# Aim-TTi TG1006 10MHz DDS Function Generator with 120MHz Counter

# Aim-TTi TG1006 10MHz DDS Function Generator with 120MHz Counter

## Key Features At A Glance

-   0.001 Hz to 10 MHz frequency range generated by direct digital synthesis (DDS) for accuracy and stability
-   Sine, square and triangle waveforms, with low-distortion, high spectral purity sine output
-   10 ppm one-year frequency accuracy at 1 mHz / 6-digit setting resolution
-   Built-in 3 Hz to >120 MHz seven-digit reciprocal frequency counter, sensitivity better than 50 mV rms
-   Phase-continuous sweep (linear or logarithmic, single or continuous) plus a manual sweep mode for quasi-analog control
-   AM, FSK and frequency-list modulation modes available on all waveforms
-   2 mV to 20 V pk-pk output from a 50 Ω or 600 Ω source impedance
-   Non-volatile storage for up to ten output frequency settings

The Aim-TTi TG1006 is a 10 MHz DDS function generator that combines waveform generation with sweep, modulation, and a 120 MHz frequency counter in a single 2U bench instrument. It was developed from the TG300 and TG550 series of analog function generators, but uses direct digital synthesis (DDS) frequency generation to provide greater accuracy and stability and to cover a wider frequency range. DDS generates waveforms digitally using a phase accumulator, a look-up table, and a DAC, so the accuracy and stability of the output are tied to that of the crystal master clock.

A function generator produces known, controllable test signals used to stimulate a device under test so that its response can be measured. On the TG1006, the manual sweep mode uses a dedicated analog control to sweep the frequency over any defined span, which is suited to setting up frequency-dependent parameters such as checking filter characteristics. The non-volatile frequency-list storage holds commonly used values that can be recalled directly or stepped through in sequence as required in a repetitive test routine.

ValueTronics stocks and holds its own inventory in a 20,000 square foot secure warehouse at 1675 Cambridge Drive in Elgin, Illinois.

## Brand Heritage

This instrument is built by Thurlby Thandar Instruments Ltd. (TTi), one of Europe's manufacturers of test and measurement instruments. Products have historically been sold under two brand names, TTi and Aim, and the full product range is transitioning to a single Aim-TTi brand. The changeover is gradual, so many products continue to carry the TTi or Aim brands.

## Product Core & Specifications

Specification

Value

Waveforms — Sine

Range

1mHz to 10MHz

Resolution

1mHz or 6 digits

Accuracy

10 ppm for 1 year; ± 1mHz below 0.2Hz

Temperature Stability

Typically <1 ppm/ºC outside 18° to 28°C

Output Level

1mV to 10Vp-p into 50Ω

Harmonic Distortion

<0.3% THD to 20kHz (typically 0.1%), <–45dBc to 300kHz, <-30dBc to 10MHz (typically <-35dBc)

Non-harmonic Spurii

<–55dBc to 1MHz, <–55dBc + 6dB/octave 1MHz to 10MHz

Waveforms — Square

Range

1mHz to 10MHz

Resolution

1mHz or 6 digits

Symmetry Control

20% to 80% 1mHz to 10MHz

Accuracy

10 ppm for 1 year; ± 1mHz below 0.2Hz

Output Level

1mV to 10Vp-p into 50Ω

Rise and Fall Times

<25ns

Aberrations

<5% + 2mV

Waveforms — Triangle

Range

1mHz to 1 MHz

Resolution

1mHz or 6 digits

Accuracy

10 ppm for 1 year; ± 1mHz below 0.2Hz

Output Level

1mV to 10Vp-p into 50Ω

Linearity Error

<0.5% to 100 kHz

Operating Modes — Continuous

Continuous

Continuous cycles of the selected waveform are output at the programmed frequency.

Operating Modes — Sweep

Carrier Waveforms

All

Sweep Mode

Manual, linear or logarithmic, single or continuous.

Sweep Width

From 0.1Hz to 10MHz in one range. Phase continuous. Independent setting of the start and stop frequency.

Sweep Time

100ms to 999s (10ms resolution).

Sweep SYNC

Start of sweep trigger available from SYNC output.

Trigger Source

The sweep may be free run or triggered from the front panel MAN TRIG key.

Manual Sweep Mode

An analogue control can be used to set any frequency between the sweep start and sweep stop frequencies.

Operating Modes — Amplitude Modulation

Carrier Frequency

From 1mHz to 10MHz.

Carrier Waveforms

All

Modulation Frequency

400Hz internal. DC to 20kHz external

External Modulation

AM/COUNT IN socket

Operating Modes — Frequency Shift Keying (FSK)

Description

Phase coherent switching between two selected frequencies at a rate defined by the switching signal source.

Carrier Frequency

From 0.1Hz to 10MHz.

Carrier Waveforms

All

Switch Repetition Rate

DC to 10kHz (internal trigger).

Switching Signal Source

Manual (front panel MAN TRIG key) or internal trigger generator

Operating Modes — Frequency List

Carrier Waveforms

All

Frequency List

Up to 10 frequencies from 1mHz to 10MHz

Switching Source

Manual from front panel MAN TRIG key

Outputs — Main Outputs

Output Impedance

50Ω and 600Ω (not independent)

Amplitude

2mV to 20V pk-pk open circuit, (1mV to 10V pk-pk into 50Ω/600Ω) in four switch selectable ranges with 20dB vernier control within each range. (Amplitude can be displayed in pk-pk or r.m.s.)

Attenuator

0, -20dB, -40dB, or -60dB.

Amplitude Flatness

±0.2dB to 500kHz; ±2dB to 10MHz.

DC Offset Range

±10V. DC offset plus signal peak limited to ±10V from 50Ω/600Ω; CLIP shows in display when offset plus signal peak exceeds ±10V. DC offset plus waveform attenuated proportionally by the attenuator.

Resolution

3 digits for both Amplitude and DC Offset.

Outputs — SYNC Out

Waveform Sync

A square wave at the main waveform frequency. Symmetry is 50% for sine and triangle waves at MAIN OUT; for square waves symmetry is the same as that of the waveform at MAIN OUT.

Sweep Sync

Outputs a trigger signal at the start of sweep to synchronize an oscilloscope.

Output Signal Level

Output impedance 50Ω nominal. Logic levels of <0.8V & >3V.

Inputs — AM In

Input Impedance

40kΩ

Input Sensitivity

Approximately 2V peak-peak for 100% modulation.

Max. Allowable Input

±10V

Inputs — Count In

Input Impedance

1MΩ/20pF

Input Sensitivity

50mVrms (sinewave)

Max. Allowable Input

30Vdc/30Vrms to 50Hz/60Hz with respect to ground, reducing to 1Vrms above 1MHz.

Display Functions — Internal Measurement Accuracy

Amplitude

Display shows peak-to-peak amplitude or rms value. Display corrected for attenuator setting. 3-digit resolution, accuracy typically ±5% of full scale.

DC Offset

3-digit resolution; accuracy typically ±2% setting ±1 digit. Display corrected for attenuator setting.

Frequency Setting

Resolution up to 6 digits, see Waveforms section for setting accuracy.

External Frequency Measurement

Frequency Range

3Hz to >120MHz.

Resolution

Up to 7 digits

Input Sensitivity

better than 50mVrms (sinewave).

Measurement Time

Automatic.

Accuracy

±1 digit ± timebase accuracy.

Timebase Accuracy

±5ppm initial error; ± 5 ppm/year ageing rate; typically < 0.1ppm/°C.

General Specifications

AC Input

110-120V AC or 220V-240V AC ±10%, 50/60Hz, 35VA max. Installation Category II.

Operating Range

+5ºC to +40ºC, 20% to 80% RH

Storage Range

\-20ºC to +60ºC

Environmental

Indoor use at altitudes up to 2000m, Pollution Degree 2.

Safety

Complies with EN61010-1

EMC

Complies with EN61326

Display

LCD, 8 digits plus annunciators

Data Entry

Frequency entry by numeric keys or by rotary control.

Stored Settings

Up to 10 output frequencies may be stored and recalled from non-volatile memory. All frequencies settings including list, FSK and sweep parameters are stored at power down and restored at switch-on.

Size

260mm (W) x 88mm (H) x 235mm (D).

Weight

1.45kg. (3.2lb.).

> **Important:** Maximum allowable input on the AM/COUNT IN socket: 30 V DC / 30 V rms to 50/60 Hz with respect to ground, reducing to 1 V rms above 1 MHz.

## About this New unit

This is a brand-new, factory-sealed unit, supplied exactly as it arrives from the manufacturer and backed by the full manufacturer warranty. For the exact configuration, available options, included accessories, or current lead time for your application, our Test Architects can confirm the details before you order.

ValueTronics supplies both new and used test and measurement equipment. New units ship factory-sealed, exactly as received from the manufacturer; every used unit is inspected and functionally verified in-house at our 20,000 sq ft secure facility in Elgin, Illinois before it ships. Our Test Architects can help you select the condition, calibration, and configuration that fit your application.

**Please review the Manufacturer's Data Sheet to verify published specifications. Feedback on this webpage is always welcome — please reach out to your Test Architect at any time for questions or concerns. Thank you, we truly appreciate you being our customer.**

## Details

- **Price:** 589.0 USD
- **Vendor:** Aim-TTi
- **Type:** Func Gen
- **Tags:** Function, Manufacturers, product_manual_group_2, Product_new, Signal Generators, Thurlby Thandar Instruments

## Variants

| Variant | Price | Available |
|---------|-------|-----------|
| Default Title | 589.00 USD | In stock |

## Images

- TG1006 Thurlby Thandar Instruments Function Generator

---

> Source: [ValueTronics](https://valuetronics.com/products/tg1006-thurlby-thandar-instruments-function-generator-new)