Skip to content
In summary, the GW Instek GPT-9503 is a 150VA multi-channel hipot tester combining AC withstanding (5kVac/30mA), DC withstanding (6kVdc/10mA), insulation resistance (50V–1000V dc, up to 10GΩ), and continuity testing — with a built-in 8-channel scanner that lets a single instrument apply the test signal to any of eight points on the DUT, eliminating the external switching matrix that was historically required for multi-point safety testing. Channel status options: H (high-voltage hot) or X (no-used). The instrument is built on a high-efficiency PWM amplifier delivering 150VA AC and 50W DC output, controlled from a 4.3" 480×272 color TFT LCD. Test Parameter Export/Import through USB Host, LCD hardcopy, ARC Detection, Open/Short Check (OSC), and a Statistics (counter) function are all standard. Multi-language operation supports Traditional Chinese, Simplified Chinese, and English. Interface coverage includes RS-232C, USB Host/Device, and Signal I/O standard.
The GPT-9503 is used by production-line manufacturers and contract assemblers to safety-test products with up to eight separate test points from a single instrument — replacing the external switching matrix that used to be required for multi-channel hipot. The integrated 8-channel scanner routes the AC/DC withstanding or insulation resistance test signal through any selected channel; each channel can be set to H (high-voltage hot) or X (no-used). This is the right model when every test point needs to be exercised against a common ground or return reference — typical for power supply isolation testing, switch-mode supply secondary-side verification, and multi-output product safety screening. Applications cover the safety test requirements of IEC, EN, UL, CSA, GB, JIS, and other safety regulations.
A conventional single-channel hipot tester applies the high-voltage test signal between one fixed HV terminal and one fixed return terminal. To test a product with multiple insulation paths or multiple test points, the operator either has to physically rewire the DUT between every test step (slow, error-prone) or integrate an external scanner/switching matrix (expensive, complex). The GPT-9503 puts the scanner inside the tester: an 8-channel scanner is built into the chassis, channel status routes as H or X, and the test sequence — including channel routing — is stored and executed from internal memory. Open/Short Check (OSC) verifies cable connection integrity before applying high voltage. ARC Detection catches incipient breakdown that might not yet trip the current limit. The Statistics (counter) function tracks Pass/Fail counts on the production line without an external PC.
The GPT-9503 delivers AC withstanding output from 0.050kV to 5.000kV ac in 1V steps with ±(1% of setting + 5V) no-load accuracy and 150VA (5kV/30mA) maximum load. DC withstanding ranges from 0.050kV to 6.000kV with 50W (5kV/10mA) maximum load. Insulation resistance testing operates from 0.050kV to 1.000kV dc over a 0.1MΩ to 10GΩ measurement range. Continuity testing applies a fixed 100mA dc current with ±(10% of reading + 2Ω) ohmmeter accuracy. AC current measurement accuracy is ±(1.5% of reading + 50μA) over the 0.001mA–30.00mA range with 1μA best resolution. The built-in 8-channel scanner with H or X channel status routing is standard. Timer accuracy across all functions is (100ppm + 20ms). The instrument is designed to meet the test requirements of IEC, EN, UL, CSA, GB, JIS, and other safety regulations.
GW Instek (Good Will Instrument Co., Ltd.) is a Taiwanese manufacturer of test and measurement instruments founded in 1975. Electrical safety testing — hipot, insulation resistance, ground bond, and continuity — is the non-discretionary verification step required for any product that connects to mains power before it can be sold under safety marks including UL, CSA, IEC 62368 (audio/video and IT equipment), IEC 60598 (luminaires), IEC 60065 (audio/video apparatus), and IEC 60335 (household appliances). For products with multiple isolation barriers or multiple test points, a multi-channel hipot tester replaces the external switching matrix that historically had to be integrated separately. The GPT-9500 Series brings that scanner inside the tester at a price point well below the integrated multi-channel platforms from other safety-test manufacturers.
The GPT-9500 Series is built around one core idea: when a production-line product has more than one safety-test point, you no longer need an external switching matrix. The built-in 8-channel scanner routes the hipot, insulation resistance, or continuity test signal through any of eight scanner channels on the DUT — and the test recipe (including channel routing) is stored on the instrument.
The GPT-9500 Series is a 3-in-1 tester providing AC withstanding, DC withstanding, and insulation resistance — plus continuity testing — at 150VA AC test capacity. The 4.3" 480 × 272 Color TFT LCD shows test conditions, channel state, measurement results, and Pass/Fail status on one screen. Open/Short Check (OSC) verifies cable connection before applying high voltage. ARC Detection catches incipient insulation breakdown that might not trip the current limit. The Statistics (counter) function tracks production yield on the unit. Test Parameter Export/Import through the front-panel USB Host means a parameter set written on one production line can be copied to identical units on every other line. Multi-language support covers Traditional Chinese, Simplified Chinese, and English. Interfaces include RS-232C, USB Host/Device, and Signal I/O — and the scanner appears as the Rear Output.
The specifications apply when the GPT-9500 is powered on for at least 30 minutes under +15°C ~ +35°C.
| Output-Voltage Range | 0.050kV ~ 5.000kV |
| Output-Voltage Resolution | 1V |
| Output-Voltage Accuracy | ±(1% of setting + 5V) [no load] |
| Maximum Rated Load | 150 VA (5kV/30mA) |
| Maximum Rated Current | 30mA ; 0.001mA ~ 10mA (0.05kV ≤ V ≤ 0.5kV) ; 0.001mA ~ 30mA (0.5kV < V ≤ 5kV) |
| Output-Voltage Waveform | Sine wave |
| Output-Voltage Frequency | 50 Hz / 60 Hz selectable |
| Voltage Regulation | ±(1% + 5V) [maximum rated load → no load] |
| Voltmeter Accuracy | ±(1% of reading + 5V) |
| Current Measurement Range | 0.001mA ~ 30.00mA |
| Current Best Resolution | 1μA (0.001mA ~ 9.999mA) ; 10μA (10.00mA ~ 30.00mA) |
| Current Measurement Accuracy | ±(1.5% of reading + 50μA) |
| Current Offset | 80μA maximum |
| ARC Detect | Yes |
| RAMP TIME (Rise Time) | 0.1s ~ 999.9s |
| FALL Time | OFF ~ 999.9s |
| WAIT Time | OFF ~ 999.9s |
| TIMER (Test Time) | CONT, 0.3s ~ 999.9s |
| TIMER Accuracy | (100ppm + 20ms) |
| GND | ON/OFF |
| Output-Voltage Range | 0.050kV ~ 6.000kV |
| Output-Voltage Resolution | 1V |
| Output-Voltage Accuracy | ±(1% of setting + 5V) [no load] |
| Maximum Rated Load | 50W (5kV/10mA) |
| Maximum Rated Current | 10mA ; 0.001mA ~ 2mA (0.05kV ≤ V ≤ 0.5kV) ; 0.001mA ~ 10mA (0.5kV < V ≤ 6kV) |
| Voltage Regulation | ±(1% + 5V) [maximum rated load → no load] |
| Voltmeter Accuracy | ±(1% of reading + 5V) |
| Current Measurement Range | 0.001mA ~ 10.00mA |
| Current Best Resolution | 0.1μA (0.1μA ~ 999.9μA) ; 1μA (1μA ~ 9.999mA) ; 10μA (10.00mA) |
| Current Measurement Accuracy | ±(1% of reading + 1μA) when I Reading < 1mA ; ±(1% of reading + 10μA) when I Reading ≥ 1mA |
| Current Offset | 5μA maximum |
| ARC Detect | Yes |
| RAMP TIME (Rise Time) | 0.1s ~ 999.9s |
| FALL Time | OFF ~ 999.9s |
| WAIT Time | OFF ~ 999.9s |
| TIMER (Test Time) | CONT, 0.3s ~ 999.9s |
| TIMER Accuracy | (100ppm + 20ms) |
| GND | ON/OFF |
| Output Voltage | 0.050kV ~ 1.000kV dc |
| Output-Voltage Resolution | 1V |
| Output-Voltage Accuracy | ±(1% of setting + 5V) [no load] |
| Resistance Measurement | 0.1MΩ ~ 10GΩ |
| Voltage Regulation | ±(1% + 5V) [maximum rated load → no load] |
| Voltmeter Accuracy | ±(1% of reading + 5V) |
| Short-Circuit Current | 10mA max. |
| Output Impedance | 2kΩ |
| RAMP TIME (Rise Time) | 0.1s ~ 999.9s |
| FALL Time | OFF ~ 999.9s |
| WAIT TIME | OFF ~ 999.9s |
| TIMER (Test Time) | 0.3s ~ 999.9s |
| TIMER Accuracy | (100ppm + 20ms) |
| GND | ON/OFF |
| Test Voltage | Measurement Range / Accuracy |
|---|---|
| 50V ≤ V < 500V |
0.1MΩ ~ 10MΩ : ±(5% of reading + 3% fs); 10.1MΩ ~ 50MΩ : ±(5% of reading + 1% fs); 50.1MΩ ~ 2GΩ : ±(10% of reading + 1% fs) |
| 500V ≤ V ≤ 1000V |
0.1MΩ ~ 10MΩ : ±(5% of reading + 3% fs); 10.1MΩ ~ 500MΩ : ±(5% of reading + 1% fs); 500.1MΩ ~ 10GΩ : ±(10% of reading + 1% fs) |
| Output-Current | 100mA dc |
| Ohmmeter Measurement Accuracy | ±(10% of reading + 2Ω), ON/OFF |
| Signal I/O | Standard |
| RS-232C | Standard |
| USB (Device) | Standard |
| USB (Host) | Standard (for Parameter/LCD Hardcopy) |
| Rear Output | Scanner |
| Display | 4.3" Color LCD (480 × 272) |
| Power Source | AC 100V ~ 240V ±10%, 50Hz/60Hz |
| Power Consumption | 400VA Max. |
| Dimensions | 320(W) x 120(H) x 435(D) mm |
| Weight | Approx. 11kg |
Specifications subject to change without notice. PT-9500GD2DS
Model No
GPT-9503
Condition
New
Manufacturer
Instek
Request A Quote