Key Features At A Glance

  • Single-phase AC output from 0 to 310 Vrms across 155 V and 310 V ranges, with a 40 to 500 Hz frequency range
  • Output power spans 500 VA to 4000 VA across the four-model AC6800B Series
  • AC and DC output modes with built-in voltage, current, and power measurements (AC, DC, and AC+DC coupling)
  • LAN/LXI Core and USB standard for automation; GPIB available as a user-installable option (AC68GPBU)
  • Built-in Web server enables remote setup, monitoring, and control through a standard browser
  • Front-panel interface consistent with Keysight DC power supplies for a familiar workflow
  • Optional user-installable analog input board (AC68BALGU) adds basic transient capability
View full details

The Keysight AC6800B Series is a family of single-phase basic AC power sources designed to deliver stable, dependable AC power for bench and production-line testing. Available in models up to 4000 VA, each source provides 0 to 310 Vrms output and a 40 to 500 Hz frequency range, engineered to keep test systems running even when grid voltage fluctuates or the device under test draws extreme inrush current or transient spikes.

These sources are built for both manual control on an R&D bench and programmatic operation in a fully automated test rack. Typical uses include simulating global AC power conditions, testing varying frequency and voltage combinations to reproduce real-world conditions, and power factor correction testing and similar AC+DC applications such as examining ripple voltage on input circuitry. Example applications include testing universal power supplies designed to work globally, supplying single-phase 400 Hz power to check avionic electronics in development, manufacturing, and repair, and testing commercial LED drivers and fluorescent ballasts up to 310 VAC — suitable for electronics rated to 277 VAC ±10%.

ValueTronics stocks and holds its own inventory in a 20,000 square-foot secure warehouse at 1675 Cambridge Drive in Elgin, Illinois.

Brand Heritage

Keysight Technologies traces its roots to the test-and-measurement business of Hewlett-Packard, which was spun off as Agilent Technologies in 1999 and subsequently established as the independent Keysight Technologies in 2014.

Compare Other Models in This Series

The AC6800B Series is a single-phase basic AC power source family spanning four models — the AC6801B, AC6802B, AC6803B, and AC6804B. All four share the same 0 to 310 Vrms output across 155 V and 310 V ranges and the same 40 to 500 Hz frequency range, while differing in maximum output power and current.

Every model in the series provides both AC and DC output modes, built-in voltage, current, and power measurements, and the same intuitive front-panel and programmatic interface. Models up to 4000 VA are available to cover a range of basic AC source requirements.

Each model below links to its own dedicated product page with condition-matched pricing. Select the model that fits your power and current requirements to view the new unit and its current availability.

The primary differences across the series are output power and current. Maximum output power steps from 500 VA on the AC6801B to 1 kVA, 2 kVA, and 4 kVA on the AC6802B, AC6803B, and AC6804B, with maximum rms current on the 310 V range rising correspondingly from 2.5 A to 5 A, 10 A, and 20 A (and double those values on the 155 V range). See the comparison table for the per-model figures.

Physical dimensions, weight, and input terminal type also scale with power — for example, weight ranges from approximately 8 kg on the AC6801B to about 31 kg on the AC6804B, and a standard power cord is included with the AC6801B. Optional accessories such as the user-installable analog (AC68BALGU), inhibit (AC68BFIU), and GPIB (AC68GPBU) interface boards, along with rack-mount flange kits, apply across the series.

Model Output Power Max AC Current (Low/High Range) DC Power Capacity
AC6804B 4000 VA 40 A/20 A 3200 W
AC6801B 500 VA 5 A/2.5 A 400 W
AC6802B 1000 VA 10 A/5 A 800 W
AC6803B 2000 VA 20 A/10 A 1600 W

Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.

Accessories Supplied

  • Ferrite core (1)
  • Cable tie (1)

Product Core & Specifications

Specification Value
Output Ratings — AC Mode (155 V/310 V range)
Rated voltage range 1 to 155 Vrms / 2 to 310 Vrms
Voltage setting accuracy 0.35% of full scale
Output phases Single
Maximum rms current 40 A / 20 A
Load power factor capability 0 to 1 (leading or lagging)
Maximum power 4 kVA
Frequency setting range 40 to 500 Hz
Frequency accuracy ± 0.02%
Output Ratings — DC Mode (155 V/310 V range)
Rated voltage range 1.4 to 219 V / 2.8 to 438 V
Voltage setting accuracy 0.3% of full scale / 0.25% of full scale
Maximum rms current 32 A / 16 A
Maximum power 3200 W
Output Voltage Stability (155 V/310 V range)
Load regulation For 40 to 100 Hz: ± 0.15 V / ± 0.3 V; for other frequencies: ± 0.5 V / ± 1 V
Measurements (155 V/310 V range)
Voltage accuracy (RMS, AC, DC) For 45 to 65 Hz and DC: ± (0.5% of reading + 0.35 V/0.7 V); typical: ± (0.25% of reading + 0.15 V/0.3 V); for all other frequencies: ± (0.7% of reading + 1.0 V/2.0 V)
Current accuracy (RMS, AC, DC) For 45 to 65 Hz and DC: ± (0.5% of reading + 0.16 A/0.08 A); typical: ± (0.25% of reading + 0.16 A/0.08 A); for all other frequencies: ± (0.7% of reading + 0.32 A/0.16 A)
AC power accuracy (45 to 65 Hz, WAC) ± (2% of reading + 4 W); typical: ± (1% of reading + 4 W)
DC power accuracy (WDC) ± (2% of reading + 4 W + 0.16 W/0.08 W per VDC); typical: ± (1% of reading + 4 W + 0.08 W/0.04 W per VDC)
Input Rating
Nominal input rating 100 to 120 Vrms / 200 to 240 Vrms, 50 Hz or 60 Hz, single-phase
Input voltage range 90 to 132 Vrms / 180 to 264 Vrms (auto detected when the power is turned on)
Input frequency range 47 Hz to 63 Hz
Apparent power 6400 VA or less
Power factor 0.9 (typical)
Input current 64 A/32 A or less (at 100 V/200 V); 53.6 A/28.0 A or less (at 120 V/230 V)
Output Ratings — AC Mode, Supplemental (155 V/310 V range)
Programmable voltage range 0 to 157.5 Vrms / 0 to 315 Vrms
Voltage setting resolution 0.1 V
Frequency setting resolution 0.1 Hz
Voltage THD 0.5% or less
Efficiency 70% or greater
Maximum repetitive peak current 120 A / 60 A
Output Ratings — DC Mode, Supplemental (155 V/310 V range)
Programmable voltage range –222.5 to 222.5 V / –445 to 445 V
Voltage setting resolution 0.1 V
Maximum repetitive peak current 96 A / 48 A
Ripple noise 0.7 Vrms / 1.4 Vrms (typical)
Output Voltage Stability, Supplemental
Temperature coefficient 100 ppm/°C (typical)
Output voltage response time 150 µs (typical)
Line regulation ± 0.15% of full scale
Measurements, Supplemental (135 V/270 V range)
Voltage measurement resolution 0.1 V
Current measurement resolution 0.01 A
Peak current measurement accuracy ± (2% of reading + 0.8 A/0.4 A) (typical)
Power measurement resolution 0.1 W, 1 W (for 1000 W or more)
Common
Isolation to ground 310 Vrms or 438 VDC
Insulation resistance (input–chassis, output–chassis, input–output) 500 Vdc, 30 MΩ or more
Withstand voltage (input–chassis, output–chassis, input–output) 1.5 kV AC for 1 minute
Earth continuity 25 A AC, 0.1 Ω or less
Electromagnetic Compatibility (EMC)
Standards Complies with EMC Directive 2004/108/EC, EN 61326-1
Safety
Standards Complies with Low Voltage Directive 2006/95/EC, EN 61010-1, Class I, Pollution Degree 2
Environment
Operating environment Indoor use, Overvoltage Category II
Operating temperature and humidity range 0 to 40 °C (32 to 104 °F), 20% to 80% R.H. non-condensing
Storage temperature and humidity range –10 to 60 °C (14 to 140 °F), 90% or less R.H. non-condensing
Altitude Up to 2000 m
Acoustic noise < 70 dBA
Physical
Dimensions (W x H x D, depth includes Barrier Block Safety Cover) 429 x 262 x 613.5 mm (16.9 x 10.3 x 24.2 in)
Weight Approx. 31 kg (68.34 lb)
Input terminal M6 terminal block
Output terminal M6 terminal block
I/O Interfaces
LAN hardware IEEE 802.3 100Base-TX or 10Base-T Ethernet; complies with LXI Specification version 1.4 Class C; IPv4, RJ-45 connector
LAN communication protocol VXI-11, HiSLIP, or SCPI-RAW
USB hardware Complies with USB 2.0; data rate 12 Mbps (full speed); USBTMC-USB488 device class
GPIB interface (Option AC68GPBU) Complies with IEEE Std 488.1-1978; SH1, AH1, T6, L4, SR1, RL1, PP0, DC1, DT1, C0, E1; primary address 1 to 30
Software protocol IEEE Std 488.2-1992
Command language SCPI Specification 1999.0
Save/recall states Save and recall up to ten instrument states in non-volatile memory
Important: EMC compliance requires the ferrite core attached on the load wires for CE-marked models, with the maximum length of all connecting cables and wires to the unit kept under 3 m.
Important: AC68xx models will not be in compliance with EMC limits unless the ferrite core is attached on the load wires (applies only to models that have CE marking on the panel).
Important: A standard power cord is included only with the AC6801B; the AC6802B, AC6803B, and AC6804B do not include a power cord.
Important: The power cord supplied with the AC6801B is based on country of use (power cords for Continental Europe, United States/Canada, or China are included); for other countries an applicable power cord can be selected.
Important: Accuracy specifications are warranted over the ambient temperature range of 0 to 40 °C after a 30-minute warm-up period and apply at the output terminals; accuracy specifications are warranted for three years.

About this New unit

This is a brand-new, factory-sealed unit, supplied exactly as it arrives from the manufacturer and backed by the full manufacturer warranty. For the exact configuration, available options, included accessories, or current lead time for your application, our Test Architects can confirm the details before you order.

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.

Model No

Keysight

Condition

Factory New

Manufacturer

Keysight Technologies

Ac_supply_freq

40 to 500 Hz

Ac_supply_phase

Single-phase

Ac_supply_va

4000 VA

Ac_supply_volts

310 V