Sensitivity: 100nV to 1V fullscale, in 1-3 steps (10nV at the METER MAG operation)
Frequency range: 0.5 Hz to 200 kHz
Input type: Differential
Input impedence: 10M // 40pF
Common-mode rejection ratio: 110dB or greater(at 1 kHz)
Input referred noise: 5nV /√Hz or lower(at 1 kHz)
Filter: LPF, HPF, BPF, automatic tuning is available.
Reference signal mode: EXT F/2F, INT F/2F
Internal oscillator: 0.5 Hz to 120 kHz
Measurement display: DATA1 : One of either A, A(dB), A% X(Acosφ), X(dB), X(%); DATA2 : One of either φ or Y (Asin φ); DATA3 : One of either ratio,reference frequency, EXT DC
Monitoring meter: Acosφ and Asinφ
Time constant: 1ms to 30s, 6dB/oct or 12dB/oct
Averaging functions: Linear or exponential averaging mode; No. of averages : 21 to 29
Interface: GPIB, RS-232C
Power requirements: AC100, 120, 220 or 240V switchable
Dimensions(mm): 432(W) X 132.5(H) X 500(D)
Weight: approx.13.5kg
Computer interface(GPIB, RS232C)
Automatic functions : AUTO SET, AUTO RANGE, AUTO NORMALIZE, PHASE SET