Skip to product information
1 of 2

Keithley DMM7512 Dual Channel 7½-Digit Sampling Multimeter

Keithley DMM7512 Dual Channel 7½-Digit Sampling Multimeter

Regular price $11,000
Sale price $11,000 Regular price $0
Shipping calculated at checkout.
Request a Quote
Estimated Delivery in
5 - 7
business days (based on availability)

Request a Quote To Get An EDU Discount

Login & Get 2% Back on Purchases
Login or Sign Up for Loyalty

Keithley DMM7512 Dual Channel 7½-Digit Sampling Multimeter

Key Features At A Glance

  • Two independent 7½-digit DMMs in a 1U-high, full-rack-width chassis that minimizes rack space in high-density systems
  • Dual 1 Msample/s 18-bit digitizers capture waveforms or identify transients on voltage or current
  • Trigger a waveform measurement on a signal level, a rising or falling edge, a window, or a digital trigger
  • Store up to 27.5 million timestamped readings on a waveform
  • Basic one-year DC voltage accuracy as low as 14 ppm (24 ppm resistance, 60 ppm DC current)
  • 10 nV, 0.1 µΩ, and 1 pA sensitivities for high-resolution low-level measurements
  • USBTMC-compliant USB, LAN LXI-compliant, and TSP-Link interfaces for automated test systems
  • Auto-calibration plus a two-year calibration interval reduce drift and life-cycle costs
View full details

The Keithley DMM7512 Dual Channel 7½-Digit Sampling Multimeter combines two full-function, high-accuracy, high-sampling-speed DMMs into a compact 1U-high, full-rack-wide chassis. Each DMM has identical functionality and operates entirely independently of the other, so the small form factor saves rack space in high-instrument-density test systems without compromising measurement performance. Each channel reaches a basic one-year accuracy as low as 14 ppm for DC volts, 24 ppm for resistance, and 60 ppm for DC current.

With dual-channel capability, fast signal sampling, high-speed data transfer rates reaching up to 100,000 measurements/s, and internal execution of program test scripts, the DMM7512 reduces space and test time in automated test systems. Its 10 nV, 0.1 µΩ, and 1 pA sensitivities support high-resolution low-level measurements such as portable device sleep currents. Using the two DMMs as synchronized digitizers, one for voltage and one for current, power can be accurately computed on time-varying waveforms, and the instrument can measure current in all operating states of a device including the sleep state and the current burst state when the device is transmitting.

Brand Heritage

Keithley Instruments operates as a Tektronix company, and its instruments are issued under that combined identity.

Accessories Supplied

  • 012178100 Test Lead Kit, Quantity 4
  • CA-18-16A LAN Crossover Cable, 40 cm (16 in), quantity 2
  • 174710700 Shielded Crossover Cable with RJ-45 Connector, 1.5 m (5 ft), quantity 1
  • CA-568-120A Green and Yellow Ground Cable, 3.0 m (120 in), quantity 1
  • 4299-13 1U High Fixed Rack Mount Kit, quantity 1
  • 0713411 Safety Precautions Document
  • 071357600 DMM7512 7½-Digit Multimeter Instrument Information

Product Core & Specifications

Specification Value
DC Voltage Accuracy (1 Year, TCAL ±5 °C) ±(ppm of reading + ppm of range)
100 mV range (10 nV res, >10 GΩ or 10 MΩ ±1%) 18 + 30
1 V range (100 nV res, >10 GΩ or 10 MΩ ±1%) 15 + 5
10 V range (1 µV res, >10 GΩ or 10 MΩ ±1%) 14 + 1.2
100 V range (10 µV res, 10 MΩ ±1%) 40 + 15 (22 + 15 within 30 days of autocalibration)
1000 V range (100 µV res, 10 MΩ ±1%) 40 + 10 (23 + 10 within 30 days of autocalibration)
DC Voltage Characteristics
ADC Linearity 1.0 ppm of reading + 1.0 ppm of range
Input Impedance 100 mV to 10 V Ranges: Selectable >10 GΩ || <400 pF (auto) or 10 MΩ ±1%; 100 V to 1000 V Ranges: 10 MΩ ±1%
Input Bias Current <50 pA at 23 °C (autozero off or input impedance 10 MΩ)
Common Mode Current <2.1 µA peak-peak in 1 MHz bandwidth; <100 nA peak-peak in 1 kHz bandwidth
Common Mode Voltage 500 VPEAK LO terminal to chassis maximum
DC Voltage Autozero Off Error For ±1 °C and ≤10 minutes, add ±(8 ppm of reading + 15 µV)
Normal Mode Rejection (line frequency ±0.1%) Line Sync On: 110 dB at 5 PLC, 90 dB at 1 PLC, 45 dB at ≤0.2 PLC; Line Sync Off: 60 dB at 5 PLC, 60 dB at 1 PLC
Common Mode Rejection (100 Ω unbalanced in LO) 140 dB at 5 PLC (Line Sync On), 140 dB at 1 PLC (On), 120 dB at 0.2 PLC (On), 80 dB at ≤0.2 PLC (Off)
Resistance Accuracy (1 Year, TCAL ±5 °C) ±(ppm of reading + ppm of range)
1 Ω (0.1 µΩ res, 10 mA test current) 50 + 60
10 Ω (1 µΩ res, 10 mA) 50 + 6
100 Ω (10 µΩ res, 1 mA) 47 + 4
1 kΩ (100 µΩ res, 1 mA) 41 + 3
10 kΩ (1 mΩ res, 100 µA) 42 + 3
100 kΩ (10 mΩ res, 10 µA) 50 + 3
1 MΩ (100 mΩ res, 10 µA) 50 + 5
10 MΩ (1 Ω res, 0.69 µA || 10 MΩ) 400 + 10
100 MΩ (10 Ω res, 0.69 µA || 10 MΩ) 2000 + 30
1 GΩ (100 Ω res, 0.69 µA || 10 MΩ) 13000 + 200
Resistance Characteristics
Maximum 4-Wire Ohms Lead Resistance 5 Ω per lead for 1 Ω range, 10% of range per lead for 10 Ω to 1 kΩ ranges; 1 kΩ per lead for all other ranges
Offset Compensation Selectable on 4-wire, 1 Ω to 100 kΩ ranges
Open Lead Detector Default is off
Input Current Limit For ±12 V to ±40 V: ±13 mA source or sink, typical; for greater than ±40 V: ±130 µA source or sink, typical
DC Current Accuracy (1 Year, TCAL ±5 °C) ±(ppm of reading + ppm of range)
10 µA (1 pA res, 15 mV burden) 125 + 40
100 µA (10 pA res, 15 mV) 100 + 15
1 mA (100 pA res, 15 mV) 100 + 15
10 mA (1 nA res, 20 mV) 100 + 15
100 mA (10 nA res, 200 mV) 400 + 30
1 A (100 nA res, 400 mV) 450 + 50
3 A (1 µA res, 1300 mV) 450 + 40
DC Current Characteristics
Effective Internal Shunt Value 10 µA: 1 kΩ; 100 µA: 100 Ω; 1 mA: 10 Ω; 10 mA: 1 Ω; 100 mA: 0.1 Ω; 1 A: 0.1 Ω; 3 A: 0.1 Ω
Autozero Off Error (±1 °C, ≤10 min, add ±(8 ppm of reading + range error)) 10 µA: 150 pA; 100 µA: 1.5 nA; 1 mA: 15 nA; 10 mA: 150 nA; 100 mA: 15 µA; 1 A: 150 µA; 3 A: 150 µA
Temperature — 4-Wire / 3-Wire RTD (Accuracy 2 Year, TCAL ±5 °C)
RTD Types 100 Ω platinum PT100, D100, F100, PT385, PT3916; or user-configurable 0 Ω to 10 kΩ
4-Wire RTD (–200° to 400 °C, 0.01 °C res) 0.09 °C
3-Wire RTD (–200° to 400 °C, 0.01 °C res) 0.75 °C
Temperature — Thermistor (Accuracy 2 Year, TCAL ±5 °C)
Thermistor Types 2.252 kΩ, 5 kΩ, and 10 kΩ
Thermistor (–80° to +150 °C, 0.01 °C res) 0.08 °C
Temperature — Thermocouple (Accuracy 2 Year, TCAL ±5 °C, simulated reference junction)
B (350 to +1820 °C, 0.1 °C res) 0.9 °C
E (−200 to +1000 °C, 0.001 °C res) 0.4 °C
J (−200 to +760 °C, 0.001 °C res) 0.4 °C
K (−200 to +1372 °C, 0.001 °C res) 0.4 °C
N (−200 to +1300 °C, 0.001 °C res) 0.4 °C
R (0 to +1768 °C, 0.1 °C res) 0.9 °C
S (0 to +1768 °C, 0.1 °C res) 0.9 °C
T (−100 to +400 °C, 0.001 °C res) 0.4 °C
Continuity (2 Year, TCAL ±5 °C)
1.0000 kΩ range (100 mΩ res, 1 mA test current, 14.0 V open circuit) 100 + 100 ±(ppm of reading + ppm of range)
Continuity High Limit User-selectable; default 10 Ω
Digitize Voltage Accuracy (1 Year, TCAL ±5 °C) ±(ppm of reading + ppm of range)
100 mV (1 µV res, >10 GΩ or 10 MΩ ±1%) 220 + 100
1 V (10 µV res) 120 + 75
10 V (0.1 mV res) 120 + 75
100 V (1 mV res, 10 MΩ ±1%) 120 + 75
1000 V (10 mV res, 10 MΩ ±1%) 120 + 75
Digitize Current DC Accuracy (1 Year, TCAL ±5 °C) ±(ppm of reading + ppm of range)
10 µA (0.1 nA res, 15 mV burden) 160 + 75
100 µA (1 nA res, 15 mV) 160 + 75
1 mA (10 nA res, 15 mV) 160 + 75
10 mA (100 nA res, 20 mV) 160 + 75
100 mA (1 µA res, 200 mV) 450 + 100
1 A (10 µA res, 400 mV) 500 + 110
3 A (100 µA res, 1300 mV) 900 + 150
Digitizer Characteristics
Maximum Resolution 18 bits
Measurement Input Coupling DC or AC (voltage only)
Sampling Rate Programmable 1 k through 1 million
Volatile Sample Memory with Timestamp 27.5 million
Minimum Record Time 1 µs
Timestamp Resolution 1 ns with standard or full buffer style; 1 µs with compact buffer style
Timestamp Accuracy Standard or full buffer style: 20 ns between adjacent readings (total buffer time <2 s); compact buffer style: 2 µs adjacent readings
Maximum Record Length 8 million
Digitize Typical (Sampling Rate / Digits / Resolution / Measurements Into Computer)
10 kS/s 5½ digits, 18-bit resolution, 9700 readings/s
20 kS/s 4½ digits, 16-bit, 19000 readings/s
50 kS/s 4½ digits, 16-bit, 44400 readings/s
100 kS/s 4½ digits, 15-bit, 80000 readings/s
1 MS/s 3½ digits, 12-bit, 108000 readings/s
Buffer Transfer Speed, Binary (Measurements into Computer per second)
USB / LAN, average for 1000 readings 280000 (USB) / 270000 (LAN)
USB / LAN, average for 1000 readings with timestamp 170000 (USB) / 140000 (LAN)
Triggering
Time Base Accuracy 25 ppm
Trigger Source Analog DCV, DCI, or any system trigger
Trigger Coupling DC
Important: Input protection: 1010 V DC on all ranges and functions for the HI and LO terminals; 350 V for the sense HI and sense LO terminals; 250 V rated for the current input terminal.
Important: The DMM7512 does not have a front panel display; the instrument has a built-in web browser with a virtual front panel that is displayed on a PC monitor through a LAN interface connection.
Usage tip: pair the DMM7512 with the Keithley 2606B System SourceMeter SMU Instrument over TSP-Link to stimulate and measure multiple DUTs as a single integrated, script-controlled test system.

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.

Model No

Keithley

Condition

Factory New

Manufacturer

Keithley

Multi_met_dcv

1000 V

Multi_met_dig

Multi_met_type