Common Mode Voltage: 300V between any terminal and chassis
Temperature Accuracy using Internal CJC: 1.0°C (see mainframe specification for details)
Totalize Input
Maximum Count: 2/32–1.
Totalize Input: 100kHz (max), rising or falling edge, programmable
Signal Level: 1Vp-p (min), 42Vpk (max)
Threshold: 0V or TTL, jumper selectable
Gate Input: TTL-Hi, TTL-Lo, or none
Count Reset: Manual or Read+Reset
Read Speed: 50/s
Analog Voltage Output
DAC 1, 2: ±12V in 1mV increments, nonisolated
Resolution: 1mV
Iout: 5mA max.
Settleing Time: 1ms to 0.01% of output
Accuracy ±(% of output + mV): 1 year ±5°C: 0.15% + 19mV; 90 day ±5°C: 0.1% + 19mV; 24 hour ±1°C: 0.04% + 19mV
Temperature Coefficient: ±(0.015% + 1mV)/°C
Digital Output
V OUT(L): <0.8V @ Iout = 400mA
V OUT(H): >2.4V @ Iout = 1mA
V OUT(H)MAX.: <42V with external open drain pull-up
Write Speed: 50/s
General
20 Channels: 20 channels of 2-pole relay input. All channels configurable to 4-pole.
Relay Type: Latching electromechanical
Actuation Time: <3ms
Firmware: Specified for Model 2700 rev. A02 or B01, 2701 rev. A01, and 2750 rev. A01 or higher
Environmental
Operating Environment: Specified for 0° to 50°C. Specified to 80% R.H. at 35°C
Storage Environment: –25° to 65°C
Weight: 0.5kg (1.1 lbs)
The Keithley 7706 plug-in module offers 20 channelsof 2-pole or 10 channels of 4-pole multiplexer switching with automatic CJC, as well as two analog output channels, 16 digital outputs,and one event counter/totalizer. The event counter/totalizer can be used to monitor and control system components, such as fixtures, limit switches, pass/fail indicators, external voltage sources, loads, door closures, revolutions, etc., while performing mixed signal measurements. The Keithley 7706 is ideal for RTD, thermistor, and thermocouple temperature applications.