Skip to product information
1 of 2
Used
lab tested

Agilent 34410A 6½-Digit High-Performance Digital Multimeter (Pre-Owned)

Agilent 34410A 6½-Digit High-Performance Digital Multimeter (Pre-Owned)

Regular price $945
Sale price $945 Regular price $0
In Stock
Shipping calculated at checkout.
Calibrations
Request a Quote
Estimated Delivery in
5 - 7
business days (based on availability)

Request a Quote To Get An EDU Discount

Login & Get 2% Back on Purchases
Login or Sign Up for Loyalty

Agilent 34410A 6½-Digit High-Performance Digital Multimeter (Pre-Owned)

Key Features At A Glance

  • 6½-digit resolution with 30 ppm 1-year basic DC accuracy
  • 10,000 readings/s at 5½ digits direct to PC; 1,000 readings/s at 6½ digits
  • LAN, USB, and GPIB interfaces standard on every unit
  • Measurement functions span DCV, ACV, DCI, ACI, 2- and 4-wire resistance, frequency, period, capacitance, temperature, continuity, and diode test
  • DC and AC current ranges extend down to 100 µA for 100 pA resolution
  • Built-in data logger with 50,000-reading non-volatile memory
  • Dual display supports simultaneous dual measurements and simplified setup
  • LXI Class C compliant for LAN-based system integration
Share on
View full details

The Agilent 34410A and 34411A are 6½-digit digital multimeters positioned as the next generation of benchtop and system multimeters from Agilent Technologies. Building on the installed base of the Agilent 34401A, these meters add improved accuracy, expanded measurement capability, faster reading speed and throughput, and modern computer interfaces. Each unit ships with a dual display that supports both dual measurement viewing and easier setup, along with LAN, USB, and GPIB connectivity standard.

An oscilloscope shows voltage over time; a digital multimeter measures discrete electrical quantities — voltage, current, resistance — with the accuracy and stability that engineering decisions and calibration work depend on. The 34410A and 34411A serve both roles a 6½-digit DMM is typically asked to fill: a precision benchtop instrument for design and characterization work, and a fast system DMM for automated test environments. The datasheet positions these meters for both bench and system use, with the 34411A adding higher-speed acquisition for waveform-style measurement applications.

ValueTronics International has stocked, tested, and shipped professional test and measurement instruments since 1992 from our 20,000 sq ft secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. Every used instrument we list passes through technician inspection and performance verification in that facility before it is graded and made available — which is what allows us to publish specific condition reports and back our pre-owned units with a ValueTronics warranty rather than the seller-description approach typical of marketplace listings.

Brand Heritage

Hewlett-Packard's test and measurement business was spun off as Agilent Technologies in 1999, and Agilent's electronic measurement division was subsequently transferred to Keysight Technologies in 2014. Instruments originally released under the Agilent brand — including the 34410A and 34411A multimeter family — trace their support, calibration, and software ecosystem through this lineage to the current Keysight organization.

Compare Other Models in This Series

The 34410A and 34411A share a common platform: 6½-digit resolution, 30 ppm 1-year basic DC accuracy, the same measurement function set, the same dual display, and the same standard LAN, USB, and GPIB interfaces. Both are LXI Class C compliant and both ship with the same accessory bundle and software set.

The difference between the two models is acquisition speed, memory depth, and triggering. The 34410A is the high-performance model intended for benchtop precision work and standard system use. The 34411A is the enhanced-performance model that adds the higher reading rate, expanded volatile memory, analog level triggering, and pre- and post-trigger control needed for waveform-style capture and higher-throughput characterization.

Each pre-owned model below is its own dedicated product page with condition-matched pricing and availability. Use the comparison below to identify which model matches the measurement requirement, then click through to the model page for the configuration details, accessories, and current options.

The 34410A delivers the speed envelope most benchtop and standard system DMM applications need: 10,000 readings per second at 5½ digits direct to a host PC, 1,000 readings per second at full 6½-digit resolution, and 50,000 readings of non-volatile data logger memory. Trigger latency and trigger jitter are each below 1 µs and bus query response is under 500 µs, which the datasheet identifies as the baseline that makes the meter suitable for fast system throughput applications.

The 34411A extends the 34410A's reading rate to 50,000 readings per second at 4½ digits direct to PC. It adds 1 million readings of volatile memory on top of the 50,000-reading non-volatile data logger memory shared with the 34410A. Triggering on the 34411A is programmable pre- and post-, with analog level triggering and a 19.9524 µs timebase resolution that the 34410A does not provide.

For applications that need to capture low-frequency waveforms, characterize device performance with high sample density, or stream large reading sets to a host computer for analysis, the datasheet positions the 34411A as the right model. For applications where 6½-digit accuracy and 10,000 readings per second at 5½ digits are sufficient — which describes most precision bench and standard system test work — the 34410A is the model the datasheet positions for that use.

Model Digits Max Reading Rate Volatile Memory
34410A 10,000 readings/s @ 5½ digits 50,000 readings
34411A 50,000 readings/s @ 4½ digits 1,000,000 readings

Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.

Accessories Supplied

  • Test Lead Kit with probes and SMT attachments
  • Test report
  • Power cord
  • USB interface cable
  • Product Reference CD-ROM with soft documentation and software

Product Core & Specifications

Category Specification Value
DC Voltage Accuracy (1 Year, % of reading + % of range) 100 mV range 0.0050 + 0.0035
1 V range 0.0035 + 0.0007
10 V range 0.0030 + 0.0005
100 V range 0.0040 + 0.0006
1000 V range 0.0040 + 0.0006
True RMS AC Voltage Accuracy (1 Year, 100 mV to 750 V) 3 Hz – 5 Hz 0.50 + 0.03
5 Hz – 10 Hz 0.10 + 0.03
10 Hz – 20 kHz 0.06 + 0.03
20 kHz – 50 kHz 0.10 + 0.05
50 kHz – 100 kHz 0.40 + 0.08
100 kHz – 300 kHz 1.20 + 0.50
Resistance Accuracy (1 Year, % of reading + % of range) 100 Ω (1 mA) 0.010 + 0.004
1 kΩ (1 mA) 0.010 + 0.001
10 kΩ (100 µA) 0.010 + 0.001
100 kΩ (10 µA) 0.010 + 0.001
1 MΩ (5 µA) 0.012 + 0.001
10 MΩ (500 nA) 0.040 + 0.001
100 MΩ (500 nA || 10 MΩ) 0.800 + 0.001
1 GΩ (500 nA || 10 MΩ) 8.000 + 0.001
DC Current Accuracy (1 Year, % of reading + % of range) 100 µA (<0.03 V) 0.050 + 0.025
1 mA (<0.3 V) 0.050 + 0.006
10 mA (<0.03 V) 0.050 + 0.020
100 mA (<0.3 V) 0.050 + 0.005
1 A (<0.8 V) 0.100 + 0.010
3 A (<2.0 V) 0.150 + 0.020
True RMS AC Current Accuracy (1 Year, 100 µA to 3 A) 3 Hz – 5 kHz 0.10 + 0.04
5 kHz – 10 kHz 0.20 + 0.04
Capacitance Accuracy (1 Year) 1 nF (500 nA) 0.50 + 0.50
10 nF (1 µA) 0.40 + 0.10
100 nF (10 µA) 0.40 + 0.10
1 µF (10 µA) 0.40 + 0.10
10 µF (100 µA) 0.40 + 0.10
Temperature Accuracy (1 Year) RTD (-200°C to 600°C) 0.06°C
Thermistor (-80°C to 150°C) 0.08°C
Frequency / Period (40 Hz – 300 kHz, 1 Year) 100 mV to 750 V 0.007 + 0.000
Continuity 1000.0 Ω range, 1 mA test current 0.010 + 0.020 (1 Year)
Diode Test 1.0000 V range, 1 mA test current 0.010 + 0.020 (1 Year)
DC Voltage Input Resistance 0.1 V, 1 V, 10 V ranges 10 MΩ or >10 GΩ (selectable)
DC Voltage Input Resistance 100 V, 1000 V ranges 10 MΩ ±1% (fixed)
DC Voltage Input Bias Current At 25°C < 50 pA
DC CMRR - 140 dB
AC Voltage Maximum Input - 400 Vdc, 1100 Vpk
AC Voltage Input Impedance - 1 MΩ ± 2% in parallel with <150 pF
AC Voltage Input Protection All ranges 750 Vrms
AC CMR - 70 dB
DC Current Input Protection - 3 A, 250 V fuse
AC Current Input Protection - 3 A, 250 V fuse
Continuity Response Time - 300 samples/sec with audible tone
Continuity Threshold - Fixed at 10 Ω
Temperature Sensors Supported Thermistor 2.2 kΩ, 5 kΩ, and 10 kΩ
Temperature Sensors Supported RTD α = 0.00385, Ro from 49 Ω to 2.1 kΩ
External Trigger Delay - < 1 µs
External Trigger Max Rate - 5,000/s
External Trigger Jitter - < 1 µs
External Trigger Min Pulsewidth - 1 µs
Voltmeter Complete Output - 3 V Logic output, 2 µs pulse with programmable edge
Nonvolatile Memory Both models 50,000 readings
Volatile Memory 34410A 50,000 readings
Volatile Memory 34411A 1,000,000 readings
Samples per Trigger 34410A 1 to 50,000
Samples per Trigger 34411A 1 to 1,000,000
Trigger Delay - 0 to 3600 s (20 µs step size)
Sample Timer Range - 0 to 3600 s (20 µs step sizes)
Sample Timer Jitter - < 100 ns
Important: Measurement category ratings: CAT II 300 V, CAT I 1000 V. Maximum input on the AC voltage function is 400 Vdc, 1100 Vpk, with 750 Vrms input protection on all ranges. Current input is protected by a 3 A, 250 V fuse.Important: Standard accuracy specifications are for 90 minute warm-up and 100 PLC integration time; capacitance specifications require 1-hour warm-up using Math Null.

About this used unit

  • Warranty included
  • Functional verification included
  • Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
  • Note that unnecessary accessories may not be included (contact Test Architect to confirm).

Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.

Model No

Agilent

Condition

Used

Manufacturer

Agilent

Ac voltage

750 V

Dc voltage

1 kV

Resolution

6.5

Type

Bench