Key Features At A Glance

  • Source and receiver covering 10 kHz to 300 MHz in a single network analyzer
  • GPIB (ANSI/IEEE 488.2), parallel I/O, and EXT TRIGGER with built-in Instrument BASIC for automated systems
  • 1 to 4 measurement channels (depends on option) with 2 to 1,601 measurement points per sweep
  • Frequency accuracy ±20 ppm at 23 ±5 °C; optional ovenized reference (Option 1D5) for ±1 ppm
  • Phase noise < -90 dBc/Hz at 10 kHz offset from a 0 dBm fundamental
  • 6.5 inch color LCD (640 x 480), built-in flexible disk drive, and VGA monitor output
  • Frequency resolution 1 mHz; output power resolution 0.1 dB
  • Measurement calibration: response, response and isolation, and 1-port 3-term
View full details

The Agilent E5100A is a network analyzer covering 10 kHz to 300 MHz, integrating a signal source and a receiver in a single instrument. It supports one to four measurement channels and performs both magnitude and phase measurements, with measurement calibration types including response, response and isolation, and 1-port 3-term.

A network analyzer characterizes how a device behaves across frequency by comparing a stimulus signal against the device's response. The E5100A offers ratio measurement modes such as A/R, R/A, and, with the appropriate input configuration, B/R, B/A, A/B, and additional pairings, alongside absolute A, B, and R measurements. Both frequency and power sweep parameters are available, using linear (step, ramp) or list sweep types over a Cartesian display format.

ValueTronics stocks and holds its own inventory in a 20,000 square foot secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. Every used unit is inspected and functionally verified in-house before it ships, while new units leave the warehouse factory-sealed exactly as received. Holding inventory directly means the instrument you order is one we physically have on hand and have prepared to the standard appropriate for its condition.

Brand Heritage

The test and measurement line carrying the Agilent name originated with Hewlett-Packard, whose test and measurement business was spun out as Agilent Technologies in 1999. That business was subsequently transferred to Keysight Technologies in 2014.

Accessories Supplied

  • Power cable (Qty 1)
  • Sample program disk (Qty 1, E5100-180X01)
  • CD-ROM (manuals) (Qty 1, E5100-905XX)
  • Function Reference (Qty 1, E5100-900X02)
  • Programming Manual (Qty 1, E5100-900X72)
  • User's Guide (Qty 1, E5100-900X12)
  • Instrument BASIC Users Handbook (Qty 1, 04155-90150)
  • Instrument BASIC Users Handbook Supplement (Qty 1, E5100-900X52)

Product Core & Specifications

Specification Value
Source — Frequency Characteristics
Range 10 kHz to 300 MHz
Accuracy (at 23 ±5 °C) ±20 ppm
Accuracy with Option E5100A-1D5 (at 0 to 55 °C, 20 minutes after power on) ±1 ppm
Stability (at 23 ±5 °C) ±5 x 10–6/day (SPC)
Stability with Option E5100A-1D5 (48 hours after power on) ±2.5 x 10–9/8 hours (SPC)
Resolution 1 mHz
Source — Output Power Characteristics (measured at RF OUT 1, RF OUT 2 terminated with 50 Ω)
Range (nominal) with Option E5100A-001 –9 dBm to +11 dBm
Range (nominal) with Option E5100A-002 –15 dBm to +5 dBm
Range (nominal) with Option E5100A-003 –12 dBm to +8 dBm
Range (nominal) with Option E5100A-801 –48 dBm to +22 dBm
Range (nominal) with Option E5100A-802 –54 dBm to +16 dBm
Range (nominal) with Option E5100A-803 –51 dBm to +19 dBm
Range (nominal) with Option E5100A-600 (at RF OUT 1) –52 dBm to +18 dBm
Range (nominal) with Option E5100A-600 (at RF OUT 2) –65 dBm to +5 dBm
Resolution 0.1 dB
Level accuracy (at 23 ±5 °C, 0 dBm output level, 50 MHz) ±1 dB
Flatness (at 23 ±5 °C, relative to 0 dBm output level at 50 MHz) +2 dB, –4 dB
Flatness with Option E5100A-803 +2.5 dB, –4.5 dB
Linearity (at 23 ±5 °C, relative to 0 dBm output level at 50 MHz) ±1 dB
Source — Spectral Purity Characteristics
Non–harmonic spurious signals (at < 300 MHz) < –45 dBc
Phase noise (at 10 kHz offset from 0 dBm fundamental) < –90 dBc/Hz
Other Source Information
Reverse power protection 20 dBm, 25 Vdc (SPC)
Output connector BNC female
Output impedance 50 Ω (nominal)
Receiver — Input Characteristics
Frequency range 10 kHz to 300 MHz
1 MΩ input for Options E5100A-703/704/707/708 10 kHz to 5 MHz
IF bandwidth (IF BW) 10 Hz to 30 kHz, 1, 1.5, 2, 3, 4, 5, 8 step (nominal)
Impedance 50 Ω (nominal)
1 MΩ input for Option E5100A-703/704/707/708 1 MΩ // 30 pF (nominal)
Return loss (at 50 Ω input, 10 kHz ≤ frequency < 100 MHz) 20 dB (SPC)
Return loss (at 50 Ω input, 100 MHz ≤ frequency ≤ 300 MHz) 15 dB (SPC)
Damage level DC 25 Vdc
Damage level AC 20 dBm
Receiver — Averaging Noise Level (magnitude, 23 ±5 °C, RF attenuator: 0 dB, 50 Ω input)
IF BW 30 kHz (at > 1 MHz) –100 dBm
IF BW 10 kHz (at > 300 kHz) –105 dBm
IF BW 3 kHz (at > 100 kHz) –110 dBm
IF BW 1 kHz (30 kHz ≤ frequency < 100 kHz) –95 dBm
IF BW 1 kHz (100 kHz ≤ frequency ≤ 300 MHz) –115 dBm
IF BW 300 Hz (10 kHz ≤ frequency < 100 kHz) –100 dBm
IF BW 300 Hz (100 kHz ≤ frequency ≤ 300 MHz) –120 dBm
IF BW 100 Hz (10 kHz ≤ frequency < 100 kHz) –105 dBm
IF BW 100 Hz (100 kHz ≤ frequency ≤ 300 MHz) –125 dBm
Receiver — Magnitude Characteristics
Absolute amplitude accuracy (at 23 ±5 °C, –30 dBm input for RF attenuator: 0 dB or –5 dBm input for RF attenuator: 25 dB, 50 Ω input) ±1 dB
Frequency response (50 Ω input, 10 kHz ≤ frequency < 100 kHz) ±1 dB
Frequency response (50 Ω input, 100 kHz ≤ frequency ≤ 100 MHz) ±0.5 dB
Frequency response (50 Ω input, 100 MHz < frequency ≤ 300 MHz) ±1 dB
Frequency response (1 MΩ input for Option E5100A-703/704/707/708, using 50 Ω feedthrough) ±3 dB
Trace noise (at 1 kHz IF BW, frequency > 305 kHz) < 10 dBm rms
Stability 0.02 dB/°C (SPC)
Receiver — Phase Characteristics
Measurement mode Normal/Expanded
Frequency response (10 kHz ≤ frequency < 100 kHz) ±5°
Frequency response (100 kHz ≤ frequency ≤ 100 MHz) +2.5°
Frequency response (100 MHz < frequency ≤ 300 MHz) ±5°
Trace noise (at 1 kHz IF BW, frequency > 305 kHz) < 50 mdeg rms
Stability 0.15 deg/°C (SPC)
General Characteristics
Operating temperature (disk drive in operation) 10 to 40 °C
Operating humidity (disk drive in operation, wet bulb ≤ 29 °C, without condensation) 15% ≤ RH ≤ 80%
Operating temperature (disk drive not in operation) 5 to 40 °C
Operating humidity (disk drive not in operation, wet bulb ≤ 29 °C, without condensation) 15% ≤ RH ≤ 80%
Operating altitude 0 to 2,000 meters
Warm-up time 30 minutes
Non-operating temperature –20 to 60 °C
Non-operating humidity (wet bulb ≤ 40 °C, without condensation) 15% ≤ RH ≤ 90%
Non-operating altitude 0 to 4,572 meters
Safety Certified by CSA-C22.2 No.1010.1-92, Based on IEC 1010-1 (1990) including Amendment 1 (1992)
EMC Complies with CISPR 11(1990)/EN 55011(1991): Group 1, Class A; IEC 801-2 (1991)/EN 55082-1(1992): 4 kV CD, 8 kV AD; IEC 801-3 (1984)/EN 55082-1(1992): 3 V/m; IEC 801-4 (1988)/EN 55082-1(1992): 1 kV power lines, 0.5 kV signal lines
Power requirement 90 to 132 V or 198 to 264 V, 47 to 63 Hz, 400 VA max
Weight (depending on option) 12 kg (SPC)
Cabinet dimensions 425(W) x 177(H) x 425(D) mm (SPC)
Supplemental — Measurement Function
Number of measurement channels 1 to 4
Display format Cartesian
Sweep parameter frequency, power
Sweep type linear (step, ramp), list
Measurement points per sweep 2 to 1,601
Measurement calibration Response, response and isolation, 1-port 3-term
Display 6.5 inch color LCD, 640 x 480 dots
Flexible disk drive 720 Kbytes/1.2 Mbytes/1.44 Mbytes, DOS format, binary or ASCII format
Flash disk 256 Kbytes
Ram disk 256 Kbytes
Programming Instrument BASIC
GPIB ANSI/IEEE 488.2 compatible
Parallel I/O port 16 bit output, 8 bit input/output, TTL level
External video monitor output VGA
Important: Damage level: 25 Vdc (DC) and 20 dBm (AC). Reverse power protection: 20 dBm, 25 Vdc.
Important: When the analyzer is equipped with Option E5100A-001 or E5100A-003, the power splitter section does not apply (deleted). When equipped with Option E5100A-600, the power-splitter insertion-loss item does not apply.

About this Pre-Owned unit

This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.

  • Warranty included
  • Functional verification included
  • Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
  • Note that unnecessary accessories may not be included (contact Test Architect to confirm).

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.

Model No

Agilent

Condition

Pre-Owned

Manufacturer

Agilent

Net_ana_freq

10 kHz to 300 MHz

2

Two RF OUT Ports , built-in power splitter

6

Parallel I/O mode B , 16 bit output , 8 bit input/output , TTL level

10

Extended output power range

E5100A-1CM

Rackmount kit

E5100A-1CN

Handle kit

E5100A-1CP

Rackmount and handle kit

E5100A-1D5

High stability frequency reference

E5100A-1F0

PC style keyboard - U.S. version

E5100A-400

300MHz , 4 Receivers , Ports R , A , B and C

E5100A-600

300MHz X'tal resonator test to use w/PI-network

E5100A-706

Type-N 50 ohm input connectors , Ports A and B

E5100A-801