Key Features At A Glance

  • Two-quadrant linear design operates as both source and sink to simulate batteries and electronic loads
  • Up to 60 W output power per channel, 0 V to 20 V, with 6 A at ≤ 6 V and 3 A above 6 V
  • Fast load recovery time of < 30 µs with minimum overshoot during demanding load changes
  • Voltage, current and power measured with up to 6 ½ digit resolution across the full range
  • USB and LAN standard; optional WLAN (NGL-K102) and GPIB/IEEE-488 (NGL-B105) for ATE integration
  • Galvanically isolated, floating output channels with relay-isolated output stages
  • Protection functions: OVP, OCP electronic fuse, OPP, overtemperature, plus settable DUT safety limits
  • Adjustable internal impedance (–50 mΩ to 100 Ω) with QuickArb and EasyRamp sequencing
View full details

The R&S®NGL200 Power Supply Series from Rohde & Schwarz is a performance-class bench and system power supply line built around a linear two-quadrant output stage. It is offered in two base configurations: the single-channel R&S®NGL201 and the two-channel R&S®NGL202, with each channel delivering up to 60 W of output power. The output channels are floating, galvanically isolated, and protected against overload and short circuits. Because the two-quadrant architecture lets each channel act as both a source and a sink, the series can supply power to a device and also absorb power from it, automatically switching to sink mode and indicating the transition with a negative current reading.

Rohde & Schwarz designed the R&S®NGL200 series for powering IoT devices and other battery-operated devices, where the load current can rise abruptly from a few microamps in sleep mode to the ampere range when the device switches to transmit mode. The two-quadrant design lets the supplies simulate batteries and loads, while the high measurement resolution makes them well suited to characterizing devices that draw very low power in standby and high current at full load — in many cases removing the need for a separate digital multimeter. The minimum residual ripple and low noise of the linear regulation support the development of sensitive circuitry such as power amplifiers and MMICs. The supplies are used in R&D labs and integrated into production test systems and ATE applications.

ValueTronics keeps its inventory on hand in a 20,000 square-foot secure warehouse at 1675 Cambridge Drive in Elgin, Illinois, where we stock and hold the instruments we sell. That combination lets us match the right instrument and condition to your application and stand behind every order.

Brand Heritage

Rohde & Schwarz is an independent, privately held company headquartered in Munich, Germany, and was founded more than 80 years ago. It has operated under the Rohde & Schwarz name throughout its history, with no major corporate renaming, acquisition, or spin-off affecting the brand.

Compare Other Models in This Series

The R&S®NGL200 series sits in Rohde & Schwarz's performance class of power supplies — the tier intended for applications where speed, accuracy, and advanced programming features matter, including DUT protection, fast programming times, and downloadable voltage and current sequences for lab and ATE use. The series comprises two base models that share the same per-channel performance but differ in channel count and total output power.

Both the single-channel R&S®NGL201 and the two-channel R&S®NGL202 use the same linear two-quadrant output stage, the same 0 V to 20 V output range, and the same protection, measurement, and programming feature set. The choice between them is primarily a question of how many isolated channels an application needs and how much total output power it requires.

Each model below has its own dedicated product page with new-matched pricing. Select the R&S NGL201 or R&S NGL202 page that matches your channel and power requirement to see the configuration and pricing for that specific new unit.

The clearest difference is channel count: the R&S®NGL201 is a single-channel supply rated at 60 W total, while the R&S®NGL202 carries two channels for up to 120 W total (60 W per channel). Because the R&S®NGL202's channels are galvanically isolated, they can run independently, be paralleled for up to 12 A (at ≤ 6 V output), or be connected in series for up to 40 V — and a FuseLink function can tie the two channels' electronic fuses together so both switch off as one.

A few connector and function details also differ. The R&S®NGL201 adds sense-line connectors on the front panel in addition to the rear, whereas the R&S®NGL202 provides front-panel safety sockets for its channels with sense connections on the rear panel. Electronic-load operation, the sense function, and the protection settings are available on both models, and on the R&S®NGL202 they apply to both channels. The comparison table below summarizes the per-model figures.

Model Output Voltage Output Current Total Output Power
NGL202 0 V to 20 V ≤ 6 V: 6 A, > 6 V: 3 A 120 W
NGL201 0 V to 20 V ≤ 6 V: 6 A, > 6 V: 3 A 60 W

Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.

Accessories Supplied

  • Set of power cables
  • Quick start guide

Product Core & Specifications

Specification Value
Outputs (channel outputs are galvanically isolated and not connected to ground)
Number of output channels 2
Total output power max. 120 W
Maximum output power per channel 60 W
Output voltage per channel 0 V to 20 V
Maximum output current per channel (≤ 6 V output voltage) 6 A
Maximum output current per channel (> 6 V output voltage) 3 A
Maximum voltage in serial operation 40 V
Maximum current in parallel operation (≤ 6 V output voltage) 12 A
Maximum current in parallel operation (> 6 V output voltage) 6 A
Adjustable output impedance –50 mΩ to 100 Ω
Adjustable output impedance, increments 1 mΩ
Voltage ripple and noise (20 Hz to 20 MHz) < 500 µV (RMS), < 2 mV (peak-to-peak) (meas.)
Current ripple and noise (20 Hz to 20 MHz) < 1 mA (RMS) (meas.)
Electronic load
Electronic load yes, both channels
Maximum sink power 120 W (60 W per channel)
Maximum sink current per channel 3 A
Sink modes constant voltage, constant current, constant resistance
Load regulation, voltage (load change: 10 % to 90 %) < 0.01 % + 0.5 mV
Load regulation, current (load change: 10 % to 90 %) < 0.01 % + 0.1 mA
Load recovery time (regulation to within ±20 mV of the set voltage) < 30 µs (meas.)
Programming resolution, voltage 1 mV
Programming resolution, current 0.1 mA
Programming accuracy, voltage < 0.02 % + 3 mV
Programming accuracy, current < 0.05 % + 2 mA
Output measurements
Measurement functions voltage, current, power, energy
Readback resolution, voltage 10 μV
Readback resolution, current 10 μA
Readback accuracy, voltage < 0.02 % + 2 mV
Readback accuracy, current < 0.05 % + 250 μA
Temperature coefficient (per °C), voltage 0.15 × specification/°C
Temperature coefficient (per °C), current 0.15 × specification/°C
Sense function yes, both channels
Maximum sense compensation 2 V
Ratings
Maximum voltage to ground 250 V DC
Maximum counter voltage (voltage with the same polarity connected to the outputs) 22 V
Maximum reverse voltage (voltage with opposite polarity connected to the outputs) 0.5 V
Maximum reverse current (for 5 minutes max.) 1 A
Remote control
Command processing time typ. < 6 ms
Protection functions
Overvoltage protection adjustable, both channels
Overvoltage protection, programming resolution 1 mV
Overpower protection adjustable, both channels
Overcurrent protection (electronic fuse) adjustable, both channels
Overcurrent protection, programming resolution 0.1 mA
Overcurrent protection, response time ((Iload > Iresp × 2) at Iload ≥ 2 A) < 1.5 ms
Fuse linking (FuseLink function) yes
Fuse delay at output-on (both channels) 0 ms to 10 s (1 ms increments)
Fuse delay time (both channels) 0 ms to 10 s (1 ms increments)
Overtemperature protection yes
Special functions
Output ramp function EasyRamp
EasyRamp time 10 ms to 10 s (10 ms increments)
Output sequencing, synchronicity typ. < 25 μs
Output sequencing, delay per channel 1 ms to 10 s (1 ms increments)
Arbitrary function QuickArb
Arbitrary function parameters voltage, current, time
Maximum number of points 4096
Dwell time 1 ms to 20 h (1 ms increments)
Repetition continuous or burst mode with 1 to 65535 repetitions
Trigger manually via the keyboard, via remote control or via optional interface
Trigger and control interfaces digital I/O, R&S®NGL-K103
Display and interfaces
Display TFT 5" 800 × 480 pixels WVGA Touch
Front panel connections 4 mm safety sockets (channels)
Rear panel connections 8-pin connector block per channel
Remote control interfaces, standard USB-TMC, USB-CDC (virtual COM port), LAN
Remote control interfaces, R&S®NGL-K102 WLAN
Remote control interfaces, R&S®NGL-B105 IEEE-488 (GPIB)
General data
Operating temperature range +5 °C to +40 °C
Storage temperature range –20 °C to +70 °C
Humidity (noncondensing) 5 % to 95 %
Mains nominal voltage 100 V/115 V/230 V (±10 %)
Mains frequency 50 Hz to 60 Hz
Maximum power consumption 400 W
Important: Maximum voltage to ground: 250 V DC.
Important: Maximum sink power for the R&S®NGL202 (120 W, 60 W per channel) is time limited at an operating temperature of > 30 °C and total power > 90 W.
Important: The WLAN function (R&S®NGL-K102 option) is not available in all regions due to country-specific regulations.
Important: The R&S®NGL-K102 (WLAN) and R&S®NGL-K103 (digital trigger I/O) options are activated using a keycode that must be ordered separately.
Recommended pairing: install the R&S NGL200 supplies in 19" racks using the R&S HZN96 rack adapter (2 HU) listed in the ordering information.

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.

Model No

Rohde

Condition

Factory New

Manufacturer

Rohde & Schwarz

Dcsup_amps

≤ 6 V: 6 A, > 6 V: 3 A

Dcsup_volts

0 V to 20 V

Dcsup_watts

120 W