Key Features At A Glance

  • 6.5-digit resolution with dc voltage accuracy of ±0.0019% and ac voltage accuracy of ±0.07%
  • Measures dc/ac voltage, dc/ac current, 2- and 4-wire resistance, plus frequency and period
  • Message-based, 1-slot C-size VXI module with GPIB readout for automated test systems
  • 1000 readings/s into internal memory at 4.5 digits, with reading storage to internal memory
  • Fast changes: 65 range changes/s and 30 function changes/s when measuring dc voltage
  • Built-in math: NULL, MIN/MAX, LIMIT, dB and dBm; limit checks can drive a TTL output
  • True RMS ac measurement plus dc:dc voltage ratio measurements
View full details

The Agilent E1412A is a 6.5-digit high-accuracy multimeter built as a 1-slot, C-size, message-based VXI module. It is identical in electrical design to the E1312A, differing only in size, and is described as delivering the widest functionality in Agilent's DMM line. It pairs that breadth with high performance at prices comparable to a 5.5-digit DMM.

A digital multimeter measures fundamental electrical quantities so an engineer can characterize a circuit or device under test. The E1412A's standard measurement set spans ac/dc voltage, ac/dc current, 2- and 4-wire resistance, and frequency/period, with advanced tests including limit checks that drive a TTL output and dc voltage ratios. As a VXI module with GPIB readout, it is suited to automated and rack-based measurement systems.

ValueTronics stocks and holds its own test-and-measurement inventory in a 20,000 sq ft secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. Every used unit is inspected and functionally verified in-house before it ships, while new units ship factory-sealed exactly as received from the manufacturer. Holding inventory directly lets our team match the right instrument and condition to each application.

Brand Heritage

Agilent Technologies was spun off from Hewlett-Packard in 1999. Agilent's electronic test and measurement business was subsequently established as Keysight Technologies in 2014.

Product Core & Specifications

Specification Value
General
Module type 1-Slot, C-size, message-based VXI module
Functions DCV, ACV, DCI, ACI, 2/4-wire Ω, frequency, period
Math/operations NULL, MIN/MAX, LIMIT, dB, dBm
dc Voltage Accuracy (1 Year, 23° C ± 5° C), ±(% of reading + % of range)
100.0000 mV 0.0050 + 0.0035
1.000000 V 0.0040 + 0.0007
10.00000 V 0.0035 + 0.0005
100.0000 V 0.0045 + 0.0006
300.0000 V 0.0045 + 0.0030
dc Voltage Characteristics
Measurement method Continuously integrating, multi-slope III A/D converter
A/D linearity 0.0002% of reading + 0.0001% of range
Input resistance (0.1 V, 1 V, 10 V ranges) Selectable 10 MΩ or 10 GΩ
Input resistance (100 V, 300 V ranges) 10 MΩ ± 1%
Input bias current <30 pA at 25° C
Input terminals Copper alloy
Input protection 300 V on all ranges
dc CMRR (60 Hz/50 Hz, 1 kΩ unbalance in LO lead) 140 dB
Resistance Accuracy (1 Year, 23° C ± 5° C), ±(% of reading + % of range)
100.0000 Ω (1 mA) 0.010 + 0.004
1.000000 kΩ (1 mA) 0.010 + 0.001
10.00000 kΩ (100 µA) 0.010 + 0.001
100.0000 kΩ (10 µA) 0.010 + 0.001
1.000000 MΩ (5 µA) 0.010 + 0.001
10.00000 MΩ (500 nA) 0.054 + 0.001
100.0000 MΩ (500 nA || 10 MΩ) 0.800 + 0.010
Resistance Characteristics
Measurement method Selectable 4-wire or 2-wire Ω (current source referenced to low input)
Max. resistance (2/4-wire Ω) 100 MΩ
Max. lead resistance (4-wire Ω) 10% of range per lead for 100 Ω and 1 kΩ, per lead on all other ranges
Input protection 300 V on all ranges
dc Current Accuracy (1 Year, 23° C ± 5° C), ±(% of reading + % of range)
10.00000 mA (<0.1 V) 0.070 + 0.020
100.0000 mA (<0.6 V) 0.070 + 0.005
1.000000 A (<1 V) 0.150 + 0.010
3.000000 A (<2 V) 0.720 + 0.020
dc Current Characteristics
Shunt resistor 0.1 Ω for 1 A and 3 A, 5 Ω for 10 mA and 100 mA
Input protection Externally accessible 3 A, 250 V fuse
True RMS ac Voltage Accuracy, 100 mV Range (1 Year, 23° C ± 5° C, Slow filter, sinewave), ±(% of reading + % of range)
3 Hz-5 Hz 1.00 + 0.04
5 Hz-10 Hz 0.35 + 0.04
10 Hz-20 kHz 0.06 + 0.04
20 kHz-50 kHz 0.12 + 0.05
50 kHz-100 kHz 0.60 + 0.08
100 kHz-300 kHz 5.00 + 0.50
True RMS ac Voltage Accuracy, 1 V to 300 V Ranges (1 Year, 23° C ± 5° C, Slow filter, sinewave), ±(% of reading + % of range)
3 Hz-5 Hz 1.00 + 0.03
5 Hz-10 Hz 0.35 + 0.03
10 Hz-20 kHz 0.60 + 0.03
20 kHz-50 kHz 0.12 + 0.05
50 kHz-100 kHz 0.06 + 0.08
100 kHz-300 kHz 5.00 + 0.50
True RMS ac Voltage Characteristics
Measurement method ac-coupled True RMS — measures the ac component of the input with up to 300 Vdc of bias on any range (Max ac+dc = 300 V rms)
Crest factor Maximum 5:1 at full scale
ac filter bandwidth (Slow) 3 Hz-300 kHz
ac filter bandwidth (Medium) 20 Hz-300 kHz
ac filter bandwidth (Fast) 200 Hz-300 kHz
Input impedance 1 MΩ ±2%, in parallel with 100 pF
Input protection 300 Vrms all ranges
ac CMRR (1 kΩ unbalance in LO lead) 70 dB
True RMS ac Current Accuracy (1 Year, 23° C ± 5° C), ±(% of reading + % of range)
1 A, 3 Hz-5 Hz 1.05 + 0.04
1 A, 5 Hz-10 Hz 0.35 + 0.04
1 A, 10 Hz-1 kHz 0.15 + 0.04
1 A, 1 kHz-50 kHz 0.40 + 0.04
3 A, 3 Hz-5 Hz 1.70 + 0.06
3 A, 5 Hz-10 Hz 0.95 + 0.06
3 A, 10 Hz-1 kHz 0.75 + 0.06
3 A, 1 kHz-50 kHz 1.00 + 0.06
True RMS ac Current Characteristics
Measurement method Direct couple to the fuse and shunt; ac-coupled True RMS measurement (measures the ac component only)
Shunt resistor 0.1 Ω for 1 A and 3 A ranges
Burden voltage (1 A range) <1 Vrms
Burden voltage (3 A range) <2 Vrms
Input protection Externally accessible 3 A, 250 V fuse
Frequency and Period Accuracy (1 Year, 23° C ± 5° C, 100 mV to 300 V), % of reading
3 Hz-5 Hz 0.10
5 Hz-10 Hz 0.06
10 Hz-40 Hz 0.03
40 Hz-300 kHz 0.01
Integration Time / Normal Mode Rejection (power-line frequency ± 0.1%)
100 PLC / 1.67 s (2 s) 60 dB
10 PLC / 167 ms (200 ms) 60 dB
1 PLC / 16.7 ms (20 ms) 60 dB
<1 PLC / 3 ms (800 µs) 0 dB
dc System Speed (4.5 digits, Delay 0, Autozero OFF)
Function change 30/s
Range change 65/s
Autorange time <30 ms
Max. internal trigger rate 1000/s
Max. external trigger rate to memory 1000/s
dc:dc Ratio
Measurement method Input HI-LO / Reference HI-LO (apply Reference HI-LO signal to Ohms 4-Wire Sense terminals)
Input HI to Input LO 100 mV to 300 V
Reference HI to Input LO <12 V on 100 mV to 10 V ranges (autoranged)
Reference LO to Input LO <2 V
Important: Agilent recommends the use of Teflon or other high-impedance, low-dielectric-absorption wire insulation for these measurements.

About this Pre-Owned unit

This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.

  • Warranty included
  • Functional verification included
  • Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
  • Note that unnecessary accessories may not be included (contact Test Architect to confirm).

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.

Model No

Agilent

Condition

Pre-Owned

Manufacturer

Agilent Technologies

Multi_met_acv

300 V

Multi_met_dcv

300 V

Multi_met_dig

6.5