Key Features At A Glance

  • Frequency coverage from 9 kHz to 8.1 GHz spanning Band 0 (0 to 3.2 GHz) and Band 1 (2.92 to 8.1 GHz)
  • Noise sidebands of ≤−100 dBc/Hz at 10 kHz offset (1 GHz) for analysis of weak signals adjacent to strong carriers
  • Average noise level of ≤−115 dBm (1 MHz to 1 GHz, RBW 1 kHz, VBW 1 Hz, RF ATT 0 dB)
  • Total level accuracy of ±1.3 dB (100 kHz to 3.1 GHz) after calibration using the internal calibration signal
  • Synthesized local oscillator with frequency counter resolution to 1 Hz and span accuracy of ±2.5% (span ≥10 kHz)
  • Built-in Measure functions: channel power, adjacent channel power, occupied bandwidth, burst average power, C/N, frequency counter, template/MASK pass-fail
  • GPIB and RS-232C interfaces standard, plus dual PCMCIA Type-I memory card slots for trace data, parameters, and bitmap screen captures
  • Compact 320 × 177 × 351 mm chassis at ≤13.5 kg (standard configuration) for portable bench and field use
Share on
View full details

The Anritsu MS2663C is a portable benchtop spectrum analyzer covering 9 kHz to 8.1 GHz, organized internally into Band 0 (0 to 3.2 GHz) and Band 1 (2.92 to 8.1 GHz) with a preselector spanning 2.92 to 8.1 GHz. The instrument uses a synthesized local oscillator and a Gaussian resolution-bandwidth filter, and is presented by Anritsu as a compact, lightweight platform for the manufacture and maintenance of radio equipment, with the 8.1 GHz upper limit providing headroom to measure spurious products well above the mobile communications bands in widest use at the time of the datasheet.

A spectrum analyzer displays signal energy across a frequency range — showing what signals are present, at what frequency, and at what power level — so engineers can verify transmitter output, hunt interference, characterize filters and amplifiers, and document compliance against published limits. Per the datasheet, the MS2663C is positioned for evaluation of radio equipment, with built-in functions including a frequency counter, C/N, channel power, adjacent channel power, occupied frequency bandwidth, burst average power, and template pass/fail comparison.

Brand Heritage

Anritsu Corporation is a Japanese precision measurement company with roots reaching back to 1895. In 1993 Anritsu acquired Wiltron Company, an established US RF and microwave instrument maker, consolidating the Wiltron product line into the modern Anritsu RF and microwave portfolio.

Accessories Supplied

  • Power cord, 2.6 m: 1 pc
  • F0013 Fuse, 5 A: 2 pcs
  • W1251AE MS2650B,MS2660B/C series operation manual: 1 copy
  • B0329G Front cover 3/4MW4U

Product Core & Specifications

Category Specification Value
Frequency Frequency range 9 kHz to 8.1 GHz
Frequency band Band 0 (0 to 3.2 GHz), band 1- (2.92 to 6.5 GHz), band 1+ (6.4 to 8.1 GHz)
Pre-selector range 2.92 to 8.1 GHz (band 1-, 1+)
Display frequency accuracy ±(display frequency x reference frequency accuracy + span x span accuracy + 100 Hz); Span: ≥10 kHz, after calibration
Marker frequency display accuracy Normal: Same as display frequency accuracy; Delta: Same as frequency span accuracy
Frequency counter Resolution: 1 Hz, 10 Hz, 100 Hz, 1 kHz; Accuracy: Display frequency x reference frequency accuracy ±1 LSD (at S/N: ≥20 dB)
Frequency span Setting range: 0 Hz, 1 kHz to 8.2 GHz; Accuracy: ±2.5% (span: ≥10 kHz), ±5% (span: <10 kHz, Option 02 installed)
Resolution bandwidth (RBW) (3 dB bandwidth) Setting range: 1 kHz, 3 kHz, 10 kHz, 30 kHz, 100 kHz, 300 kHz, 1 MHz, 3 MHz; Bandwidth accuracy: ±20% (1 kHz to 1 MHz), ±30% (3 MHz); Selectivity (60 dB:3 dB): ≤15:1
Video bandwidth (VBW) 1 Hz to 3 MHz (1-3 sequence), OFF
Noise sideband, stability Noise sidebands: ≤-100 dBc/Hz (1 GHz, 10 kHz offset); Residual FM: ≤20 Hzp-p/0.1 s (1 GHz, span: 0 Hz); Frequency drift: ≤200 Hz/min
Amplitude Measurement range Average noise level to +30 dBm
Maximum input level +30 dBm (CW average power, RF ATT: ≥10 dB), ±0 Vdc
Average noise level ≤-115 dBm (1 MHz to 1 GHz, band 0); ≤-115 dBm + 1.5f [GHz] dB (1 to 3.1 GHz); ≤-115 dBm + 0.5f [GHz] dB (2.92 to 8.1 GHz, band 1)
Residual response ≤-100 dBm (RF ATT: 0 dB, input: 50 Ω terminated, 1 MHz to 8.1 GHz)
Total level accuracy ±1.3 dB (100 kHz to 3.1 GHz, band 0), ±2.3 dB (2.92 to 8.1 GHz, band 1)
Reference level setting range Log scale: -100 to +30 dBm; Linear scale: 224 µV to 7.07 V
Frequency response ±0.5 dB (100 kHz to 3.2 GHz, band 0); ±1.5 dB (9 to 100 kHz, band 0); ±1.5 dB (2.92 to 8.1 GHz, band 1)
Spurious response 2nd harmonic distortion: ≤-75 dBc (0.2 to 1.3 GHz, mixer input: -30 dBm); 3rd order intermodulation: ≤-80 dBc (0.1 to 8.1 GHz)
Sweep Sweep time Setting range: 20 ms to 1000 s; Accuracy: ±15% (20 ms to 100 s), ±45% (110 to 1000 s)
Sweep mode Continuous, single
Time domain sweep mode Analog zero span, digital zero span
Zone sweep Sweeps only in frequency range indicated by zone marker
Tracking sweep Sweeps while tracing peak points within zone marker
Display/Functions Number of data points 501
Detection mode NORMAL, POS PEAK, NEG PEAK, SAMPLE
Display Color TFT-LCD, Size: 5.5", Number of colors: 17
Storage functions NORMAL, VIEW, MAX HOLD, MIN HOLD, AVERAGE, CUMULATIVE, OVER WRITE
Input connector N-J, 50 Ω
Reference Oscillator Frequency 10 MHz
Aging rate 2 x 10⁻⁶/year (typical)
Temperature characteristics 1 x 10⁻⁵ (typical, 0 to 50°C)
General EMC EN61326:1997/A2:2001 (Class A), EN61000-3-2:2000 (Class A)
LVD EN61010-1:2001 (Pollution Degree 2)
Vibration Meets MIL-STD-810D
Power (operating range) 85 to 132/170 to 250 Vac (automatic voltage switching), 47.5 to 63 Hz, 380 to 420 Hz (85 to 132 V only), ≤330 VA
Dimensions and mass 320 (W) x 177 (H) x 351 (D) mm, ≤13.5 kg (without option)
Ambient temperature 0 to +50°C (operate), -40 to +75°C (storage)
Important: Maximum input level: +30 dBm CW average power (RF ATT ≥10 dB), ±0 Vdc.Important: Option 08 (Pre-amplifier) and Option 20 (Tracking generator) cannot be installed simultaneously.Important: Option 10 (Centronics interface) and the standard GPIB interface cannot be installed simultaneously.Important: Option 12 (QP detector) requires Option 02 (Narrow resolution bandwidth).Important: Option 04 (High-speed time domain sweep) must be used with Option 06 (Trigger/gate circuit).Important: Option 21 and Option 24 (Television monitor) require Option 08 (Pre-amplifier).Important: Maximum input level: +30 dBm (CW average power, RF ATT ≥10 dB), ±0 Vdc. Exceeding this rating may damage the instrument.

About this used unit

  • Warranty included
  • Functional verification included
  • Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
  • Note that unnecessary accessories may not be included (contact Test Architect to confirm).

Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.

Model No

Anritsu

Condition

Used

Manufacturer

Anritsu

Frequency

8.1 GHz