In summary, the B&K Precision 5493C is a 6½-digit (1,200,000-count) bench digital multimeter built around accuracy, repeatability, and ease of use. It delivers 0.0035% (35 ppm) basic DC voltage accuracy, measurement speeds up to 1000 readings per second, an internal storage buffer holding 10,000 readings, and 12 measurement functions including DCV, ACV, DCI, ACI, 2-wire and 4-wire resistance, capacitance, frequency, diode, temperature, continuity, and DCV Ratio. The instrument features a high-contrast 4.3-inch full color LCD with histogram, bar meter, and trend chart display modes, dual measurement display, and USB (USBTMC-compliant), RS232, and LAN interfaces. Rear input terminals are included for rack-mount use.
The B&K Precision 5493C is used by engineering, production, and service teams for general-purpose precision measurement at the bench. Its 6½-digit resolution and 35 ppm basic DC voltage accuracy place it in the precision tier alongside the Keysight 34461A and Fluke 8845A — appropriate for calibration verification, R&D characterization, automated test systems, and any application where 5½-digit resolution leaves measurement noise floor on the table. The instrument's 12 measurement functions and 4.3-inch color display make it suitable as a primary bench DMM in a small lab or as a programmable measurement node in an automated test system.
A handheld 3½-digit or 4½-digit DMM resolves 2,000 to 20,000 counts and typically delivers 0.5% to 0.05% basic DC accuracy — adequate for field troubleshooting but not for measurements that feed design data or quality records. The 5490C Series is a bench precision instrument: 6½-digit (1,200,000-count) resolution (1,200,000 counts), 0.0035% (35 ppm) basic DC accuracy, programmable control via USB (USBTMC-compliant), RS232, and LAN, a 10,000-reading internal buffer for data logging, and dual measurement display for simultaneous viewing of two measurements such as AC voltage and frequency on the same screen. A bench DMM and a handheld DMM are complementary instruments — the 5490C Series is the bench reference, not a field replacement.
The 5493C delivers 0.0035% (35 ppm) basic DC voltage accuracy, a temperature coefficient of 5 ppm/°C on the 10 V range, measurement speeds up to 1000 readings per second, a 10,000-reading internal storage buffer, and 2-wire/4-wire resistance measurements down to 10 Ω with 0.1 mΩ resolution on the 10 Ω range (10 µΩ resolution at 6½ digits). The frequency function covers 3 Hz to 1 MHz with sine wave input, and capacitance measurement spans 1 nF to 10 mF in eight ranges. Temperature measurement is supported via Pt100 (DIN/IEC 751) RTD probes or 5 kΩ thermistor probes. The 4.3-inch full color LCD supports histogram, bar meter, and trend chart display modes with simultaneous min/max/peak-peak/average/standard deviation statistics over the captured buffer. Power consumption is ≤ 30 VA from 110 V/220 V ±10% AC mains; the instrument weighs 5.51 lbs (2.5 kg) in a 8.85" × 3.93" × 13.97" half-rack form factor.
B&K Precision Corporation has provided reliable, value-priced test and measurement instruments worldwide for more than 60 years. Headquartered in Yorba Linda, California, the company is ISO9001:2015 registered (NSF-ISR, certificate 6Z241-IS8) and operates global subsidiaries in France (Sefram), Taiwan (B+K Precision Taiwan), and Brazil (B&K Brasil), with independent service support in Singapore for the broader Asia-Pacific market. The 5490C Series sits in the bench precision DMM category — instruments designed for calibration verification, automated test, and R&D measurement where 5½ to 6½ digits of resolution and sub-0.01% basic accuracy are required, and where programmable control via standard interfaces (USB, LAN, RS232, GPIB) is non-negotiable for system integration. Provided LabVIEW™ drivers simplify integration into existing measurement systems.
Standard accessories shipped in the box with every 5490C Series multimeter:
| Item | Description |
|---|---|
| Power cord | Region-appropriate AC power cord |
| USB cable | USB cable for USBTMC-compliant remote control and data transfer |
| Test leads | Multimeter test leads for front-panel input terminals |
| Spare fuses | Replacement fuses for the current input |
| Test report | Factory test report documenting measurement performance |
| Certificate of calibration | Factory calibration certificate |
The bench precision multimeter is the measurement reference at the center of an engineer's bench. It is the instrument that converts unknown electrical quantities — DC voltage, AC voltage, current, resistance, capacitance, frequency, temperature — into trustworthy numbers that other instruments and analyses depend on. Within the precision DMM category, the difference between a 3½-digit handheld and a 6½-digit bench instrument is not just decimal places; it is the difference between a tool for finding problems and a tool that contributes to design records, calibration verifications, and quality documentation. The B&K Precision 5493C sits in the 6½-digit precision tier — the resolution class required for calibration verification, automated characterization, and any measurement that feeds design or quality data.
The 5490C Series consolidates twelve measurement functions into a single bench instrument with intuitive front-panel switching between modes. The function set is comprehensive enough that for most bench applications the 5490C Series replaces a stack of dedicated single-function instruments:
Conventional bench DMMs present readings as a single line of numbers — leaving the engineer to track stability, drift, and distribution manually over time. The 5490C Series replaces that with a 4.3-inch high-contrast color LCD that supports four display modes including numeric readout, bar meter with upper/lower limit setpoints and PASS/FAIL indication, trend chart showing live measurement values plotted against time, and histogram showing the distribution of measurements across the captured buffer with mean, standard deviation, and sample count overlaid. Bar meter and limit mode together create an inline production-test station: set the upper and lower limits, watch the PASS/FAIL counter, and the operator sees go/no-go status without interpreting the reading. Trend chart and histogram replace what would otherwise require exporting data to a PC and plotting in spreadsheet software.
The dual measurement feature displays two measurement results simultaneously — AC voltage with frequency, DC voltage with AC voltage, or any other compatible pair — so the engineer reads both quantities from one screen without function switching. Probe hold mode captures stable readings to an on-screen list of up to eight values, designed for situations where the engineer is moving probe tips between test points and needs a hands-free record of each captured value. The statistics display presents minimum, maximum, peak-to-peak, average, and standard deviation over the captured buffer simultaneously — over a buffer that holds up to 10,000 data samples for trend analysis, characterization sweeps, or production logging.
The 5490C Series is built for system integration as well as bench use. Every model includes standard USB (USBTMC-compliant), RS232, and LAN interfaces (the 5493CGPIB variant adds GPIB). The USB connection is USBTMC-compliant for seamless integration with VISA-based control software. A LabVIEW™ driver is provided to simplify automated test system development and instrument control. The front panel USB host port accepts USB drives for direct save and recall of instrument settings, screen capture export, and measurement-data file export — all logged with date and time stamp from the multimeter's internal real-time clock. The provided B&K Precision operating software supports remote instrument control, live measurement monitoring, and data logging from a PC.
The 5490C Series spans the full bench-measurement range across each function:
Accuracy: ± (% of reading + % of range). 24-hour specifications are typical relative to calibration standards (TCAL ±1 °C). 90-day and 1-year specifications valid for TCAL ±5 °C. Add temperature coefficient for every °C outside TCAL ±5 °C. 60-minute warm-up, 10 or 100 NPLC integration, auto zero enabled.
| Range | Resolution | Full Scale | Accuracy | Temperature Coefficient / °C | ||
|---|---|---|---|---|---|---|
| 24 Hours | 90 Days | 1 Year | ||||
| 100 mV | 0.1 µV | 119.999 mV | 0.0030 + 0.0030 | 0.0040 + 0.0035 | 0.0050 + 0.0035 | 0.0005 + 0.0005 |
| 1 V | 1 µV | 1.19999 V | 0.0020 + 0.0006 | 0.0030 + 0.0007 | 0.0040 + 0.0007 | 0.0005 + 0.0001 |
| 10 V | 10 µV | 11.9999 V | 0.0015 + 0.0004 | 0.0020 + 0.0005 | 0.0035 + 0.0005 | 0.0005 + 0.0001 |
| 100 V | 100 µV | 119.999 V | 0.0020 + 0.0006 | 0.0035 + 0.0006 | 0.0045 + 0.0006 | 0.0005 + 0.0001 |
| 1000 V | 1 mV | 1050.00 V | 0.0020 + 0.0006 | 0.0035 + 0.0010 | 0.0045 + 0.0010 | 0.0005 + 0.0001 |
20% overrange on all ranges except 1000 V (5% overrange only).
| Range | Burden Voltage | Resolution | Full Scale | Accuracy | Temp. Coef. / °C | ||
|---|---|---|---|---|---|---|---|
| 24 Hours | 90 Days | 1 Year | |||||
| 100 µA | < 0.011 V | 0.1 nA | 119.999 µA | 0.010 + 0.020 | 0.040 + 0.025 | 0.050 + 0.025 | 0.0020 + 0.0030 |
| 1 mA | < 0.11 V | 1 nA | 1.19999 mA | 0.010 + 0.006 | 0.030 + 0.006 | 0.050 + 0.006 | 0.0020 + 0.0005 |
| 10 mA | < 0.05 V | 10 nA | 11.9999 mA | 0.010 + 0.020 | 0.030 + 0.020 | 0.050 + 0.020 | 0.0020 + 0.0020 |
| 100 mA | < 0.5 V | 0.1 µA | 119.999 mA | 0.010 + 0.004 | 0.030 + 0.005 | 0.050 + 0.005 | 0.0020 + 0.0005 |
| 1 A | < 0.7 V | 1 µA | 1.19999 A | 0.050 + 0.006 | 0.080 + 0.010 | 0.100 + 0.010 | 0.0050 + 0.0010 |
| 3 A | < 2.0 V | 1 µA | 3.15000 A | 0.180 + 0.020 | 0.200 + 0.020 | 0.200 + 0.020 | 0.0050 + 0.0020 |
| 10 A | < 0.5 V | 10 µA | 11.9999 A | 0.050 + 0.010 | 0.120 + 0.010 | 0.120 + 0.010 | 0.0050 + 0.0010 |
10 A range available on front panel only.
Specifications apply to 2-wire and 4-wire resistance measurements using NULL for offset. Without NULL, add 0.2 Ω additional error for 2-wire measurements.
| Range | Test Current | Resolution | Full Scale | Accuracy | Temp. Coef. / °C | ||
|---|---|---|---|---|---|---|---|
| 24 Hours | 90 Days | 1 Year | |||||
| 10 Ω | 10 mA | 10 µΩ | 11.9999 Ω | 0.0050 + 0.0040 | 0.0080 + 0.0060 | 0.0100 + 0.0080 | 0.0006 + 0.0008 |
| 100 Ω | 10 mA | 0.1 mΩ | 119.999 Ω | 0.0030 + 0.0020 | 0.0080 + 0.0030 | 0.0100 + 0.0040 | 0.0006 + 0.0005 |
| 1 kΩ | 1 mA | 1 mΩ | 1.19999 kΩ | 0.0020 + 0.0005 | 0.0080 + 0.0010 | 0.0100 + 0.0010 | 0.0006 + 0.0001 |
| 10 kΩ | 100 µA | 10 mΩ | 11.9999 kΩ | 0.0020 + 0.0005 | 0.0080 + 0.0010 | 0.0100 + 0.0010 | 0.0006 + 0.0001 |
| 100 kΩ | 50 µA | 100 mΩ | 119.999 kΩ | 0.0020 + 0.0005 | 0.0080 + 0.0010 | 0.0100 + 0.0010 | 0.0006 + 0.0001 |
| 1 MΩ | 5 µA | 1 Ω | 1.19999 MΩ | 0.0020 + 0.0010 | 0.0080 + 0.0010 | 0.0100 + 0.0010 | 0.0010 + 0.0002 |
| 10 MΩ | 500 nA | 10 Ω | 11.9999 MΩ | 0.0150 + 0.0010 | 0.0200 + 0.0010 | 0.0400 + 0.0010 | 0.0030 + 0.0004 |
| 100 MΩ | 500 nA || 10 MΩ | 100 Ω | 119.999 MΩ | 0.3000 + 0.0100 | 0.8000 + 0.0100 | 0.8000 + 0.0100 | 0.1500 + 0.0002 |
20% overrange on all ranges. Shielded test cables are recommended for resistance measurements above 100 kΩ.
Ranges: 100 mV, 1 V, 10 V, 100 V, and 750 V. Valid for sine wave input > 0.3% of range and greater than 1 mVrms. Three filters available for low-frequency performance: 3 Hz, 20 Hz, 200 Hz.
| Range | Resolution | Full Scale Reading |
|---|---|---|
| 100 mV | 1 µV | 119.999 mV |
| 1 V | 10 µV | 1.19999 V |
| 10 V | 100 µV | 11.9999 V |
| 100 V | 1 mV | 119.999 V |
| 750 V | 10 mV | 787.50 V |
| Frequency | Accuracy | Temp. Coef. / °C | ||
|---|---|---|---|---|
| 24 Hours | 90 Days | 1 Year | ||
| 3 Hz to 5 Hz | 1.00 + 0.02 | 1.00 + 0.02 | 1.00 + 0.03 | 1.00 + 0.03 |
| 5 Hz to 10 Hz | 0.35 + 0.02 | 0.35 + 0.03 | 0.35 + 0.03 | 0.35 + 0.03 |
| 10 Hz to 20 kHz | 0.04 + 0.02 | 0.05 + 0.03 | 0.06 + 0.03 | 0.07 + 0.03 |
| 20 kHz to 50 kHz | 0.10 + 0.04 | 0.11 + 0.05 | 0.12 + 0.05 | 0.13 + 0.05 |
| 50 kHz to 100 kHz | 0.55 + 0.08 | 0.60 + 0.08 | 0.60 + 0.08 | 0.60 + 0.08 |
| 100 kHz to 300 kHz | 4.00 + 0.50 | 4.00 + 0.50 | 4.00 + 0.50 | 4.00 + 0.50 |
20% overrange on all ranges except 750 V (5% overrange only). 750 V range limited to 8 × 107 Volt-Hz.
Three filters available for low-frequency performance: 3 Hz, 20 Hz, 200 Hz. Valid for sinewave input > 1% of range and greater than 10 µA AC. 20% overrange on all ranges except 3 A (5% overrange only). 10 A range available on front panel only.
| Range & Burden | Frequency | Accuracy | Temp. Coef. / °C | ||
|---|---|---|---|---|---|
| 24 Hours | 90 Days | 1 Year | |||
| 100 µA, < 0.011 V 1 mA, < 0.11 V 10 mA, < 0.05 V 100 mA, < 0.5 V |
3 Hz to 5 Hz | 1.00 + 0.02 | 1.00 + 0.02 | 1.00 + 0.04 | 0.100 + 0.006 |
| 5 Hz to 10 Hz | 0.30 + 0.02 | 0.30 + 0.03 | 0.30 + 0.04 | 0.035 + 0.006 | |
| 10 Hz to 5 kHz | 0.10 + 0.04 | 0.10 + 0.04 | 0.10 + 0.04 | 0.015 + 0.006 | |
| 5 kHz to 10 kHz | 0.10 + 0.04 | 0.10 + 0.04 | 0.10 + 0.04 | 0.030 + 0.006 | |
| 1 A, < 0.7 V | 3 Hz to 5 Hz | 1.00 + 0.02 | 1.00 + 0.02 | 1.00 + 0.04 | 0.100 + 0.006 |
| 5 Hz to 10 Hz | 0.30 + 0.02 | 0.30 + 0.03 | 0.30 + 0.04 | 0.035 + 0.006 | |
| 10 Hz to 5 kHz | 0.10 + 0.04 | 0.10 + 0.04 | 0.10 + 0.04 | 0.015 + 0.006 | |
| 5 kHz to 10 kHz | 0.30 + 0.04 | 0.30 + 0.04 | 0.30 + 0.04 | 0.030 + 0.006 | |
| 3 A, < 2.0 V | 3 Hz to 5 Hz | 1.00 + 0.02 | 1.00 + 0.02 | 1.00 + 0.04 | 0.100 + 0.006 |
| 5 Hz to 10 Hz | 0.35 + 0.02 | 0.35 + 0.03 | 0.35 + 0.04 | 0.035 + 0.006 | |
| 10 Hz to 5 kHz | 0.23 + 0.04 | 0.23 + 0.04 | 0.23 + 0.04 | 0.015 + 0.006 | |
| 5 kHz to 10 kHz | 0.30 + 0.04 | 0.30 + 0.04 | 0.30 + 0.04 | 0.030 + 0.006 | |
| 10 A, < 0.5 V | 3 Hz to 5 Hz | 1.00 + 0.02 | 1.00 + 0.02 | 1.00 + 0.04 | 0.100 + 0.006 |
| 5 Hz to 10 Hz | 0.30 + 0.02 | 0.35 + 0.03 | 0.30 + 0.04 | 0.035 + 0.006 | |
| 10 Hz to 5 kHz | 0.15 + 0.04 | 0.15 + 0.04 | 0.15 + 0.04 | 0.015 + 0.006 | |
| 5 kHz to 10 kHz | 0.15 + 0.04 | 0.15 + 0.04 | 0.15 + 0.04 | 0.030 + 0.006 | |
| Range | Full Scale | Accuracy | Temp. Coef. / °C | ||
|---|---|---|---|---|---|
| 24 Hours | 90 Days | 1 Year | |||
| 1 nF | 1.19999 nF | 0.5 + 0.10 | 0.5 + 0.40 | 1.0 + 0.50 | 0.02 + 0.001 |
| 10 nF | 11.9999 nF | 0.2 + 0.05 | 0.5 + 0.10 | 1.0 + 0.10 | 0.02 + 0.001 |
| 100 nF | 119.999 nF | 0.2 + 0.05 | 0.5 + 0.10 | 1.0 + 0.10 | 0.02 + 0.001 |
| 1 µF | 1.19999 µF | 0.2 + 0.05 | 0.5 + 0.05 | 1.0 + 0.10 | 0.02 + 0.001 |
| 10 µF | 11.9999 µF | 0.2 + 0.05 | 0.5 + 0.05 | 1.0 + 0.10 | 0.02 + 0.001 |
| 100 µF | 119.999 µF | 0.2 + 0.05 | 0.5 + 0.05 | 1.0 + 0.10 | 0.02 + 0.001 |
| 1 mF | 1.19999 mF | 0.2 + 0.05 | 0.5 + 0.05 | 1.0 + 0.20 | 0.02 + 0.001 |
| 10 mF | 11.9999 mF | 0.5 + 0.20 | 0.5 + 0.30 | 1.0 + 0.50 | 0.02 + 0.001 |
20% overrange on all ranges.
Applies to sine wave input and gate time set to 1 second. For sine and square wave input larger than 100 mV; for 10 mV to 100 mV input, multiply % reading error by 10. Square input specified for 10 Hz to 300 kHz.
| Frequency Range | Accuracy ± (% of reading) | Temp. Coef. / °C | ||
|---|---|---|---|---|
| 24 Hours | 90 Days | 1 Year | ||
| 2 Hz to 10 Hz | 0.1 | 0.100 | 0.100 | 0.0002 |
| 10 Hz to 100 Hz | 0.030 | 0.030 | 0.030 | 0.0002 |
| 100 Hz to 1 kHz | 0.003 | 0.008 | 0.010 | 0.0002 |
| 1 kHz to 300 kHz | 0.002 | 0.006 | 0.010 | 0.0002 |
| 300 kHz to 1 MHz | 0.002 | 0.006 | 0.010 | 0.0002 |
| Square wave | 0.001 | 0.006 | 0.010 | 0.0002 |
Additional gate time errors ± (% of reading):
| Frequency | 1 second | 0.1 second | 0.01 second |
|---|---|---|---|
| 3 Hz to 40 Hz | 0 | 0.200 | 0.200 |
| 40 Hz to 100 Hz | 0 | 0.060 | 0.200 |
| 100 Hz to 1 kHz | 0 | 0.020 | 0.200 |
| 1 Hz to 300 kHz | 0 | 0.004 | 0.030 |
| Square wave | 0 | 0 | 0 |
| Function | Range | Test Current | Accuracy | Temp. Coef. / °C | ||
|---|---|---|---|---|---|---|
| 24 Hours | 90 Days | 1 Year | ||||
| Continuity | 1 kΩ | 1 mA | 0.002 + 0.030 | 0.008 + 0.030 | 0.010 + 0.030 | 0.0010 + 0.0020 |
| Diode | 5 V | 1 mA | 0.002 + 0.030 | 0.008 + 0.030 | 0.010 + 0.030 | 0.0010 + 0.0020 |
| Temperature Sensor Type | Accuracy |
|---|---|
| Pt100 (DIN / IEC 751) | Probe accuracy + 0.05 °C |
| 5 kΩ thermistor | Probe accuracy + 0.10 °C |
| AC Input | 110 V / 220 V ±10%, 50/60 Hz |
| Power Consumption | ≤ 30 VA |
| I/O Interfaces | RS232, USB host, USB (USBTMC-compliant) device, LAN |
| Operating Temperature | 32 °F to 104 °F (0 °C to 40 °C) |
| Storage Temperature | −40 °F to 158 °F (−40 °C to 70 °C) |
| Humidity | Indoor use, ≤ 95% |
| Safety | EN 6110-1:2010, Low Voltage Directive (LVD) 2014/35/EU |
| Electromagnetic Compatibility | EN 61326-1:2013, EMC Directive 2014/30/EU |
| Dimensions (W × H × D) | 8.85" × 3.93" × 13.97" (225 mm × 100 mm × 355 mm) |
| Weight | 5.51 lbs (2.5 kg) |
| Input Terminals | Front and rear (rear terminals on 6½-digit models for rack use) |
| Warranty | 3 years (manufacturer) |
| Quality Management | B&K Precision Corporation is ISO9001:2015 registered (Certification body NSF-ISR, certificate number 6Z241-IS8) |
Specifications are based on a one-year calibration cycle and a calibration temperature TCAL of 23 °C. Accuracy specifications are expressed as ± (% of reading + % of range), valid for TCAL ±5 °C. For DC measurements, technical specifications are valid after a 60-minute warm-up with integration time set to 10 or 100 NPLC and auto zero enabled. For AC measurements, technical specifications are valid after a 60-minute warm-up with slow AC filtering enabled and a sine wave input signal.
This webpage was written by a human with an A.I. "Intern", which may contain errors. Please review the Manufacturers Data Sheet to verify published specifications. Feedback on this webpage is always welcome, please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.
Model No
5493C
Condition
New
Manufacturer
B&K Precision
Ac voltage
1 kV
Dc voltage
750 V
Resolution
6.5
Type
Bench