The Agilent E1475A is a 75 Ω RF multiplexer expander designed to inexpensively increase system switching capacity in C-size VXI test platforms. It is a register-based, A16 slave module occupying a single C-size slot, and it operates exclusively as a companion to either an E1472A or E1474A RF multiplexer installed in the same system — the E1475A is not a standalone switching instrument but an expander that adds six additional 1x4 multiplexer banks to its host multiplexer.
Within the VXI switching architecture, the E1475A is positioned for video and telecommunications signal routing where 75 Ω impedance matching is required. The expander supports signal frequencies up to 1.3 GHz, and the front panel terminates in 75 Ω SMB male connectors that share the same physical size as their 50 Ω counterparts — a practical detail for users migrating between impedance environments. Up to two E1475A expanders can be daisy-chained to a single host multiplexer, producing a total of eighteen 1x4 banks, and 75 Ω E1475A expanders can be mixed with 50 Ω E1473A expanders under the E1474A and E1472A multiplexers.
ValueTronics International stocks and supports legacy VXI switching modules such as the E1475A from our 20,000 sq ft secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. Each pre-owned unit passing through our facility is inspected and performance-tested before shipment, and we are positioned to source matching companion modules — such as the E1472A or E1474A host multiplexer required to operate the E1475A — when customers are restoring or expanding existing VXI-based switching systems rather than building from a single transaction.
The E1475A originated in the test and measurement division of Hewlett-Packard, which was spun off as Agilent Technologies in 1999. Agilent's electronic measurement business was further separated as Keysight Technologies in 2014, and Keysight is the current corporate successor for legacy HP and Agilent VXI instrumentation of this generation.
| Product Specifications | |
|---|---|
| Input | |
| Maximum voltage (center or shield-to-center, shield or chassis) | 42 V |
| Maximum current per channel or common (DC) | 1 A |
| Maximum current per channel or common (AC rms) | 1 A |
| Maximum power per channel or common (DC) | 24 W |
| Maximum power per channel or common (AC) | 24 VA |
| DC | |
| Maximum thermal offset | 6 µV |
| Closed channel resistance (typical) | <1 Ω initial |
| Insulation resistance (between any two terminals) | >10E8 Ω ≤40 °C, ≤65% RH |
| AC (Note: For AC performance, ZL=ZS=ZO, ≤40 °C, RH ≤95%) | |
| Characteristic impedance (Zo) | 75 Ω |
| Insertion loss <10 MHz | <0.3 dB |
| Insertion loss <100 MHz | <0.4 dB |
| Insertion loss <500 MHz | <0.8 dB |
| Insertion loss <1.3 GHz | <1.0 dB |
| Crosstalk (channel-to-channel) <10 MHz | <–85 dB |
| Crosstalk (channel-to-channel) <100 MHz | <–75 dB |
| Crosstalk (channel-to-channel, one channel closed or channel-to-common, terminated) <500 MHz | <–60 dB |
| Crosstalk (channel-to-channel, one channel closed or channel-to-common, terminated) <1.3 GHz | <–42 dB |
| VSWR <10 MHz | <1.05 |
| VSWR <100 MHz | <1.15 |
| VSWR <500 MHz | <1.35 |
| VSWR <1.3 GHz | <1.5 |
| Risetime | <300 ps |
| Signal delay | <3 ns |
| General Characteristics | |
| Relays | Non-latching armature |
| Power up/down state | All open |
| Minimum relay life (No load) | 5x10E6 operations |
| Minimum relay life (Rated load) | 10E5 operations |
| VXI Characteristics | |
| VXI device type | Register based, A16, slave only |
| Size | C |
| Slots | 1 |
| Connectors | P1 |
| Shared memory | None |
| VXI busses | None |
| Module Current | |
| +5 V (IPM / IDM) | 0 / 0 |
| +12 V (IPM / IDM) | 0.36 / 0 |
| –12 V (IPM / IDM) | 0 / 0 |
| +24 V (IPM / IDM) | 0 / 0 |
| –24 V (IPM / IDM) | 0 / 0 |
| –5.2 V (IPM / IDM) | 0 / 0 |
| –2 V (IPM / IDM) | 0 / 0 |
| Cooling/Slot | |
| Watts/slot | 1.00 |
| ΔP mm H2O | 0.02 |
| Air Flow liter/s | 0.10 |
| Instrument Drivers | |
| Command module firmware | Downloadable |
| Command module firmware rev | A.02 |
| I-SCPI Win 3.1 | Yes |
| I-SCPI Series 700 | Yes |
| C-SCPI LynxOS | Yes |
| C-SCPI Series 700 | Yes |
| Panel Drivers | Yes, with E1474A |
| VXIplug&play Win Framework | Yes |
| VXIplug&play Win95/NT Framework | Yes |
| VXIplug&play HP-UX Framework | No |