Measurement Mode: Measure R series or parallel; L and Q series or parallel; C and D series or parallel. All measurement modes are pushbutton selectable
Displays: LED-type numerical display with automatically positioned decimal points and illumination of units. For R LC, five digits (99999) and simultaneously for DQ, four digits (9999)
Measurement Speed: Greater than 3 measurements per second
Test frequencies: Pushbutton selection between 2. Accuracy re panel legends: +2%, -.01%, Actual frequencies: for 1657-9700, 1020.0 Hz +/- .01% (panel legend "1kHz") and 120.0 Hz +/- 0.1%
Applied Voltage: 0.3 V rms maximum
Ranges: Pushbutton selection with automatic front-panel guidance. Three basic ranges (best accuracy, see table) of 2 decades each, for each parameter. Automatic extensions to min and max, as tabulated