Input Specifications
- Voltage: 115/230 VAC ± 10%, user selectable
- Frequency: 50/60 Hz ± 5%
- Fuse: 3.15 A, fast acting 250 VAC
Dielectric Withstand Test Mode
Output Rating
- 5000 V @ 20 mA AC; 6000 V @ 7.5 mA DC
Voltage Setting
- Range: 0 - 5.00 KV AC; 0 - 6.00 KV DC
- Resolution: 0.01 KV
- Accuracy: ± (2% of setting + 5 V)
Maximum Limit
- AC Range: 0.00 - 20.00 mA
- Resolution: 0.01 mA
- DC Range: 0 - 7500 μA
- Resolution: 1 μA
- Accuracy: AC and DC ± (2% of setting + 2 counts)
Minimum Limit
- AC Range: 0.000 - 9.999 mA
- Resolution: 0.001 mA
- DC Range: 0.0 - 999.9 μA
- Resolution: 0.1 μA
- Accuracy: AC and DC ± (2% of setting + 2 counts)
Arc Detection
Ground Fault Interrupt
- GFI Trip Current: 450 μA max (AC or DC)
- HV Shut Down Speed: < 1ms
Current Display Auto Range
- AC Range 1: 0.000 - 3.500 mA
- Range 2: 3.00 - 20.00 mA
- DC Range 1: 0.0 μA - 350.0 μA
- Range 2: 0.300 mA - 3.500 mA
- Range 2: 3.00 mA - 7.50 mA
- Accuracy All Ranges ± (2% of reading + 2 counts)
DC Output Ripple
- ≤5% Ripple RMS at 6 KV DC @ 7.5 mA, Resistive Load
Discharge Time
The maximum capacitive load vs output voltage:
- 0.20 μF < 1 KV 0.050 μF < 4 KV
- 0.10 μF < 2 KV 0.040 μF < 5 KV
- 0.06 μF < 3 KV 0.015 μF < 6 KV
AC Voltage Waveform
- Sine Wave, Crest Factor = 1.3 - 1.5, <2% THD
Output Frequency
- Range: 50 or 60 HZ, User Selectable
Output Voltage Regulation
- ± (1% of output + 5 V) from no load to full load and over input voltage range.
Dwell Timer
- Range: AC 0, 0.3 - 999.9 sec (0 = Constant)
- DC 0, 0.4 - 999.9 sec (0 = Constant)
- Accuracy: ± (0.1% of reading + 0.05 sec)
Ramp Timer
- Range: Ramp-Up: 0.1 - 999.9 sec
- Ramp-Down: AC 0.0 - 999.9 sec
- DC 1.0 - 999.9 sec (0=OFF)
- Accuracy: ± (0.1% of reading + 0.05 sec)
Ground Continuity Current
Ground Continuity Maximum Limit, Minimum Limit
- Range: 0.0 Ω - 1.50 Ω
- Resolution: 0.01 Ω
- Accuracy: ± (3% of setting + 0.02 Ω)
Ground Continuity Auto Offset
- Range: 0.0 Ω - 0.50 Ω
- Resolution: 0.01 Ω
- Accuracy: ± (3% of setting + 0.02 Ω)
Insulation Resistance Test Mode
Voltage Setting
- Range: 30 - 1000 VDC
- Accuracy: ± (2% of setting + 5 V)
Resistance Display
- Range: 1 - 9999 MΩ (4 Digit, Auto Ranging)
- Resolution: 500 VDC - 1000 VDC
MΩ MΩ
0.001 1.000 - 9.999
0.01 10.00 - 99.99
0.1 100.0 - 999.9
1 1000 - 9999
Accuracy
- ± (2% of reading +2 counts) at test voltage; 500 - 1000 V and 1 - 999.9 MΩ
- ± (5% of reading +2 counts) at test voltage; 500 - 1000 V and 1000 - 9999 MΩ
- ± (8% of reading +2 counts) at test voltage; 30 - 500 V and 1 - 1000 MΩ
Maximum Limit
- Range: 0, 1 - 9999 MΩ (0=OFF)
- Accuracy: Same as Resistance Display
Minimum Limit
- Range: 1 - 9999 MΩ
- Accuracy: Same as Resistance Display
Ramp Timer
- Range: Ramp-Up: 0.1 - 999.9 sec
- Ramp-Down: 1.0 - 999.9 sec (0=OFF)
- Accuracy: ± (0.1% of reading + 0.05 sec)
Delay Timer
- Range: 0, 0.5 - 999.9 sec (0 = Constant)
- Accuracy: ± (0.1% of reading + 0.05 sec)
Ground Fault Interrupt
- GFI Trip Current: 450 μA max
- HV Shut Down Speed: < 1 ms
General Specifications
Remote Control & Signal Output
The following input and output signals are provided through two 9 pin D type connectors:
1. Remote control: Test, Reset, and Remote Interlock
2. Remote recall of memory program #1, #2, and #3
3. Outputs: Pass, Fail, Test-in-process, and Reset
- Program Memory: 10 Memories, 3 steps per memory
- Security: Key Lock capability to avoid unauthorized access to all test parameters and memory locations
- Mechanical: Bench or rack mount with tilt up feet
- Dimensions: (W x H x D) 8.46 x 3.5 x 14.57 in. (215 x 89 x 370 mm)
- Calibration: Traceable to National Institute of Standards & Technology (NIST)
- Calibration controlled by software
- Adjustments are made through front panel keypad in a restricted access calibration mode
- Calibration information stored in non-volatile memory