Key Features At A Glance

  • 3-slot mainframe with built-in 6½-digit (22-bit) internal DMM and 0.003% basic 1-year DC voltage accuracy
  • Scalable from 4 to 120 channels using nine switch, RF, and control plug-in modules
  • Measures 14 input signal types: temperature, DC/AC volts and current, resistance, frequency, capacitance, and strain
  • Scan rates up to 450 channels/s with 1-million-reading nonvolatile, time-stamped memory
  • LAN and USB 2.0 standard for PC connectivity and automation; DAQ973A model adds built-in GPIB (IEEE-488)
  • 4.3-inch color display with numeric, bar graph, trend chart, and histogram measurement views
  • Code compatible with 34970A/34972A; SCPI with C, C#, Visual Basic, and Visual Studio drivers
  • BenchVue DAQ software included for PC-based logging and time/frequency-domain analysis
View full details

The Keysight DAQ970A/DAQ973A is a data acquisition system built around a compact 3-slot mainframe with a built-in 6½-digit (22-bit) internal DMM. The measurement engine from a high-performance benchtop DMM is embedded inside the mainframe, combining proven measurement performance, universal inputs with built-in signal conditioning, and modular flexibility in a single low-cost package. It delivers 0.003% basic DC voltage accuracy, ultra-low reading noise, and scan rates of up to 450 channels per second, giving an engineer both the speed and the accuracy needed for demanding acquisition work.

Configured with a 20-channel relay multiplexer, the system becomes a powerful, low-cost data logger for characterization applications, and it also serves as a data acquisition front end for automated test. Typical data-logging and monitoring uses include environmental chamber monitoring, component inspection, benchtop testing, process troubleshooting, and temperature profiling. It supports both the R&D engineer characterizing a new design and the manufacturing engineer building a test system or troubleshooting a process. The 40-channel single-ended multiplexer adds density for applications such as battery test and component characterization.

ValueTronics stocks the instruments it sells from its own 20,000 sq ft secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. Every pre-owned unit is inspected and functionally verified in-house by our technicians before it ships, while new units ship factory-sealed exactly as received from the manufacturer. That combination of real inventory on the shelf and hands-on verification of used equipment is the difference between buying from a stocking specialist and gambling on an unaccountable listing.

Brand Heritage

Keysight Technologies traces its roots to the test and measurement business of Hewlett-Packard. That business was spun out as Agilent Technologies in 1999, and in 2014 Agilent's electronic measurement division became the independent company Keysight Technologies. Equipment from across that history is frequently encountered under all three names.

Compare Other Models in This Series

The DAQ970A/DAQ973A family pairs a common 6½-digit (22-bit) measurement engine with a 3-slot modular mainframe, giving the simple operation and low per-channel cost of a data logger alongside the flexibility and performance of a modular data acquisition system. Both mainframes share the same internal DMM, the same 4.3-inch color display, and the same compatibility with all nine plug-in switch, RF, and control modules.

The two models differ in their standard remote-control interfaces. The DAQ970A provides built-in Gigabit LAN and USB 2.0, while the DAQ973A carries the same LAN and USB connectivity plus a built-in GPIB interface for ATE environments that depend on IEEE-488 control.

Each model in the DAQ970A/DAQ973A family below links to its own dedicated product page, where pre-owned pricing and availability are matched to that exact configuration. Choose the mainframe and module combination that fits your channel count and connectivity requirements.

The defining difference across the family is connectivity: the DAQ970A ships with built-in LAN and USB, and the DAQ973A adds a built-in GPIB interface to that same LAN and USB set. Both mainframes accept up to three of the nine available plug-in modules in any combination and deliver identical measurement performance, channel density, and 1-million-reading memory.

Beyond connectivity, configuration differences come from module choice rather than the mainframe—multiplexer type, scan speed, voltage and current ratings, and switching topology all vary by module. The comparison table below summarizes the interface and configuration differences across the family.

Model Type Channels Max Voltage
DAQM907A Multifunction module (DIO/totalizer/analog out) — 42 V
DAQ970A DAQ Mainframe (USB, LAN) Up to 120 300 V
DAQ973A DAQ Mainframe (USB, LAN, GPIB) Up to 120 300 V
DAQM900A 2-wire solid state (4-wire selectable) 20 120 V

Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.

Product Core & Specifications

Specification Value
DAQM907A Multifunction Module — Functionality
Digital I/O (Port 1, 2) 16 bits (two 8-bit ports), input or output, non-isolated
Totalizer 100 kHz input, 26-bit event counter
Analog outputs 2 channels, ± 12 V or ± 24 mA
Max input voltage 42 V
Max current 400 mA
Analog Output (DAC) Characteristics
Resolution 100 µV, 0.2 µA
Voltage accuracy (1 year ± 5 °C) ± 0.027% of output + 4.4 mV
Current accuracy (1 year ± 5 °C) ± 0.115% of output + 4.4 µA
Voltage mode current (IOUT) 15 mA max per channel
Current mode compliance 12 V
Settling time 1 ms to 0.01% of output
Max total output per frame 40 mA
Digital I/O Characteristics
Vin(L) < 0.8 V (TTL)
Vin(H) > 2.0 V (TTL)
Vout(L) < 0.8 V @ Iout = -400 mA
Vout(H) > 2.4 V @ Iout = 1 mA
Vin(H) max < 42 V with external open drain pull-up
Alarming Maskable pattern match or state change
Alarm sampling speed 4 ms (max)
Latency to alarm output 5 ms (typical)
Read/write speed 95/s
Totalizer Characteristics
Max count 2^26 - 1
Totalize input 100 kHz (max), rising or falling edge, programmable, min H or L time 5 µs
Signal level 1 Vp-p (min), 42 Vpk (max)
Threshold 0 V or TTL, jumper selectable
Gate input TTL-Hi, TTL-Lo, or none
Count reset Manual or Read + Reset
Read speed 85/s
Important: The internal DMM input section is optically isolated and shielded from the earth-referenced circuitry and computer interface, providing up to 300 V of input isolation.
Important: The DAQ970A/DAQ973A mainframe requires at least one low-frequency multiplexer module to make measurements: only the DAQM900A, DAQM901A, DAQM902A, and DAQM908A modules connect to the internal DMM. Modules are ordered separately ('buy only what you need—and add more modules later as your application grows').
Important: Up to three modules, in any combination, can be inserted into a single mainframe.
Important: The DAQM905A RF Multiplexer uses SMB connectors; a standard set of (10) BNC-to-SMB adapter cables is provided with each RF module for BNC connections (2 GHz bandwidth via SMB, 1 GHz via the BNC-to-SMB adapters).
Important: On-module screw terminals accept wire sizes from 16-gage to 22-gage; twenty-gage wire is recommended for high channel count applications.
Recommended pairing: listed accessories include the 34307A J-type thermocouples and 34308A 10 kΩ thermistors for temperature measurements and the 11062A Kelvin clip set for resistance measurements.

About this Pre-Owned unit

This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.

  • Warranty included
  • Functional verification included
  • Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
  • Note that unnecessary accessories may not be included (contact Test Architect to confirm).

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.

Model No

Keysight

Condition

Pre-Owned

Manufacturer

Keysight Technologies

Plugin_type

Multifunction Module

Keysight DAQM903A 20-Channel Actuator/General Purpose Switch Module for DAQ970A/DAQ973A (Pre-Owned)

DAQM903A

Keysight DAQM904A 4x8 Two-Wire Matrix Switch Module for DAQ970A/DAQ973A (Pre-Owned)

DAQM904A

Keysight DAQM901A 20-Channel Armature Multiplexer Module for DAQ970A/DAQ973A (Pre-Owned)

DAQM901A

34307A Agilent Accessory Used

34307A

11062A Agilent Accessory Used

11062A