Hioki MR8791 Pulse Generator Unit

Hioki MR8791 Pulse Generator Unit

Regular price $1,778
Sale price $1,778 Regular price $0
Shipping calculated at checkout.
Request a Quote
Estimated Delivery in
5 - 7
business days (based on availability)

🎓 Request a Quote To Get An EDU Discount

Login & Get 2% Back on Purchases
Login or Sign Up for Loyalty

Hioki MR8791 Pulse Generator Unit

Key Features At A Glance

  • Plug-in generator units that expand the Memory HiCorder MR8847A with waveform or pulse generation alongside recording
  • Channel configurations from 2 to 8 channels per unit across the series
  • Isolated, floating outputs allow connection to devices at different potentials
  • SMB output terminals with maximum rated voltage to ground of 33 V AC rms or 70 V DC
  • Built-in self-test function monitors output voltage and current
  • Complies with EN61010 safety and EN61326 EMC, with 350 V AC dielectric withstand voltage
View full details

The U8793 Arbitrary Waveform Generator Unit installs in the Memory HiCorder MR8847A to combine generating and recording functions in a single instrument. With the unit installed, the MR8847A provides a function generator mode, an arbitrary waveform generator mode, and a waveform measurement mode together, so an engineer can generate and record at the same time with the same instrument and observe waveforms while varying output conditions such as amplitude and frequency.

The combination is applied to automobile and railroad transport-equipment testing, where actual waveforms are used to test control boards for engine control, airbags, brake systems, power steering, and active suspension. Additional listed applications include vibration reproducibility testing using vibration waveforms recorded in the field, power anomaly simulation that recreates supply dips, instantaneous interruptions, and voltage fluctuations for immunity testing, inverter motor performance testing for VVVF control and PWM waveforms, power steering endurance testing, and vehicle sensor inspection — including 14 V DC pulse generation to simulate a car battery.

ValueTronics stocks and holds its own inventory in a 20,000-square-foot secure warehouse at 1675 Cambridge Drive in Elgin, Illinois.

Compare Other Models in This Series

The U8793 Arbitrary Waveform Generator Unit, the MR8790 Waveform Generator Unit, and the MR8791 Pulse Generator Unit are plug-in generator units that expand the practical applications of the Memory HiCorder. They are used in various combinations with measurement units so that a single Memory HiCorder can both generate stimulus signals and record results.

Each unit shares the same compact form factor of roughly 106 mm wide by 19.8 mm high by 196.5 mm deep, an SMB output terminal, a maximum rated voltage to ground of 33 V AC rms or 70 V DC, and compliance with EN61010 for safety and EN61326 for EMC. They differ in channel count, output type, and the kinds of waveforms each generates.

Each new unit below is its own dedicated product page with condition-matched pricing. Select the model that matches your channel count and output requirements, and review that page for its specific new availability and price.

The U8793 is the arbitrary-waveform member of the group, adding a function generator mode, an arbitrary waveform generator mode, and a waveform measurement mode to the MR8847A, and generating isolated output of up to 15 V without an external amplifier.

The U8793 provides 2 channels per unit with an output frequency range of 0 Hz to 100 kHz, a maximum output voltage of -10 V to 15 V, and full function-generator, arbitrary-waveform, sweep, and program capability. The MR8790 Waveform Generator Unit provides 4 channels per unit producing sine waves up to 20 kHz and DC voltage output with a high-precision DC output accuracy of ±0.6 mV and a maximum output voltage of ±10 V. The MR8791 Pulse Generator Unit provides 8 channels per unit with pulse and pattern output, selectable TTL-level logic output or open-collector output, and a minimum pulse width of 1 µs.

Note that the MR8790 Waveform Generator Unit and the MR8791 Pulse Generator Unit do not have phase control or synchronize control modes. The comparison table below summarizes the channel counts, output types, and key output specifications for each unit.

Model Output Type Output Channels Max Output Voltage
MR8791 Pulse / Pattern (logic / open collector) 8 ch/unit 0 V to 5 V (logic)
U8793 Arbitrary / Function Generator 2 ch/unit -10 V to 15 V
MR8790 Sine wave / DC voltage 4 ch/unit ±10 V

Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.

Product Core & Specifications

Specification Value
Pulse output specifications
Output frequency Setting range: 0 Hz to 20 kHz (Can be set for each of the 8 channels); Setting resolution: 0.1 Hz; Frequency accuracy: Refer to the time axis accuracy of the Memory HiCorder in which the MR8791 is installed.
Duty Setting range: 0.1% to 99.9%, 0, 100% (DC); Setting resolution: 0.1%; Duty accuracy: Refer to the time axis accuracy of the Memory HiCorder in which the MR8791 is installed.
Min. pulse width 1 μs
Number of output channels 8ch/unit (isolated between unit and output) (Not isolated between each channel (common ground)) (Output connector frame not isolated (main unit grounded))
Output mode 1 Pattern output/Pulse output (common 8-channel switching)
Output mode 2 Logic output/Open collector output (Can be set for each of the 8 channels)
Logic output Output voltage level: 0 V - 5 V (H level: 3.8 V or more, L level: 0.8 V or less); Rated current: ±5 mA
Open collector output Absolute maximum rated voltage for collector/emitter: 50 V; Overcurrent protection: 100 mA
Output mode 3 Output/Open (= self diagnostic) (common 8-channel switching)
Open collector output regulation (startup time (10% - 90%)) 5 μs (max.) (Load capacity: 1000 pF, Pull-up resistance: 1 kΩ)
Self-test function Detected voltage: H level: 3.4 V or more, L level: 1.6 V or less
Relay switch time 5 ms or less (Logic/Open collector switch, Output/Open (self-diagnostic) switch)
Pattern output specifications
Clock frequency Range: 0 Hz to 120 kHz (common to 8 channels); Setting resolution: 10 Hz; Frequency accuracy: Refer to the time axis accuracy of the Memory HiCorder in which the MR8791 is installed.
Memory (pattern) 2048 word (16384 bit=2048 word×8 bit/word)
General specifications
Maximum rated voltage to ground 33 V AC rms or 70 V DC (between each output channel and the main unit). Expected transient overvoltage: 330 V
Dielectric withstand voltage 350 V AC (sensed current: 1 mA) (Between each output channel and the main unit, and between the output units)
Standard compliance Safety: EN61010; EMC: EN61326
Operating temperature and humidity range According to the Memory HiCorder installed with the MR8791
Operating environment According to the Memory HiCorder installed with the MR8791
Storage temperature and humidity -20°C to 50°C (-4°F to 122°F), and 90% rh or less (no condensation)
External dimensions Approx. 106 mm (4.17 in) W × 19.8 mm (0.78 in) H × 196.5 mm (7.74 in) D (not including protrusions)
Weight Approx. 230 g (8.1 oz)
Accuracy guarantee Accuracy guaranteed for 1 year; Post-adjustment accuracy guaranteed for 1 year
Important: Caution: MR8847 waveform data can be used for output, but only the MR8847A can be used for waveform generation.
Important: The Memory HiCorder by itself cannot create arbitrary waveforms; use the Waveform Maker SF8000 (bundled on the Application Disk included with the Memory HiCorder).
Important: The MR8790 Waveform Generator Unit and the MR8791 Pulse Generator Unit do not have phase control or synchronize control modes.
Important: These generator units are plug-in units "For the MR8847A or other" Memory HiCorder and require a Hioki Memory HiCorder mainframe (MR8847A) to operate; they cannot be used standalone.
Important: Connection cables are not part of the unit; the appropriate connection cable (L9795-01 SMB-to-alligator clip or L9795-02 SMB-to-BNC, both 1.5 m, max. rated voltage to ground 33 V AC rms or 70 V DC) must be ordered separately as an Option to connect the SMB output terminals to a device under test.
Recommended pairing: install the U8793 in the Memory HiCorder MR8847A and use the bundled Waveform Maker SF8000 software to create, edit, and output arbitrary waveforms.

About this New unit

This is a brand-new, factory-sealed unit, supplied exactly as it arrives from the manufacturer and backed by the full manufacturer warranty. For the exact configuration, available options, included accessories, or current lead time for your application, our Test Architects can confirm the details before you order.

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.

Model No

Hioki

Condition

Factory New

Manufacturer

Hioki

Plugin_type

Pulse Generator Unit