Skip to content

The MS710C/D/E/F is a family of high-performance microwave spectrum analyzers spanning 10 kHz to 23 GHz, with an optional measurement range up to 140 GHz when external mixers are used. The design pairs a simplified PLL synthesizer local oscillator with a two-channel digital memory and supports both easy manual operation and fully automatic operation over GPIB. Operator-focused design choices include a grouped key layout with key sizes scaled to functional importance, an operation guide display for complex SHIFT and MARKER sequences, and a preset memory that holds up to 10 sets of measurement conditions.
The series is applied to spectrum analysis of microwave devices and components; spurious emission and spectral distribution measurements of analog and digital communications transmitters; interference measurements for radio stations, satellite earth stations, and similar installations; spectrum analysis in basic research such as nuclear physics and radio-astronomy; and spurious measurements for home-use satellite broadcast receivers and related equipment.
ValueTronics stocks and holds its own inventory in a 20,000-square-foot secure warehouse at 1675 Cambridge Drive in Elgin, Illinois, where we have specialized in new and used test and measurement equipment since 1992. Every pre-owned unit is inspected and functionally verified in-house before it ships, while new units ship factory-sealed exactly as received from the manufacturer. Holding our own stock lets us match the right instrument and condition to each application and ship promptly from our own shelves.
The MS710C/D/E/F is a four-model microwave spectrum analyzer series that shares a common architecture — the same resolution bandwidth range, marker and memory functions, display, and GPIB automation — while differing in frequency coverage and frequency accuracy. All four models cover the 100 kHz to 2 GHz and 1.7 to 23 GHz bands; the variations between models concern the low-frequency band, the external-mixer extension to 140 GHz, and whether the model includes high frequency accuracy.
Because the core measurement functions are common across the family, the choice between models comes down to the required frequency span and accuracy class rather than a difference in operating workflow. The comparison table below summarizes how each model is configured.
Each pre-owned model below is its own dedicated product page with condition-matched pricing. Select the specific model that matches your required frequency coverage and accuracy, then choose the pre-owned listing that fits your budget and lead-time needs.
The MS710C covers the widest frequency range in the series: it is the only model that adds the 10 kHz to 30 MHz band, and it combines high frequency accuracy with external-mixer support that extends measurement coverage up to 140 GHz.
The four models differ along three axes. The MS710C provides the 10 kHz to 30 MHz band, high frequency accuracy, and external-mixer extension to 140 GHz. The MS710D covers 100 kHz to 23 GHz with external-mixer extension to 140 GHz. The MS710E covers 100 kHz to 23 GHz with high frequency accuracy.
High frequency accuracy is provided on the MS710C and MS710E, while external-mixer operation up to 140 GHz is supported on the MS710C and MS710D. Refer to the comparison table for the per-model frequency coverage and accuracy details.
| Model | Frequency Range | External Mixer Extension | High Frequency Accuracy |
|---|---|---|---|
| MS710F | 100 kHz to 23 GHz | — | No |
| MS710C | 10 kHz to 23 GHz | 18 to 140 GHz | Yes |
| MS710D | 100 kHz to 23 GHz | 18 to 140 GHz | No |
| MS710E | 100 kHz to 23 GHz | — | Yes |
Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.
| Specification | Value |
|---|---|
| Frequency — Center Frequency (100 kHz to 2 GHz and 1.7 to 23 GHz band) | |
| Measuring range | MS710C/E: 100 Hz to 2 GHz, 1.7 to 23 GHz; MS710D/F: 0 MHz to 2 GHz, 1.7 to 23 GHz |
| Readout resolution | MS710C/E: 10 kHz (10 MHz to 2 GHz), 10 kHz (1.7 to 23 GHz); MS710D/F: 100 kHz (0 MHz to 2 GHz), 1 MHz (1.7 to 23 GHz) |
| Readout accuracy | ±(30 kHz (0 MHz to 2 GHz, 1.7 to 6.5 GHz), 60 kHz (6.5 to 12.5 GHz), 90 kHz (12.5 to 18.5 GHz), 120 kHz (18.5 to 23 GHz) + 2% of frequency span + 10% of resolution bandwidth) |
| Setting | Number/unit keys, data knob, peak center key, or half-screen shift key |
| Frequency — Frequency Span | |
| Setting range and resolution | 0 Hz (fixed tuning); 1 kHz/div to 200 kHz/div in 1 kHz increments; 210 kHz/div to 2 MHz/div in 10 kHz increments; 2.1 to 20 MHz/div in 100 kHz increments; 21 to 200 MHz/div in 1 MHz increments; span up/down keys: 1 kHz/div to 200 MHz/div in 1-2-5-10 sequence |
| Readout accuracy | ±5% (6 kHz/div to 200 MHz/div), ±10% (1 to 5 kHz/div) |
| Setting | Number/unit keys, data knob, or span up/down keys |
| Frequency — Start/Stop Frequency | |
| Range | In each band (span ≥10 kHz), Min. 10 kHz (varies with span settings); span = stop frequency − start frequency |
| Readout accuracy | ±(center frequency accuracy + 2.5% of span) |
| Setting | Number/unit keys or data knob |
| Resolution Bandwidth (6 dB bandwidth) | |
| Setting range | 100 Hz to 3 MHz in 1-3-10 sequence |
| Setting | Selectable manually or automatically coupled to frequency span |
| Selectivity (60 dB/3 dB) | ≤10 : 1 (resolution bandwidth ≥1 kHz) |
| Residual FM | ≤200 Hzp-p/0.1 s (center frequency ≤6.5 GHz, span ≤100 kHz/div) |
| Noise sidebands | ≤−75 dB (1 kHz resolution bandwidth, 10 Hz video bandwidth, 30 kHz from signal, center frequency ≤6.5 GHz) |
| Amplitude | |
| Measuring range | Average noise level to +30 dBm |
| Display | Vertical 8 divisions, reference level is top line of graticule |
| Graticule (LOG) | 10 dB/div: 0 to −70 dB; 5 dB/div: 0 to −40 dB; 2 dB/div: 0 to −16 dB; 1 dB/div: 0 to −8 dB from reference level |
| Graticule (LIN) | 12.5%/div |
| Linearity | ±0.2 dB/1 dB, ±1.5 dB/70 dB |
| Reference level setting range | −109 to +30 dBm |
| Calibration output accuracy | −10 dBm ±0.3 dB (100 MHz ±10 kHz) |
| Reference level accuracy | ±2.0 dB (reference level −99 to −10 dBm, frequency 100 MHz, 0 dB input attenuator, after calibrated using CAL OUTPUT) |
| Input attenuator setting range | 0 to 70 dB, 10 dB steps, selected manually or automatically coupled to reference level |
| Input attenuator error between steps | ±1 dB (0 to 60 dB, 100 kHz to 2 GHz), ±2 dB (0 to 40 dB, 100 kHz to 23 GHz) |
| Maximum accumulation error | ±2.2 dB (0 to 60 dB, 100 kHz to 2 GHz), ±3 dB (0 to 40 dB, 100 kHz to 23 GHz) |
| Frequency response | ±2.5 dB (100 kHz–10 MHz), ±1.5 dB (10 MHz–2 GHz), ±2.5 dB (1.7–5.48 GHz), ±3 dB (5.48–12.52 GHz), ±4 dB (12.52–23 GHz); 10 dB input attenuator after preselector peak adjustment |
| Dynamic Range | |
| 2nd harmonic distortion | ≤−60 dB (100 kHz–10 MHz, −40 dBm); ≤−70 dB (10–200 MHz, −30 dBm); ≤−80 dB (200–850 MHz, −30 dBm); ≤−100 dB (850 MHz–11.5 GHz [1.7–23 GHz band], −10 dBm) — input level less input attenuator value |
| Two-signal 3rd intermodulation distortion | ≤−80 dB (100 kHz–2 GHz, Δf ≥2.5 MHz, total −30 dBm); ≤−100 dB (1.7–12.5 GHz, Δf ≥70 MHz, −10 dBm); ≤−100 dB (12.5–23 GHz, Δf ≥100 MHz, −10 dBm) |
| Residual response | ≤−90 dBm (0 dB input attenuator, 10 MHz to 6.5 GHz fundamental mixing, 50 Ω termination) |
| Average noise level | ≤−95 dBm (100 kHz–1 MHz), ≤−115 dBm (1 MHz–2 GHz), ≤−110 dBm (1.7–6.5 GHz), ≤−100 dBm (6.5–12.5 GHz), ≤−95 dBm (12.5–18.5 GHz), ≤−88 dBm (18.5–23 GHz); 1 kHz resolution bandwidth, 0 dB input attenuator, 3 Hz video bandwidth |
| Video bandwidth | 1 Hz to 3 MHz, 1-3-10 sequence; selected manually or automatically coupled to frequency span |
| Input | |
| Connector | N-type (nominal 50 Ω) |
| Maximum input level | +30 dBm, ±0 Vdc |
| Marker | |
| Normal | Frequency and level at marker displayed |
| Δ (delta) | Frequency and level difference at two markers displayed |
| Peak | Marker always tracks peak position and shows frequency and level |
| MKR→CF | Marker frequency set to center frequency |
| CRT Display | |
| Display | CRT; display area 80 mm × 100 mm |
| Display item | Graticule, signal traces, function setting value, error message, title, frequency band list, shift function list, and memory list |
| Display Memory | |
| Signal traces / memory capacity | Horizontal 501 points, vertical 801 points, A and B traces, backed up by battery |
| Display | NORMAL, MAX HOLD, AVERAGE, A→B, A↔B |
| Function Setting Memory | |
| Capacity | Up to 10 sets of each function setting can be saved or recalled; the memory list can be displayed on the CRT, backed up by battery |
| Store/recall | Up to 9 sets of display (title, function settings, signal trace) can be saved or recalled |
| Sweep | |
| Sweep time | 2 ms/div to 10 s/div, selected manually or automatically coupled to frequency span, resolution bandwidth, and video bandwidth; for 0 Hz frequency span, 2 µs/div to 10 s/div with manual setting; when (stop frequency − start frequency) >2 GHz the previous time is set and time cannot be set manually |
| Trigger | Signal, free run, line, video, and external trigger |
| General | |
| Remote control | GPIB (IEEE488, IEC625-1, 24 pins); all front-panel functions (except power switch, CRT intensity, level calibration, and trigger level adjustment knob) can be remote-controlled |
| Direct plotting | CRT information can be plotted by the specified plotter or printer |
| Power | AC 100 V +10/−15 %, 50/60 Hz, ≤200 VA |
| Dimensions and mass | 426 (W) × 177 (H) × 451 (D) mm, ≤27 kg |
Important: Maximum input level: +30 dBm, ±0 Vdc.
Important: 18 to 140 GHz measurements on the MS710C and MS710D require an external waveguide mixer, which is not supplied with the instrument and must be ordered separately. When using a 3-port external mixer, an additional amplifier (such as the MP11975A) is required for local signal amplification.
Important: Local output for external mixer: 3 to 6 GHz, ≥+7 dBm; IF input for external mixer: 521.4 MHz.
Important: Power cord plug type must be specified at time of order.
Recommended pairing: the MH680A1 Tracking Generator and MH648A Pre-amplifier are offered as application equipment for the MS710 series.
This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.
ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.
Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.
Model No
Anritsu
Condition
Pre-Owned
Manufacturer
Anritsu
Spec_ana_freq
100 kHz to 23 GHz
Request A Quote