Key Features At A Glance

  • Frequency coverage from 300 kHz to 3 GHz across the two-model ENA-L series
  • Type-N test ports available in 50 Ω or 75 Ω configurations
  • Up to 4 independent measurement channels with 4 data and 4 memory traces each
  • 10.4-inch TFT color LCD (VGA); hardkey, optional touch screen, or mouse control
  • GPIB (IEEE-488), 10/100BaseT LAN, USB, and Handler I/O; LXI Class C compliant
  • System dynamic range up to 120 dB at 10 Hz IF bandwidth (1 MHz to 3 GHz)
  • Calibration types include response, one-port, enhanced response, and optional full 2-port
View full details

The Agilent ENA-L RF Network Analyzers comprise two models — the E5061A, covering 300 kHz to 1.5 GHz, and the E5062A, covering 300 kHz to 3 GHz. Both are vector network analyzers built on a common platform that employs a graphical user interface based on the Windows operating system, operable through a hardkey interface, an optional touch screen, or a mouse. Each analyzer provides up to four independent measurement channels, each with its own display window and up to four data traces and four memory traces.

A network analyzer characterizes how a device or network reflects and transmits signals across frequency. The ENA-L measures S-parameters — S11 and S21 in the transmission/reflection configuration, and the full S11, S21, S12, and S22 set with the full 2-port option — and converts those parameters into reflection and transmission impedance, reflection and transmission admittance, and 1/S. Available data formats include log and linear magnitude, phase, expanded phase, positive phase, group delay, SWR, real, imaginary, Smith chart, and polar. Fault-location transformation to the distance and time domains is available as an option, alongside limit-line definition for pass/fail testing.

ValueTronics stocks and holds its own test-and-measurement inventory in a 20,000-square-foot secure warehouse at 1675 Cambridge Drive in Elgin, Illinois, and every pre-owned unit is fulfilled directly from that facility. Used units are inspected and functionally verified in-house before they ship, while new units ship factory-sealed exactly as received from the manufacturer.

Brand Heritage

The ENA-L line carries Agilent Technologies branding. Agilent was established in 1999 as a spin-off of Hewlett-Packard, and its electronic test-and-measurement business was subsequently established as Keysight Technologies in 2014.

Compare Other Models in This Series

The ENA-L family is offered as two models sharing a common platform: the E5061A and the E5062A. Both provide the same measurement architecture — up to four channels, Type-N 50 Ω or 75 Ω test ports, the same selectable calibration types, and the same Windows-based interface — and differ primarily in their upper frequency limit.

The E5061A covers 300 kHz to 1.5 GHz, while the E5062A extends coverage to 3 GHz. Measurement-parameter scope (S11/S21 versus the full 2-port S11/S21/S12/S22 set) and capabilities such as fault location, extended power range, and touch screen are selected by option code rather than by base model.

Each model below is its own dedicated product page with pre-owned-matched pricing. Select the model whose frequency range and option configuration matches your measurement requirement, then open the listing for the condition you need.

The principal difference between the two models is frequency span: the E5061A reaches 1.5 GHz and the E5062A reaches 3 GHz, as shown in the comparison table. Test port output frequency range follows the model accordingly — 300 kHz to 1.5 GHz for the E5061A and 300 kHz to 3 GHz for the E5062A.

Beyond frequency range, the two models share corrected-performance specifications, connector options, calibration types, channel count, and interfaces. Measurement-parameter scope and features such as fault location, extended power range, and touch screen are determined by option codes that apply across both models.

Model Frequency Range Max Output Power Connector Type
E5061A 300 kHz to 1.5 GHz 10 dBm Type-N, 50 Ω or 75 Ω
E5062A 300 kHz to 3 GHz 10 dBm Type-N, 50 Ω or 75 Ω

Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.

Product Core & Specifications

Specification Value
System Dynamic Range (no averaging, 23 °C ±5 °C; applies to units with serial prefix MY442 and above)
300 kHz to 1 MHz, IF bandwidth = 3 kHz 90 dB
1 MHz to 3 GHz, IF bandwidth = 3 kHz 95 dB
300 kHz to 1 MHz, IF bandwidth = 10 Hz 115 dB
1 MHz to 3 GHz, IF bandwidth = 10 Hz 120 dB (130 dB typical)
Corrected System Performance — Type-N 50 Ω, 85032F cal kit, Full 2-Port (IF BW = 10 Hz, 23 °C ±5 °C); values shown as 300 kHz–1.5 GHz / 1.5–3 GHz
Directivity 49 dB / 46 dB
Source match 41 dB / 40 dB
Load match 49 dB / 46 dB
Reflection tracking ±0.011 dB / ±0.021 dB
Transmission tracking ±0.015 dB / ±0.018 dB
Corrected System Performance — Type-N 50 Ω, 85032F cal kit, Enhanced Response; 300 kHz–1.5 GHz / 1.5–3 GHz
Directivity 49 dB / 46 dB
Source match 41 dB / 40 dB
Load match 15 dB / 15 dB
Reflection tracking ±0.011 dB / ±0.021 dB
Transmission tracking ±0.015 dB / ±0.018 dB
Corrected System Performance — Type-N 75 Ω, 85036E cal kit, Full 2-Port; 300 kHz–1.5 GHz / 1.5–3 GHz
Directivity 48 dB / 44 dB
Source match 41 dB / 35 dB
Load match 48 dB / 44 dB
Reflection tracking ±0.010 dB / ±0.019 dB
Transmission tracking ±0.015 dB / ±0.029 dB
Corrected System Performance — Type-N 75 Ω, 85036E cal kit, Enhanced Response; 300 kHz–1.5 GHz / 1.5–3 GHz
Directivity 48 dB / 44 dB
Source match 41 dB / 35 dB
Load match 15 dB / 15 dB
Reflection tracking ±0.010 dB / ±0.019 dB
Transmission tracking ±0.015 dB / ±0.029 dB
Uncorrected System Performance (correction off, 23 °C ±5 °C, 300 kHz to 3 GHz)
Directivity 25 dB
Source match 25 dB
Load match 15 dB
Transmission tracking ±1.0 dB
Reflection tracking ±1.0 dB
Test Port Output Frequency
Range (E5061A) 300 kHz to 1.5 GHz
Range (E5062A) 300 kHz to 3 GHz
Resolution 1 Hz
Source stability ±5 ppm (5 °C to 40 °C)
CW accuracy ±5 ppm, 23 °C ±5 °C
Test Port Output Power (300 kHz to 3 GHz)
Level accuracy (23 °C ±5 °C) ±0.8 dB (at 0 dBm, 50 MHz absolute); ±1.0 dB (at 0 dBm, relative to 50 MHz reference)
Level linearity (23 °C ±5 °C) ±0.75 dB (at –5 to 10 dBm)
Range (standard) –5 to 10 dBm
Range (extended power) –45 to 10 dBm (non-harmonic spurious may limit power range)
Sweep range (without extended power range) –5 to 10 dBm
Level resolution 0.05 dB
Test Port Output Signal Purity
Harmonics (2nd or 3rd), 10 MHz to 2 GHz < –25 dBc (at 5 dBm)
Non-harmonic spurious, 10 MHz to 3 GHz < –30 dBc (at 5 dBm)
Test Port Input Levels (300 kHz to 3 GHz)
Maximum test port input level +10 dBm
Damage level +20 dBm, ±30 VDC
Crosstalk –110 dB
Test Port Input — Trace Noise (IF BW = 3 kHz, source power +10 dBm, 23 °C ±5 °C)
Trace noise magnitude, 300 kHz to 1 MHz 8 mdB rms
Trace noise magnitude, 1 MHz to 3 GHz 5 mdB rms
Trace noise phase, 300 kHz to 1 MHz 0.05° rms
Trace noise phase, 1 MHz to 3 GHz 0.03° rms
Test Port Input — Stability (23 °C ±5 °C)
Stability magnitude, 3 MHz to 3 GHz 0.01 dB/°C
Stability phase, 3 MHz to 3 GHz 0.1°/°C
System Bandwidths
IF bandwidth range 10 Hz to 30 kHz
Nominal IF bandwidth settings 10, 30, 100, 300, 1 k, 3 k, 10 k, 30 k
Front Panel Information
RF connectors Type-N, female; 50 Ω or 75 Ω
Display size 10.4 in TFT color LCD
Display resolution VGA (640 x 480)
Rear Panel Information
External trigger connector BNC, female; LOW threshold 0.5 V / HIGH threshold 2.1 V; input range 0 to +5 V; pulse width ≥ 2 µsec; negative (downward) polarity only
External reference signal input BNC, female; 10 MHz ±10 ppm; 0 dBm ±3 dB
Internal reference signal output BNC, female; 10 MHz ±10 ppm; sine wave; 0 dBm ±3 dB into 50 Ω; output impedance 50 Ω
VGA video output 15-pin mini D-Sub, female; drives VGA compatible monitors
GPIB 24-pin D-Sub (type D-24), female; compatible with IEEE-488
Parallel port 36-pin D-Sub (type 1284-C), female
USB port Type A, female (Vcc 4.75 to 5.25 VDC, 500 mA maximum)
LAN 10/100 BaseT Ethernet, 8-pin; auto selects data rate
Handler I/O port 36-pin Centronics, female
Line Power (a third-wire ground is required)
Frequency 47 Hz to 63 Hz
Voltage 90 to 132 VAC, or 198 to 264 VAC (automatically switched)
Important: Maximum test port input level: +10 dBm. Damage level: +20 dBm, ±30 VDC.
Important: A third-wire ground is required for the line power connection.
Important: Level accuracy for 75 Ω analyzers is not a specification for frequencies > 2 GHz; it is a typical characteristic.
Important: System dynamic range specifications apply only to units with serial prefix MY442 and above.
Important: LXI Class C compliance applies only to units shipped with firmware revision A.03.00 or later.
Important: Corrected system performance is achieved using an external calibration kit (85032F for 50 Ω Type-N, 85036E for 75 Ω Type-N); the calibration kit is a separate product and is required to perform full 2-port or enhanced response calibration.

About this Pre-Owned unit

This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.

  • Warranty included.
  • Functional verification included.
  • Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
  • Note that unnecessary accessories may not be included (contact Test Architect to confirm).

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.

Model No

Agilent

Condition

Pre-Owned

Manufacturer

Agilent

Net_ana_freq

300 kHz to 1.5 GHz

E5061B-020

Standard Data Storage

E5061B-115

Transmission/Reflection test set , 100 kHz to 1.5 GHz , 50 ohm system impedance

E5061B-117

Transmission/Reflection test set , 100 kHz to 1.5 GHz , 75 ohm system impedance

E5061B-135

Transmission/Reflection test set , 100 kHz to 3 GHz , 50 ohm system impedance

E5061B-137

Transmission/Reflection test set , 100 kHz to 3 GHz , 75 ohm system impedance

E5061B-1CM

Rack mount kit

E5061B-1CN

Front handle kit

E5061B-1CP

Rack mount and front handle kit

E5061B-1E5

High stability timebase

E5061B-215

S-parameter test set , 100 kHz to 1.5 GHz , 50 ohm system impedance

E5061B-217

S-parameter test set , 100 kHz to 1.5 GHz , 75 ohm system impedance

E5061B-235

S-parameter test set , 100 kHz to 3 GHz , 50 ohm system impedance

E5061B-237

S-parameter test set , 100 kHz to 3 GHz , 75 ohm system impedance

E5061B-3L3

LF-RF network analyzer with DC bias source , 5 Hz to 500 MHz

E5061B-3L4

LF-RF network analyzer with DC bias source , 5 Hz to 1.5 GHz

E5061B-3L5

LF-RF network analyzer with DC bias source , 5 Hz to 3 GHz

E5061B-720

Add 50 ohm resistor set

E5061B-810

Add keyboard

E5061B-820

Add mouse