Skip to content

The Rigol DSA1000 Series is a general-purpose spectrum analyzer covering 9 kHz to 3 GHz. Built around an all-digital IF core, it is designed to provide stable performance with a broad, easy-to-use set of test functions intended to meet most spectrum-related measurement and application requirements. Rigol positions it as a highly cost-effective instrument for stages spanning research and development through manufacturing and maintenance.
The series is aimed at maintenance, production, and education applications, and it supports work from research and development through manufacturing and maintenance. A spectrum analyzer of this type displays signal amplitude across frequency, letting an engineer characterize spectral content, identify individual signals, and measure their levels.
ValueTronics stocks and holds its own test-equipment inventory in a 20,000-square-foot secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. Every used unit is inspected and functionally verified in-house before it ships, while new units leave the warehouse factory-sealed exactly as received from the manufacturer. Because we hold inventory directly, the analyzer you order sits on our own shelves, ready to be prepared and shipped.
The DSA1000 Series comprises four ordering configurations of the same 9 kHz to 3 GHz spectrum analyzer: the DSA1030, the DSA1030-TG, the DSA1030A, and the DSA1030A-TG. All share the all-digital IF architecture, the 8.5-inch widescreen 800×480 display, and the standard Quasi-Peak detector with EMI filter.
The configurations differ along two axes the ordering guide makes explicit: whether a 3 GHz tracking generator is fitted (the -TG suffix) and whether the preamplifier and advanced measurement functions are included as standard (the A models) or remain options (the non-A models).
Each model below is its own dedicated product page with pre-owned-condition pricing. Select the configuration that matches the options you need, and review that page for its specific pre-owned availability and specifications.
The DSA1030A ships with the preamplifier and advanced measurement functions as standard. Its specified headline figures are a minimum 10 Hz resolution bandwidth, typical −88 dBc/Hz phase noise at a 10 kHz offset, displayed average noise level down to −148 dBm with the preamplifier on, and total amplitude uncertainty under 1.0 dB.
Across the family, the A configurations specify finer performance than the non-A configurations: a minimum resolution bandwidth of 10 Hz versus 100 Hz, typical phase noise of −88 dBc/Hz versus −80 dBc/Hz at a 10 kHz offset, and a lower displayed average noise level. The preamplifier and advanced measurement functions are standard on the A models and are offered as the PA-DSA1030 and AMK-DSA1000 options on the DSA1030 and DSA1030-TG.
The -TG suffix adds a 3 GHz tracking generator (10 MHz to 3 GHz output, −20 dBm to 0 dBm) to either base model. The comparison table below summarizes which capabilities are standard, optional, or absent for each ordering number.
| Model | Min RBW | DANL (Preamplifier On) | Preamplifier |
|---|---|---|---|
| DSA1030A | 10 Hz | -148 dBm (typical, 10 MHz to 2.5 GHz) | Standard |
| DSA1030A-TG | 10 Hz | -148 dBm (typical, 10 MHz to 2.5 GHz) | Standard |
| DSA1030 | 100 Hz | -138 dBm (typical, 10 MHz to 2.5 GHz, with optional preamplifier) | Optional (PA-DSA1030) |
| DSA1030-TG | 100 Hz | -138 dBm (typical, 10 MHz to 2.5 GHz, with optional preamplifier) | Optional (PA-DSA1030) |
Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.
| Specification | Value |
|---|---|
| Frequency | |
| Frequency Range | 9 kHz to 3 GHz |
| Frequency Resolution | 1 Hz |
| Internal Frequency Reference — Reference Frequency | 10 MHz |
| Aging Rate | <3 ppm/year |
| Temperature Drift | <3 ppm, 20 ℃ to 30 ℃ |
| Marker Resolution | span/(sweep points-1) |
| Marker Uncertainty | ± (frequency indication × frequency reference uncertainty + 1% × span + 10% × resolution bandwidth + marker resolution) |
| Marker Frequency Counter Resolution | 1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz |
| Marker Frequency Counter Uncertainty | ±(frequency indication × frequency reference uncertainty + counter resolution) |
| Frequency Span | |
| Range | 0 Hz, 100 Hz to 3 GHz |
| Uncertainty | ±span/(sweep points-1) |
| SSB Phase Noise (typical fc = 500 MHz, RBW ≤ 1 kHz, sample detector, trace average ≥ 50) | |
| Carrier Offset 10 kHz | <-88 dBc/Hz, typical |
| Carrier Offset 100 kHz | <-100 dBc/Hz, typical |
| Carrier Offset 1 MHz | <-110 dBc/Hz, typical |
| Bandwidths | |
| Resolution Bandwidth (-3 dB) | 10 Hz to 1 MHz, in 1-3-10 sequence |
| Bandwidth (-6 dB) | 200 Hz, 9 kHz, 120 kHz, 1 MHz |
| RBW Uncertainty | <5%, nominal |
| Resolution Filter Shape Factor (60 dB : 3 dB) | <5, nominal |
| Video Bandwidth (-3 dB) | 1 Hz to 3 MHz, in 1-3-10 sequence |
| Amplitude — Measurement Range | |
| 10 MHz to 3 GHz | DANL to +30 dBm |
| 1 MHz to 10 MHz | DANL to +21 dBm |
| 9 kHz to 1 MHz | DANL to +17 dBm |
| Maximum Rated Input Level (protection switch on when input level >33 dBm) | |
| DC Voltage | 50 V |
| CW RF Power | 30 dBm (1 W) (RF attenuation ≥ 20 dB) |
| Max. Damage Level | 40 dBm (10 W) |
| 1 dB Gain Compression (Total Power at Input Mixer, fc ≥ 50 MHz, preamplifier off) | >0 dBm |
| Displayed Average Noise Level (DANL) — Preamplifier Off (0 dB RF Attenuation, RBW=100 Hz, VBW=1 Hz, RMS Average Detector, Trace Average ≥ 50, Input Impedance=50 Ω, TG Off) | |
| 100 kHz to 10 MHz | <-85 dBm-3 × (f/1 MHz) dB, typical -125 dBm |
| 10 MHz to 2.5 GHz | <-127 dBm+3 × (f/1 GHz) dB, typical -130 dBm |
| 2.5 GHz to 3 GHz | <-115 dBm |
| Displayed Average Noise Level (DANL) — Preamplifier On | |
| 100 kHz to 1 MHz | <-103 dBm |
| 1 MHz to 10 MHz | <-103 dBm-3 × (f/1 MHz) dB, typical -143 dBm |
| 10 MHz to 2.5 GHz | <-145 dBm+3 × (f/1 GHz) dB, typical -148 dBm |
| 2.5 GHz to 3 GHz | <-133 dBm |
| Level Display | |
| Log Scale | 1 dB to 200 dB |
| Linear Scale | 0 to Reference Level |
| Number of Display Points | Normal: 601; Full Screen: 751 |
| Number of Traces | 3 + Math Trace |
| Trace Detectors | Normal, Positive-peak, Negative-peak, Sample, RMS, Voltage Average, Quasi-Peak |
| Trace Functions | Clear Write, Max Hold, Min Hold, Average, Freeze, Blank |
| Scale Units | dBm, dBmV, dBμV, V, W |
| Frequency Response (10 dB RF Attenuation, relative to 50 MHz, 20 ℃ to 30 ℃) | |
| Preamplifier Off, 100 kHz to 3 GHz | <0.7 dB |
| Preamplifier On, 1 MHz to 3 GHz | <1.0 dB |
| Input Attenuation Switching Uncertainty | |
| Setting Range | 0 to 50 dB, in 1 dB step |
| Switching Uncertainty (fc=50 MHz, relative to 10 dB, 20 ºC to 30 ºC) | < (0.3 + 0.01 × attenuator setting) dB |
| Absolute Amplitude Uncertainty | |
| Uncertainty (fc=50 MHz, peak detector, preamplifier off, 10 dB RF attenuation, input signal=-10 dBm, 20 ºC to 30 ºC) | ±0.4 dB |
| RBW Switching Uncertainty | |
| 10 Hz to 1 MHz, relative to 1 kHz RBW | <0.1 dB |
| Reference Level | |
| Range | -100 dBm to +30 dBm, in 1 dB step |
| Resolution | Log Scale: 0.01 dB; Linear Scale: 5 digits |
| Level Measurement Uncertainty (95% confidence level, S/N>20 dB, RBW=VBW=1 kHz, preamplifier off, 10 dB RF attenuation, -50 dBm | |
| Level Measurement Uncertainty | <1.0 dB, nominal |
| RF Input VSWR (10 dB RF Attenuation) | |
| 100 kHz to 10 MHz | <1.8, nominal |
| 10 MHz to 2.5 GHz | <1.5, nominal |
| 2.5 GHz to 3 GHz | <1.8, nominal |
| Intermodulation | |
| Second Harmonic Intercept (SHI) | +35 dBm |
| Third-order Intermodulation (TOI) | fc >30 MHz: +7 dBm |
| Spurious Responses | |
| Image Frequency | <-60 dBc |
| Intermediate Frequency | <-60 dBc |
| Spurious Response | <-88 dBm, typical |
| System-related Sideband | <-60 dBc |
Important: Maximum rated input: 50 V DC and 30 dBm (1 W) CW RF power at ≥20 dB RF attenuation; maximum damage level 40 dBm (10 W). When the input level exceeds 33 dBm, the protection switch engages.
Important: USB Disk for the mass memory is not supplied with the instrument.
Important: RF test cables, adaptors, antennas and attenuators are offered as Optional Accessories and are not included in the standard delivery (standard accessories: Front Panel Cover, Quick Guide, CDROM, USB Data Cable, and Power Cable). Order the appropriate RF cable/adaptor for your measurement connection separately.
Important: GPIB (IEC/IEEE Bus, IEEE 488.2) remote control requires the optional USB-GPIB interface converter.
Important: The Advanced Measurement Kit is an option for the DSA1030 and DSA1030-TG and is standard on the DSA1030A and DSA1030A-TG; the preamplifier is standard on the DSA1030A/DSA1030A-TG and an option (PA-DSA1030) for the DSA1030/DSA1030-TG.
Usage tip: the DSA1030 and DSA1030-TG can add the advanced measurement functions through the optional Advanced Measurement Kit (AMK-DSA1000) and a preamplifier through the PA-DSA1030 option.
This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.
ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.
Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.