Key Features At A Glance

  • Frequency coverage from 10 kHz to 23 GHz across the series
  • External waveguide mixer support extends measurement coverage up to 140 GHz on supported models
  • Resolution bandwidth selectable from 100 Hz to 3 MHz in a 1-3-10 sequence, manual or span-coupled
  • Two-channel digital memory for dual-trace display, trace subtraction, MAX HOLD, and AVERAGING
  • Five marker functions: Normal, Delta, Peak, Marker→CF, and Signal Track
  • GPIB (IEEE-488) remote control of all front-panel functions for automated measurement systems
  • N-type 50 Ω input with +30 dBm maximum input level
View full details

The MS710C/D/E/F is a family of high-performance microwave spectrum analyzers spanning 10 kHz to 23 GHz, with an optional measurement range up to 140 GHz when external mixers are used. The design pairs a simplified PLL synthesizer local oscillator with a two-channel digital memory and supports both easy manual operation and fully automatic operation over GPIB. Operator-focused design choices include a grouped key layout with key sizes scaled to functional importance, an operation guide display for complex SHIFT and MARKER sequences, and a preset memory that holds up to 10 sets of measurement conditions.

The series is applied to spectrum analysis of microwave devices and components; spurious emission and spectral distribution measurements of analog and digital communications transmitters; interference measurements for radio stations, satellite earth stations, and similar installations; spectrum analysis in basic research such as nuclear physics and radio-astronomy; and spurious measurements for home-use satellite broadcast receivers and related equipment.

ValueTronics stocks and holds its own inventory in a 20,000-square-foot secure warehouse at 1675 Cambridge Drive in Elgin, Illinois, where we have specialized in new and used test and measurement equipment since 1992. Every pre-owned unit is inspected and functionally verified in-house before it ships, while new units ship factory-sealed exactly as received from the manufacturer. Holding our own stock lets us match the right instrument and condition to each application and ship promptly from our own shelves.

Compare Other Models in This Series

The MS710C/D/E/F is a four-model microwave spectrum analyzer series that shares a common architecture — the same resolution bandwidth range, marker and memory functions, display, and GPIB automation — while differing in frequency coverage and frequency accuracy. All four models cover the 100 kHz to 2 GHz and 1.7 to 23 GHz bands; the variations between models concern the low-frequency band, the external-mixer extension to 140 GHz, and whether the model includes high frequency accuracy.

Because the core measurement functions are common across the family, the choice between models comes down to the required frequency span and accuracy class rather than a difference in operating workflow. The comparison table below summarizes how each model is configured.

Each pre-owned model below is its own dedicated product page with condition-matched pricing. Select the specific model that matches your required frequency coverage and accuracy, then choose the pre-owned listing that fits your budget and lead-time needs.

The MS710C covers the widest frequency range in the series: it is the only model that adds the 10 kHz to 30 MHz band, and it combines high frequency accuracy with external-mixer support that extends measurement coverage up to 140 GHz.

The four models differ along three axes. The MS710C provides the 10 kHz to 30 MHz band, high frequency accuracy, and external-mixer extension to 140 GHz. The MS710D covers 100 kHz to 23 GHz with external-mixer extension to 140 GHz. The MS710E covers 100 kHz to 23 GHz with high frequency accuracy.

High frequency accuracy is provided on the MS710C and MS710E, while external-mixer operation up to 140 GHz is supported on the MS710C and MS710D. Refer to the comparison table for the per-model frequency coverage and accuracy details.

Model Frequency Range External Mixer Extension High Frequency Accuracy
MS710C 10 kHz to 23 GHz 18 to 140 GHz Yes
MS710D 100 kHz to 23 GHz 18 to 140 GHz No
MS710E 100 kHz to 23 GHz Yes
MS710F 100 kHz to 23 GHz No

Additional differences in specifications beyond the few shown above are not listed here — see each model's full specifications below.

Accessories Supplied

  • J0104A Coaxial cord, 1 m (BNC-P • RG-55/U • N-P): 1 pc
  • J0017 Power cord, 1 m (plug type must be specified): 1 pc
  • F0013 (F0011) Fuse, 5 A or 2 A: 2 pcs
  • F0010 Fuse, 1.6 A: 1 pc
  • F0011 Fuse, 2 A: 1 pc
  • F0012 Fuse, 3.15 A: 1 pc
  • W0087AE MS710[ ] operation manual: 1 copy
  • W0087BE MS710[ ] service manual: 1 copy

Product Core & Specifications

Specification Value
10 kHz to 30 MHz Band (MS710C)
Measuring frequency 10 kHz to 30 MHz
Center frequency setting range 0 kHz to 30 MHz
Center frequency readout resolution 1 kHz
Center frequency readout accuracy ±(3 kHz + 2% of frequency span + 10% of resolution bandwidth)
Frequency span setting range and resolution 0 Hz (fixed tuning); 1 to 200 kHz/div in 1 kHz increments; 210 kHz/div to 2 MHz/div in 10 kHz increments; 2.1 to 3 MHz/div in 100 kHz increments; span up/down keys: 1 kHz/div to 2 MHz/div in 1-2-5-10 sequence and 3 MHz/div
Frequency response ±1.5 dB (10 kHz start frequency, 30 MHz stop frequency, 10 dB input attenuator)
2nd harmonic distortion ≤−60 dB (10 to 300 kHz, −40 dBm); ≤−70 dB (300 kHz to 15 MHz, −30 dBm) — input level less input attenuator value
Two-signal 3rd intermodulation distortion ≤−70 dB (10 to 100 kHz, Δf ≥2.5 MHz, total −30 dBm)
Residual response ≤−90 dBm
Average noise level ≤−95 dBm (100 kHz to 1 MHz), ≤−115 dBm (1 to 30 MHz); 1 kHz resolution bandwidth, 0 dB input attenuator, 3 Hz video bandwidth
Note Other specifications are the same as the 100 kHz to 2 GHz and 1.7 to 23 GHz band
18 to 140 GHz Band (with external mixer)
Frequency band and harmonic number 18.0–26.5 GHz: 6; 22.0–33.0 GHz: 6; 26.5–40.0 GHz: 8; 40.0–60.0 GHz: 10; 60.0–90.0 GHz: 16; 90.0–140.0 GHz: 26
Center frequency resolution 100 kHz (18 to 60 GHz), 1 MHz (60 to 140 GHz)
Center frequency readout accuracy 30 kHz × harmonic number
Frequency span setting range and resolution 0 Hz (fixed tuning); 1 kHz×n/div to 200 kHz×n/div in 1 kHz×n increments; 210 kHz×n/div to 2 MHz×n/div in 10 kHz×n increments; 2.1 MHz×n/div to 20 MHz×n/div in 100 kHz×n increments; 21 MHz×n/div to 200 MHz×n/div in 1 MHz×n increments; span up/down keys: 1 kHz×n/div to 200 MHz×n/div (n: harmonic number)
Frequency span readout accuracy ±5% (6 kHz×n/div to 200 MHz×n/div), ±10% (1 kHz×n/div to 5 kHz×n/div)
Resolution bandwidth (6 dB bandwidth) setting range 100 Hz to 3 MHz in 1-3-10 sequence
Selectivity (60 dB/6 dB) ≤10 : 1 (resolution bandwidth ≤1 kHz)
Amplitude measuring range Average noise level to +30 dBm
Reference level setting range −105 to +30 dBm (LOG), −9.5 to +30 dBm (LIN)
Calibration output accuracy −10 dBm ±0.3 dB (100 MHz ±10 kHz)
Reference level accuracy ±2.0 dB (reference level −99 to −10 dBm, frequency 100 MHz, 0 dB input attenuator, after calibration using CAL OUTPUT)
Frequency response Depends on external mixer
Average noise level Depends on external mixer (−100 dBm typical with 30 dB external mixer conversion loss, 1 kHz resolution bandwidth)
Video bandwidth 1 Hz to 3 MHz, 1-3-10 sequence
Local output for external mixer 3 to 6 GHz, ≥+7 dBm
IF input for external mixer 521.4 MHz
Note Other specifications are the same as the 100 kHz to 2 GHz and 1.7 to 23 GHz band
Important: Maximum input level: +30 dBm, ±0 Vdc.
Important: 18 to 140 GHz measurements on the MS710C and MS710D require an external waveguide mixer, which is not supplied with the instrument and must be ordered separately. When using a 3-port external mixer, an additional amplifier (such as the MP11975A) is required for local signal amplification.
Important: Local output for external mixer: 3 to 6 GHz, ≥+7 dBm; IF input for external mixer: 521.4 MHz.
Important: Power cord plug type must be specified at time of order.
Recommended pairing: the MH680A1 Tracking Generator and MH648A Pre-amplifier are offered as application equipment for the MS710 series.

About this Pre-Owned unit

This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.

  • Warranty included
  • Functional verification included
  • Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
  • Note that unnecessary accessories may not be included (contact Test Architect to confirm).

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.

Model No

Anritsu

Condition

Pre-Owned

Manufacturer

Anritsu

Spec_ana_freq

10 kHz to 23 GHz