The HP 4142B Modular DC Source/Monitor is a user-configurable DC parametric tester from Hewlett-Packard with an open architecture mainframe accepting plug-in measurement modules. The mainframe provides eight slots that accept a mix of Source/Monitor Units (SMUs), High Current Units (HCUs), High Voltage Units (HVUs), Voltage Source/Voltage Monitor Units (VS/VMUs), and the Analog Feedback Unit (AFU). The instrument is positioned by HP for DC characterization of semiconductor devices, process control in semiconductor manufacturing, and device modeling in research and development.
Applications listed for the pre-owned HP 4142B include verifying process quality in semiconductor manufacturing, developing sub-micron and high-power technologies, characterizing new materials, and evaluating finished components. Device types named on the datasheet include GaAs FETs and diodes, bipolar transistors, MOS and CMOS structures, power bipolar and MOSFETs, thyristors and SCRs, operational amplifiers, resistive networks, and smart power devices. The pulse capability is presented as a method for reducing thermal drift error and enabling tests on high-power devices, including at the wafer level.
Every {{CONDITION}} HP 4142B sold by ValueTronics is processed through our 20,000 sq ft secure warehouse at 1675 Cambridge Drive, Elgin, Illinois, where our technical team verifies each mainframe and its installed modules before shipment. This in-house handling is the structural difference between a verified instrument from ValueTronics and an unverified marketplace listing — and it is the reason engineers and procurement teams have relied on us as a single source for legacy HP, Agilent, and Keysight parametric test equipment since 1992.
Hewlett-Packard's test and measurement division was spun off as Agilent Technologies in 1999, and the electronic measurement business of Agilent became Keysight Technologies in 2014. Instruments originally branded HP — including the 4142B family documented here — are part of the lineage now supported under the Keysight name.
| Specification | Value |
|---|---|
| Chassis | |
| Plug-in Slots | 8 slots for plug-in measurement units |
| AFU Limit | No more than one HP 41425A Analog Feedback Unit per mainframe |
| Maximum Common to Ground Voltage | ±42V |
| Maximum Output Power | |
| Total Instantaneous Output Power | The sum of all measurement unit's instantaneous output power cannot exceed 32W |
| Ground Unit (GNDU) | |
| Connection | 0 Volt, Kelvin |
| Maximum offset voltage | ±500µV |
| Current range | ±1.6A |
| Maximum cable resistance (FORCE) | ≤1Ω |
| Maximum cable resistance (SENSE) | ≤10Ω |
| Maximum capacitive load | 10µF |
| Control Unit (CTLU) | |
| Module Selector Output Voltage | 24V |
| Module Selector Current Limit | 30mA |
| Module Selector Typical Control Speed | 30msec |
| 16-Bit Relay Max Voltage | 20V |
| 16-Bit Relay Saturation Voltage | 0.7V (at sink current = 50mA) |
| 16-Bit Relay Pull-up Voltage/Resistor | 4.5V/10kΩ |
| 16-Bit Relay Typical Control Speed | 10ms |
| Memory | |
| Program memory | Stores approx. 2000 program steps, which can be grouped into 99 subroutines |
| Data memory | Maximum of 4095 data values (binary); Maximum of 1023 data values (ASCII) |
| Interfaces | |
| External trigger input | TTL level negative logic, minimum pulse width 100µsec |
| External trigger output | TTL level negative logic, approx pulse width 100µsec |
| HP-IB interface | SH1, AH1, T6, L4, SR1, RL1, PP0, DC1, DT1, C0, E1 |
| General Reference | |
| Self-Test | At power-up the HP 4142B checks operational status. The self-test can be performed at any time via HP-IB |
| Auto-Calibration | Offset errors in each measurement unit are automatically calibrated every 30 minutes |
| Environmental | |
| Operating temperature | 5°C to 40°C |
| Allowable temperature drift | ±3°C (after auto-calibration) |
| Operating humidity | 5% to 80% RH |
| Storage temperature | -40°C to 65°C |
| Storage humidity | ≤90% RH at 65°C |
| Operating inclination | ±20° from horizontal |
| Power Requirements | |
| Line Voltage Tolerance | 100/120/220V: ±10%; 240V: -10% to +5% |
| Maximum VA | 750 (48–66 Hz) |
| Physical | |
| Dimensions | 426mm W × 235mm H × 676mm D |
| Weight (mainframe) | 23 kg |
| Typical Measurement Times (HP 9000 Series 332) | |
| HP 41420A/41421B SMU Force I or V | 3.5msec |
| HP 41420A/41421B SMU Measure I or V | 4.0msec |
| HP 41425A AFU (Vt at I0 = 1µA) | 12msec |
| HP-IB Data Transfer (ASCII) | 1300µsec/point |
| HP-IB Data Transfer (Binary) | 450µsec/point |
| Software Included | |
| Parameter Measurement Library | Resistance, MOS, and bipolar transistor algorithm sets |
| Test Instruction Set | Initialize, Force, Measure, Pulse, Sweep, Graphics, and Data Storage |
Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.