  • AC Withstanding Voltage in a single bench instrument
  • 200VA AC test capacity with PWM switching amplifier for high-stability output
  • AC withstanding: 0.100–5.000 kV (2V steps), 40 mA max current
  • 240 × 64 Ice Blue dot-matrix LCD with function-key navigation
  • 100 MANU + 100 AUTO memory blocks, each step individually named
  • RS-232C, USB, Signal I/O standard; GPIB optional
  • Zero-crossing turn-on, safety interlock, 150µs abnormality cutoff, 200ms DUT discharge
  • Meets test requirements for IEC, EN, UL, CSA, GB, JIS safety standards
Share on
View full details

In summary, the GW Instek GPT-9801 is a benchtop electrical safety tester that performs AC withstanding (hipot) testing only. It is built on a high-efficiency PWM amplifier platform delivering 200 VA of AC output capacity, with high-resolution measurement (1µA current resolution, 2V voltage steps) and a 240 × 64 dot-matrix LCD for clear readout of conditions, parameters, and results. The instrument is designed to meet the test requirements of IEC, EN, UL, CSA, GB, JIS, and other safety standards applied to electrical products and components.

What is the GW Instek GPT-9801 used for?

The GPT-9801 is used for production-line safety testing, quality-assurance verification, and safety-standard compliance pre-qualification of electrical products. Typical devices under test include power cords, home appliances, information technology equipment, and medical equipment. Operators apply the tester at the manufacturing, regulation-compliance, and R&D stages to confirm that insulation barriers between live circuits and accessible surfaces are intact and that protective earth connections are reliable under fault current.

Which safety tests does the GPT-9801 perform?

It performs AC withstanding (hipot) testing — applying a sinusoidal high voltage between the live circuit of the DUT and its accessible surfaces or protective ground to verify the integrity of the electrical insulation barrier. All major test functions are performed under high-stability voltage output with high-resolution measurement results. Test duration, ramp-up time, and upper/lower limits are fully adjustable to accommodate a wide variety of safety tests.

The GPT-9801 delivers the following measurable performance against datasheet specifications: AC withstanding 0.100–5.000 kV at 200 VA, 40 mA max, 2V resolution, ±(1% of setting + 5V) accuracy; Ramp time 0.1–999.9 s, test time off or 0.5–999.9 s, automatic DUT discharge within 200 ms; 100 MANU memory blocks plus 100 AUTO memory blocks with up to 16 steps per AUTO sequence. Output regulation is held to less than 1% under load variation, and the PWM amplifier topology achieves up to 98% efficiency — reducing internal heating and improving service lifetime compared to conventional transformer + class A/B designs.

The GPT-9801 addresses the electrical-safety regulatory framework that governs mains-connected products in global markets. The instrument is designed to support testing to IEC 60601-1 (medical electrical equipment), IEC 60335 (household appliances), IEC 62368-1 and IEC 60950 (information technology and audio/video equipment), and the corresponding UL, CSA, EN, GB, and JIS derivatives applied at certification and in production. GW Instek (Good Will Instrument Co., Ltd., established 1975, Taipei) is a global supplier of test and measurement instruments distributed through authorized partners worldwide.

Common applications

  • Single-test AC hipot stations where only AC withstanding is mandated
  • Production-line safety screening of mains-powered consumer products
  • Compliance pre-qualification labs that already own separate IR and ground bond instruments

Accessories Supplied

The GPT-9801 ships from the factory with the following standard accessories:

Item Qty
Quick Start Guide 1
Power cord 1
CD (complete user manual) 1
Interlock key 1
Remote terminal male plug 1
Test lead GHT-114 1

Optional Accessories

Part Number Description
Opt.1 GPIB card
GHT-113 High Voltage Test Pistol
GHT-205 High Voltage Test Probe
GTL-232 RS-232C cable, 9-pin female to 9-pin null modem, for computer
GTL-247 USB cable, A-A type, approx. 1.8 m
GTL-248 GPIB cable, approx. 2 m
GRA-402 Rack adapter panel (19", 4U)

Product Core & Specifications

The GPT-9801 is built on a high-efficiency PWM amplifier platform with up to 98% efficiency, reducing internal heat versus conventional variable-transformer + class A/B designs. Output regulation stays under 1%. Hardware and software protections include power-on self-check, zero-crossing turn-on, output cutoff within 150µs of an abnormal condition, automatic DUT discharge within 200 ms, and a hardware safety interlock.

After "Start" is pressed, a 100 ms initialization applies ~100V to detect short-circuit defects before full test voltage ramps up. The 240 × 64 Ice Blue LCD displays conditions, parameters, measurement values, and pass/fail simultaneously, with multi-color status LEDs that flash whenever output voltage is present.

AC Withstanding Specifications

Output Voltage Range 0.100kV – 5.000kV ac
Output Voltage Resolution 2V/step
Output Voltage Accuracy ±(1% of setting + 5V) [no load]
Maximum Rated Load 200 VA (5kV/40mA)
Maximum Rated Current 40mA (0.5kV
Output Voltage Waveform Sine wave
Output Voltage Frequency 50Hz / 60Hz selectable
Voltage Regulation ±(1% of reading + 5V) [full load → no load]
Voltmeter Accuracy ±(1% of reading + 5V)
Current Measurement Range 0.001mA – 40.0mA
Current Best Resolution 0.001mA / 0.01mA / 0.1mA
Current Measurement Accuracy ±(1.5% of reading + 30µA)
Window Comparator Method Yes
ARC Detect Yes
RAMP (Ramp-Up Time) 0.1s – 999.9s
TIMER (Test Time) OFF, 0.5s – 999.9s
GND RETURN / GUARD

Note on probes & accessories: The optional GHT-113 High Voltage Test Pistol and GHT-205 High Voltage Test Probe support manual contact testing of larger assemblies. Probe choice depends on fixture geometry, voltage range, and operator-safety procedures — consult your Test Architect to match the right accessory to the application.

Important — high-voltage safety: The GPT-9801 produces lethal voltages at its output terminals during testing. The hardware interlock key must be installed in the Signal I/O port for the tester to send high voltage; never bypass it. Calibration documentation is available as an optional service where formal accredited testing requires NIST-traceable verification.

General & Interface Specifications

Memory — Single Step (MANU) 100 blocks
Memory — Automatic Testing (AUTO) 100 blocks, menu per auto: 16
Interface — RS-232C Standard
Interface — USB Standard
Interface — GPIB Option
Remote Terminal (Front) Standard
Signal I/O Standard
Display 240 × 64 Ice Blue dot-matrix LCD
Power Source AC 100V / 120V / 220V / 230V ±10%, 50/60Hz
Dimensions (W × H × D) 330 × 150 × 460 mm
Weight Approx. 15kg max.

About This Pre-Owned Unit

  • Warranty included.
  • Functional verification included.
  • Calibration available as an optional add-on service.
  • Unnecessary accessories may not be included — contact your Test Architect to confirm anything specific your application requires.

This webpage was written by a human with an A.I. "Intern", which may contain errors. Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.

Model No

GPT-9801

Condition

Used

Manufacturer

Instek