Key Features At A Glance

  • Test frequency range of 1 MHz to 3 GHz with 100 kHz resolution
  • Basic measurement accuracy of ±1.0% across the impedance measurement range
  • Measurement speed of 9 msec per point at maximum speed for high-throughput production testing
  • List-sweep capability with up to 32 points per sweep for multi-frequency characterization
  • RF I-V measurement method delivers accurate impedance measurement over a wider impedance range than network analyzers
  • GPIB, LAN, and opto-isolated handler interfaces standard for automated component handler integration
  • Built-in Rdc (DC resistance) measurement function for contact check during automated handler testing
  • Multi-function comparator with 13 bins and 4 limit values per bin for production sorting
Share on
View full details

The Agilent 4287A is a high performance RF LCR meter designed for production line testing of devices such as SMD inductors and EMI filters, where impedance testing at high frequencies is required. The instrument covers a test frequency range of 1 MHz to 3 GHz with 100 kHz resolution and offers measurement parameters including |Z|, θz, |Y|, θy, X, G, B, Ls, Lp, Cs, Cp, Rs, Rp, Q, and D with five-digit display resolution. The 4287A uses the RF I-V measurement method, which measures the current flowing through the device under test and the voltage applied across it, enabling accurate characterization over the full 1 MHz to 3 GHz range.

The 4287A is built for production line testing of RF SMD inductors and EMI filters, research and development of next-generation passive components, and reliable quality control of wireless communication equipment components. The instrument supports automated component handler integration through its opto-isolated handler interface, and the small test head with 1 m test cable extends the measurement plane to the handler stage without introducing additional error. The multi-function comparator with 13 bins and four limit values per bin supports bin sorting workflows where conditions such as frequency and measurement parameters can be set independently in each column.

Brand Heritage

Agilent Technologies was spun out of Hewlett-Packard in 1999 when HP separated its test and measurement business from its computing business. In 2014, Agilent split again, with the electronic measurement instruments business — including the 4287A product line — transferring to Keysight Technologies. Instruments carrying the Agilent brand on their front panels are part of the same engineering lineage as today's Keysight Technologies products.

Accessories Supplied

  • Test head with 1 m test cable
  • N (m) - SMA (f) Adapter
  • Wrench for 3.5/SMA connector
  • CD-ROM (Operation manual, Programming manual and Sample Program)
  • Power cord

Product Core & Specifications

Parameter Specification
Test frequency 1 MHz - 3 GHz with 100 kHz resolution. With list-sweep, up to 32 points per sweep is available.
Impedance parameters |Z|, θz (rad), θz (deg), |Y|, θy (rad), θy (deg), X, G, B, Ls, Lp, Cs, Cp, Rs, Rp, Q, D
Display resolutions 5 digits
Test signal level V (open condition): 4.47 mVrms - 502 mVrms (447 mVrms when >1 GHz); I (short condition): 0.0894 mArms - 10 mArms (8.94 mArms when >1 GHz)
Basic accuracy ± 1.0%
Measurement range 200 mΩ to 3 kΩ (@1 MHz, accuracy ≤10%)
Measurement time 9 msec per point (max. speed)
Measurement terminal 3.5 mm (female)
Calibration and compensation Open/short/load/low-loss capacitor calibration, fixture electrical length compensation, open/short compensation
Rdc measurement function For contact check (on/off selectable)
Data storage devices About 18 Gbyte internal hard disk and 1.44 Mbyte floppy disk
Interface GPIB, LAN (10base-T/100base-TX automatically switched), and Opto-isolated handler interface
Display 8.4 inch color LCD
List sweep points Up to 256 multiple frequency points (32 points max./table × 8 tables max.)
Comparator bins 13 bins, with four limit values for each bin
Important: Maximum test signal level is 502 mVrms (447 mVrms above 1 GHz) in the open condition and 10 mArms (8.94 mArms above 1 GHz) in the short condition. Measurement accuracy is specified at the test head.Important: A keyboard and a mouse are required for initial setup of the 4287A (available as options 4287A-810 keyboard and 4287A-820 mouse).Important: Test fixture is not included with the base unit; an appropriate 7-mm test fixture (e.g., 16196A/B/C/D, 16197A, 16192A, or 16194A) must be ordered separately based on the DUT size, shape, and application.Important: Measurement accuracy is specified at the test head. When using a test fixture, perform open/short/load calibration with a working load standard at the measurement plane for accurate results.Important: Manuals are not furnished as standard for the SMD test fixture accessories. For 16196A/B/C/D, magnifying lens and tweezers are not furnished as standard.Recommended pairing: use the 16196A, 16196B, 16196C, or 16196D SMD test fixture matched to your chip size (1608 through 0402 mm) for measurements to 3 GHz, or the 16200B external DC bias adapter when applying DC bias to EMI filters up to 1 GHz.

About this used unit

  • Warranty included
  • Functional verification included
  • Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
  • Note that unnecessary accessories may not be included (contact Test Architect to confirm).

Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.

Model No

Agilent

Condition

Used

Manufacturer

Agilent

Frequency

3 GHz