# Keysight U3606B Multimeter | DC Power Supply (Pre-Owned)

# Keysight U3606B Multimeter | DC Power Supply (Pre-Owned)

## Key Features At A Glance

-   Two instruments in one box: a 5½-digit digital multimeter combined with a 30-W DC power supply in a single unit
-   120,000-count resolution with 0.025% basic DC voltage accuracy (1-year specification)
-   10 DMM functions: DCV, ACV, DCI, ACI, 2- and 4-wire resistance, frequency, continuity, diode, and capacitance
-   4-wire milliohm measurement with 0.001 mΩ resolution for low-resistance characterization
-   8 built-in math functions
-   OVP and OCP load protection to safeguard the device under test
-   Ramp and scan sweep (up to 100 scan steps, 10,000 ramp steps) plus square-wave output to 4.8 kHz
-   GPIB (IEEE-488) and USB 2.0 interfaces for remote control and automated test systems
-   Kensington lock slot for physical security against theft or misplacement

The Keysight U3606B Multimeter | DC Power Supply is a one-box source-and-measure device that pairs a full-featured 5½-digit digital multimeter with a built-in 30-W DC power supply. Because a single unit can power the device under test while simultaneously measuring voltage and current, it performs two test functions in the same instrument, occupying a compact footprint in place of two separate boxes. The combined unit is lighter than a standalone DMM and power supply together, making it straightforward to move between benches.

The U3606B is suited to applications across education, commercial electronics, semiconductors, sensors, and research and development. Its ramp and scan sweep functions simplify device characterization for multilevel DC bias testing such as DC motor testing, transistor gain test, relay control, and margin tests. The square-wave output serves uses including pulse-width modulation (PWM), adjustable voltage control, and synchronous clock generation, and supports checking and calibrating flow-meter displays, tachometers, LEDs, sensors, oscilloscopes, frequency converters, frequency transmitters, and other frequency-input devices.

ValueTronics stocks and holds its own inventory in a 20,000 square-foot secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. Every pre-owned unit is inspected and functionally verified in-house before it ships, while new units ship factory-sealed exactly as received from the manufacturer. Holding equipment on-site lets our team support buyers directly and turn orders around quickly.

## Brand Heritage

Keysight Technologies traces its test-and-measurement roots to Hewlett-Packard, whose instrument business was spun off as Agilent Technologies in 1999. Agilent's electronic measurement division was then established as Keysight Technologies in 2014, the brand under which this product is offered today.

## Accessories Supplied

-   Certificate of Calibration
-   U8201A Combo Test Lead Kit
-   USB 2.0 High-Speed Type-A to Type-B cable
-   AC power cord
-   Quick Start Guide
-   Product Reference CD
-   Agilent IO Library Suite

## Product Core & Specifications

Specification

Value

DC Voltage Accuracy (1 Year, 23 ºC ± 5 ºC), ± (% of reading + % of range)

19.9999 mV range

0.025 + 0.04

100.000 mV range

0.025 + 0.008

1.00000 V range

0.025 + 0.005

10.0000 V range

0.025 + 0.005

100.000 V range

0.025 + 0.005

1000.00 V range

0.025 + 0.005

DC Current Accuracy (1 Year, 23 ºC ± 5 ºC), ± (% of reading + % of range)

10.0000 mA range (burden < 0.2 V)

0.05 + 0.015

100.000 mA range (burden < 0.2 V)

0.05 + 0.005

1.00000 A range (burden < 0.3 V)

0.15 + 0.007

3.0000 A range (burden < 0.7 V)

0.15 + 0.007

Resistance Accuracy (1 Year, 23 ºC ± 5 ºC), ± (% of reading + % of range)

100.000 Ω range (0.83 mA)

0.05 + 0.008

1000.00 Ω range (0.83 mA)

0.05 + 0.005

10.0000 kΩ range (100 µA)

0.05 + 0.005

100.000 kΩ range (10 µA)

0.05 + 0.005

1.00000 MΩ range (900 nA)

0.06 + 0.005

10.0000 MΩ range (205 nA)

0.25 + 0.005

100.000 MΩ range (205 nA || 10 MΩ)

2.00 + 0.005

Low-Resistance (4-Wire) Accuracy (1 Year, 23 ºC ± 5 ºC), ± (% of reading + % of range)

100 mΩ range

0.25 + 0.05

1000 mΩ range

0.25 + 0.03

10 Ω range

0.09 + 0.03

100 Ω range

0.09 + 0.03

1000 Ω range

0.09 + 0.03

Continuity / Diode Accuracy (1 Year, 23 ºC ± 5 ºC), ± (% of reading + % of range)

Continuity, 1.0000 kΩ range (0.83 mA)

0.05 + 0.005

Diode, 1.0000 V range (0.83 mA)

0.05 + 0.005

Capacitance Accuracy (1 Year, 23 ºC ± 5 ºC), ± (% of reading + % of range)

1.000 nF range (0.75 µA current source)

2.0 + 0.8

10.00 nF range (0.75 µA)

1.0 + 0.5

100.00 nF range (8.3 µA)

1.0 + 0.5

1.000 µF range (83 µA)

1.0 + 0.5

10.00 µF range (83 µA)

1.0 + 0.5

100.0 µF range (83 µA)

1.0 + 0.5

1000 µF range (0.83 mA)

1.0 + 0.5

10000 µF range (0.83 mA)

2.0 + 0.5

AC True RMS Voltage Accuracy (1 Year, 23 ºC ± 5 ºC), ± (% of reading + % of range)

100.000 mV range, 20 to 45 Hz

1 + 0.1

100.000 mV range, 45 Hz to 10 kHz

0.2 + 0.1

100.000 mV range, 10 to 30 kHz

1.5 + 0.3

100.000 mV range, 30 to 100 kHz

5 + 0.3

1.00000 V to 750.00 V, 20 to 45 Hz

1 + 0.1

1.00000 V to 750.00 V, 45 Hz to 10 kHz

0.2 + 0.1

1.00000 V to 750.00 V, 10 to 30 kHz

1 + 0.1

1.00000 V to 750.00 V, 30 to 100 kHz

3 + 0.2

AC True RMS Current Accuracy (1 Year, 23 ºC ± 5 ºC), ± (% of reading + % of range)

10.0000 mA to 3.0000 A, 20 to 45 Hz

1.5 + 0.1

10.0000 mA to 3.0000 A, 45 Hz to 1 kHz

0.5 + 0.1

10.0000 mA to 3.0000 A, 1 to 10 kHz

2 + 0.2

Frequency Accuracy (1 Year, 23 ºC ± 5 ºC), ± (% of reading + % of range)

Voltage path 100 mV to 750 V, < 2 Hz

0.18 + 0.003

Voltage path 100 mV to 750 V, < 20 Hz

0.04 + 0.003

Voltage path 100 mV to 750 V, 20 Hz to 100 kHz

0.02 + 0.003

Voltage path 100 mV to 750 V, 100 to 300 kHz

0.02 + 0.003

Current path 10 mA to 3 A, < 2 Hz

0.18 + 0.003

Current path 10 mA to 3 A, < 20 Hz

0.04 + 0.003

Current path 10 mA to 3 A, 20 Hz to 10 kHz

0.02 + 0.003

Duty Cycle and Pulse Width

Duty cycle range / resolution / accuracy of full scale

100.000% / 0.001% / 0.3% + 0.2% per kHz

Pulse width range / resolution

199.999 ms (0.001 ms) and 1999.99 ms (0.01 ms)

DMM Measurement Characteristics

Maximum display counts, 5½ digits

120,000

Maximum display counts, 4½ digits

12,000

CMRR (1 kΩ unbalanced in LO lead)

DC: 140 dB; AC: 70 dB

NMRR, 60 Hz ± 0.1%

5½ digit: 65 dB; 4½ digit: 0 dB

NMRR, 50 Hz ± 0.1%

5½ digit: 55 dB; 4½ digit: 0 dB

DC voltage maximum input

1000 Vdc on all ranges

DC voltage input impedance

10 MΩ ± 2% range (typical) in parallel with capacitance < 120 pF

DC voltage input protection

1000 Vrms on all ranges

AC voltage maximum input

750 Vrms / 1200 Vpeak / 3 x 10^7 V-Hz of product

AC voltage input impedance

1 MΩ ± 2% range (typical) in parallel with capacitance < 120 pF

DC/AC current maximum input

10 mA to 3.0 A

Current input protection

Protected with 3.15 A/500 V, FF fuse

DC Power Supply Output Ratings

S1S2 (AUTO)

30 V/1 A and 8 V/3 A

S1

30 V/1 A

S1m

100 mA/30 V

S2

8 V/3 A

S2m

1000 mV/3 A

DC Power Supply Performance (1 Year, 23 ºC ± 5 ºC), ± (% of output + offset)

Programming accuracy, voltage

0.05% + 5 mV

Programming accuracy, current

0.15% + 3 mA

Readback accuracy, voltage

0.05% + 5 mV

Readback accuracy, current

0.15% + 3 mA

Ripple and noise, normal mode voltage (20 Hz to 1 MHz)

< 2 mVrms; < 30 mVpp

Ripple and noise, normal mode current

< 1 mArms

Front terminal load regulation, voltage

< 3 mV + (6 mV/A)

Front terminal load regulation, current

< 0.03% + 0.3 mA

Rear terminal load regulation, voltage

< 0.01% + 3 mV

Rear terminal load regulation, current

< 0.03% + 0.3 mA

Line regulation, voltage

3 mV typical

Line regulation, current

1.5 mA typical

Programming / readback / front panel resolution, voltage

1 mV (0.1 mV for S2m)

Programming / readback / front panel resolution, current

0.1 mA (0.01 mA for S1m)

> **Important:** Safety class I instrument: the protective earth terminal must be connected to an earth ground through a power source with a 3-wire ground receptacle.

> **Important:** Any current measurement greater than 500 mA will have a temporary thermo-effect. If you wish to measure a lower current or offset current immediately after a high-current measurement, ensure that the U3606B has cooled down.

> Recommended pairing: the 11059A Kelvin probe set or 11062A Kelvin clip set listed among the accessories supports the four-wire low-resistance (milliohm) measurement function.

> **Important:** For resistance measurements above 100 kΩ, a shielded test cable is recommended to eliminate noise interference that may be induced by the test leads.

> **Important:** For 2-wire resistance measurements without using the Null math operation, add a 0.2 Ω error.

> Recommended pairing: the U8202A Electronic Test Lead Kit and 34133A Precision Electronic Test Leads listed as accessories for the DMM function.

## About this Pre-Owned unit

This pre-owned unit is inspected and functionally verified in-house before it ships and is backed by our pre-owned warranty. To confirm its exact condition, firmware revision, installed options, or included accessories for your application before ordering, contact our Test Architects.

-   Warranty included
-   Functional verification included
-   Standard Calibration Upgrade Options: No Calibration Required, NIST Traceable, Z540.1 or ISO 17025 with Data, Z540.3 Guardbanding with Data.
-   Note that unnecessary accessories may not be included (contact Test Architect to confirm).

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 0.0 USD - 270.0 USD
- **Vendor:** Keysight
- **Type:** Multi Met
- **Tags:** Agilent, Agilent Keysight, Manufacturers, Meters & DAQ, Multimeter, product_manual_group_1, Product_used

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| None | 0.00 USD | In stock |
| NIST Traceable | 190.00 USD | In stock |
| NIST Traceable with Full Data | 225.00 USD | In stock |
| ISO IEC 17025 Accredited | 270.00 USD | In stock |