Key Features At A Glance

  • DC-coupled current probe for simultaneous AC/DC current measurements
  • DC to 100 MHz bandwidth with the AM 503B/TM502A amplifier system
  • Measures up to 20 A DC + peak AC and 50 A peak pulse current
  • Minimum sensitivity of 1 mA/div for high-sensitivity current measurements
  • Split-core construction enables easy connection to the circuit under test
  • Maximum barewire voltage of 300 V (Category I)
  • 100 x 10⁻⁶ amp-second product; ≤3.5 ns risetime with the AM 503B
  • Carries UL, CSA, and CE marking
View full details

The A6312 is a DC-coupled current probe rated for DC to 100 MHz bandwidth, designed for use with the AM 503, AM 503A, AM 503B, and AM 5030 current probe amplifiers. Its split-core construction permits easy connection to the circuit under test, and it may be ordered either separately as the A6312 or as Option 05 to the AM 503S current measurement system.

A current probe clamps around a conductor to measure the current flowing through it without breaking the circuit, presenting the result as a voltage waveform for display on an oscilloscope. The combination of an A6312, an AM 503B, and a TM502A forms a current-measurement tool that can display currents up to 20 amps DC + peak AC and 50 amps peak pulse on any oscilloscope. Because the probe is DC-coupled, it measures both the DC and AC components of a current simultaneously.

ValueTronics stocks and holds its own inventory in a 20,000-square-foot secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. Pre-owned instruments are inspected and functionally verified in-house before they ship, while new units arrive and leave factory-sealed exactly as received.

Brand Heritage

Tektronix was founded in 1946 and has operated under the Tektronix name ever since. The company was acquired by Danaher Corporation in 2007 and became part of Fortive Corporation when Fortive was spun off from Danaher in 2016.

Accessories Supplied

  • Instruction Manual (P/N 070-9603-00)

Product Core & Specifications

Specification Value
Standard Specifications (A6312/AM 503B/TM502A)
Bandwidth DC to 100 MHz
Maximum DC + Peak AC Current 20 Amps
Maximum Peak Pulse Current 50 Amps
Minimum Sensitivity 1 mA/div
DC Accuracy ±3% (±1.5% typical)
Maximum Barewire Voltage 300 Volts (Category I)
Typical Characteristics
Insertion Impedance 0.1 Ohms at 1 MHz; 0.2 Ohms at 50 MHz; 1.0 Ohm at 100 MHz
Propagation Delay 17 ns
Maximum Conductor Diameter 3.8 mm (0.15 in.)
Electrostatic Immunity (per IEC 801-2) >8 kV
Amp * Sec Product 100 x 10–6 Amp * sec
Risetime ≤3.5 ns
Physical Characteristics
Probe Weight 0.25 kg (0.56 lbs.)
Cable Length 2 m (6.6 ft.)
Power Requirements
Power Source Powered by the AM 503B/TM502A
Safety Certifications
UL UL 3111-1
CSA CSA 1010.1
IEC IEC 1010-1, 1010-2-032
System Specifications
Bandwidth (A6312/AM 503B) DC to 100 MHz
Bandwidth (A6312/AM 5030) DC to 100 MHz
Bandwidth (A6312/AM 503A) DC to 80 MHz
Bandwidth (A6312/AM 503) DC to 100 MHz
Risetime, 10% to 90% (A6312/AM 503B) ≤3.5 ns
Risetime, 10% to 90% (A6312/AM 5030) ≤3.5 ns
Risetime, 10% to 90% (A6312/AM 503A) ≤4.4 ns
Risetime, 10% to 90% (A6312/AM 503) ≤3.5 ns (requires adjustment of R364 HFComp in amplifier)
Aberrations, typical (all combinations) 7% p-p
DC Gain Accuracy (A6312/AM 503B) ≤3% (<1.5% typical)
DC Gain Accuracy (A6312/AM 5030) ≤3%
DC Gain Accuracy (A6312/AM 503A) ≤3%
DC Gain Accuracy (A6312/AM 503) ≤3%
Noise, RMS, typical (all combinations) ≤250 µA
Important: Maximum barewire voltage: 300 V (Category I).
Important: The A6312 is powered by the AM 503B/TM502A; the listed performance specifications apply to this combination of products. A current probe amplifier (AM 503, AM 503A, AM 503B, or AM 5030) and a TM502A mainframe are required to operate the probe and are not included with the A6312.
Important: For the A6312/AM 503 combination, the specified risetime and aberrations require adjustment of R364 (HFComp) in the amplifier.

About this Pre-Owned unit

This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.

  • Warranty included
  • Functional verification included
  • Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
  • Note that unnecessary accessories may not be included (contact Test Architect to confirm).

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.

Model No

Tektronix

Condition

Pre-Owned

Manufacturer

Tektronix

Curr_prob_freq