The HP 6002A is a laboratory-type DC power supply that employs an extended range technique to furnish maximum output power over a wide range of voltage and current combinations. As described on the datasheet, the supply provides a full 200 watts of output from 20V/10A up to 50V/4A, with the output completely adjustable through the entire operating range of 0 to 50 volts and 0 to 10 amps via front panel voltage and current controls. The unit is a Constant Voltage / Constant Current (CV/CC) type and is packaged in a System II modular enclosure.
According to the datasheet, the 6002A is intended for general laboratory use where a single supply needs to cover a wide range of voltage and current combinations rather than being restricted to a single rectangular operating locus. Documented operating modes include local front panel control, remote voltage sensing at the load, remote programming by either resistance or voltage source, auto-parallel operation for greater current capacity, auto-series operation for greater voltage capability (up to three units), and auto-tracking operation where two or three supplies share a common negative bus and are simultaneously programmed from a master. With the optional HP-IB Interface (Option 001), output voltage or current can be programmed digitally from a calculator, computer, or other controller.
Hewlett-Packard's test and measurement business was spun off as Agilent Technologies in 1999, and Agilent's electronic measurement business was subsequently transferred to Keysight Technologies in 2014. Instruments originally manufactured under the HP brand — including the 6002A — therefore trace through Agilent to Keysight as the current corporate successor for legacy HP T&M product support.
The HP 6002A is documented in this datasheet as a single-model laboratory power supply with one optional configuration: Option 001, which adds an HP-IB Interface for digital control of output voltage or current from a calculator, computer, or other controller. The base unit and the Option 001 unit share the same DC output characteristics — 0 to 50V and 0 to 10A with 200W maximum output power — and the same regulation, PARD, and overvoltage protection specifications.
The Option 001 modification involves the addition of three internal PC boards (A100 Input Board, A200 Digital Board, A300 Analog Board), two mode switches (CV/CC/LOCAL), and an HP-IB connector on the rear panel. Option 001 hardware does not change the supply's output performance — it adds a programming path. The datasheet specifies HP-IB programming speed of approximately 100msec (no load) to 200msec (full load) for output transitions between 0 and 100% or 100% to 0.1% of maximum rated output.
Each pre-owned 6002A configuration shown is its own product page with condition-matched pricing. Select the base 6002A or the HP-IB-equipped Option 001 variant based on whether your application requires digital programming from a controller, then choose the condition tier that matches your quality system and budget requirements.
The functional difference between the base 6002A and the Option 001 variant is the addition of HP-IB digital programming capability. The base unit is controlled exclusively via front panel knobs or rear terminal analog remote programming (resistance or voltage source). The Option 001 unit retains those local and analog control paths and adds HP-IB control through a 24-pin micro-ribbon connector on the rear panel, with a pair of CV/CC/LOCAL mode switches selecting the active control mode.
Programming resolution differs between the supply's two HP-IB programmable ranges (Option 001 only). The datasheet specifies high-range resolution of 50mV (CV) and 10mA (CC) and low-range resolution of 10mV (CV) and 2mA (CC); accuracy is 0.2% plus 25mV or 25mA depending on range and mode. Customers building automated test systems around legacy HP-IB infrastructure should specify Option 001; customers using the supply as a manually controlled bench instrument can use either configuration.
| Model | Output Voltage | Output Current | Output Power |
|---|---|---|---|
| 6002A | 0 to 50 V | 0 to 10 A | 200 W |
Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.
| Specification | Value |
|---|---|
| Input Power | 100/120/220/240 Vac (-13%, +6%), 48-63 Hz. 3-wire detachable line cord supplied. |
| DC Output | Adjustable from 0-50V and 0-10A. Maximum output power is 200W from 20-50V. |
| Load Effect (Load Regulation), Constant Voltage | Less than 0.01% of output plus 1mV for a load change equal to the maximum current rating of the supply. |
| Load Effect (Load Regulation), Constant Current | Less than 0.01% of output plus 1mA for a load change equal to the maximum voltage rating of the supply. |
| Source Effect (Line Regulation), Constant Voltage | Less than 0.01% of output plus 1mV for any line voltage change within rating. |
| Source Effect (Line Regulation), Constant Current | Less than 0.01% of output plus 1mA for any line voltage change within rating. |
| PARD (Ripple and Noise), Constant Voltage | Less than 1mV rms and 10mV p-p (20Hz to 20MHz). |
| PARD (Ripple and Noise), Constant Current | Less than 5mA rms. |
| Temperature Coefficient, Constant Voltage | Less than 0.02% plus 200µV change in output per degree Celsius change in ambient following a 30-minute warmup. |
| Temperature Coefficient, Constant Current | Less than 0.02% plus 5mA change in output per degree Celsius change in ambient following a 30-minute warmup. |
| Drift (Stability), Constant Voltage | Less than 0.05% of output plus 1mV (over 8-hour interval under constant line, load, and ambient temperature following a 30-minute warmup). |
| Drift (Stability), Constant Current | Less than 0.05% of output plus 5mA. |
| Load Transient Recovery Time | Less than 100µsec required for output voltage recovery (in constant voltage operation) to within 15mV of the nominal output following a change in output current from 50% to 100% or 100% to 50% of maximum current rating. |
| Remote Resistance Control | CV: 1KΩ/V ±7% CC: 100Ω/A ±7% |
| Remote Voltage Control | CV: 1V/V ±20mV (-50mV offset) CC: 50mV/A ±10% 7.5 mV offset |
| Response Time (UP) | No Load - 100msec; Full Load - 100msec |
| Response Time (DOWN) | No Load - 400msec; Full Load - 200msec |
| Overvoltage Protection | Trip voltage adjustable from 3.5V to 60V. Minimum setting above output voltage to avoid false tripping is 1 volt. |
| DC Output Isolation | 150 Vdc |
| Temperature Ratings | Operating: 0 to 55°C; Storage: -40 to +75°C |
| Resolution | CV: 10mV CC: 10mA |
| Output Impedance | Typical value is 0.5mΩ in series with 1µH. |
| Options | 001 HP-IB Interface |
| Net Weight | 14.5 kg (32 lb) |
| Shipping Weight | 15.9 kg (35 lb) |
Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.
Model No
HP
Condition
Used
Manufacturer
Agilent
Current
10 A
Voltage
50 V
Power
200 W
1
HP-IB Interface