Skip to product information
1 of 1

RIGOL DM3068 6½ Digit Digital Multimeter

RIGOL DM3068 6½ Digit Digital Multimeter

Regular price $882
Sale price $882 Regular price $0
Shipping calculated at checkout.
Request a Quote
Estimated Delivery in
5 - 7
business days (based on availability)

Request a Quote To Get An EDU Discount

Login & Get 2% Back on Purchases
Login or Sign Up for Loyalty

RIGOL DM3068 6½ Digit Digital Multimeter

Key Features At A Glance

  • Real 6½ digit resolution (2,200,000 count) for high-precision multifunction measurement
  • Measurement speed up to 10k readings/s with 512k-reading volatile memory
  • True-RMS AC voltage to 750 V and AC current to 10 A measurement
  • Functions include DC/AC volts and current, 2/4-wire resistance, capacitance, frequency, diode, continuity, temperature
  • USB Device/Host, LAN (LXI-C), RS-232 and GPIB standard with SCPI control for automated test systems
  • Built-in thermocouple cold-junction compensation; supports TC, RTD and thermistor sensors plus UltraSensor software
  • Real-time Trend and Histogram displays with statistics (max, min, average, standard deviation)
  • Measurement CAT I 1000 V / CAT II 300 V rated; LXI Class C web remote control
View full details

The DM3068 is a 6½ digit digital multimeter built for high-precision, multifunction, and automatic measurement. It delivers a real 6½ digit reading resolution of 2,200,000 counts and combines a broad set of measurement functions with analysis capabilities in a single bench instrument, targeting users who need both accuracy and measurement flexibility from one DMM.

The DM3068 is aimed at the requirements of the research, education, industrial electronics, consumer electronics, and automotive industries. Across these settings it serves as a general-purpose precision multimeter, covering DC and AC voltage and current, resistance, capacitance, frequency and period, diode, continuity, and temperature measurements from a single front panel.

ValueTronics stocks and holds its own inventory in a 20,000 sq ft secure warehouse at 1675 Cambridge Drive, Elgin, Illinois.

Accessories Supplied

  • Power Cord conforming to the standard of the country
  • Two Test Leads (black and red)
  • Two Alligator Clips (black and red)
  • USB Cable (CB-USB-150)
  • Four Spare Fuses (two kinds): 2 AC, 250 V, T250 mA fuses and 2 AC, 250 V, T125 mA fuses
  • Quick Guide
  • Resource CD (User's Guide and Application Software)

Product Core & Specifications

Specification Value
DC Voltage Accuracy (1 Year, TCAL °C ± 5 °C), ±(% of reading + % of range)
200 mV 0.0040 + 0.0025
2 V 0.0035 + 0.0006
20 V 0.0040 + 0.0005
200 V 0.0050 + 0.0006
1000 V 0.0055 + 0.0010
DC Current Accuracy (1 Year, TCAL °C ± 5 °C), ±(% of reading + % of range)
200 µA 0.050 + 0.015
2 mA 0.050 + 0.003
20 mA 0.050 + 0.015
200 mA 0.050 + 0.003
2 A 0.100 + 0.020
10 A 0.150 + 0.010
Resistance Accuracy (1 Year, TCAL °C ± 5 °C), ±(% of reading + % of range)
200 Ω 0.010 + 0.004
2 kΩ 0.010 + 0.001
20 kΩ 0.010 + 0.001
200 kΩ 0.010 + 0.001
1 MΩ 0.012 + 0.001
10 MΩ 0.040 + 0.001
100 MΩ 0.800 + 0.010
Diode Test (1 Year, TCAL °C ± 5 °C), ±(% of reading + % of range)
2.0000 V range (1 mA test current) 0.010 + 0.020
Continuity Test (1 Year, TCAL °C ± 5 °C), ±(% of reading + % of range)
2000.0 Ω range (1 mA test current) 0.010 + 0.020
AC (True-RMS) Voltage Accuracy (1 Year, TCAL °C ± 5 °C), ±(% of reading + % of range)
200 mV, 3 Hz - 5 Hz 1.00 + 0.04
200 mV, 5 Hz - 10 Hz 0.35 + 0.04
200 mV, 10 Hz - 20 kHz 0.06 + 0.04
200 mV, 20 kHz - 50 kHz 0.12 + 0.05
200 mV, 50 kHz - 100 kHz 0.60 + 0.08
200 mV, 100 kHz - 300 kHz 4.00 + 0.50
2 V, 3 Hz - 5 Hz 1.00 + 0.03
2 V, 5 Hz - 10 Hz 0.35 + 0.03
2 V, 10 Hz - 20 kHz 0.06 + 0.03
2 V, 20 kHz - 50 kHz 0.12 + 0.05
2 V, 50 kHz - 100 kHz 0.60 + 0.08
2 V, 100 kHz - 300 kHz 4.00 + 0.50
20 V, 3 Hz - 5 Hz 1.00 + 0.04
20 V, 5 Hz - 10 Hz 0.35 + 0.04
20 V, 10 Hz - 20 kHz 0.08 + 0.04
20 V, 20 kHz - 50 kHz 0.15 + 0.05
20 V, 50 kHz - 100 kHz 0.60 + 0.08
20 V, 100 kHz - 300 kHz 4.00 + 0.50
200 V, 3 Hz - 5 Hz 1.00 + 0.03
200 V, 5 Hz - 10 Hz 0.35 + 0.03
200 V, 10 Hz - 20 kHz 0.08 + 0.03
200 V, 20 kHz - 50 kHz 0.15 + 0.05
200 V, 50 kHz - 100 kHz 0.60 + 0.08
200 V, 100 kHz - 300 kHz 4.0 + 0.50
750 V, 3 Hz - 5 Hz 1.00 + 0.03
750 V, 5 Hz - 10 Hz 0.35 + 0.03
750 V, 10 Hz - 20 kHz 0.08 + 0.03
750 V, 20 kHz - 50 kHz 0.15 + 0.05
750 V, 50 kHz - 100 kHz 0.60 + 0.08
750 V, 100 kHz - 300 kHz 4.0 + 0.50
AC (True-RMS) Current Accuracy (1 Year, TCAL °C ± 5 °C), ±(% of reading + % of range)
200 µA, 3 Hz - 5 Hz 1.10 + 0.06
200 µA, 5 Hz - 10 Hz 0.35 + 0.06
200 µA, 10 Hz - 5 kHz 0.15 + 0.06
200 µA, 5 kHz - 10 kHz 0.35 + 0.70
2 mA, 3 Hz - 5 Hz 1.00 + 0.04
2 mA, 5 Hz - 10 Hz 0.30 + 0.04
2 mA, 10 Hz - 5 kHz 0.12 + 0.04
2 mA, 5 kHz - 10 kHz 0.20 + 0.25
20 mA, 3 Hz - 5 Hz 1.10 + 0.06
20 mA, 5 Hz - 10 Hz 0.35 + 0.06
20 mA, 10 Hz - 5 kHz 0.15 + 0.06
20 mA, 5 kHz - 10 kHz 0.35 + 0.70
200 mA, 3 Hz - 5 Hz 1.00 + 0.04
200 mA, 5 Hz - 10 Hz 0.30 + 0.04
200 mA, 10 Hz - 5 kHz 0.10 + 0.04
200 mA, 5 kHz - 10 kHz 0.20 + 0.25
2 A, 3 Hz - 5 Hz 1.10 + 0.06
2 A, 5 Hz - 10 Hz 0.35 + 0.06
2 A, 10 Hz - 5 kHz 0.15 + 0.06
2 A, 5 kHz - 10 kHz 0.35 + 0.70
10 A, 3 Hz - 5 Hz 1.10 + 0.10
10 A, 5 Hz - 10 Hz 0.35 + 0.10
10 A, 10 Hz - 5 kHz 0.15 + 0.10
Frequency and Period Accuracy (1 Year, TCAL °C ± 5 °C), ±(% of reading); Range 200 mV to 750 V
3 Hz - 5 Hz 0.07
5 Hz - 10 Hz 0.04
10 Hz - 40 Hz 0.02
40 Hz - 300 kHz 0.007
300 kHz - 1 MHz 0.007
Capacitance Accuracy (1 Year, TCAL °C ± 5 °C), ±(% of reading + % of range)
2 nF (200 nA test current) 2 + 2.5
20 nF (2 µA) 1 + 0.3
200 nF (10 µA) 1 + 0.3
2 µF (100 µA) 1 + 0.3
20 µF (1 mA) 1 + 0.3
200 µF (1 mA) 1 + 0.3
2 mF (1 mA) 1 + 0.3
20 mF (1 mA) 1 + 0.3
100 mF (1 mA) 3 + 0.2
Temperature Accuracy (1 Year, TCAL °C ± 5 °C)
RTD, α = 0.00385 (-200 °C to 660 °C) 0.16 °C
RTD, α = 0.00389 (-200 °C to 660 °C) 0.17 °C
RTD, α = 0.00391 (-200 °C to 660 °C) 0.14 °C
RTD, α = 0.00392 (-200 °C to 660 °C) 0.15 °C
Thermal Resistance 2.2 kΩ (-40 °C to 150 °C) 0.08 °C
Thermal Resistance 3 kΩ (-40 °C to 150 °C) 0.08 °C
Thermal Resistance 5 kΩ (-40 °C to 150 °C) 0.08 °C
Thermal Resistance 10 kΩ (-40 °C to 150 °C) 0.08 °C
Thermal Resistance 30 kΩ (-40 °C to 150 °C) 0.08 °C
Thermocouple B (0 °C to 1820 °C) 0.76 °C
Important: Measurement category rating: CAT I 1000 V / CAT II 300 V.
Recommended pairing: add the optional RM-DM-3 Rack Mount Kit to install the DM3068 in a standard equipment rack.

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.

Model No

RIGOL

Condition

Factory New

Manufacturer

RIGOL

Multi_met_acv

750 V

Multi_met_dcv

1000 V

Multi_met_dig

Multi_met_type

True-RMS