Skip to content

The Hioki PW3335 is a single-phase AC/DC power meter designed for high-accuracy measurement spanning standby power through full operating power. The instrument provides a wide measurable range from 10 µA to 30 A on the current input and 60 mV to 1000 V on the voltage input, with basic accuracy specified at ±0.1% for voltage, current, and power. Frequency bandwidth covers DC and 0.1 Hz to 100 kHz, with built-in harmonic measurement to the 50th order.
Applications listed in the datasheet include measurement of AC adapter standby power on both primary-side AC and secondary-side DC, characterization of power supply conversion devices such as inverters and thyristors, and measurement of solar panels and power converters using the 1000 V range. The PW3335 is also positioned for evaluation and development of power sources for home appliances, for input and output measurement of secondary batteries in EVs, and for measuring the sold power of solar panels, with power consumption and regeneration power integrated separately.
ValueTronics stocks the PW3335 family alongside Hioki current sensors, sensor units, and the connection accessories required for full system deployment. All inventory is housed and processed at our 20,000 sq ft secure warehouse at 1675 Cambridge Drive, Elgin, Illinois, where instruments are inspected and verified before shipment. Unlike pass-through brokers and drop-shippers, we maintain the instrument in-house, which means availability questions, configuration confirmations, and accessory compatibility checks are answered against units we can physically reference.
Hioki E.E. Corporation has operated as an independent Japanese precision measurement manufacturer since 1935, headquartered in Ueda, Nagano. The company has not been absorbed into a larger conglomerate, and the Hioki brand has remained continuous on the PW-series power meter line through its successive generations.
The PW3335 family is a single product platform offered in five configurations that differ only in installed interface and input options. The base PW3335 includes LAN and RS-232C. The PW3335-01 substitutes GP-IB for RS-232C. The PW3335-02 adds D/A output to the base LAN and RS-232C configuration.
All five variants share the same measurement engine: identical voltage and current ranges, identical accuracy specifications, identical frequency bandwidth, identical harmonic measurement capability, and identical synchronized control of up to 8 units. The selection decision is driven entirely by the interface and input requirements of the user's bench or automated test system.
Each new PW3335 variant below links to its own product page with condition-matched pricing. The model number is the determinant of which interfaces and inputs ship with the unit; the underlying measurement instrument is the same across the family.
The differences between the five variants are interface-specific. LAN is included on every model. RS-232C is included on every model except the PW3335-01 (which substitutes GP-IB). GP-IB is included only on the PW3335-01 and PW3335-04.
Buyers building automated test systems on GP-IB infrastructure typically select the PW3335-01 or PW3335-04. Buyers requiring analog output to a data logger or Memory HiCorder typically select the PW3335-02 or PW3335-04. Buyers measuring current above the 30 A direct-input range — using Hioki clamp-on or AC/DC current sensors for measurements up to 5000 A AC — require the PW3335-03 or PW3335-04. The PW3335-04 is the all-options variant for users who need GP-IB, D/A output, and external current sensor input together.
| Model | LAN | RS-232C | GP-IB |
|---|---|---|---|
| PW3335-03 | Yes | Yes | No |
| PW3335 | Yes | Yes | No |
| PW3335-01 | Yes | No | Yes |
| PW3335-02 | Yes | Yes | No |
Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.
| Specification | Value |
|---|---|
| Input Specifications | |
| Measurement line type | Single-phase 2-wire (1P2W) |
| Voltage input method | Isolated input, resistive voltage divider method |
| Current input method | Isolated input, shunt input method |
| Voltage measurement ranges | AUTO / 6.0000 V / 15.000 V / 30.000 V / 60.000 V / 150.00 V / 300.00 V / 600.00 V / 1.0000 kV |
| Current measurement ranges | AUTO / 1.0000 mA / 2.0000 mA / 5.0000 mA / 10.000 mA / 20.000 mA / 50.000 mA / 100.00 mA / 200.00 mA / 500.00 mA / 1.0000 A / 2.0000 A / 5.0000 A / 10.000 A / 20.000 A |
| Power ranges | Depends on combination of voltage and current ranges; from 6.0000 mW to 20.000 kW (also applies to VA, var) |
| Voltage input resistance | Approx. 2 MΩ |
| Current input resistance | 1 mA to 100 mA range: 520 mΩ or less; 200 mA to 20 A range: 15 mΩ or less |
| Basic Measurement Specifications | |
| Measurement method | Simultaneous voltage and current digital sampling, zero-cross simultaneous calculation |
| Sampling frequency | Approx. 700 kHz |
| A/D converter resolution | 16-bit |
| Frequency bandwidth | DC, 0.1 Hz to 100 kHz (values within 0.1 Hz ≤ f < 10 Hz are for reference only) |
| Synchronization sources | U, I, DC (fixed to 200 ms) |
| Basic accuracy (DC) | Voltage / current / power: ±0.1% rdg, ±0.1% f.s. |
| Basic accuracy (45 Hz to 66 Hz) | Voltage / current / power: ±0.1% rdg, ±0.05% f.s. |
| Effect of power factor | ±0.1% f.s. (45 to 66 Hz, at power factor = 0) |
| Effect of common mode voltage | ±0.01% f.s. or less (600 V, 50/60 Hz) |
| Maximum effective peak voltage | ±600% of each voltage range; for 300 V, 600 V, and 1000 V ranges, ±1500 V peak |
| Maximum effective peak current | ±600% of each current range; for 20 A range, ±60 A peak |
| Guaranteed accuracy period | 1 year |
| Temperature coefficient | ±0.03% f.s. per °C or less (±0.06% f.s. per °C for 1 mA range) |
| Harmonic Measurement | |
| Measurement method | Zero-cross simultaneous calculation; IEC 61000-4-7:2002 compliant when synchronization frequency is 45 Hz to 66 Hz |
| FFT processing | 32-bit word length, 4096 points |
| Window function | Rectangular |
| Maximum analysis order | Up to 50th |
| Analysis order upper limit setting | 2nd to 50th |
| Frequency Measurement | |
| Number of channels | 2 (voltage, current) |
| Measurement ranges | 100 Hz / 500 Hz / 5 kHz / 100 kHz (linked to zero-cross filter) |
| Measurement accuracy | ±0.1% rdg ±1 dgt (±0.2% rdg ±1 dgt for 1 mA range) |
| Effective measuring range | 0.1 Hz to 100 kHz |
| Integration | |
| Integration time | 1 min. to 10000 hr., settable in 1 min. blocks |
| Integration time accuracy | ±0.01% rdg ±1 dgt |
| Display resolution | 999999 (6 digits + decimal point) |
| Display | |
| Display type | 7-segment LED |
| Number of display parameters | 4 (display area a, b, c, and d) |
| Display resolution | 99999 count (5 digits); integrated values 999999 count (6 digits) |
| Display update rate | 200 ms ±50 ms (approx. 5 updates/sec) to 20 s |
| Synchronized Control | |
| Terminal | BNC terminal Ă— 1 (non-isolated) |
| Number of units | Up to 7 slaves per master (total of 8 units including PW3336/PW3337 series) |
| LAN Interface (all models) | |
| Connector | RJ-45 Ă— 1 |
| Electrical specifications | Compliant with IEEE802.3 |
| Transmission method | 10Base-T / 100Base-TX (automatic detection) |
| Protocol | TCP/IP |
| RS-232C Interface (PW3335, PW3335-02, PW3335-03, PW3335-04) | |
| Connector | D-sub 9-pin Ă— 1 |
| Communication method | Full duplex, start-stop synchronization; 1 stop bit (fixed); 8 data bits (fixed); no parity |
| Communication speed | 9600 bps / 38400 bps |
| GP-IB Interface (PW3335-01, PW3335-04) | |
| Method | Compliant with IEEE488.1 1987, referenced to IEEE488.2 1987 |
| Address | 00 to 30 |
| D/A Output (PW3335-02, PW3335-04) | |
| Number of output channels | 7 channels |
| Configuration | 16-bit D/A converter (polarity + 15 bits) |
| Maximum output voltage | Approx. ±12 V DC |
| Output resistance | Approx. 100 Ω |
| External Current Sensor Input (PW3335-03, PW3335-04) | |
| Terminal | Isolated BNC terminals |
| Current sensor type switching | Off / TYPE.1 / TYPE.2 |
| Current measurement range | Auto / 1 A / 2 A / 5 A |
| Power range configuration | From 24.000 W to 5.0000 MW (also applies to VA, var) |
| General Specifications | |
| Product warranty period | 1 year |
| Operating environment | Indoors, altitude up to 2000 m (6562 ft), pollution degree 2 |
| Operating temperature and humidity | 0°C to 40°C (32°F to 104°F), 80% RH or less (no condensation) |
| Storage temperature and humidity | -10°C to 50°C (14°F to 122°F), 80% RH or less (no condensation) |
| Dielectric strength | 4290 V rms AC (current sensitivity: 1 mA) |
Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.
Model No
Hioki
Condition
New
Manufacturer
Hioki
Request A Quote