
Noise Floor, 10 KHz IF BW -113 dBm
Crosstalk -134 dB
Dynamic Range @ 10 Hz 134 dB @ 0.1dB compression to noise floor
Dynamic Accuracy (-40 dBm reference, over range set by compression and noise floor @ IF Frequencies)
-10 dBm 0.037 dB
-20 dBm 0.024 dB
-30 dBm 0.016 dB
-40 dBm 0.010 dB
-50 dBm 0.013 dB
-60 dBm 0.021 dB
-70 dBm 0.032 dB
-80 dBm 0.041 dB
-90 dBm 0.049 dB
-100 dBm 0.057 dB
-110 dBm 0.072 dB
-120 dBm 0.188 dB
LO output 2 (Option 108)
Frequency Stability +/- 0.05 ppm, -10 to 70C, +/- 0.1ppm/yr max
Frequency Accuracy +/- 1 ppm
Frequency Range 10 MHz to 26.5 GHz
Frequency Switching Speed < 100 microsecond/point
Frequency Resolution 1 Hz
Power Flatness +/- 1.0 dB
Power Output +10 dBm
2nd Harmonics, 20 MHz to 2.0 GHz -23 dBc
2nd Harmonics, 2.0 GHz to 5.0 GHz -28 dBc
2nd Harmonics, 5.0 GHz to 23.0 GHz -35 dBc
2nd Harmonics, 23.0 GHz to 26.5 GHz -27 dBc
3rd Harmonics, 30 MHz to 8.0 GHz -32 dBc
3rd Harmonics, 8.0 GHz to 15.0 GHz -38 dBc
3rd Harmonics, 15.0 GHz to 26.5 GHz -48 dBc
Phase Noise, 10 MHz to 500 MHz (1/10/100 kHz, 1 MHz offset) -80 / -85 / -76 / -113 dBc/Hz
Phase Noise, 500 MHz to 1 GHz (1/10/100 kHz, 1 MHz offset) -90 / -110 / -106 / -115 dBc/Hz
Phase Noise, 1 GHz to 2 GHz (1/10/100 kHz, 1 MHz offset) -85 / -105 / -101 / -110 dBc/Hz
Phase Noise, 2 GHz to 4 GHz (1/10/100 kHz, 1 MHz offset) -80 / -100 / -96 / -105 dBc/Hz
Phase Noise, 4 GHz to 8 GHz (1/10/100 kHz, 1 MHz offset) -74 / -94 / -90 / -99 dBc/Hz
Phase Noise, 8 GHz to 16 GHz (1/10/100 kHz, 1 MHz offset) -68 / -88 / -84 / -93 dBc/Hz
Phase Noise, 16 GHz to 26.5 GHz (1/10/100 kHz, 1 MHz offset) -62 / -82 / -78 / -87 dBc/Hz
10 MHz Reference
Reference In Connector BNC, female
Reference In Input Frequency 10 MHz ± 10 ppm, typical
Reference In Input Level -15 dBm to +20 dBm, typical
Reference In Input Impedance 200 Ω, nom.
Reference Out Connector BNC, female
Reference Out Output Frequency 10 MHz ± 1 ppm, typical
Reference Out Signal Type Sine Wave, typical
Reference Out Output Level +10 dBm ± 4 dB into 50 Ω
Reference Out Output Impedance 50 Ω, nominal
Reference Out Harmonics <-40 dBc, typical
External Monitor Information
VGA Video Output Connector 15-pin mini D-Sub; Drives VGA compatible monitors
Flat Panel (TFT) Resolutions 1024 X 768, 800 X 600, 640 X 480
Flat Panel (DSTN) Resolutions 800 X 600, 640 X 480
CRT Monitor Resolutions 1280 X 1024, 1024 X 768, 800 X 600, 640 X 480
Test Set IO 25-pin D-Sub connector, available for external test set control
Power IO 9-pin D-Sub, female; analog and digital IO
Handler IO 36-pin parallel I/O port; all input/output signals are default set to negative logic; can be reset to positive logic via GPIB command
Trigger Information
Trigger In/Meas Trigger, Nominal Input Impedance 5K Ohms
Trigger In/Meas Trigger, Minimum Pulse Width 1 us
Trigger In/Meas Trigger, DC Damage Level 5.5 volts
Trigger In/Meas Trigger, Drive Voltage TTL (0, +5.0) Volts
Trigger Out/Meas Trigger Ready, Nominal Input Impedance 5K Ohm
Trigger Out/Meas Trigger Ready, Pulse Width = Data acquisition
Trigger Out/Meas Trigger Ready, Polarity Selectable with sweep or point mode
Trigger Out/Meas Trigger Ready, Drive Voltage TTL (0, +5.0) Volts
Trigger Inputs/Outputs (Aux. 1 & 2) BNC(f), TTL/CMOS compatible
GPIB 24-pin D-sub (Type D-24), female; compatible with IEEE-488 (two ports - dedicated controller and dedicated talker/listener)
Parallel Port (LPT1) 25-pin D-Sub miniature connector, female; provides connection to printers or any other parallel port peripherals
Serial Port (COM 1) 9-pin D-Sub, male; compatible with RS-232
USB Port Four ports on front panel (all Host) and five ports (four hosts and one Device) on rear panel. Type A configuration (eight hosts) and Type B configuration (one Device), USB 2.0 compatible
LAN 10/100BaseT Ethernet, 8-pin configuration; auto selects between the two data rates
Line Power
Frequency, Voltage 50/60 Hz for 100 240 VAC (auto switching)
Max 450 watts
Important: External IF input damage levels: +23 dBm RF and 1 VDC.
Important: Simultaneous operation of the internal and external displays is allowed, but with 640 X 480 resolution only. If you change resolution, you can only view the external display (internal display will "white out").
Important: All specifications and characteristics apply over a 25 °C ±5 °C range (unless otherwise stated) and 90 minutes after the instrument has been turned on.
Important: When configuring the N5264A Option 108 as the LO source, you may improve system measurement sensitivity by using a method of AM noise suppression.
Important: For Regulatory and Environmental information, refer to the PNA Series Installation and Quick Start Guide, located online at http://cp.literature.agilent.com/litweb/pdf/E8356-90001.pdf.

About this Pre-Owned unit

This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.

  • Warranty included
  • Functional verification included
  • Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
  • Note that unnecessary accessories may not be included (contact Test Architect to confirm).

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.

Model No

Agilent

Condition

Pre-Owned

Manufacturer

Agilent