Skip to content

The Stanford Research Systems SR630 is a 16-channel thermocouple monitor designed to read, scan, print and log temperatures or voltages. Using any one of seven standard thermocouple types, it reads temperatures from −200 °C to +1700 °C and displays results in °C, °F, K, mV or V on the front panel. Each of the 16 channels can be set independently to monitor either temperature or voltage, giving the instrument broad configuration flexibility for multi-point monitoring tasks.
The SR630 is built for remote monitoring applications, where it can time-stamp and store up to 2,000 readings in non-volatile memory for later analysis. Its isolated differential inputs and 250 V breakdown level let it tackle difficult applications such as temperature profiling of electrically live equipment. The four analog outputs can drive chart recorders or provide a feedback signal for proportional temperature control systems, and the integrated 1:15 analog multiplexer makes the unit useful in ATE systems and many other monitoring applications.
ValueTronics supplies the SR630 from our own 20,000 square-foot secure warehouse at 1675 Cambridge Drive in Elgin, Illinois, where we stock and hold our inventory rather than sourcing to order. Every pre-owned unit is supported by that in-house operation: pre-owned units are inspected and functionally verified by our technicians before they ship, while new units ship factory-sealed exactly as received from the manufacturer.
| Specification | Value |
|---|---|
| Thermocouple | |
| Channels | 16 |
| Thermocouple types | B, E, J, K, R, S, T |
| Display units | °C, °F and K |
| Display resolution | 0.1 °C |
| Temperature displays | Actual, Nominal, or Offset |
| Accuracy | 0.5 °C for J, K, E and T; 1.0 °C for R, S and B. Errors are for the SR630 only. Standard errors for thermocouple wire are 2 to 5 times the error due to the SR630. |
| Inputs | |
| Channels | 16 |
| Input type | Independent, floating and differential |
| Input resistance | 10 MΩ between + and − terminals, >1 GΩ to ground |
| Input capacitance | 0.001 µF |
| Input bias current | <100 pA |
| Input protection | 250 Vrms |
| Open circuit check | 250 µA |
| Common mode | ±200 VDC |
| Voltmeter | |
| Full-scale display | (±9.999, ±99.99 or ±999.9) mVDC, (±9.999 or ±99.99) VDC |
| Range select | Automatic |
| Resolution | ±1 of least significant displayed digit |
| Offset | ±2 of least significant displayed digit |
| Gain accuracy | 0.05 % |
| Conversion rate | 10/s for 50 Hz line, 12/s for 60 Hz |
| Line rejection | >100:1 |
| Scanning and Data Logging | |
| Scanning | Selected channels will be scanned. Dwell time between scans is set from 10 s to 9999 s. |
| Alarm | Temperature or voltage limit for each channel |
| Scan enable | All channels may be scanned or skipped |
| Printer output | Voltages, temperatures, time and date as a list or in a graphical format |
| Data memory | Last 2000 measurements in battery backed-up memory |
| General | |
| Analog outputs | Four analog voltage outputs proportional to temperature of channels 1, 2, 3 and 4 |
| Relay output | Switching 10 A |
| Store and recall | Nine locations for instrument set-up |
| Interfaces | RS-232, GPIB and Centronics ports (standard). All instrument functions may be set and read via RS-232 or GPIB. |
| Power | 10 W, 100/120/220/240 VAC, 50/60 Hz |
| Dimensions | 8.5" × 3.5" × 13" (WHD) |
| Weight | 9 lbs. |
| Warranty | One year parts and labor on defects in materials and workmanship |
Important: Maximum common-mode input: ±200 VDC. Input protection: 250 Vrms.
Important: Accuracy errors are for the SR630 only. Standard errors for thermocouple wire are 2 to 5 times the error due to the SR630.
Important: Thermocouples are not included with the SR630; K-type thermocouple probes are listed separately in the ordering information (O630KF1, O630KF2, O630KT1, O630KT2) and must be ordered to use the instrument.
Recommended pairing: the SRS O630-series K-type thermocouples (O630KF1/KF2 fiberglass or O630KT1/KT2 teflon, 5 ft and 10 ft) and the O630RMS/O630RMD rack-mount kits listed in the ordering information.
This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.
ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.
Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.
Model No
Stanford
Condition
Pre-Owned
Manufacturer
Stanford Research Systems
Request A Quote