Fluke Calibration 6531-20M E-DWT Electronic Deadweight Tester Kit, 2 to 20 MPa

Fluke Calibration 6531-20M E-DWT Electronic Deadweight Tester Kit, 2 to 20 MPa

Regular price $37,371.37
Sale price $37,371.37 Regular price $0
Shipping calculated at checkout.
Request a Quote
Estimated Delivery in
5 - 7
business days (based on availability)

Request a Quote To Get An EDU Discount

Login & Get 2% Back on Purchases
Login or Sign Up for Loyalty

Fluke Calibration 6531-20M E-DWT Electronic Deadweight Tester Kit, 2 to 20 MPa

Key Features At A Glance

  • Pressure ranges available to 200 MPa (30,000 psi)
  • ±0.02% of reading total one-year measurement uncertainty; ±0.025% for two years
  • Low-torque variable volume enables pressure generation and control up to 200 MPa (30,000 psi) with minimal physical effort
  • Built-in priming system fills the system with test fluid and purges unwanted air for smooth operation
  • Unique test port design eliminates the need for PTFE tape or wrenches; adapters included for 1/8, 1/4, 3/8 and 1/2 in NPT, the same sizes in BSP, plus M20×1.5 and M14×1.5
  • Compatible with a wide range of liquids including water, sebacate, silicone oils, propylene glycol, fluorinated liquids, isopropyl alcohol and mineral oil
  • AutoTest, AutoRange, ready/not-ready indicator, onboard data capture and storage, overpressure alarm, and RS-232 remote interface
  • 6532 kits add a second Q-RPT to extend percent-of-reading performance from 1% to 100% of device full scale
View full details

The Fluke Calibration 6531 and 6532 E-DWT Electronic Deadweight Tester Kits are complete hydraulic pressure calibration systems built around the E-DWT-H electronic deadweight tester. Each kit pairs an E-DWT-H configuration with the accessories required to perform high-level hydraulic pressure calibration directly out of the box, including a vacuum liquid fill kit and a test station quick-mount fitting with adapters for the most common pressure connection sizes. The 6531 kits use a single quartz reference pressure transducer (Q-RPT) to deliver percent-of-reading performance from 10% to 100% of device full scale, while the 6532 kits add a second Q-RPT to extend that performance from 1% to 100% of device full scale for maximum workload coverage.

These kits are designed to replace traditional hydraulic deadweight testers in metrology and calibration labs, on the production floor, and in the field. Because the E-DWT-H is an electronic calibrator, there are no weights to load, unload, transport, or send out for calibration, no piston-cylinder changes required, and no significant sensitivity to level or vibration. The instrument provides real-time pressure indication with no need to correct for local gravity or ambient temperature, and it can set and read any pressure value exactly rather than being limited to the nominal increments of available masses — a particular advantage for applications such as analog gauge calibration where a specific pressure must be set on the device under test and measured directly. Operators can switch between units of measure as needed.

Brand Heritage

Fluke Calibration is the precision measurement division of Fluke Corporation. The Fluke Calibration pressure product line traces its engineering lineage through DH Instruments, which Fluke acquired in 2007 and integrated into the Fluke Calibration brand; the E-DWT-H platform reflects that pressure-calibration heritage. Fluke Corporation itself is part of the Fortive Corporation group following Fortive's 2016 spin-off from Danaher.

Compare Other Models in This Series

The 6531 and 6532 kit family is built around a common E-DWT-H electronic deadweight tester platform and differs primarily in the Q-RPT (quartz reference pressure transducer) configuration and the pressure range that configuration covers. The 6531 kits use a single Q-RPT and are identified by their full-scale pressure: 7M, 14M, 20M, 40M, 70M, 140M, and 200M-B. The 6532 kits add a second, lower-range Q-RPT to extend the percent-of-reading performance band downward.

Across the family, the mechanical hardware — variable volume, priming system, DH500 test port, adapter set, reservoir, and communication ports — is the same. What changes between models is the upper pressure ceiling and, for the 6532 models, the addition of a second Q-RPT that improves accuracy on the lower portion of the workload.

Each model below is its own product page with condition-matched pricing. new listings are quoted on the specific unit's verified configuration, Q-RPT arrangement, and included accessories. Confirm pressure range, included adapters, and test medium history with the Test Architect before ordering.

The 6531 single-Q-RPT kits deliver ±0.02% of reading performance from 10% to 100% of device full scale. Stepping through the range: 6531-7M covers 0.7 to 7 MPa (100 to 1,000 psi), 6531-14M covers 1.4 to 14 MPa (200 to 2,000 psi), 6531-20M covers 2 to 20 MPa (300 to 3,000 psi), 6531-40M covers 4 to 40 MPa (600 to 6,000 psi), 6531-70M covers 7 to 70 MPa (1,000 to 10,000 psi), 6531-140M covers 14 to 140 MPa (2,000 to 20,000 psi), and 6531-200M-B covers 40 to 200 MPa (6,000 to 30,000 psi).

The 6532 dual-Q-RPT kits combine a high-range and a low-range Q-RPT to broaden the percent-of-reading workload. 6532-70M (A70M/A7M) covers 0.7 to 70 MPa (100 to 10,000 psi), 6532-140M (A140M/A14M) covers 1.4 to 140 MPa (200 to 20,000 psi), and 6532-200M-B (A200M/A20M) covers 4 to 200 MPa (600 to 30,000 psi). See the comparison table below for the full mapping of kit model to E-DWT-H configuration and percent-of-reading accuracy band.

Model E-DWT-H Configuration Pressure Range (MPa) Pressure Range (psi)
6531-20M E-DWT-H A20M 2 to 20 MPa 300 to 3,000 psi
6531-7M E-DWT-H A7M 0.7 to 7 MPa 100 to 1,000 psi
6531-14M E-DWT-H A14M 1.4 to 14 MPa 200 to 2,000 psi
6531-40M E-DWT-H A40M 4 to 40 MPa 600 to 6,000 psi

Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.

Accessories Supplied

  • E-DWT-H electronic deadweight tester with power supply and operating manual
  • Vacuum liquid fill kit
  • Test station quick-mount fitting
  • NPT adapters: 1/8 in, 1/4 in, 3/8 in and 1/2 in
  • BSP adapters: 1/8 in, 1/4 in, 3/8 in and 1/2 in
  • Metric adapters: M20 x 1.5 and M14 x 1.5

Product Core & Specifications

General Specifications
Power requirements To RPM4-E-DWT: 12 V dc 1.2 A. To ac to dc power supply: 100 V ac to 240 V ac, 50/60 Hz
Operating temperature range 10 °C to 40 °C (50 °F to 104 °F)
Storage temperature range -20 °C to 70 °C (-4 °F to 158 °F) (must be above freezing point and below boiling point of test medium)
Relative operating humidity 0 % to 70 %
Relative storage humidity 0 % to 100 %
Weight E-DWT with 1 Q-RPT: 12 kg (26 lb) approximate. E-DWT with 2 Q-RPTs: 14 kg (30 lb) approximate
Dimensions E-DWT footprint: 41.4 cm W x 37.1 cm D (16.3 in x 14.6 in). E-DWT height: 26.9 cm (10.6 in), 33.6 cm (13.2 in) to max variable volume handle height
Pressure range Up to 200 MPa (30k psi), depending on configuration
Operating medium All 6531 and 6532 E-DWT Kits are shipped dry, standard preparation. Can be filled with di-ethyl-hexyl sebacate, silicon oils, propylene glycol, fully fluorinated liquids, partially fluorinated liquids, isopropyl alcohol and distilled water or mineral oil.
Reservoir capacity 300 cc (18 in³)
Variable volume displacement 3 cc (0.18 in³), 200 MPa (30 000 psi) maximum
Filling and priming pump displacement 3.7 cc (0.23 in³)
TEST pressure connection DH500 female, 200 MPa (30k psi) maximum working pressure. Adapters included for 1/8 in, 1/4 in, 3/8 in and 1/2 in NPT, 1/8 in, 1/4 in, 3/8 in and 1/2 in BSP, M20 x 1.5 and M14 x 1.5. Adapter maximum working pressure is 140 MPa (20k psi). Note: DH500 is a gland and collar type fitting for 6 mm (1/4 in) coned and left hand threaded tubes equivalent to AE F250C, HIP HF4, 9/16-18 UNF, etc.
Pressure limits Maximum working pressure: Range of RPM4-E-DWT monitor's Hi Q-RPT. Maximum priming pump pressure: 700 kPa (100 psi). Maximum working pressure with Lo Q-RPT selected: Range of RPM4-E-DWT monitor's Lo Q-RPT
Communication ports RS-232 (COM1, COM2)
Pressure Measurement
Warm up time 15 minute temperature stabilization recommended from cold power up
Normal operating temperature range 10 °C to 40 °C (50 °F to 104 °F)
Resolution Default: 0.01 % of active range. User adjustable to 1 ppm of Q-RPT maximum or 10 ppm of active AutoRange, whichever is larger
Precision 200 MPa (30,000 psi) ranges: ± 0.018 % of reading or 0.0036 % of Q-RPT span, whichever is greater. All other ranges: ± 0.018 % of reading or 0.0018 % of Q-RPT span, whichever is greater. (Combined linearity, hysteresis, and repeatability. Precision does not include stability or calibration reference uncertainty.)
Predicted Stability One year: ± 0.0075 % of reading. Two year: ± 0.015 % of reading. (Predicted Q-RPT measurement stability limit (k=2) assuming regular use of AutoZero function and short term stability between rezeroing.)
Measurement uncertainty (One year) 200 MPa (30,000 psi) ranges: ± 0.02 % of reading or 0.004 % of Q-RPT span, whichever is greater. All other ranges: ± 0.02 % of reading or 0.002 % of Q-RPT span, whichever is greater.
Measurement uncertainty (Two year) 200 MPa (30,000 psi) ranges: ± 0.025 % of reading or 0.005 % of Q-RPT span, whichever is greater. All other ranges: ± 0.025 % of reading or 0.0025 % of Q-RPT span, whichever is greater.
Important: Maximum working pressure on the included thread adapters is 140 MPa (20,000 psi); the DH500 TEST pressure connection itself is rated to 200 MPa (30,000 psi). Maximum priming pump pressure is 700 kPa (100 psi).Important: All 6531 and 6532 E-DWT Kits are shipped dry, standard preparation. They must be filled by the user with a compatible test medium (di-ethyl-hexyl sebacate, silicon oils, propylene glycol, fully fluorinated liquids, partially fluorinated liquids, isopropyl alcohol and distilled water or mineral oil) before use. Sebacate oil is available separately as an accessory.Important: Storage temperature range (-20 °C to 70 °C) must be above freezing point and below boiling point of the test medium.Important: Adapters supplied with the kit have a maximum working pressure of 140 MPa (20k psi). For pressures above this limit, use the DH500 female TEST connection directly.Important: 15 minute temperature stabilization is recommended from cold power up before performing measurements.Recommended pairing: order Fluke Calibration Sebacate oil (1 quart) as the test medium with the kit if your calibration practice uses sebacate. The kits ship dry as standard preparation and must be filled before use.

Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.

Model No

Fluke

Condition

New

Manufacturer

Fluke