Aim-TTi PSA1302 1.3GHz Portable RF Spectrum Analyzer

Aim-TTi PSA1302 1.3GHz Portable RF Spectrum Analyzer

Regular price $1,411
Sale price $1,411 Regular price $0
Shipping calculated at checkout.
Request a Quote
Estimated Delivery in
5 - 7
business days (based on availability)

🎓 Request a Quote To Get An EDU Discount

Login & Get 2% Back on Purchases
Login or Sign Up for Loyalty

Aim-TTi PSA1302 1.3GHz Portable RF Spectrum Analyzer

Key Features At A Glance

  • Frequency range from 1 MHz up to 1.3 GHz (PSA1302) or 2.7 GHz (PSA2702)
  • Genuinely handheld at 560 g (20 oz) and 92 mm wide, with ruggedised buffered casing
  • 4.3-inch colour TFT touch-screen with three-row hierarchical menu and hard-key control
  • More than 8 hours continuous Li-ion battery operation; recharges in under 3 hours
  • USB host for Flash drives and USB device for direct PC file transfer
  • Selectable resolution bandwidths of 1 MHz, 280 kHz or 15 kHz
  • Typically -96 dBm displayed noise floor (at -20 dBm reference level, 15 kHz RBW)
  • Optional U01 firmware upgrade adds limits comparator, data logging and compensation tables
View full details

The PSA Series II is a range of portable RF spectrum analyzers from Aim-TTi, comprising the PSA1302, covering 1 MHz to 1.3 GHz, and the PSA2702, covering 1 MHz to 2.7 GHz. It builds on the earlier PSA-T series to offer a lightweight, truly handheld instrument with an extensive feature set, an improved user interface, and a ruggedised casing with rubberised buffers top and bottom plus a removable screen protector and sun-shield for field use.

A spectrum analyzer displays signal amplitude across a range of frequencies, letting an engineer observe signals in the frequency domain to reveal information that cannot be seen by looking in the time domain with an oscilloscope. The PSA Series II is suited to both bench-top RF work and field work. Listed traditional field applications include antenna alignment, signal strength mapping, interference analysis, covert transmitter detection, spurious emissions, and EMC evaluations, with its wide span supporting investigation of equipment emissions for compliance with EMC requirements relating to CE or FCC approvals. Listed development and servicing uses include RF amplifier block design, oscillator and mixer design, power measurements, impedance matching measurement, IF section alignment, RFID measurements, site evaluations, and transmitted power measurements.

ValueTronics supports the PSA Series II from its own 20,000 sq ft secure warehouse at 1675 Cambridge Drive in Elgin, Illinois, where we stock and hold our own inventory rather than sourcing to order.

Brand Heritage

Aim-TTi is the brand name under which Thurlby Thandar Instruments Ltd, a United Kingdom test and measurement manufacturer, designs and markets its products. The company was formed through the merger of Thurlby Electronics and Thandar Electronics.

Compare Other Models in This Series

The PSA Series II comprises two models that share a common handheld platform and differ in their upper frequency limit: the PSA1302 covers 1 MHz to 1.3 GHz and the PSA2702 covers 1 MHz to 2.7 GHz. Both use the same 4.3-inch colour touch-screen, ruggedised casing, Li-ion battery, 1.8 GB internal memory, USB host and device interfaces, and selectable 1 MHz, 280 kHz, and 15 kHz resolution bandwidths.

Beyond frequency coverage, both models share the same amplitude measurement architecture, marker and trace functions, zero span AM/FM demodulation, and accessory ecosystem. The optional U01 firmware upgrade applies equally to either model, and USC variants bundle that upgrade with field accessories.

Each model below is its own dedicated product page with new-matched pricing. Select the model whose frequency coverage matches your application, then choose the condition that fits your budget and requirements.

The defining difference between the two models is the upper frequency limit and corresponding maximum span: the PSA1302 reaches 1.3 GHz with a maximum span of 1299 MHz, while the PSA2702 reaches 2.7 GHz with a maximum span of 2699 MHz. The comparison table that follows summarises these per-model values.

Both models otherwise carry identical specifications and are available as standard products or as USC versions (PSA1302USC and PSA2702USC) that add the pre-installed U01 firmware upgrade, the WB-ANT telescopic antenna, the PSA-VC vehicle charger, and the PSA2-SC travel case. Choose the 2.7 GHz model when coverage above 1.3 GHz is required; otherwise the 1.3 GHz model covers the same lower range.

Model Frequency Range Maximum Span Resolution Bandwidth
PSA1302 1 MHz to 1300 MHz 1299 MHz 1 MHz, 280 kHz, or 15 kHz
PSA2702 1 MHz to 2700 MHz 2699 MHz 1 MHz, 280 kHz, or 15 kHz
PSA1302USC 1 MHz to 1300 MHz 1299 MHz 1 MHz, 280 kHz, or 15 kHz
PSA2702USC 1 MHz to 2700 MHz 2699 MHz 1 MHz, 280 kHz, or 15 kHz

Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.

Accessories Supplied

  • Universal voltage mains adaptor/charger
  • USB connection lead
  • Spare stylus
  • BNC converter for trigger input
  • Multi-language Short Guide
  • Full users manual (English only)
  • Support CD containing PDF versions of manuals and support software

Product Core & Specifications

Specification Value
Frequency Measurement
Frequency Range 1 MHz to 1300 MHz (PSA1302); 1 MHz to 2700 MHz (PSA2702)
Setting Modes Centre frequency plus Span, or Start plus Stop frequencies
Maximum Span 1299 MHz (PSA1302); 2699 MHz (PSA2702)
Minimum Span 270 kHz, or Zero Span with demodulation
Setting Resolution 1 kHz at any frequency
Setting Accuracy Reference Frequency Accuracy for Start, Stop & Centre (Zero-Span) frequencies
Reference Frequency Accuracy
Initial Accuracy Better than ± 10 ppm at 20 °C
Stability Better than ± 10 ppm over 10 °C to 30 °C
Ageing Better than ± 3 ppm per year
Phase Noise
Phase Noise Phase noise at 100kHz offset at 500MHz typically -90dBc/Hz
Resolution Bandwidth
RBW Selectable 1 MHz, 280 kHz, or 15 kHz
Video Filtering Selectable independently of RBW setting
Frequency Markers
No. of Markers One, Two (or None)
Resolution 0.1 kHz at all frequencies
Marker Accuracy 1/270th of Span ± 0.1kHz plus reference frequency accuracy.
Readout The frequencies at the marker points and the frequency difference are displayed
Amplitude Range
Units Selectable as dBm or dBµV
Display Range 85 dB from reference level
Magnification x2, x5 or x10
Reference Level Selectable as -20 dBm or 0 dBm (87 dBµV or 107 dBµV)
Amplitude Accuracy
Calibration Level Accuracy Better than ± 1 dB at 10dB below ref. level @ 50MHz (20 °C ± 5 °C)
Flatness Better than ± 1.5 dB over the range 1 MHz to 1300 MHz (PSA1302); 1 MHz to 2700 MHz (PSA2702)
Linearity Better than ± 1 dB over 50dB from the reference level
Noise Floor Better than -93 dBm average displayed noise floor (typically -96 dBm) (reference level = -20 dBm, RBW = 15 kHz)
Distortion and Spurii
3rd Order Intermodulation < –60dBc for two signals at 10dB below reference level (500MHz and 502MHz)
Harmonic < –60dBc at 10dB below reference level (100MHz)
1st & 2nd Image <– 55dBc, typically <– 60dBc
Residual Spurii better than 3dB above noise floor
Amplitude Markers
No. of Markers One, Two (or None)
Resolution 0.1 dB
Readout The level at the marker points and difference are displayed.
Signal Input
Input Connector SMA connector, 50 Ω
VSWR 1.5 : 1 typical
Maximum Level + 20 dBm, (127 dBµV); +/-50V DC
Sweep
Sweep Method Peak detection for 270 points per sweep. The amplitude and frequency of the peak level found within each sub-span is stored (sub-span = span/270).
Sweep Time Set automatically by Span and RBW.
Sweep Modes Repeat (continuous) or Single Shot
Trace Writing Modes Normal (overwrite), Peak Hold, Average (2 to 48 sweeps).
Demodulation (Zero Span mode)
Demod. Modes AM or FM
Display Carrier amplitude only (horizontal line).
Audio Internal loudspeaker.
Audio Out 30 mW into 32 Ω mono or stereo headphones, adjustable volume, 3.5mm jack socket
Audio Filter Switchable 3kHz Low Pass Filter.
Display
Display Type 4.3 inch (10.9 cm) backlit TFT LCD, 480 x 272 pixels total, 16 colours, touch screen.
Trace Area 232 x 272 pixels.
Graticule 8.5 x 10 divisions, light grey graticule.
Displayed Points 271 points per sweep (peak detected).
Live Trace Dot-joined trace from current sweep.
View Trace Buffered “instance” of the live trace.
Reference Trace Stored trace recalled from a trace file.
Memory Storage
Internal Disk 1.8GB of internal memory.
External Storage USB host interface for removable USB Flash drives.
Store Trace Up to 999 traces can be stored under either default file names or user entered file names. Traces are stored as tables of amplitude versus frequency and can be imported into other programs, as well as being recalled to the screen.
Recall Trace Recalls any stored trace to the reference trace of the display.
Store Set-up Up to 999 instrument set-ups can be stored under either default file names or user entered file names. All settings of the instrument are saved.
Recall Set-up Recalls any stored set-up, overwriting the existing settings of the instrument.
Store Screen This function copies the whole screen area to memory as a bit-map. Up to 999 screens can be stored under either default file names or user entered file names.
Recall Screen Recalls any stored screen as an image.
Connectors
RF Input Standard SMA connector.
DC Power 1.3 mm power socket for external power supply/charger
USB Host Standard USB type A connector for connection of USB Flash drives.
USB Device Mini USB connector for connection to a PC.
Audio Out 3.5 mm jack socket for demodulated audio out (accepts mono or stereo plugs).
Important: Maximum input level: +20 dBm (127 dBuV); +/-50 V DC.
Important: The Trigger In/Out connectors are for use with Option U01 only.
Important: Logging files, limit patterns and compensation tables are only available with Option U01.
Important: Option U01 is pre-installed within the PSA1302USC and PSA2702USC products.
Important: The telescopic antenna (WB-ANT) is intended for general purpose applications where absolute measurements of field strength are not required; its useful bandwidth is specified by the manufacturer as 30MHz to 1.8GHz.

About this New unit

This is a brand-new, factory-sealed unit, supplied exactly as it arrives from the manufacturer and backed by the full manufacturer warranty. For the exact configuration, available options, included accessories, or current lead time for your application, our Test Architects can confirm the details before you order.

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.

Model No

Aim-TTi

Condition

Factory New

Manufacturer

Aim-TTi

Spec_ana_freq

1300 MHz