Key Features At A Glance

  • Bandwidth from 600 MHz to 1 GHz across the series, with up to 4 GSa/s sample rate on interleaved channels
  • Available as 2-channel, 4-channel, 2+16-channel, or 4+16-channel mixed-signal configurations
  • MegaZoom deep memory: 2 Mpts per channel standard (4 Mpts max interleaved), expandable to 64 Mpts per channel (128 Mpts max interleaved)
  • Mixed-signal models add 16 digital channels with selectable thresholds (TTL, CMOS, ECL, PECL, user-defined) for time-correlated analog and digital debug
  • Windows XP Pro open platform with 10/100 LAN, GPIB, USB, web-enabled remote control, and e-mail-on-trigger connectivity
  • Advanced triggering including edge, glitch, pattern, state, delay-by-time, delay-by-events, TV, setup/hold, and transition violation modes
  • Optional protocol decode, jitter analysis, compliance test, and FPGA dynamic probe applications expand the platform for serial and embedded work
Share on
View full details

The Agilent Infiniium 54830 Series is a family of digital storage and mixed-signal oscilloscopes built for engineers developing high-speed digital and communications designs. The series spans 600 MHz and 1 GHz bandwidth models in 2-channel, 4-channel, 2+16-channel, and 4+16-channel configurations, combining real-time scope acquisition with integrated logic timing channels on the mixed-signal versions. Every model in the family shares the same MegaZoom deep-memory architecture, the same Windows-based user interface, and the same advanced trigger set, so the selection decision comes down to bandwidth, channel count, and whether digital timing channels are required.

Engineers reach for the 54830 Series when they need to verify and debug embedded systems, characterize high-speed digital buses, and analyze the interaction between analog signals and digital control. The mixed-signal models are positioned for designs that include complex digital buses such as PCI or SDRAM, where the engineer needs to trigger on a specific bus condition and simultaneously view the analog characteristics of the data and clock lines. The data sheet also calls out communications mask testing against ITU-T G.703, ANSI T1.102, and IEEE 802.3 templates, Ethernet electrical compliance testing for 10BaseT, 100BaseT, and 1000BaseT, USB 2.0 signal-integrity compliance testing, and FPGA internal-signal probing for Xilinx devices as supported application areas through optional software.

ValueTronics International has been a trusted source for test and measurement instruments since 1992. Each Agilent 54830 Series oscilloscope handled by ValueTronics is processed through our 20,000 square foot secure warehouse at 1675 Cambridge Drive in Elgin, Illinois, where our technicians inspect, test, and prepare every unit before it ships. We are not a drop-shipper or a pass-through broker; we hold the instruments we sell, we know their condition, and we stand behind them. Customers who need a 54830 Series scope can speak directly with a specialist who can confirm what is in inventory, what accessories are included, and what calibration options are available.

Brand Heritage

Hewlett-Packard's test and measurement business was spun out as Agilent Technologies in 1999. In 2014, Agilent's electronic measurement business was transferred to Keysight Technologies, which is the current corporate home for the Infiniium oscilloscope line. Instruments such as the 54830 Series that carry Agilent branding from this era continue to be supported through the Keysight service infrastructure.

Compare Other Models in This Series

The Agilent Infiniium 54830 Series is a family of eight oscilloscope models sharing a common platform: the same MegaZoom deep-memory architecture, the same Windows XP Pro user interface, the same advanced trigger set, and the same connectivity options. The models differ along three axes called out on the datasheet's specification page: analog bandwidth (600 MHz or 1 GHz), analog channel count (2 or 4), and whether 16 digital channels are included to make the instrument a mixed-signal oscilloscope.

All eight models share the same 4 GSa/s maximum interleaved sample rate (2 GSa/s per channel), the same 250 GSa/s equivalent-time sampling on repetitive signals, and the same memory-upgrade ladder reaching 128 Mpts interleaved through options 040, 080, 160, 320, and 640. The choice between models is driven by the customer's bandwidth requirement, the number of analog signals to view simultaneously, and whether digital bus debug is part of the application.

Each model below is its own product page with pre-owned-matched pricing, condition reporting, accessory listings, and warranty terms. Customers should select the page that matches both the model they need and the condition state they are working with.

The bandwidth split runs across the series: the 54830B, 54831B, 54830D, and 54831D are 600 MHz instruments with a calculated rise time of 583 ps, while the 54832B, 54833A, 54832D, and 54833D are 1 GHz instruments with a calculated rise time of 350 ps. The channel-count split separates 2-channel models (54830B, 54833A, 54830D, 54833D) from 4-channel models (54831B, 54832B, 54831D, 54832D). The MSO designation, indicated by the D suffix, adds 16 digital channels with their own trigger engine and threshold groupings to the base scope.

Standard acquisition memory is 2 Mpts per channel (4 Mpts interleaved) on every model except the 54833A, which starts at 500 kpts per channel (1 Mpts interleaved). All other specifications including sample rate, trigger set, measurement library, mask testing, FFT range, and connectivity are common across the family. The comparison table that follows summarizes the model-level differences side by side.

Model Bandwidth Analog Channels Digital Channels
54832B 1 GHz 4 0
54830D 600 MHz 2 16
54831D 600 MHz 4 16
54832D 1 GHz 4 16

Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.

Accessories Supplied

  • Optical USB Mouse
  • Condensed Keyboard
  • User's Quick Start Guide (English)
  • Documentation CD (Service Guide, Programmer's Guide)
  • Information System (English)
  • Accessory pouch (54810-68701)
  • Front panel cover
  • Power cord

Product Core & Specifications

Category Specification Value
Vertical: Analog Channels Input Channels 54830B/54833A: 2 analog; 54830D/54833D: 2 analog + 16 digital; 54831B/54832B: 4 analog; 54831D/54832D: 4 analog + 16 digital
Vertical: Analog Channels Analog Bandwidth @50 Ω (-3 dB) 54830B/D, 54831B/D: 600 MHz; 54832B/D, 54833A/D: 1 GHz
Vertical: Analog Channels Calculated Rise Time @50 Ω 54830B/D, 54831B/D: 583 ps; 54832B/D, 54833A/D: 350 ps
Vertical: Analog Channels Input Impedance 1 MΩ ± 1% (13 pF typical), 50 Ω ± 1.5%
Vertical: Analog Channels Sensitivity 1 mV/div to 5 V/div (1 MΩ); 1 mV/div to 1 V/div (50 Ω)
Vertical: Analog Channels Input Coupling 1 MΩ: AC, DC; 50 Ω: DC
Vertical: Analog Channels Hardware Bandwidth Limit 20 MHz
Vertical: Analog Channels Vertical Resolution 8 bits, ≥12 bits with averaging
Vertical: Analog Channels Channel-to-Channel Isolation DC to 50 MHz: 50 dB; >50 MHz to 500 MHz: 40 dB; >500 MHz to 1 GHz: 30 dB
Vertical: Analog Channels DC Gain Accuracy ± 1.25% of full scale at full resolution channel scale
Vertical: Analog Channels Maximum Input Voltage (1 MΩ) 150 V RMS or DC, CAT I; ± 250 V (DC + AC) in AC coupling
Vertical: Analog Channels Maximum Input Voltage (50 Ω) 5 Vrms, CAT I
Vertical: Analog Channels Dynamic Range ± 8 div from center screen (1 MΩ); ± 12 div from center screen (50 Ω)
Vertical: Digital Channels (MSO only) Number of Channels 16 Digital – labeled D15 – D0
Vertical: Digital Channels (MSO only) Threshold Groupings Pod 1: D7 – D0; Pod 2: D15 – D8
Vertical: Digital Channels (MSO only) Threshold Selections TTL, 5.0V CMOS, 3.3V CMOS, 2.5V CMOS, ECL, PECL, User Defined
Vertical: Digital Channels (MSO only) User-Defined Threshold Range ±8.00 V in 10 mV increments
Vertical: Digital Channels (MSO only) Maximum Input Voltage ±40 V peak CAT I
Vertical: Digital Channels (MSO only) Threshold Accuracy ±(100 mV + 3% of threshold setting)
Vertical: Digital Channels (MSO only) Input Dynamic Range ±10 V about threshold
Vertical: Digital Channels (MSO only) Minimum Input Voltage Swing 500 mV peak-to-peak
Vertical: Digital Channels (MSO only) Input Impedance 100 kΩ ± 2% (~ 8 pF) at probe tip
Vertical: Digital Channels (MSO only) Channel-to-Channel Skew 2 ns typical, 3 ns maximum
Vertical: Digital Channels (MSO only) Glitch Detect ≥ 2.5 ns
Vertical: Digital Channels (MSO only) Resolution 1 bit
Horizontal Main Time Base Range 54830B/D, 54831B/D: 500 ps/div to 20 s/div; 54832B/D, 54833A/D: 200 ps/div to 20 s/div
Horizontal Horizontal Position Range 0 to ± 200 s
Horizontal Delayed Sweep Range 1 ps/div to current main time base setting
Horizontal Resolution 4 ps
Horizontal Timebase Accuracy 15 ppm (±0.0015%)
Horizontal Channel-to-Channel Deskew Range –100 µs to 100 µs
Horizontal Modes Main, Delayed, Roll
Horizontal Reference Positions Left, Center, Right
Acquisition: Analog Real Time Sample Rate (Max), 2 Channels Interleaved 4 GSa/s
Acquisition: Analog Real Time Sample Rate (Max), Each Channel 2 GSa/s
Acquisition: Analog Equivalent Time Sample Rate (Max) 250 GSa/s
Acquisition: Analog Memory Depth Standard (Interleaved/each channel) 4 M / 2 M (1 M / 500 K for 54833A)
Acquisition: Analog Memory Depth Option 040 8 M / 4 M
Acquisition: Analog Memory Depth Option 080 16 M / 8 M
Acquisition: Analog Memory Depth Option 160 32 M / 16 M
Acquisition: Analog Memory Depth Option 320 64 M / 32 M
Acquisition: Analog Memory Depth Option 640 128 M / 64 M
Important: Maximum input voltage: 150 V RMS or DC at the 1 MΩ inputs (CAT I), ±250 V (DC + AC) in AC coupling, and 5 Vrms at the 50 Ω inputs (CAT I). Digital channels (D-series models): ±40 V peak CAT I.Important: Passive probes are not included with the 54832B, 54833A, 54832D, and 54833D models. A passive probe option (001, 002, or 004) or a recommended 1130A or 1156A active probe must be ordered separately.Important: The 1130A InfiniiMax probe amplifier does not include probe heads. For a complete probing solution, also order a connectivity kit or individual probe head(s); E2675A differential browser probe head is recommended.Important: Segmented Memory acquisition mode is a standard feature with version A.03.70 system software or higher, but only available on 54830B/31B/32B models with serial number MY41003401 or greater, 54830D/31D/32D models with MY42001701 or greater, or 54833A/33D models with MY43000601 or greater.Important: After-purchase acquisition memory upgrade options 160, 320, and 640 are only available on 54830B/31B/32B models with serial number MY41003601 or greater, 54830D/31D/32D models with MY42001901 or greater, or 54833A/33D models with MY43000801 or greater.Important: The N5383A Infiniium Performance Upgrade Kit is required to upgrade system software from A.02.xx (Windows 98) to A.03.xx or higher (Windows XP Pro).Recommended pairing: the Agilent 81100A series pulse pattern generators are identified on the datasheet as a complement to the 54830 Series for building a complete stimulus-and-response measurement system.

About this used unit

  • Warranty included
  • Functional verification included
  • Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
  • Note that unnecessary accessories may not be included (contact Test Architect to confirm).

Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.

Model No

Agilent

Condition

Used

Manufacturer

Agilent

Channels

4

Frequency

1 GHz

Record length

4 MPts

Sampling rate

4 GS/s