Skip to content

The Keysight N3304A is a DC electronic load module that installs in a slot of an N3300A or N3301A mainframe and absorbs power from a power supply, battery charger or DC to DC converter under test, in constant current, constant voltage or constant resistance mode. Its 16-bit voltage, current and power measurement system uses individual but linked current and voltage measurement systems.
The N3300 Series loads operate with full stability down to zero volts. For manufacturing test, downloaded “Lists” of command sequences execute independent of the computer, rising and falling slew rates are controlled separately, and the built-in transient generator operates in all modes. On the bench, the standard connector accepts an unterminated wire and may be hand-tightened.
Conditions: at 25 °C ± 5 °C
| Parameter | Value |
|---|---|
| Input voltage rating | 0–60 V |
| Input current rating | 0–60 A |
| Maximum power @ 40 °C | 300 W |
| Maximum power derating (Table A-1 note 1) | Maximum continuous power available is derated linearly from 100% of maximum at 40 °C, to 75% of maximum at 55 °C. |
| Parameter | Value | Conditions |
|---|---|---|
| Specified current @ low voltage operation, 2.0 V | 60 A | at 25 °C ± 5 °C |
| Specified current @ low voltage operation, 1.5 V | 45 A | at 25 °C ± 5 °C |
| Specified current @ low voltage operation, 1.0 V | 30 A | at 25 °C ± 5 °C |
| Specified current @ low voltage operation, 0.5 V | 15 A | at 25 °C ± 5 °C |
| Specified current @ low voltage operation, 0 V | 0 A | at 25 °C ± 5 °C |
| Typical minimum operating voltage @ full scale current | 1.2 V | typically, under normal operating conditions |
Conditions: at 25 °C ± 5 °C; accuracy is ± (% of reading + fixed offset), 1000 samples
| Parameter | Value |
|---|---|
| Constant current, low range/high range | 6 A/60 A |
| Constant current regulation | 10 mA |
| Constant current low range accuracy | 0.1% + 7.5 mA |
| Constant current high range accuracy | 0.1% + 15 mA |
Conditions: at 25 °C ± 5 °C; accuracy is ± (% of reading + fixed offset), 1000 samples
| Parameter | Value |
|---|---|
| Constant voltage, low range/high range | 6 V/60 V |
| Constant voltage regulation | 10 mV |
| Constant voltage low range accuracy | 0.1% + 3 mV |
| Constant voltage high range accuracy | 0.1% + 8 mV |
Conditions: at 25 °C ± 5 °C
| Parameter | Value |
|---|---|
| Constant resistance range 1 (I > 10% of current rating) | 0.033–2 Ω |
| Constant resistance range 2 (I > 1% of current rating) | 1.8–20 Ω |
| Constant resistance range 3 (I > 0.1% of current rating) | 18–200 Ω |
Conditions: at 25 °C ± 5 °C; accuracy is ± (% of reading + fixed offset), 1000 samples; DC current accuracy applies 30 seconds after input current is applied
| Parameter | Value |
|---|---|
| Current measurement, low range/high range | 6 A/60 A |
| Current measurement low range accuracy | 0.05% + 5 mA |
| Current measurement high range accuracy | 0.05% + 10 mA |
| Voltage measurement, low range/high range | 6 V/60 V |
| Voltage measurement low range accuracy | 0.05% + 3 mV |
| Voltage measurement high range accuracy | 0.05% + 8 mV |
| Power measurement accuracy | 0.1% + 0.6 W |
Conditions: typical (supplemental characteristic)
| Parameter | Value |
|---|---|
| Current slew rate, slow band | 1 k–50 kA/s |
| Current slew rate, fast band ≥3 V | 100 k–5 MA/s |
| Current slew rate, fast band <3 V | 100 k–500 kA/s |
| Voltage slew rate, slow band | 1 k–50 kV/s |
| Slew rate bands (Table A-2 note 1) | Slew rate bands are the ranges of programmable slew rates available. |
| Below 3 V (Table A-2 note 1) | Below 3 volts, the maximum bandwidth of the electronic load is reduced by a factor of ten to one. |
Conditions: typical (supplemental characteristic)
| Parameter | Value |
|---|---|
| Voltage ripple and noise (rms), 20 Hz–10 MHz | 6 mV rms |
Conditions: typical (supplemental characteristic)
| Parameter | Value |
|---|---|
| Programmable short, maximum | 33 mΩ |
| Programmable short, typical | 20 mΩ |
| Programmable open | ≥ 20 kΩ |
| Parameter | Value |
|---|---|
| Control | GPIB, RS232, and manual use of the front panel |
| Command set | SCPI (Standard Commands for Programmable Instruments) |
| Analog programming and monitoring ports | for those applications that utilize nonstandard interfaces, require custom waveforms, or utilize process control signals |
| Lists | Custom waveforms can also be created by downloading a “List” of load parameters. |
| Command processing time (average time for the output voltage to change after a GPIB command) | 3 ms for discrete commands; 1 ms for list commands |
| List dwell range | 0 - 10 s |
| List dwell resolution | 1 ms |
| List dwell accuracy | 5 ms |
| Software driver | VXIplug&play |
| Parameter | Value |
|---|---|
| Transient generator | built-in transient generator, which operates in all modes |
| Transient generator frequency range | 0.25 Hz - 10 kHz |
| Transient generator frequency accuracy | 0.5% |
| Transient generator duty cycle range | 3 to 97% (0.25 Hz - 1 kHz); 6 to 94% (1 kHz - 10 kHz) |
| Transient generator duty cycle accuracy | 1% |
| Transient generator pulse width | 50 μs ± 1% to 4 seconds ± 1% |
| Model | Input voltage rating | Input current rating | Maximum power @ 40 °C | Mainframe slots required | Net weight |
|---|---|---|---|---|---|
| N3302A | 0–60 V | 0–30 A | 150 W | one slot | 2.7 kg (6 lb) |
| N3303A | 0–240 V | 0–10 A | 250 W | one slot | 2.7 kg (6 lb) |
| N3304A (this page) | 0–60 V | 0–60 A | 300 W | one slot | 2.7 kg (6 lb) |
| N3305A | 0–150 V | 0–60 A | 500 W | 2 slots | 4.6 kg (10 lb) |
| N3306A | 0–60 V | 0–120 A | 600 W | 2 slots | 4.6 kg (10 lb) |
| N3307A | 0–150 V | 0–30 A | 250 W | one slot | 2.7 kg (6 lb) |
The Keysight N3304A accepts 0–60 V and 0–60 A, with 300 W maximum power at 40 °C; maximum continuous power is derated linearly from 100% of maximum at 40 °C, to 75% of maximum at 55 °C. The source's voltage and current must both sit inside these ratings, and their product inside the power rating.
The Keysight N3304A specified current at low voltage operation is 60 A at 2.0 V, 30 A at 1.0 V, 15 A at 0.5 V and 0 A at 0 V. Typically, under normal operating conditions, the load can sink its maximum current down to 1.2 V, and the N3300 Series loads operate with full stability down to zero volts.
The Keysight N3304A programmable current slew rate is 1 k–50 kA/s in the slow band. In the other band it is 100 k–5 MA/s at 3 V and above and 100 k–500 kA/s below 3 V, because below 3 volts the maximum bandwidth of the electronic load is reduced by a factor of ten to one. Rising and falling slew rates are controlled separately, and the built-in transient generator operates in all modes.
The Keysight N3304A programmable short is 33 mΩ maximum and 20 mΩ typical, and its programmable open is ≥ 20 kΩ.
The Keysight N3304A standard connector accepts an unterminated wire and may be hand-tightened, and the data sheet describes it as designed for bench applications and short-term automated tests. Option UJ1 8 mm screw connectors take wires terminated with insulated ring terminals, up to AWG 4 wire, with multiple wires on each screw terminal for parallel load connections, and are designed for test systems.
Voltage, current and power envelope. DC power = voltage × current. The required voltage and current, including any start-up peak, must both lie inside the stated ranges, and their product inside the stated power — on autoranging outputs the power limit is lower than maximum voltage × maximum current. Rule of thumb, not a specification.
Example with the Keysight N3304A: The Keysight N3304A accepts 0–60 V and 0–60 A with 300 W maximum power; 60 V × 60 A = 3,600 W, more than its 300 W rating, so at 60 V the current is limited to about 5 A (300 W ÷ 60 V).
The Keysight N3304A options and accessories, and the other mainframes and modules of the series, are listed on the Options tab.
| Parameter | Value |
|---|---|
| Mainframe slots required | one slot |
| N3300A mainframe | full rack width; 6 slots |
| N3301A mainframe | half rack width; 2 slots |
| Module mix | Any assortment of the 6 different modules can be configured into these mainframes, up to the slot capacity. |
| Parallel operation | Electronic load modules can be operated in parallel to provide additional current sinking capability. |
| Parameter | Value |
|---|---|
| DC floating voltage (output terminals from chassis ground) | up to ± 300 VDC |
| Remote sensing | 5 V DC between sense and load input |
Conditions: typical (supplemental characteristic)
| Parameter | Value |
|---|---|
| Programming resolution, constant current mode (low/high range) | 0.1/1 mA |
| Programming resolution, constant voltage mode (low/high range) | 0.1/1 mV |
| Readback resolution, current (low/high range) | 0.1/1 mA |
| Readback resolution, voltage (low/high range) | 0.1/1 mV |
Conditions: typical (supplemental characteristic)
| Parameter | Value |
|---|---|
| Resistance range 1 slew rate, slow band | 22–560 Ω/s |
| Resistance range 1 slew rate, fast band ≥3 V | 1120–17 kΩ/s |
| Resistance range 1 slew rate, fast band <3 V | 1120–1.7 kΩ/s |
| Resistance range 2 slew rate, slow band | 220–5600 Ω/s |
Conditions: typical (supplemental characteristic)
| Parameter | Value |
|---|---|
| External analog programming, voltage programming accuracy (all ranges) | 0.5% + 12 mV |
| External analog programming, current programming accuracy (all ranges) | 0.25% + 9 mA |
| External monitor port, voltage monitor accuracy | 0.25% + 12 mV |
| External monitor port, current monitor accuracy | 0.1% + 9 mA |
| Parameter | Value |
|---|---|
| Analog programming bandwidth | 10 kHz rms (-3 db frequency) |
| Analog programming voltage | Voltage: 0 - 10 V; Current: 0 - 10 V |
| Analog monitor ports | Voltage: 0 - 10 V; Current: 0 - 10 V |
| Number of samples or points | Measurement time | Additional fixed offset (in addition to measurement accuracy) |
|---|---|---|
| 1000 samples (default) | 20 ms | none |
| 200 samples | 10 ms | < 6% |
| 100 samples | 9 ms | < 10% |
| 20 points | 7 ms | < 30% |
| < 20 points | 7 ms | > 30% |
| Parameter | Value |
|---|---|
| Digital/Trigger inputs | Vil = 0.9 V max at Iil = -1 mA; Vih - 3.15 V min (pull-up resistor on input) |
| Digital/Trigger outputs | Vol = 0.72 V max at Iol = 1 mA; Voh = 4.4 V min at Ioh = -20 μA |
Conditions: at 25 °C ± 5 °C; accuracy is ± (% of reading + fixed offset), 1000 samples
| Parameter | Value |
|---|---|
| N3304A-J01 input voltage rating | 0–80 V |
| N3304A-J01 input current rating | 0–60 A |
| N3304A-J01 maximum power @ 40 °C | 300 W |
| N3304A-J01 constant voltage, low range/high range | 8 V/80 V |
| N3304A-J01 constant voltage low range accuracy | 0.1% + 5 mV |
| N3304A-J01 constant voltage high range accuracy | 0.1% + 12 mV |
| Specification | Where in the data sheet |
|---|---|
| Constant resistance range 4 (I > 0.01% of current rating) | Data sheet p.5, Table A-1 |
| Programming resolution, constant resistance mode | Data sheet p.6, Table A-2 |
| Voltage slew rate, fast band ≥3 V | Data sheet p.6, Table A-2 |
| Resistance range 2 slew rate, fast bands | Data sheet p.6, Table A-2 |
| Resistance range 3 and range 4 slew rates | Data sheet p.6, Table A-2 |
| Current ripple and noise (rms/peak to peak), 20 Hz–10 MHz | Data sheet p.7, Table A-2 |
| Parameter | Value |
|---|---|
| Net weight | 2.7 kg (6 lb) |
| Shipping weight | 4.1 kg (9 lb) |
The Keysight N3304A input voltage rating is 0–60 V, and its constant voltage low range/high range is 6 V/60 V.
The Keysight N3304A input current rating is 0–60 A, and its constant current low range/high range is 6 A/60 A.
The Keysight N3304A maximum power is 300 W at 40 °C. Maximum continuous power is derated linearly from 100% of maximum at 40 °C, to 75% of maximum at 55 °C.
The Keysight N3304A operates in constant current, constant voltage and constant resistance modes and has a programmable short and a programmable open; its constant resistance range 1 (I > 10% of current rating) is 0.033–2 Ω.
Yes, the Keysight N3304A has a built-in transient generator, which operates in all modes, with a frequency range of 0.25 Hz - 10 kHz and a pulse width of 50 μs ± 1% to 4 seconds ± 1%; custom waveforms can also be created by downloading a “List” of load parameters.
The Keysight N3304A measures current with an accuracy of 0.05% + 5 mA on the low range and 0.05% + 10 mA on the high range, and voltage with 0.05% + 3 mV and 0.05% + 8 mV; its power measurement accuracy is 0.1% + 0.6 W. Each figure is ± (% of reading + fixed offset) with 1000 samples, at 25 °C ± 5 °C.
The Keysight N3304A loads power supplies, battery chargers, multiple output DC power supplies and DC to DC converters under test, in manufacturing test systems or on the bench, and measures their voltage, current and power.
The Keysight N3304A needs an N3300 Series mainframe, the N3300A (full rack width, 6 slots) or the N3301A (half rack width, 2 slots), in which it requires one slot. The mainframes and the other modules are listed on the Options tab.
The data sheet lists N3304A-J01 as a special order option of the Keysight N3304A, with input ratings of 0–80 V, 0–60 A and 300 W at 40 °C, constant voltage ranges of 8 V/80 V and voltage ripple and noise of 8 mV rms. Its other figures are in the special order option table on this page.
The Keysight N3304A is rated 0–60 V, 0–60 A and 300 W and requires one slot; the other N3300 Series modules differ in these ratings and slot needs, as the compare table on this page shows.
The Keysight N3304A absorbs power from a source under test; a DC power supply does the opposite and delivers power to a device.
Specifications are from the manufacturer's data sheet (Keysight N3300 Series DC Electronic Loads Data Sheet, 5980-0232E, August 2, 2014). Questions? Contact your Test Architect.
Updated 2026-11-25
Model No
Keysight
Condition
Pre-Owned
Dc_load_volts
0–60 V
Load_amps
0–60 A
Load_watts
300 W
Manufacturer
Keysight Technologies
The N3300A mainframe is full rack width. It has 6 slots.
The N3301A mainframe is half rack width. It has 2 slots.
UJ1
8 mm screw terminal connector.
N3304A-J01
Special order option.
The N3302A (150 watts), N3303A (250 watts), N3307A (250 Watts) and N3304A (300 watts) each require one slot.
The N3302A (150 watts), N3303A (250 watts), N3307A (250 Watts) and N3304A (300 watts) each require one slot.
The N3305A (500 watts) and the N3306A (600 watts) each require 2 slots.
The N3305A (500 watts) and the N3306A (600 watts) each require 2 slots.
The N3302A (150 watts), N3303A (250 watts), N3307A (250 Watts) and N3304A (300 watts) each require one slot.