Skip to content

The HP 4339A High-Resistance Meter is a precision instrument for making high-resistance measurements on capacitors, relays, switches, connectors, materials, cables, and PC boards. It measures DC resistance (R), DC current (I), surface resistivity (ps), and volume resistivity (pv), with a measurement range from 1 × 10³ Ω to 1.6 × 10¹⁶ Ω and a basic accuracy of 0.6%.
High-resistance measurement is used to characterize insulation integrity, dielectric properties, and leakage behavior in passive components and materials. The HP 4339A is designed for measurement of capacitors, relays, switches, connectors, materials, cables, and PC boards. The grounded device-under-test capability supports measurement of cables and transformers under grounded conditions, which is required for evaluating components in their installed reference configuration.
ValueTronics stocks, tests, and ships precision component-measurement instruments like the HP 4339A from our 20,000 sq ft secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. Every pre-owned unit is inspected and functionally verified by our in-house technical team before it leaves the building, so the instrument arriving on your bench performs as the documentation describes — not as a distant listing photo suggests.
Hewlett-Packard's test and measurement business was spun out as Agilent Technologies in 1999 and then transferred to Keysight Technologies in 2014. Instruments originally branded HP, like the 4339A, are part of the lineage that today carries the Keysight name.
| Specification | Value |
|---|---|
| Measurement Parameters | R (dc resistance), I (dc current), ps (surface resistivity), pv (volume resistivity) |
| Measurement Range (I) | 60 fA to 100 µA |
| Measurement Range (R) | 1 x 10^3 Ω to 1.6 x 10^16 Ω |
| Basic Accuracy (I) | ±0.4% |
| Basic Accuracy (R) | ±0.6% |
| Measurement Time (Short) | 10 ms (typical) |
| Measurement Time (Medium) | 30 ms (typical) |
| Measurement Time (Long) | 390 ms (typical) |
| Display | 5 digits (max) |
| Mathematical Functions | Deviation and percent deviation |
| Ranging | Auto and manual |
| Trigger | Internal, external, manual, and HP-IB |
| Delay Time | 0 to 9999 ms in 1 ms steps |
| Test Cable Lengths | 2 m maximum |
| Averaging | 1 to 256 |
| Test Voltage | 0.1 to 1000 Vdc, 0.1 V steps @ 0.1 to 200 V, 1 V steps @ 200 to 1000 V |
| Voltage Accuracy | (0.16% + 100 mV) @ ≤ 200 V; (0.16% + 500 mV) @ > 200 V |
| Maximum Current | 10 mA @ ≤ 100 V, 5 mA @ ≤ 250 V, 2 mA @ ≤ 500 V, 1 mA @ ≤ 1 kV |
| Current Compliance Setting | 0.5 mA, 1 mA, 2 mA, 5 mA, 10 mA |
| Output Resistance | 1 kΩ ±10% |
| Input Resistance | 1 kΩ ±10% |
| Correction Function | Zero OPEN: Eliminates measurement errors due to stray parasitic resistance in the test fixtures |
| Test Sequence Program | Controls a series of resistance measurements. Charge time, measurement internal time, and measurement number can be programmed |
| Comparator Function | HIGH/IN/LOW for the measurement parameter |
| Contact Check Function | Contact failure between the test fixture and device can be detected. Available DUT type: Capacitive DUT's only |
| DUT Capacitance | ≥0.5 pF + 5% of residual stray capacitance |
| Residual Stray Capacitance of the Fixture | ≤50 pF |
| Additional Measurement Time for Contact Check | 2 ms |
| Save/Recall | Ten instrument setups can be saved/recalled from the internal nonvolatile memory |
| Continuous Memory Capability | If the instrument is turned off, or if a power failure occurs, instrument settings are automatically memorized (≥ 72 hours at 23 ±5° C) |
| HP-IB Interface | All control settings, measured values, and comparator information |
| Handler Interface | All output signals are negative-logic, optically isolated open collectors. Output signals include: HIGH/IN/LOW, no contact, index, end of measurement, and alarm. Input signals include: high voltage off, keylock, and external trigger |
| Power Requirements | 90 to 132 V or 198 to 264 V, 47 to 66 Hz, 45 VA maximum |
| Operating Temperature | 0 to 55° C |
| Size | 320 mm W x 100 mm H x 450 mm D (12.6 in x 3.94 in x 17.72 in) |
| Weight | 6.5 kg (14.3 lb) |
Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.