Skip to content

The Agilent 8514B is an S-parameter test set designed for vector network analyzer systems operating from 45 MHz to 20 GHz. It provides the signal separation, switching, and reference sampling hardware required to perform full two-port S-parameter measurements, with 3.5 mm male connectors at both test ports and a 50 ohm nominal port impedance.
The 8514B is built around two test ports configured for transmission and reflection measurements across its full frequency range. Each test port carries a nominal operating power level of 2 to -6 dBm depending on band, and the test ports support DC bias up to 500 mA and 40 Vdc maximum for measurements that require an applied bias on the device under test. The rear-panel RF input connector is a 3.5 mm female with a maximum input power rating of +16 dBm.
The 8514B originated within Hewlett-Packard's test and measurement business. HP spun out its test and measurement instruments as Agilent Technologies in 1999, and that electronic measurement business was subsequently transferred to Keysight Technologies in 2014. Instruments carrying the original HP brand, the Agilent brand shown on this datasheet, and current Keysight branding share the same engineering lineage for this product family.
| Specification | Value |
|---|---|
| General | |
| Frequency Range | 0.045 to 20 GHz |
| Test Ports (port 1 or 2) | |
| Nominal Operating Power Level | 2 to -6 dBm |
| Connector Type | 3.5 mm, male |
| Impedance, DC Bias | 50 ohm nominal, 500 mA, 40 Vdc maximum |
| Attenuation Range (incident signal) | |
| Range | 0 to 90 dB, in 10 dB steps |
| RF Input Connector (rear panel) | |
| Maximum Input Power | +16 dBm |
| Connector Type | 3.5 mm, female |
| Dynamic Range (for transmission measurements) | |
| Frequency Range (GHz) | .045 to 2 / 2 to 8 / 8 to 20 |
| Maximum Power Measured at Port 2 (dBm) | +20 / +11 / +10 |
| Reference Power at Port 1, nominal (dBm) | +2 / -2 / -6 |
| Minimum Power Measured at Port 2 (dBm) | -66 / -95 / -95 |
| Receiver Dynamic Range (dB) | 86 / 106 / 105 |
| System Dynamic Range (dB) | 68 / 93 / 89 |
| Measurement Port Characteristics — Residual (dB) | |
| Frequency Range (GHz) | .045 to 2 / 2 to 8 / 8 to 20 |
| Directivity | 48 / 44 / 44 |
| Source Match | 40 / 33 / 31 |
| Load Match | 48 / 44 / 44 |
| Reflection Tracking | ±0.003 / ±0.003 / ±0.006 |
| Transmission Tracking | ±0.017 / ±0.044 / ±0.084 |
| Crosstalk | 89 / 115 / 110 |
| Raw (typical) (dB) (before calibration/error correction) | |
| Directivity | 23 / 23 / 14 |
| Source Match | 17 / 15 / 11 |
| Load Match | 17 / 15 / 11 |
Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.
Model No
Agilent
Condition
Used
Manufacturer
Agilent
Request A Quote