The HP 3437A System Voltmeter is a microprocessor-controlled, 3½-digit successive approximation voltmeter designed for system-level data acquisition applications requiring high sampling rates. The instrument samples voltages at rates up to 5,700 samples per second and provides chassis-isolated input terminals, a wideband input amplifier, auto-zero, auto-polarity, sample and hold, and 100% overrange on each of three voltage ranges (0.1 volt, 1 volt, and 10 volt). Floating measurement capability (±20 V) is supported over the frequency range of DC through 1.0 MHz on the 10V range.
Applications listed for the 3437A include fast multipoint data acquisition, repetitive-waveform analysis, low-frequency transient characterization, and low-frequency true RMS measurements. Additional system measurement applications cover waveform analysis (harmonic content, amplitude characteristics) and transient characterization. The instrument is positioned for bench measurements such as variable sample-rate DVM operation, oscilloscope accuracy enhancement (time-selective DVM), delay generation, and burst trigger generation.
ValueTronics International tests, stocks, and ships every unit out of our 20,000 sq ft secure warehouse at 1675 Cambridge Drive, Elgin, Illinois. Operating since 1992 and serving over 35,000 customers with more than 8,000 units in stock across 28 authorized new equipment partners and over 90 legacy brands, we provide a level of inventory depth and verification rigor that drop-shippers and pass-through brokers cannot match. Each instrument is professionally evaluated before listing — not merely forwarded from a third-party warehouse.
The 3437A was originally manufactured by Hewlett-Packard. HP's test and measurement business was spun out as Agilent Technologies in 1999, and Agilent's electronic measurement business subsequently became Keysight Technologies in 2014. Calibration support and service documentation for legacy HP test instruments routes through the Keysight service network where it remains available.
| Voltage Measurement Characteristics | |
|---|---|
| Range | 0.1 V, 1 V, 10 V |
| Bandwidth (3 dB) - 10 V Range | 1.0 MHz |
| Bandwidth (3 dB) - 1 V Range | 1.1 MHz |
| Bandwidth (3 dB) - 0.1 V Range | 40 kHz |
| Display - 10 V Range | ±19.98 (max), ±99.99 (ovrld) |
| Display - 1 V Range | ±1.998, ±9.999 |
| Display - 0.1 V Range | ±.1998, ±.9999 |
| Static Accuracy (90 days, 23°C ±5°C) - 10 V Range | ±0.05% of Reading ±1.6 Digits |
| Static Accuracy (90 days) - 1 V Range | ±0.03% of Reading ±1.6 Digits |
| Static Accuracy (90 days) - 0.1 V Range | ±0.06% of Reading ±1.8 Digits |
| Static Accuracy (1 year, 23°C ±5°C) - 10 V Range | ±0.05% of Reading ±2.0 Digits |
| Static Accuracy (1 year) - 1 V Range | ±0.03% of Reading ±2.0 Digits |
| Static Accuracy (1 year) - 0.1 V Range | ±0.06% of Reading ±2.2 Digits |
| Static Accuracy Temperature Coefficient (0°C to 50°C) | ±0.002% reading/°C ±0.05 digits/°C |
| Dynamic Accuracy - 10V Range, 10V Step | ±200 mV within 700 ns; ±30 mV within 7.5 µs |
| Dynamic Accuracy - 1V Range, 1V Step | ±20 mV within 700 ns; ±3 mV within 1.5 µs |
| Dynamic Accuracy - 0.1V Range, 0.1V Step | ±200 µV within 25 µs |
| Delay Characteristics | |
| Delay (NRDGS = 0 or 1) | 0 to 0.9999999 sec in 100 ns steps |
| Delay (NRDGS > 1, ASCII) | 277.8 µs to 0.9999999 sec |
| Delay (NRDGS > 1, Packed) | 175.4 µs to 0.9999999 sec |
| Offset (actual delay with 0 delay programmed) | 100 ns ±25 ns |
| Delay Accuracy | ±0.008% Delay + Delay offset |
| Repeatability (Jitter) - Delay 0 or 100 ns | 2 ns |
| Repeatability (Jitter) - Delay 200 ns to 50 ms | 10 ns + 0.002% of Delay |
| Repeatability (Jitter) - Delay > 50 ms | 110 ns |
| Number of Readings (per trigger) | 0 to 9999 |
| Input Characteristics | |
| Input Impedance - 10 V Range | 1 MΩ (±20%), <75 pF |
| Input Impedance - 1 V Range | >108 Ω, <75 pF |
| Input Impedance - 0.1 V Range | >108 Ω, <75 pF |
| Maximum Input Voltage HI to LO | <±30 V Peak |
| Maximum Input Voltage LO to CHASSIS | <±42 V Peak |
| Common Mode Rejection Ratio | ≥75 dB (1 kΩ unbalance in low input lead at 60 Hz) |
| Programmability | |
| Interface | HP-IB (IEEE 488-1975) |
| HP-IB Capability | AH1, C0, DC1, DT1, L4, PP0, RL1, SR1, SH1, T5 |
| Maximum Reading Rate | |
| ASCII Format | 3600 Readings per second |
| Packed Format | 5700 Readings per second |
| General | |
| Power Requirements | 100 V, 120 V, 220 V, 240 V (+5% -10%, 48-440 Hz), ≤42 VA |
| Operating Temperature | 0°C to 50°C |
| Storage Temperature | -40°C to +75°C |
| Humidity Range | <95% RH (0°C to 40°C) |
| Dimensions | 212.7 mm wide × 88.9 mm high × 527.1 mm deep |
| Net Weight | 5.6 kg |
| Shipping Weight | 7.6 kg |
Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.