# Aim-TTI TGR6000 6 GHz RF Signal Generator

# Aim-TTI TGR6000 6 GHz RF Signal Generator

## Key Features At A Glance

-   10 MHz to 6000 MHz CW output, −110 dBm to +7 dBm in 0.1 dB steps with 10 Hz frequency resolution
-   Full remote control via RS-232, USB, GPIB (IEEE-488) and LAN (Ethernet) interfaces for ATE integration
-   Low phase noise: typ <−110 dBc/Hz at 20 kHz offset on a 500 MHz carrier
-   Stepped and list sweep of frequency and amplitude, up to 1000 points, dwell times from 10 ms to 10 s
-   User trim table of up to 100 amplitude/frequency pairs to calibrate an entire test set-up
-   Internal 1 ppm timebase plus 10 MHz external reference input/output for higher precision
-   Compact half-rack 2U case; 19-inch rack mount kit available as an option
-   Type N 50 Ω output with 50 V DC reverse protection

The Aim-TTi TGR6000 is a 6 GHz RF signal generator covering 10 MHz to 6000 MHz. It is intended for CW (carrier wave) applications where modulation is not required, an approach that avoids the costs associated with adding modulation and keeps it significantly lower in cost than other 6 GHz generators. Output level is adjustable from −110 dBm to +7 dBm in 0.1 dB steps, with frequency set to 10 Hz resolution and entered in dBm, µV/mV, or dBµV.

An RF signal generator produces a known, controllable signal used to stimulate a device under test, against which an engineer measures the response. The TGR6000 serves CW and swept or stepped CW applications: a single carrier can be held steady, or frequency and amplitude can be swept across a defined range or driven through a user-defined list to match virtually any required test pattern. Its trim function lets the output level be adjusted to calibrate an entire test set-up.

ValueTronics stocks and holds its own inventory in a 20,000 square foot secure warehouse at 1675 Cambridge Drive in Elgin, Illinois. Operating since 1992, we support engineers and labs with both new and used test and measurement equipment across a broad range of categories.

## Brand Heritage

Aim-TTi is the brand under which Thurlby Thandar Instruments Ltd. designs and builds its test instruments in Europe, based in Huntingdon, Cambridgeshire, United Kingdom.

## Product Core & Specifications

Specification

Value

Frequency Setting

Frequency Range

10MHz to 6000MHz

Setting Resolution

10Hz

Accuracy/Stability

See Reference Frequency section.

Phase Noise

500MHz Carrier

<-110dBc/Hz (typ) @ 20kHz offset; <-120dBc/Hz (typ) @ 100kHz offset

3GHz Carrier

<-95dBc/Hz (typ) @ 20kHz offset; <-110dBc/Hz (typ) @ 100kHz offset

6GHz Carrier

<-89dBc/Hz (typ) @ 20kHz offset; <-104dBc/Hz (typ) @ 100kHz offset

Residual FM

Residual FM

12 Hz @ 500MHz. Equivalent peak deviation in a 300Hz to 3.4kHz bandwidth.

Switching Speed

Settling Time

<8ms to settle within 100Hz or 0.1ppm of final frequency, if greater.

Frequency Reference

Reference Accuracy

<± 1ppm, 15ºC to 30ºC; <± 2ppm, 5ºC to 40ºC

Reference Stability

<1ppm/year

Ext. Reference In

10MHz, 50Ω input impedance, 2 to 5Vpp. Automatic detection and selection when external reference signal is present and Reference IN is selected from front panel. Front panel indicator shows when External Reference is active.

Int. Reference Out

10MHz, 50Ω output impedance, >2Vpp into 50Ω. Signal present when Reference OUT is selected from front panel.

Level Setting

Level Range

–110dBm to +7dBm

Setting Parameters

dBm, dBµV or µV/mV

Setting Resolution

0.1dB, 0.01uV to 1mV

Setting Accuracy

± 2dBm

Signal Purity

Harmonically Related

<-25dBc @ +7dBm, <-30dBc@levels <=0dBm 30 to 6000MHz; <-25dBc@levels <=0dBm 10 to 30MHz

Sub-harmonic

<3000MHz: None. >3000MHz: <–40dBc (typ) @ +7dBm

Non-harmonic Spurii

<–50dBc >10kHz offset 10 to 3000MHz (Note 1); <-44dBc >10kHz offset 3000MHz to 6000MHz (Note 2). Note 1 - <-45dBc >10kHz offset 1900–2150MHz. Note 2 - <-39dBc >10kHz offset 3800–4300MHz.

Output Connection

Output Impedance

50Ω

Output Connector

Type N

Reverse Protection

50V DC

Output Switch

RF OUT On/Off switch with indication of ON status

Step Sweep

Max Points

1000

Formula Specifies

Frequency start/stop. Amplitude start/stop. Dwell time at each step – programmable 0.01 to 10.000sec

Sweep Run

Continuous or single. Sweep up or down

Step Spacing

Linear or logarithmic

Sweep Triggering

Manual, Ext. signal, timed (0.1 – 999.9sec) or via remote interface.

Sync Signal

(Output Stable) available during dwell time. Programmable to be high or low

List Sweep

Max Points

1000

List Storage

Up to 16 Sweep Lists can be stored permanently within the instrument

Point Trigger

Point Triggering

Manual, Ext. signal or via remote interface.

Trim (User Level Compensation Tables)

Max. Points

100

TRIG IN

Input Threshold

1.65V nominal. Trigger polarity can be set to Negative Edge or Positive Edge.

Input Protection

Maximum/minimum external applied voltage is +6V or –1V.

SYNC Out

Active Output Level

+5V (Active state set to 'Pos') or 0V (Active state set to 'Neg')

Output Impedance

50Ω. Minimum load impedance is also 50Ω.

Output Protection

Output will withstand accidental short circuit to ground and applied external voltages up to +5V.

Digital Interfaces

RS-232

9-pin D connector. Variable Baud rate can be set between 1200 and 115200 Baud maximum.

USB

USB 2.0 connection (backwards compatible with USB 1.x). Operates as a virtual COM port.

GPIB (IEEE-488)

The interface conforms with IEEE-488.1 and IEEE-488.2.

Ethernet (LAN)

Standard 10/100 base-T hardware connection. ICMP and TCP/IP Protocol for connection to Local Area Network or direct connection to a single PC.

General Specifications

AC Input

110-240VAC ±10% 50/60Hz; 100-120VAC ±10% 400Hz; 60VA max. Installation Category II.

Operating Range

+5ºC to +40ºC, 20% to 80% RH

Storage Range

\-20ºC to +60ºC

Environmental

Indoor use at altitudes up to 2000m, Pollution Degree 2.

Safety

Complies with EN61010-1

EMC

Complies with EN61326

Display

20 character x 4 row backlit alphanumeric LCD

Data Entry

Keyboard selection of all major parameters. Value entry by character scroll using rotary control or up/down keys, or value stepping in user-selected increment values using rotary control or up/down keys

Stored Settings

Up to 12 complete set-ups. Up to 16 sweep lists

Location

Built-in tilt feet for bench-top use. Rack mountable with optional mount.

Size

86.5mm (2U) height; 213.5mm (½-rack) width; 350mm long

Weight

3.65kg (8lb)

> **Important:** Output reverse protection: 50 V DC.

## About this New unit

This is a brand-new, factory-sealed unit, supplied exactly as it arrives from the manufacturer and backed by the full manufacturer warranty. For the exact configuration, available options, included accessories, or current lead time for your application, our Test Architects can confirm the details before you order.

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 6289.01 USD
- **Vendor:** Aim-TTI
- **Type:** Rf Gen
- **Tags:** Manufacturers, product_manual_group_2, Product_new, RF, Signal Generators, Thurlby Thandar Instruments

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| Default Title | 6289.01 USD | In stock |

## Images

- TGR6000 Thurlby Thandar Instruments RF Generator

---

> Source: [ValueTronics](https://valuetronics.com/products/tgr6000-thurlby-thandar-instruments-rf-generator-new)