Skip to product information
1 of 1
Used
lab tested

33120A Agilent Arbitrary Waveform Generator Used

33120A Agilent Arbitrary Waveform Generator Used

Regular price $895
Sale price $895 Regular price
In Stock
Shipping calculated at checkout.
Calibrations
Request a Quote
Login & Get 2% Back on Purchases
Login or Sign Up for Loyalty

Features:

  • 15 MHz sine & square waveforms
  • Ramp, triangle, noise, sin(x)/x, exponential rise, exponential fall, cardiac, DC volts
  • 12-bit, 40 MSa/s, 16,000 points arbitrary waveforms
  • AM (int/ext), FM (int), FSK (internal/external), burst (int/ext)
  • Linear & logarithmic sweeps and burst operation
  • IntuiLink software plus GPIB & RS-232 interfaces included
Share on
View full details

Arbitrary Waveform Specifications:

  • Waveform length: 8 to 16,000 points
  • Amplitude resolution: 12 bits (including sign)
  • Sample rate: 40 MSa/s
  • Non-volatile memory: Four (4) 16,000 waveforms

Frequency Characteristics:

  • Sine: 100 μHz - 15 MHz
  • Square: 100 μHz - 15 MHz
  • Triangle: 100 μHz - 100 kHz
  • Ramp: 100 μHz - 100 kHz
  • White noise: 10 MHz bandwidth
  • Resolution: 10 μHz or 10 digits
  • Accuracy: 10 ppm in 90 days, 20 ppm in 1 year, 18°C - 28°C
  • Temp: Coeff < 2 ppm/°C
  • Aging: < 10 ppm/yr

Sinewave Spectral Purity (Harmonic Distortion):

  • DC to 20 kHz: -70 dBc
  • 20 kHz to 100 kHz: -60 dBc
  • 100 kHz to 1 MHz: -45 dBc
  • 1 MHz to 15 MHz: -35 dBc

Sinewave Spectral Purity (Spurious):

  • DC to 1 MHz: < -65 dBc
  • 1 MHz to 15 MHz: < -65 dBc + 6 dB/octave

Sinewave Spectral Purity (Total Harmonic Distortion):

  • DC to 20 kHz: <0.04%
  • Phase noise: <-55 dBc in a 30 kHz band

Signal Characteristics:

Squarewave

  • Rise/Fall time: < 20 ns
  • Overshoot: 4%
  • Asymmetry: 1% + 5ns
  • Duty cycle: 20% to 80% (to 5 MHz), 40% to 60% (to 15 MHz)

Triangel, Ramp, Arb

  • Rise/Fall time: 40 ns (typical)
  • Linearity: <0.1% of peak output
  • Setting Time: <250 ns to 0.5% of final value
  • Jitter: <25ns

Output Characteristics

  • Amplitude (into 50 ohms): 50 mVpp - 10 Vpp
  • Accuracy (at 1 kHz): ± 1% of specified output
  • Flatness (sinewave relative to 1 kHz):
  • < 100 kHz ± 1% (0.1 dB)
  • 100 kHz to 1 MHz ± 1.5% (0.15 dB)
  • 1 Mz to 15 MHz ± 2% (0.2 dB) Ampl, 3Vrms ± 3.5% (0.3 dB) Ampl < 3Vrms
  • Output Impedance: 50 ohms (fixed)
  • Offset (into 50½): + 5 Vpk AC + DC
  • Accuracy: ± 2% of setting + 2 mV
  • Resolution: 3 digits, amplitude and offset
  • Units: Vpp, Vrms, dBm
  • Isolation: 42 Vpk maximum to earth
  • Protection: Short circuit protected ± 15 Vpk overdrive < 1 minute

Modulation:

AM

  • Carrier -3dB Freq: 10 MHz (typical)
  • Modulation: any internal waveform including Arb
  • Frequency: 10 MHz - 20 kHz
  • Depth: 0% - 120%
  • Source: Internal/External 

FM

  • Modulation: any internal waveform including Arb
  • Frequency: 10 MHz - 10 kHz
  • Deviation: 10 MHz - 15 MHz
  • Source: Internal only

FSK

  • Internal rate: 10 MHz - 50 kHz
  • Frequency Range10 mHz - 15 MHz
  • Source: Internal/External (1 MHz max.)

Burst

  • Carrier Freq: 5 MHz max.
  • Count: 1 to 50,000 cycles or infinite
  • Start Phase: -360° to +360°
  • Internal Rate: 10 mHz - 50 kHz ± 1%
  • Gate Source: Internal/External Gate
  • Trigger: Single, External or Internal Rate

Sweep

  • Type: Linear or Logarithmic
  • Direction: Up or Down
  • Start F/Stop F: 10 mHz - 15 MHz
  • Speed: 1 ms to 500 s ± 0.1%
  • Trigger: Single, External, or Internal

Uncompromising performance for standard waveforms

The HP 33120A function/arbitrary waveform generator uses direct digital-synthesis techniques to create a stable, accurate output signal for clean, low-distortion sine waves. It also gives you fast rise and fall-time square wave, and linear ramp waveforms down to 10 MHz.

Custom waveform generation

Use the HP 33120A to generate complex custom waveforms such as a heartbeat or the output of a mechanical transducer. With 12-bit resolution and a sampling rate of 40 MSa/s, the HP 33120A gives you the flexibility to create any waveform you need. It also lets you store up to found 16,000-deep waveforms in nonvolatile memory.

Easy-to-use functionality

The front-panel operation of the HP 33120A is straightforward and intuitive. You can access any of ten major functions with a single key press or two, then use a simple knob to adjust frequency, amplitude and offset. To save time, you can enter voltage values directly in Vp-p, Vrms or dBm. Internal AM, FM, FSK and burst modulation make it easy to modulate waveforms without the need for a separate modulation source. Linear and log sweeps are also built in, with sweep rates selectable from 1ms to 500s. HP-IB and RS-232 interfaces are both standard, plus you get full programmability using SCPI commands.

Optional phase-lock capability

The Option 001 phase lock/TCXO timebase gives you the ability to generate synchronized phase-offset signals. An external clock input/output lets you synchronize with up to three other HP 33120As or with an external 10-MHz clock. Option 001 also gives you a TCXO timebase for increased frequency stability. With an accuracy of 4 ppm/yr, the TCXO timebase makes an HP 33120A ideal for frequency calibrations and other demanding applications. With Option 001, new commands let you perform phase changes on the fly, via the front panel or from a computer, allowing precise phase calibration and adjustment. Link the HP 33120A to your PC To further increase your productivity, use the HP 33120A in conjunction with HP 34811A BenchLink Arb software. The Windows® -based program lets you create and edit waveforms on your PC and download them to your HP 33120A with the click of a mouse. Create complex waveforms in a math or statistics program--or use the freehand drawing tool--then pass them into HP BenchLink Arb. Use in conjunction with HP BenchLink Scope, the software also lets you capture a waveform with your HP oscilloscope or DMM and send it to your HP 33120A for output.

Model No

33120A

Condition

Used

Manufacturer

Agilent

Frequency

15 MHz

1

Phase Lock/TCXO Timebase

STANDARD

Standard 10 MHZ timebase