Skip to product information
1 of 1

Fluke Calibration 7526A Precision Process Calibrator

Fluke Calibration 7526A Precision Process Calibrator

Regular price $11,175
Sale price $11,175 Regular price $0
Shipping calculated at checkout.
Request a Quote
Estimated Delivery in
5 - 7
business days (based on availability)

Request a Quote To Get An EDU Discount

Login & Get 2% Back on Purchases
Login or Sign Up for Loyalty

Fluke Calibration 7526A Precision Process Calibrator

Key Features At A Glance

  • Sources and measures DC voltage from 0 to 100 V with 30 ppm output accuracy (1-year)
  • Sources DC current 0 to 100 mA and measures DC current 0 to 50 mA via isolated input
  • Simulates and measures nine RTD types and thirteen thermocouple types (B, C, E, J, K, L, N, R, S, T, U, XK, BP)
  • Precision pressure measurement when used with Fluke 700 or 525A-P Series Pressure Modules
  • Includes 24 V DC transmitter loop power supply with measurement of 4-20 mA loop current
  • Accepts ITS-90 coefficients for accurate SPRT measurements
  • Standard RS-232 and IEEE-488 (GPIB) interfaces for automated calibration systems
  • Compatible with MET/CAL Calibration Software and includes automated switch-test function
View full details

The Fluke Calibration 7526A Precision Process Calibrator is a benchtop calibration instrument designed for temperature and pressure process instrumentation. The datasheet positions the 7526A as a single-box tool that sources and measures DC voltage, current, and resistance, simulates and measures RTDs and thermocouples, and accepts external pressure modules for precision pressure measurement. It is offered as a balance of economy and accuracy for process calibration work.

Per the datasheet, the 7526A is used to calibrate RTD and thermocouple readouts, pressure gauges, temperature transmitters, digital process simulators, data loggers, and multimeters. The instrument simulates and measures nine RTD types and thirteen thermocouple types, sources DC voltage from 0 to 100 V, sources DC current from 0 to 100 mA, and measures 4-20 mA loop current using the included 24 V DC transmitter loop power supply. With Fluke 525A-P Series Pressure Modules, the 7526A also performs pressure measurement to within 0.008 % of reading as stated by the datasheet.

ValueTronics International has supplied new and used test and measurement instrumentation since 1992 from our 20,000 sq ft secure warehouse at 1675 Cambridge Drive, Elgin, Illinois. Every {{CONDITION}} 7526A that passes through our facility is inspected, performance-verified, and graded by our in-house technical team before it ships, with calibration options available to match your quality system requirements.

Brand Heritage

Fluke Calibration is the calibration-instrument division of Fluke Corporation, which itself sits within the broader Fortive group following the 2016 spin-off from Danaher. The Fluke Calibration product line traces back through several decades of precision metrology development and includes process, electrical, RF, temperature, pressure, and flow calibration instruments under a single brand.

Product Core & Specifications

DC Voltage, Output
Range Absolute Uncertainty, ± (ppm of output) Absolute Uncertainty, ± (µV), 1-yr Resolution
0 to 100 mV 30 3 1 µV
0 to 1 V 30 10 10 µV
0 to 10 V 30 100 100 µV
0 to 100 V 30 1 mV 1 mV
TC Output and Input: -10 mV to 75 mV 30 2 10 Ω
All outputs are positive only unless otherwise noted
DC Voltage, Isolated Input
Range Absolute Uncertainty, ± (ppm of reading + mV), 1-yr Resolution
0 to 10 V 50 + 0.2 100 µV
10 V to 100 V 50 + 2.0 1 mV
DC Current, Output
Range Absolute Uncertainty, ± (ppm of reading + µA), 1-yr Resolution
0 to 100 mA 50 + 1 1 µA
For line voltages less than 95 V, ±100 ppm of reading
DC Current, Isolated Input
Range Absolute Uncertainty, ± (ppm of reading + µA), 1-yr Resolution
0 mA to 50 mA 100 + 1 0.1 µA
0 mA to 24 mA (Loop Power) 100 + 1 0.1 µA
Loop Power: 24 V ±10%. HART Resistor: 250 Ω ± 3%
Resistance, Output
Range Absolute Uncertainty, tcal ± 5 °C, ± Ohms, 1-yr Resolution Nominal Current
5 Ω to 400 Ω 0.015 0.001 Ω 1 to 3 mA
5 Ω to 4 kΩ 0.3 0.01 Ω 100 µA to 1 mA
Resistance, Input
Range Absolute Uncertainty, ± (ppm of reading + Ω), 1-yr Resolution
0 Ω to 400 Ω 20 + 0.004 0.001 Ω
0 Ω to 4 kΩ 20 + 0.04 0.01 Ω
Sample Thermocouple Accuracy, Input/Output
TC Type Min Temperature (°C) Max Temperature (°C) Absolute Uncertainty, tcal ±5 °C, ± (°C), 1-yr
J -210 1200 0.09
K -250 1372 0.1
S -50 1767 0.29
T -250 400 0.11
See extended specifications for all TC types (B, C, E, J, K, L, N, R, S, T, U, XK, BP). Best accuracy within specified TC temperature range.
Sample RTD and Thermistor, Output
RTD Type Min Temperature (°C) Max Temperature (°C) Absolute Uncertainty, tcal ±5 °C, ± (°C), 1-yr
Pt 385, 100 Ω -200 630 0.05
YSI 400 15 50 0.007
See extended specifications for all RTD types: Pt-100 (385, 3926, 3916), Pt-200, Pt-500, Pt-1000, Ni-120, Cu-427, SPRT
Sample RTD and Thermistor, Input
RTD Type Min Temperature (°C) Max Temperature (°C) Absolute Uncertainty, tcal ±5 °C, ± (°C), 1-yr
Pt 385, 100 Ω -80 100 0.020
Pt 385, 100 Ω 100 300 0.024
YSI 400 15 50 0.007
See extended specifications for all RTD types: Pt-100 (385, 3926, 3916), Pt-200, Pt-500, Pt-1000, Ni-120, Cu-427, SPRT
General Specifications
Standard Interface RS-232, IEEE-488 (GPIB)
Temperature Performance Operating: 0 °C to 50 °C; Calibration (tcal): 18 °C to 28 °C; Storage: -20 °C to 70 °C
Electromagnetic Compatibility CE: Conforms to EN61326; operation in controlled EM environments
Temperature Coefficient Temperature coefficient for temperatures outside tcal 5 °C is 10% of the 90-day specification (or 1 year if applicable) per °C
Relative Humidity Operating: <80% to 30 °C; <70% to 40 °C; <40% to 50 °C
Altitude Operating: 3,000 m (9,800 ft) max; Non-operating: 12,200 m (40,000 ft) max
Safety EN/IEC 61010-1:2010 3rd Edition, UL 61010-1:2012, CAN/CSA 22.2 No. 61010-1-12
Analog Low Isolation 20 V
Line Power 120 V~: 100 V to 120 V; 240 V~: 220 V to 240 V
Line Frequency 47 Hz to 63 Hz
Line Voltage Variation ± 10 % about setting
Power Consumption 15 VA maximum
Dimensions Height: 14.6 cm (5.75 in); Width: 44.5 cm (17.5 in); Depth: 29.8 cm (11.75 in)
Weight (without options) 4.24 kg (9.35 lb)
Important: Maximum input voltage rating for the isolated input is specified by the datasheet at 0 to 100 V DC. The 24 V DC loop power supply specification assumes a HART resistor of 250 Ω ±3%.Important: Precision pressure measurement requires Fluke 700 Series or 525A-P Series pressure modules (sold separately); pressure measurement accuracy is determined by the pressure module used.Important: Floor specification for DC voltage output increases with range. See Specifications for more information.Important: For DC current output, line voltages less than 95 V result in ±100 ppm of reading uncertainty.Important: Loop Power isolated input specification assumes 24 V ±10% Loop Power and a HART Resistor of 250 Ω ± 3%.Recommended pairing: use the 7526A with Fluke 700 Series or 525A-P Series Pressure Modules as named in the datasheet for precision pressure measurement to within 0.008 % of reading.

Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.

Model No

Fluke

Condition

New

Manufacturer

Fluke