Skip to content

The Keysight DAQ970A/DAQ973A is a data acquisition system built around a compact 3-slot mainframe with a built-in 6½-digit (22-bit) internal DMM. The measurement engine from a high-performance benchtop DMM is embedded inside the mainframe, combining proven measurement performance, universal inputs with built-in signal conditioning, and modular flexibility in a single low-cost package. It delivers 0.003% basic DC voltage accuracy, ultra-low reading noise, and scan rates of up to 450 channels per second, giving an engineer both the speed and the accuracy needed for demanding acquisition work.
Configured with a 20-channel relay multiplexer, the system becomes a powerful, low-cost data logger for characterization applications, and it also serves as a data acquisition front end for automated test. Typical data-logging and monitoring uses include environmental chamber monitoring, component inspection, benchtop testing, process troubleshooting, and temperature profiling. It supports both the R&D engineer characterizing a new design and the manufacturing engineer building a test system or troubleshooting a process. The 40-channel single-ended multiplexer adds density for applications such as battery test and component characterization.
That combination of real inventory on the shelf and hands-on verification of used equipment is the difference between buying from a stocking specialist and gambling on an unaccountable listing.
Keysight Technologies traces its roots to the test and measurement business of Hewlett-Packard. That business was spun out as Agilent Technologies in 1999, and in 2014 Agilent's electronic measurement division became the independent company Keysight Technologies. Equipment from across that history is frequently encountered under all three names.
The DAQ970A/DAQ973A family pairs a common 6½-digit (22-bit) measurement engine with a 3-slot modular mainframe, giving the simple operation and low per-channel cost of a data logger alongside the flexibility and performance of a modular data acquisition system. Both mainframes share the same internal DMM, the same 4.3-inch color display, and the same compatibility with all nine plug-in switch, RF, and control modules.
The two models differ in their standard remote-control interfaces. The DAQ970A provides built-in Gigabit LAN and USB 2.0, while the DAQ973A carries the same LAN and USB connectivity plus a built-in GPIB interface for ATE environments that depend on IEEE-488 control.
Each model in the DAQ970A/DAQ973A family below links to its own dedicated product page, where new pricing and availability are matched to that exact configuration. Choose the mainframe and module combination that fits your channel count and connectivity requirements.
The defining difference across the family is connectivity: the DAQ970A ships with built-in LAN and USB, and the DAQ973A adds a built-in GPIB interface to that same LAN and USB set. Both mainframes accept up to three of the nine available plug-in modules in any combination and deliver identical measurement performance, channel density, and 1-million-reading memory.
Beyond connectivity, configuration differences come from module choice rather than the mainframe—multiplexer type, scan speed, voltage and current ratings, and switching topology all vary by module. The comparison table below summarizes the interface and configuration differences across the family.
| Model | Type | Channels | Max Voltage |
|---|---|---|---|
| DAQ970A | DAQ Mainframe (USB, LAN) | Up to 120 | 300 V |
| DAQ973A | DAQ Mainframe (USB, LAN, GPIB) | Up to 120 | 300 V |
| DAQM900A | 2-wire solid state (4-wire selectable) | 20 | 120 V |
| DAQM901A | 2-wire armature (4-wire selectable) | 20 + 2 | 300 V |
Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.
| Specification | Value |
|---|---|
| DC Voltage Accuracy ± (% of reading + % of range) (1 Year, Tcal ± 5 ºC) | |
| 100 mV | 0.0050 + 0.0060 |
| 1 V | 0.0035 + 0.0006 |
| 10 V | 0.0030 + 0.0004 |
| 100 V | 0.0040 + 0.0006 |
| 300 V | 0.0040 + 0.0020 |
| Resistance Accuracy ± (% of reading + % of range) (1 Year, Tcal ± 5 ºC) | |
| 100 Ω | 0.0060 + 0.0060 |
| 1 kΩ | 0.0040 + 0.0007 |
| 10 kΩ | 0.0040 + 0.0005 |
| 100 kΩ | 0.0040 + 0.0005 |
| 1 MΩ | 0.0070 + 0.0005 |
| 10 MΩ | 0.025 + 0.001 |
| 100 MΩ | 0.350 + 0.001 |
| 1000 MΩ | 3.500 + 0.001 |
| DC Current Accuracy ± (% of reading + % of range) (1 Year, Tcal ± 5 ºC) | |
| 1 µA (typical) | 0.050 + 0.005 |
| 10 µA (typical) | 0.050 + 0.002 |
| 100 µA (typical) | 0.050 + 0.001 |
| 1 mA | 0.050 + 0.005 |
| 10 mA | 0.050 + 0.020 |
| 100 mA | 0.050 + 0.005 |
| 1 A | 0.080 + 0.010 |
| Diode Test (Typical, 1 Year) | |
| 5 V | 0.010 + 0.010 |
| True RMS AC Voltage Accuracy ± (% of reading + % of range) (1 Year) — 100 mV, 1 V, 10 V, 100 V ranges | |
| 3 Hz – 5 Hz | 0.50 + 0.02 |
| 5 Hz – 10 Hz | 0.10 + 0.02 |
| 10 Hz – 20 kHz | 0.05 + 0.02 |
| 20 kHz – 50 kHz | 0.07 + 0.03 |
| 50 kHz – 100 kHz | 0.15 + 0.05 |
| 100 kHz – 300 kHz | 1.00 + 0.1 |
| True RMS AC Voltage Accuracy ± (% of reading + % of range) (1 Year) — 300 V range | |
| 3 Hz – 5 Hz | 0.50 + 0.06 |
| 5 Hz – 10 Hz | 0.10 + 0.06 |
| 10 Hz – 20 kHz | 0.05 + 0.06 |
| 20 kHz – 50 kHz | 0.07 + 0.09 |
| 50 kHz – 100 kHz | 0.15 + 0.15 |
| 100 kHz – 300 kHz | 1.00 + 0.3 |
| True RMS AC Current Accuracy ± (% of reading + % of range) (1 Year) — 100 µA, 1 mA, 10 mA, 100 mA, 1 A ranges | |
| 3 Hz – 5 kHz | 0.10 + 0.04 |
| 5 kHz – 10 kHz (typical) | 0.10 + 0.04 |
| Capacitance (Typical) ± (% of reading + % of range) | |
| 1.0000 nF | 0.50 + 0.50 |
| 10.000 nF | 0.40 + 0.10 |
| 100.00 nF | 0.40 + 0.10 |
| 1.0000 µF | 0.40 + 0.10 |
| 10.000 µF | 0.40 + 0.10 |
| 100.00 µF | 0.40 + 0.10 |
| Frequency Accuracy (% of reading) (1 Year) — 100 mV, 1 V, 10 V, 100 V, 300 V ranges | |
| 3 Hz – 10 Hz | 0.070 |
| 10 Hz – 100 Hz | 0.030 |
| 100 Hz – 1 kHz | 0.007 |
| 1 kHz – 300 kHz | 0.007 |
| Square wave | 0.006 |
| Temperature — Thermocouple (Ref junction and DMM accuracy) | |
| B (1100 ºC to 1820 ºC) | 1.2 ºC |
| E (-150 ºC to 1000 ºC) | 1.0 ºC |
| J (-150 ºC to 1200 ºC) | 1.0 ºC |
| K (-100 ºC to 1200 ºC) | 0.9 ºC |
| N (-100 ºC to 1300 ºC) | 1.0 ºC |
| R (300 ºC to 1760 ºC) | 0.5 ºC |
| S (400 ºC to 1760 ºC) | 1.2 ºC |
| T (-100 ºC to 400 ºC) | 0.9 ºC |
| Temperature — RTD / Thermistor (DMM accuracy) | |
| RTD (PT100 DIN/IEC 751, PT1000) | 0.05 ºC |
| Thermistor (2.2 K, 5 K, 10 K) | 0.1 ºC |
| A/D Linearity (Measured) | |
| Linearity | 0.0001% of reading + 0.0001% of range |
| DC Voltage Measurement Characteristics | |
| Measurement method | Keysight patented continuously integrating multi-slope IV A/D converter |
| Input resistance (0.1 V, 1 V, 10 V range) | Selectable 10 MΩ or >10 GΩ |
| Input resistance (100 V, 300 V range) | 10 MΩ ± 1% |
| Input bias current | < 30 pA at 25 ºC |
| Input protection | 300 V on all ranges |
| True RMS AC Voltage Measurement Characteristics | |
| Measurement type | AC-coupled True RMS. Measures the AC component of the input. |
| Measurement method | Digital sampling with anti-alias filter |
| Maximum input | 250 DCV, 300 Vrms |
| Input impedance | 1 MΩ ± 1%, in parallel with < 100 pF |
| Input protection | 300 Vrms |
Important: The internal DMM input section is optically isolated and shielded from the earth-referenced circuitry and computer interface, providing up to 300 V of input isolation.
Important: The DAQ970A/DAQ973A mainframe requires at least one low-frequency multiplexer module to make measurements: only the DAQM900A, DAQM901A, DAQM902A, and DAQM908A modules connect to the internal DMM. Modules are ordered separately ('buy only what you need—and add more modules later as your application grows').
Important: Up to three modules, in any combination, can be inserted into a single mainframe.
Important: The DAQM905A RF Multiplexer uses SMB connectors; a standard set of (10) BNC-to-SMB adapter cables is provided with each RF module for BNC connections (2 GHz bandwidth via SMB, 1 GHz via the BNC-to-SMB adapters).
Important: On-module screw terminals accept wire sizes from 16-gage to 22-gage; twenty-gage wire is recommended for high channel count applications.
Recommended pairing: listed accessories include the 34307A J-type thermocouples and 34308A 10 kΩ thermistors for temperature measurements and the 11062A Kelvin clip set for resistance measurements.
ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.
Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.
Model No
Keysight
Condition
Factory New
Manufacturer
Keysight Technologies
Keysight DAQM905A Dual 4-Channel 50 Ohm RF Multiplexer Module for DAQ970A/DAQ973A
DAQM905A
Keysight DAQM902A 16-Channel Reed Multiplexer Module for DAQ970A/DAQ973A
DAQM902A
Keysight DAQM903A 20-Channel Actuator/General Purpose Switch Module for DAQ970A/DAQ973A
DAQM903A
Keysight DAQM904A 4x8 Two-Wire Matrix Switch Module for DAQ970A/DAQ973A
DAQM904A
Keysight DAQM901A 20-Channel Armature Multiplexer Module for DAQ970A/DAQ973A
DAQM901A
34307A Keysight Accessory New
34307A
Keysight 34194A Dual-Lock Link Kit for 53200 Series Frequency Counter/Timers
34194A
34308A Keysight Accessory New
34308A
11062A Keysight Accessory New
11062A