Skip to product information
1 of 1

Agilent E8663D PSG RF Analog Signal Generator (Pre-Owned)

Agilent E8663D PSG RF Analog Signal Generator (Pre-Owned)

Calibrations

🎓 Request a Quote To Get An EDU Discount

Login & Get 2% Back on Purchases
Login or Sign Up for Loyalty

Agilent E8663D PSG RF Analog Signal Generator (Pre-Owned)

Key Features At A Glance

  • Frequency coverage from 100 kHz to 3.2 GHz (Option 503) or 9 GHz (Option 509)
  • GPIB (IEEE-488.2), RS-232, and 10BaseT LAN interfaces for ATE integration
  • 0.001 Hz CW frequency resolution
  • Optional ultra-low phase noise configurations (Options UNX and UNY)
  • Optional analog ramp sweep (Option 007), compatible with the Agilent 8757D scalar network analyzer
  • Optional AM, FM, phase, and pulse modulation with LF output (Option UNT)
  • SCPI 1997.0 command set; code compatible with the previous PSG model E8663B
  • Recommended 24-month calibration cycle
View full details

The Agilent E8663D PSG is a fully synthesized RF analog signal generator built around high output power, low phase noise, and optional ramp sweep capability. Frequency coverage is set by the installed frequency option: Option 503 spans 100 kHz to 3.2 GHz, and Option 509 spans 100 kHz to 9 GHz. CW frequency resolution is 0.001 Hz, with 0.01 Hz resolution across the sweep modes, and the internal timebase reference supports an external 10 MHz reference input as well as a 10 MHz reference output.

A signal generator supplies a known, controllable stimulus used to drive a device or system under test so that its response can be measured. The E8663D supports basic swept measurements when used with the Agilent 8757D scalar network analyzer, and it offers two-tone master/slave operation in which two PSGs synchronously track each other with independent control of start and stop frequencies. For automated environments, it emulates the command sets of several earlier signal generators, providing general compatibility with ATE systems and with the E5500 phase noise system.

ValueTronics stocks and holds its own inventory in a 20,000 square foot secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. Every used unit is inspected and functionally verified in-house before it ships, while new units ship factory-sealed exactly as received from the manufacturer. Holding our own stock lets us match the right instrument and configuration to your application and ship it promptly.

Brand Heritage

Hewlett-Packard's test and measurement business was established as Agilent Technologies in 1999. Agilent's electronic measurement division was subsequently spun off as Keysight Technologies in 2014.

Product Core & Specifications

Specification Value
Frequency
Range (Option 503) 100 kHz to 3.2 GHz
Range (Option 509) 100 kHz to 9 GHz
Resolution, CW 0.001 Hz
Resolution, all sweep modes 0.01 Hz
CW switching speed (Standard) < 11 ms (typ); 7 ms (nom)
CW switching speed (Opt UNX) < 11 ms (typ); 7 ms (nom)
CW switching speed (Opt UNY) < 26 ms (typ); < 22 ms (nom)
Phase offset Adjustable in nominal 0.1° increments
Internal Timebase Reference Oscillator
Aging rate < ± 3 x 10–8/year or < ± 2.5 x 10–10/day after 30 days
Initial achievable calibration accuracy < ± 4 x 10–8
Temperature effects (typ) < ± 4.5 x 10–9 from 0 to 55 °C
Line voltage effects (typ) < ± 2 x 10–10 for ± 10% change
External Reference
Frequency 10 MHz only
Lock range ± 1.0 ppm
Reference output frequency 10 MHz
Reference output amplitude > +4 dBm into 50 Ω load (typ)
External reference input amplitude 5 dBm ± 5 dB
External reference input impedance 50 Ω (nom)
Step (Digital) Sweep
Operating modes Step sweep of frequency, amplitude, or both (start to stop); List sweep of frequency, amplitude, or both (arbitrary list)
Dwell time 1 ms to 60 s
Number of points, step sweep 2 to 65535
Number of points, list sweep 2 to 1601 per table
Triggering Auto, external, single, or GPIB
Settling time, frequency (Standard) < 9 ms (typ)
Settling time, frequency (Opt UNY) < 24 ms (typ)
Settling time, amplitude < 5 ms (typ)
Ramp (Analog) Sweep (Option 007)
Sweep span range Settable from minimum 3 to full range
Maximum sweep rate (250 kHz to < 0.5 GHz) 25 MHz/ms
Maximum sweep rate (0.5 to < 1 GHz) 50 MHz/ms
Maximum sweep rate (1 to < 2 GHz) 100 MHz/ms
Maximum sweep rate (2 to < 3.2 GHz) 200 MHz/ms
Maximum sweep rate (≥ 3.2 GHz) 400 MHz/ms
Frequency accuracy ± 0.05% of span ± timebase (at 100 ms sweep time)
Sweep time, manual mode Settable 10 ms to 200 seconds
Markers 10 independent continuously variable frequency markers
Output
Minimum settable output power (Standard) −20 dBm
Minimum settable output power (Option 1E1 step attenuator) −135 dBm
Maximum output power, > 0.4 to 3.2 GHz (Standard) +15 dBm
Maximum output power, > 0.4 to 3.2 GHz (Option 1EU) +21 dBm (+23 typ)
Maximum output power, > 3.2 to 9 GHz (Standard) +15 dBm
Maximum output power, > 3.2 to 9 GHz (Option 1EU) +22 dBm (+23 typ)
Step attenuator range (Option 1E1) 0 dB and 5 dB to 115 dB in 10 dB steps
Level accuracy resolution 0.01 dB
Temperature stability 0.02 dB/°C (typ)
Output impedance 50 Ω (nom)
SWR, 250 kHz to 2 GHz < 1.4:1 (typ)
SWR, > 2 GHz to 9 GHz < 1.6:1 (typ)
Maximum reverse power 1/2 Watt, 0 VDC
Spectral Purity
Harmonics, 1 to < 10 MHz −25 dBc
Harmonics, 10 MHz to < 60 MHz −28 dBc
Harmonics, 0.06 to 2 GHz −30 dBc
Harmonics, 0.06 to 2 GHz with Option 1EH filters on −55 dBc
Harmonics, > 2 to 9 GHz −55 dBc
Sub-harmonics, 100 kHz to 9 GHz None
Residual FM, CW mode (RMS, 50 Hz to 15 kHz BW) < N x 6 Hz (typ)
Residual FM, CW mode with Option UNX or UNY < N x 4 Hz (typ)
Broadband noise, 10 MHz to 9 GHz (offsets > 10 MHz) < −148 dBc/Hz (typ)
SSB Phase Noise (dBc/Hz, CW, 20 kHz offset from carrier)
250 kHz to 250 MHz −130 (typ −134)
> 250 MHz to 500 MHz −134 (typ −138)
> 500 MHz to 1 GHz −130 (typ −134)
> 1 to 2 GHz −124 (typ −128)
> 2 to 3.2 GHz −120 (typ −124)
> 3.2 to 9 GHz −110 (typ −113)
Frequency Modulation (Option UNT)
Maximum deviation, 250 kHz to 250 MHz 2 MHz
Maximum deviation, > 250 to 500 MHz 1 MHz
Maximum deviation, > 500 MHz to 1 GHz 2 MHz
Maximum deviation, > 1 GHz to 2 GHz 4 MHz
Maximum deviation, > 2 GHz to 3.2 GHz 8 MHz
Maximum deviation, > 3.2 GHz to 9 GHz 16 MHz
Resolution 0.1% of deviation or 1 Hz, whichever is greater
Deviation accuracy < ± 3.5% of FM deviation + 20 Hz (1 kHz rate, deviations < N x 800 kHz)
Distortion < 1% (1 kHz rate, deviations < N x 800 kHz)
Sensitivity ± 1 Vpeak for indicated deviation
Phase Modulation (Option UNT)
Maximum deviation, 250 kHz to 250 MHz (100 kHz BW mode) 20 rad
Maximum deviation, > 3.2 GHz to 9 GHz (100 kHz BW mode) 160 rad
Resolution 0.1% of set deviation
Deviation accuracy < ± 5% of deviation + 0.01 radians
Sensitivity ± 1 Vpeak for indicated deviation
Amplitude Modulation (Option UNT) (typical)
Maximum depth, linear mode (ALC on) > 90%
Maximum depth, ALC off with power search or ALC on with deep AM > 95%
Settable depth 0 to 100%
Sensitivity 0 to 100%/V
Resolution 0.1%
Distortion, 30% AM < 1.5%
Distortion, 60% AM < 2%
Important: Maximum reverse power into the RF output: 1/2 Watt, 0 VDC.
Important: During storage below -20 °C, instrument states may be lost.
Important: Phase noise specifications are not warranted in a vibrating environment.
Important: For master/slave (two-tone) operation, use Agilent part number 8120-8806 master/slave interface cable.
Important: When executing power search with ALC off, RF power will be present for typically 10 to 50 ms; the step attenuator (Option 1E1) can be set to automatically switch to maximum attenuation to protect sensitive devices.
Recommended pairing: use the E8663D with the Agilent 8757D scalar network analyzer for swept measurements, per the ramp sweep compatibility listed for Option 007.

About this Pre-Owned unit

This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.

  • Warranty included
  • Functional verification included
  • Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
  • Note that unnecessary accessories may not be included (contact Test Architect to confirm).

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.

Model No

Agilent

Condition

Pre-Owned

Manufacturer

Keysight Technologies

Rf_gen_freq

100 kHz to 9 GHz