Skip to product information
1 of 1

Rohde & Schwarz SMB100B RF Signal Generator (Pre-Owned)

Rohde & Schwarz SMB100B RF Signal Generator (Pre-Owned)

Request a Quote To Get An EDU Discount

Login & Get 2% Back on Purchases
Login or Sign Up for Loyalty

Rohde & Schwarz SMB100B RF Signal Generator (Pre-Owned)

Key Features At A Glance

  • Frequency from 8 kHz up to 1, 3, or 6 GHz set by the installed frequency option (SMBB-B101/-B103/-B106)
  • 0.001 Hz frequency setting resolution; phase offset adjustable from -36000° to +36000°
  • Level setting range from -145 dBm up to +20 dBm standard, extendable to +30/+36 dBm with output power options
  • SSB phase noise < -126 dBc (-132 dBc typ.) at 1 GHz, 20 kHz offset (CW, 1 Hz BW, +10 dBm)
  • Standard reverse power protection switches off the RF output if reverse power exceeds the rated limit
  • 5-inch capacitive touch color display; LAN (10/100/1000BASE-T) standard for control
  • GPIB (IEEE-488.2) and USB-TMC remote control via SMBB-B86 option for ATE integration
  • Optional analog AM/FM/phase modulation, pulse modulation, pulse generator, and stereo/RDS coder (licensed options)
View full details

The R&S SMB100B is an RF signal generator from Rohde & Schwarz covering a frequency range that begins at 8 kHz and extends up to 1 GHz, 3 GHz, or 6 GHz depending on the installed frequency option (SMBB-B101, SMBB-B103, or SMBB-B106). It produces continuous-wave and modulated RF signals with a frequency setting resolution of 0.001 Hz, and is operated through a 5-inch color display with capacitive touch functionality. Phase offset is adjustable from -36000° to +36000° in 0.01° steps.

A signal generator provides a known, controllable stimulus signal used to drive a device under test so that its response can be measured. The SMB100B supports user-defined pulse and pulse-sequence scenarios — including jittered or staggered pulse patterns of the type used in radar applications — through its pulse generator and pulse train options. With the stereo/RDS coder option it can act as a stereo and RDS broadcast source across the 66 MHz to 110 MHz range, generating L/R stereo, pilot tone, and ARI/RDS subcarrier signals.

ValueTronics stocks the instruments it sells from its own 20,000 sq ft secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. Every pre-owned unit is handled accordingly: pre-owned units are inspected and functionally verified in-house before they ship, while new units ship factory-sealed exactly as received from the manufacturer.

Brand Heritage

Rohde & Schwarz is an independent, privately held electronics company headquartered in Munich, Germany, founded more than 80 years ago. It has operated continuously under the Rohde & Schwarz name without a major corporate rename or acquisition-driven brand change.

Accessories Supplied

  • Power cable
  • Quick start guide

Product Core & Specifications

Specification Value
Frequency
Range (R&S®SMBB-B101) 8 kHz to 1 GHz
Range (R&S®SMBB-B103) 8 kHz to 3 GHz
Range (R&S®SMBB-B106) 8 kHz to 6 GHz
Resolution of setting 0.001 Hz
Resolution of synthesis (f = 1 GHz) 0.163 μHz (nom.)
Settling time < 1.1 ms (0.8 ms meas.)
Phase offset setting range and resolution –36000° to +36000°, 0.01° resolution
Frequency Sweep
Operating mode digital sweep in discrete steps
Step size (linear) full frequency range, min. 0.001 Hz
Step size (logarithmic) 0.01 % to 100 %
Dwell time setting range 5 ms to 100 s
Dwell time resolution 0.1 ms
Reference Frequency
Frequency error at time of calibration (standard) < 1 × 10–7
Frequency error at calibration (with B1 or B1H) < 1 × 10–8
Aging (standard) ≤ 1 × 10–6/year
Aging (with B1) ≤ 1 × 10–9/day, ≤ 1 × 10–7/year
Aging (with B1H) ≤ 5 × 10–10/day, ≤ 3 × 10–8/year
Temperature effect, 0 °C to +55 °C (standard) ±2 × 10–6
Temperature effect (with B1) ±1 × 10–7
Temperature effect (with B1H) ±1 × 10–8
Warm-up time (with B1 or B1H) ≤ 10 min
Reference input level range 0 dBm to +16 dBm
Reference input impedance 50 Ω (nom.)
Reference output level +7 dBm to +13 dBm, +10 dBm (typ.)
Level Settings
Setting range, standard (8 kHz ≤ f < 100 kHz) –145 dBm to +8 dBm
Setting range, standard (100 kHz ≤ f < 300 kHz) –145 dBm to +13 dBm
Setting range, standard (300 kHz ≤ f < 1 MHz) –145 dBm to +18 dBm
Setting range, standard (1 MHz ≤ f < 6 GHz) –145 dBm to +20 dBm
Setting range with R&S®SMBB-K31 (1 MHz ≤ f ≤ 6 GHz) –145 dBm to +30 dBm
Setting range with R&S®SMBB-B32 and -K31 (1 MHz ≤ f ≤ 6 GHz) –145 dBm to +36 dBm
Setting resolution 0.01 dB
Level Performance
Level accuracy (level > –90 dBm, 8 kHz ≤ f ≤ 200 kHz) < 1.2 dB
Level accuracy (level > –90 dBm, 200 kHz < f ≤ 3 GHz) < 0.5 dB
Level accuracy (level > –90 dBm, f > 3 GHz) < 0.7 dB
Additional level error (ALC state Off (Table)) < 0.5 dB
Level settling time < 1 ms (0.7 ms meas.)
Reverse Power
Max. permissible RF power (1 MHz < f ≤ 1 GHz) 50 W
Max. permissible RF power (1 GHz < f ≤ 2 GHz) 25 W
Max. permissible RF power (2 GHz < f ≤ 6 GHz) 10 W
Max. permissible DC voltage 50 V (nom.)
VSWR
Output impedance VSWR in 50 Ω system (f > 200 kHz) < 1.8
Spectral Purity
Harmonics (1 MHz < f ≤ 6 GHz; level ≤ 13 dBm) < –30 dBc
Nonharmonics (f ≤ 750 MHz) < –80 dBc
Nonharmonics (750 MHz < f ≤ 1500 MHz) < –76 dBc
Nonharmonics (1500 MHz < f ≤ 3000 MHz) < –70 dBc
Nonharmonics (3.00 GHz < f ≤ 6.00 GHz) < –64 dBc
Subharmonics (3.00 GHz < f ≤ 6.00 GHz) < –75 dBc (< –90 dBc measured)
Wideband noise (15 MHz < f ≤ 6 GHz, 30 MHz offset) < –146 dBc, < –153 dBc (typ.)
SSB phase noise (20 kHz offset, f = 100 MHz) < –142 dBc, –150 dBc (typ.)
SSB phase noise (f = 1 GHz) < –126 dBc, –132 dBc (typ.)
SSB phase noise (f = 2 GHz) < –120 dBc, –126 dBc (typ.)
SSB phase noise (f = 3 GHz) < –116 dBc, –123 dBc (typ.)
SSB phase noise (f = 4 GHz) < –114 dBc, –120 dBc (typ.)
SSB phase noise (f = 6 GHz) < –110 dBc, –117 dBc (typ.)
Residual FM (f = 1 GHz, 0.03 kHz to 23 kHz) < 4 Hz, 1.9 Hz (typ.)
Residual AM (RMS 0.03 kHz to 20 kHz, level = 12 dBm, f > 10 MHz) < 0.02 %
List Mode
Max. number of steps (learned mode) 10000
Dwell time setting range 1 ms to 100 s
Dwell time setting resolution 0.1 ms
Internal Modulation Generator
Signal types sine
Frequency setting range 0.1 Hz to 1 MHz
Frequency setting resolution 0.01 Hz
Pulse Modulation (R&S®SMBB-K22 option)
On/off ratio > 80 dB, > 92 dB (typ.)
Rise/fall time (transition type: fast) < 15 ns, < 5 ns (meas.)
Minimum pulse width (transition type: fast) < 20 ns
Pulse repetition frequency 0 Hz to 25 MHz
Remote Control
Interfaces (standard) Ethernet/LAN 10/100/1000BASE-T
Interfaces (with R&S®SMBB-B86 option) IEC 60625 (GPIB IEEE-488.2), USB 2.0 (VISA USB-TMC), Serial (RS-232)
Command set SCPI 1999.5 or compatible command sets
Connectors
RF output (front) N female
REF IN / REF OUT (rear) BNC female
REF 1 GHz IN / OUT (with B3 option) SMA female
LAN RJ-45
IEEE 488 (with B86 option) 24-pin Amphenol series 57 female
General Data
Operating temperature range 0 °C to +55 °C
Storage temperature range –40 °C to +71 °C
Altitude (operating) up to 4600 m (15000 ft)
Rated voltage 100 V to 240 V (± 10 %)
Rated frequency 50 Hz to 60 Hz (± 5 %), 400 Hz (± 5 %)
Important: The base unit must be ordered together with one frequency option — R&S SMBB-B101 (1 GHz), SMBB-B103 (3 GHz), or SMBB-B106 (6 GHz).
Important: Only one of the R&S®SMBB-B1 or R&S®SMBB-B1H reference oscillator options can be installed.
Important: The R&S®SMBB-B32 ultra high output power option requires the R&S®SMBB-K31 option.
Important: The R&S®SMBB-K27 pulse train option requires the R&S®SMBB-K23 option.
Important: The R&S®SMBB-B5 stereo/RDS coder option requires the R&S®SMBB-K720 option.
Important: Serial (RS-232) remote control requires the recommended extra R&S®TS-USB1.
Important: The R&S®SMB100B includes reverse power protection as standard, which switches off the RF output signal if reverse power exceeds the limit.
Recommended pairing: use an R&S NRP-series power sensor (for example the NRP18S-20 with the NRP-ZKU USB cable) for external power measurement and instrument level adjustment via the USB connector.

About this Pre-Owned unit

This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.

  • Warranty included.
  • Functional verification included.
  • Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
  • Note that unnecessary accessories may not be included (contact Test Architect to confirm).

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.

Model No

Rohde

Condition

Pre-Owned

Manufacturer

Rohde & Schwarz

Rf_gen_freq

8 kHz to 6 GHz