The HIOKI CT6844, CT6845, and CT6846 are clamp-on AC/DC current probes built around an ergonomic clamp design that delivers single-handed, one-touch opening and closing. The compact form lets the clamp reach into tangled wiring, while the larger-diameter models are sized to clamp even thick cables and double wiring. Across the series the sensors are engineered for consistent, high-precision high-current measurement over a wide temperature range, with reduced sensitivity to where the conductor sits inside the clamp core and to current flowing in nearby wires.
These current probes are positioned for EV/HEV and PV (photovoltaic) work and for the operational evaluation of devices inside equipment that is subject to extreme temperature changes. Because they hold high-current accuracy over a broad temperature span, they are well suited to measurements taken in thermostatic chambers and engine rooms. The clamp's tolerance to conductor position and to noise from nearby conductors makes it practical for measuring in crowded, real-world wiring rather than only on an isolated bench setup.
ValueTronics stocks and holds its own inventory in a 20,000 sq ft secure warehouse at 1675 Cambridge Drive in Elgin, Illinois, so the instruments we list are backed by equipment we physically keep on hand.
The CT6844, CT6845, and CT6846 form a single AC/DC current probe series that shares an ergonomic, one-touch clamp design and a basic amplitude accuracy of +/-0.3% rdg. +/-0.01% f.s. at 50/60 Hz. The models differ in rated primary current, frequency characteristics, and the maximum conductor diameter the clamp will accept, allowing a measurement task to be matched to the right combination of current range, bandwidth, and cable size.
Within the series the CT6844 pairs 500 A AC/DC with the widest bandwidth and the smallest clamp opening, the CT6845 keeps 500 A AC/DC with a larger clamp opening at a lower bandwidth, and the CT6846 raises the rated current to 1000 A AC/DC with the narrowest bandwidth. Selecting between them is a question of how much current, how much bandwidth, and how large a conductor the application requires.
Each model below is its own dedicated product page with condition-matched pricing for the new unit. Choose the model that matches your required current rating, bandwidth, and conductor diameter, then select the listing for the condition you need.
The CT6844 is the wide-bandwidth member of the series, rated 500 A AC/DC from DC to 200 kHz with a clamp sized for conductors up to 20 mm. It also holds the tightest conductor-position effect in the series at +/-0.1% rdg. or less and the lowest external-magnetic-field effect at 100 mA or less, in the most compact 153W x 67H x 25D mm, 400 g body.
The three models separate primarily on current rating, bandwidth, and conductor size. The CT6844 covers 500 A AC/DC, DC to 200 kHz, and conductors up to 20 mm; the CT6845 covers 500 A AC/DC, DC to 100 kHz, and conductors up to 50 mm; and the CT6846 covers 1000 A AC/DC, DC to 20 kHz, and conductors up to 50 mm. Output voltage is 4 mV/A on the CT6844 and CT6845 and 2 mV/A on the CT6846.
Secondary characteristics also track with the model. Conductor-position effect is +/-0.1% rdg. or less on the CT6844 versus +/-0.2% rdg. or less on the CT6845 and CT6846, and external-magnetic-field effect is 100 mA or less on the CT6844 versus 150 mA or less on the larger two. The comparison table below lists these values side by side to support a direct selection.
| Model | Rated Current | Frequency Bandwidth | Conductor Diameter |
|---|---|---|---|
| CT6845 | 500 A AC/DC | DC to 100 kHz | φ50 mm (1.97 in) |
| CT6844 | 500 A AC/DC | DC to 200 kHz | φ20 mm (0.79 in) |
| CT6846 | 1000 A AC/DC | DC to 20 kHz | φ50 mm (1.97 in) |
| CT6844-05 | 500 A AC/DC | DC to 200 kHz | φ20 mm (0.79 in) |
Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.
Other probes in the same AC/DC current probe series include the CT6841 (20 A) and CT6843 (200 A), extending the available current ranges within one connector ecosystem.
| Specification | Value |
|---|---|
| Current Sensing | |
| Rated primary current | 500 A AC/DC |
| Frequency characteristics | DC to 100 kHz |
| Measurable conductor diameter | φ50 mm (1.97 in) or less |
| Output voltage | 4 mV/A |
| Connector type | HIOKI PL23 |
| Accuracy (guaranteed for 1 year) | |
| Basic accuracy (50/60 Hz) | Amplitude accuracy: ±0.3% rdg. ±0.01% f.s.; Phase accuracy: ±0.1 deg |
| Basic accuracy (DC) | Amplitude accuracy: ±0.3% rdg. ±0.02% f.s. |
| Effect of conductor position | ±0.2% rdg. or less |
| Effect of external magnetic fields | 150 mA or less (Scaled value, in a DC and 60 Hz magnetic field of 400 A/m) |
| Magnetic susceptibility | 75 mA or less (Scaled value, after 500 A AC/DC input) |
| Effect of common-mode voltage | 0.05% f.s. or less (1000 Vrms, 50/60 Hz DC) |
| Temperature and Humidity | |
| Temperature and humidity range of guaranteed accuracy | 0℃ to +40℃ (32 °F to 104 °F), 80% RH or less (no condensation) |
| Operating temperature and humidity | -40℃ to +85℃ (-40 °F to 185 °F), 80% RH or less (no condensation) |
| General | |
| Dimensions | 238W × 116H × 35D mm (9.37"W × 4.57"H × 1.38"D) |
| Mass | 860 g (30.3 oz) |
| Cord length | 3 m (9.84 ft) |
| Accessories | Instruction manual ×1 (Japanese, English and Chinese), Mark band ×6, Carrying case ×1 |
Important: Requires SENSOR UNIT 9555-10 to supply power to the current sensor when the probe is used alone.
Important: These AC/DC current probes are sensor units that require a compatible HIOKI power meter to supply power and read their output: the PL23-connector models (CT6844/CT6845/CT6846) work with the PW6001, 3390 and 3193-10; the ME15W-connector models (CT6844-05/CT6845-05/CT6846-05) work with the 3334-10. The power meter is sold separately.
Important: For output to an oscilloscope or recorder, the Connection Cord L9217 (insulated BNC connectors at both ends, 1.6 m / 5.25 ft) is required and is ordered separately.
Important: Use the CT9900 (PL23 to ME15W) or CT9901 (ME15W to PL23) conversion cable to adapt between connector types.
This is a brand-new, factory-sealed unit, supplied exactly as it arrives from the manufacturer and backed by the full manufacturer warranty. For the exact configuration, available options, included accessories, or current lead time for your application, our Test Architects can confirm the details before you order.
ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.
Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.
Model No
HIOKI
Condition
Factory New
Manufacturer
Hioki
Curr_prob_freq
DC to 100 kHz
Plugin_type
PL23
Request A Quote